Introduction

In recent decades, photonic crystal fiber (PCF) has supplanted traditional fiber optics, and with technological progress, it has increasingly integrated with surface plasmon resonance (SPR) to offer exceptional sensing capabilities and user-friendly functionality [1]. PCF-based SPR sensors have some advantages as compared to the traditional prism-based sensors. In prism-based sensors, SPR takes place at specific incident angles of incoming light, necessitating various mechanical components, whether fixed or movable, to achieve these conditions [2, 3]. This mechanical complexity adds bulkiness to the complete experimental setup. To overcome these limitations, in 1993, Jorgenson introduced PCF-SPR-based bioinstrumentation [4]. On the contrary, PCF offers several advantages, including cost-effectiveness, compactness, and robustness making it a desired choice for sensing detection and higher sensitivity as compared to the fiber and prism based SPR sensors. Furthermore, PCF sensors offer the significant advantage of easy resonance wavelength (RW) tuning by adjusting various parameters (such as air hole size and pitch distance) during fabrication [5]. Selecting the right plasmonic material is also crucial to achieve strong excitation of plasmon waves on the PCF surface. In the literature, gold is often favored as the greatest striking plasmonic material in PCF-based sensors due to its high chemical stability and biocompatibility [6]. Gold exhibits a larger resonance wavelength shift than other plasmonic materials. However, it is worth noting that this choice can lead to broader resonance peaks which gives low sensing range. Silver can be used as a noble plasmonic material to address this issue, offering narrower resonance peaks and attaining higher sensitivity with better detection accuracy (DA) [7]. Nonetheless, silver is susceptible to rapid oxidation in humid environments, potentially compromising the accuracy of analyte detection. To overcome the rapid oxidation of the silver and to enhance surface to volume ratio, a few nanometers thick TiO2 coating has been incorporated on the top of the silver layer. By harnessing the unique properties of PCF, SPR-based sensors employ a range of sensing mechanisms. These include slotted-based, nanowire-based, and both external and internal metal coating-based approaches [8, 9]. Among these techniques, internal and external metal coating-based sensing methods are the most generally employed [9,10,11].

Lately, this approach has gained extensive popularity for photonic crystal fiber-surface plasmon resonance sensing applications due to its simple and highly practical identification capabilities [12]. When assessing sensitivity, it is crucial to contemplate a design that enables a streamlined manufacturing process while attaining outstanding high-performance features [13]. In a study by Wang et al. a PCF-SPR design based on a D-shaped biosensor demonstrated a wavelength sensitivity (WS) of 3340 nm/refractive index unit (RIU) [14]. However, achieving such high sensitivity often requires excessive surface polishing, which can pose manufacturing challenges. In another study focused on external detection, Hasan et al. reported a sensor with a maximum amplitude sensitivity (AS) of 371.5 RIU–1 and a WS of 4600 nm-RIU–1, although, in this study, the AS was not as high compared to some other designs. Recently, there has been research on surface plasmon resonance-based diamond ring optical fiber sensors. This study introduced a diamond ring PCF-SPR sensor, demonstrating a maximum WS of 6000 nm/RIU within a sensing range spanning from 1.33 to 1.39 RIU [14]. Additionally, another open ring channel-based PCF-SPR biosensor was studied in the same study, which showcased its capability to detect small variations in refractive indices (RI) of real-time analytes, ranging from 1.23 to 1.29 RIU. This sensor exhibited an average WS of 5500 nm-RIU−1 and an AS of 333.8 RIU–1 [15]. The envisioned exhibition of the PCF holds significant promise for refractive index (RI) sensing applications, thanks to its noteworthy sensing capabilities. Biosensors play a crucial role in various fields such as healthcare, environmental monitoring, and food safety by providing real-time and sensitive detection of biological and biochemical substances. However, existing biosensor technologies often face limitations in terms of sensitivity, practicality, and ease of fabrication [16]. Addressing these limitations while introducing novel features is essential for advancing biosensor capabilities. The proposed dual-core photonic crystal fiber-based plasmonic biosensor addresses these limitations through several novel features and design choices like Plasmonic Enhancement, Compact and Practical Design, Simplified Fabrication, High Sensitivity, and Versatility. Further, several numerical methods have been developed for biosensing applications. Various advantages like ease of operation and design, a lower identification threshold, provide electrochemical biosensor. To identify the change in mass due to biomolecular interaction proposes piezoelectric sensor using the piezoelectric effect. To check the thermodynamic changes that emerge as a result of the cascade effect of physiological effect can be utilized as temperature biosensors. Recently, optical biosensing has gained significant attention as a technique in the field of biosensing. The operation of optical biosensors relies on the principle that the detection of an analyte molecule initiates a series of events leading to either the emission or absorption of light. Surface plasmon resonance (SPR), luminescence, absorption, and fluorescence are various techniques for biosensing in optical biosensors.

Milk infections are currently an important threat to public health, particularly among babies [17]. Lactoglobulin (β-LG) is a compact, globular-form allergenic protein present in the milk of various mammals. Therefore, considering β-LG, a key allergen, for diagnosis is a preferred choice. The β-LG monomer has a standard size of diameter 3 nm, comprising 162 amino acid residues, and possesses a molar mass of 18.40 k Da. Numerous biosensors to identify milk proteins, particularly β-LG, have been proposed [18]. The present paper explores a theoretical analysis of a SPR sensor based on DC-PCF. The sensor design presents a hexagonal lattice structure with circular air holes, which contributes to its simplicity. Sensor’s performance has been investigated employing the Finite Element Method (FEM) with a circular Perfectly Matched Layer (PML) boundary condition [19]. The sensor identifies lower concentration of the biomolecules with the property of PCF-SPR in considerably limiting the electric field in the narrower small RI core area. Accomplish all the investigation using β-LG molecules in a liquid state as a particular case; however, the outcomes may be applied to identify other tiny molecules as well. The core switched to β-LG molecules, creating an adlayer of the biomolecule. Hence, the RW shifts spectrally because of β-LG molecule concentration. The present system depends on the very precise spectral interrogation technique, which quantifies variations in the biosamples refractive index and thickness of affinity layer by monitoring spectral shifts in the RW. However, a more realistic approach has been devised and used to measure the refractive index of an adlayer produced via surface sensing. Through optimization of structural parameters, enhanced sensitivity in terms of amplitude sensitivity, wavelength sensitivity, and Figure of Merit (FOM) have been achieved. Present analysis is geared toward maximizing these key sensor characteristics. The highly sensitive nature of this sensor, combined with its straightforward structure, offers promising prospects for practical implementation in various optical sensing applications [20]. By employing the FEM numerical approach, the optical and dielectric characteristics of the offered biosensor as it guides electromagnetic waves in the Vis–NIR spectrum within the fiber core have been analyzed. The “Introduction” section presents the state of art of the present work. A brief outline of the device’s geometry and the fundamental physical principles governing its operation is presented in the “Design and Theoretical Backdrop” section. The “Formulation and Theoretical Interpretation” section provides a detailed explanation of the specific FEM techniques utilized in this study. Moving forward, the “Optimization of Sensor” section includes results and discussion, which is devoted to the analysis and discourse on the obtained outcomes. The “Performance analysis of the sensor” section shows experimental fabrication possibility. Finally, the concluding remark is presented in the “Fabrication Feasibility and Applicability” section.

Design and Theoretical Backdrop

Design Considerations

The sensor is designed using the Finite Element Method (FEM) in COMSOL Multiphysics 5.1 [21]. Upon launching the program, a two-dimensional model (2D) is selected for the space dimension. The chosen physics is Electromagnetic Wave, Frequency Domain (ewfd), and the study type is set to mode analysis. Subsequently, the geometries, materials, and mesh type are specified to measure the sensor’s confinement loss based on the imaginary portion of the effective refractive index (neff) of the fundamental mode. The cross-sectional view of the proposed sensor with different layers is shown in Fig. 1a. The two hexagonal rings are organized by lattice air holes, maintaining consistent center-to-center distances referred to as the pitch (p). The phase-matching properties can be made easier by a smaller air hole with a diameter of (dc) in the middle. To implement the dual-core scheme, two of the inner ring’s air holes are eliminated. Regular air holes have a (d) diameter. Four scaled-down (d2) and two scaled-up (d1) air holes in the outlying ring help to create a robust coupling between the surface plasmon polaritons (SPPs) wave. These air holes result adequate space to propagate light effectively through the metal-dielectric surface and improve sensitivity [6].

Fig. 1
figure 1

a 2D illustrative cross-sectional view and b developmental layout of the offered sensor

For practical sensing, a schematic experimental setup is depicted in Fig. 1. An optical tunable source (OTS) is required to launch the incoming light, and a polarizer accompanying a polarizer controller is required to transmit the light. Through single-mode fiber (SMF), linearly polarized light has been launched into the proposed sensor. A splicing method can be initiated in the proposed sensor to couple the SMF. The inlet and outlet of the liquid analyte are aided using an analyte flow tube that lies at the external border. It is possible to preserve the analyte intake and outflow by employing a pump. The change in the effective refractive index of the plasmonic mode owes shift in the signal’s wavelength, which occurs with the interaction of the analyte and ligand. The optical spectrum analyzer (OSA) is employed to compute the transmitted light, linked with a sensor by other SMF. With the help of an output spectrum, the shifting of wavelength can be noticed and examined on the computer. Blue shift occurs when the RW shifts toward the lower wavelength; on the contrary, shift toward the large wavelength is known as red shift.

Selection of RI

Fused silica has been employed as the background material whose wavelength-dependent refractive index can be evaluated employing the following Sellmeier Eq. (1) [22].

$${{n}_{s}}^{2}\left(\lambda \right)=1+\frac{{B}_{1}{\lambda }^{2}}{{\lambda }^{2}-{C}_{1}}+\frac{{{B}_{2}\lambda }^{2}}{{\lambda }^{2}-{C}_{2}}+\frac{{{B}_{3}\lambda }^{2}}{{\lambda }^{2}-{C}_{3}}$$
(1)

where ns (\(\lambda\)) is refractive index at a specified wavelength (\(\lambda\)) in (µm) and \(\lambda\) represents the wavelength of light source. B(i=1,2,3) and C(i=1,2,3) denote material-particular coefficients that can be taken from reference [23]. The variation of RI of fused silica changes 1.28 × 10−5 (per degree Celsius) only in response to temperature. Consequently, in typical environments with no significant temperature fluctuations, the impact of temperature can be disregarded.

The sensor’s plasmonic component consists of a small silver layer. The Drude model is suitable for determining the dielectric permittivity of silver metal. This model analytically approximates the dielectric function associated with intra-band transitions of conduction electrons as they interact with an electromagnetic wave. When experienced in an electric field, the dielectric function is further modified by the polarization of electrons in the innermost coatings of noble metals, indicating optical shift from the valence bands to conduction bands. The expression of the Drude model for a thin silver film is followed by Eq. (2) [7]:

$${\epsilon }_{{\text{Ag}}}={\epsilon }_{\infty }-\frac{{\omega }_{D}^{2}}{\omega \left(\omega +j{\gamma }_{D}\right)}-\frac{\Delta \epsilon {\Omega }_{L}^{2}}{\left({\omega }^{2}-{\Omega }_{L}^{2}\right)+j{\Gamma }_{L}\omega }$$
(2)

The symbol denotes \({\epsilon }_{{\text{Ag}}}\) the dielectric constant of silver,\({\epsilon }_{\infty }=5.9673\) high frequency permitivity, \({\omega }_{D}\) plasma frequency, \(\omega\) angular frequency and \({\gamma }_{D}\) damping frequency, \(\Delta \epsilon\) weight coefficient, \({\Omega }_{L}\) oscillator strength, and \({\Gamma }_{L}\) Lorentz oscillator’s frequency width. The merits have been used from the reference.

Between the Ag layer and the silica glass, a thin TiO2 coating has been utilized. The Ag has weak adherence without the TiO2 coating and easily flakes off with light pressure. The required adherence can be provided by a coating of TiO2. Additionally, TiO2 film has a maximum refractive index and functions as a transition metal that can produce a significant number of electrons at the surface. This generates a potent evanescent wave that pulls the core mode into the SPP mode and intensifies their interaction. The RI of TiO2 film can be obtained employing the Eq. (3) [24].

$${{n}_{{{\text{TiO}}}_{2}}}^{2}=5.913+\frac{2.441\times {10}^{7}}{{\uplambda }^{2}-0.803\times {10}^{7}}$$
(3)

where \({n}_{{{\text{TiO}}}_{2}}\) denotes the RI of \({{\text{TiO}}}_{2}\) and \(\lambda\) represents the operating wavelength (µm). On the outer surface of the PCF, it is crucial for the biosensor to deposit consecutively Ag and \({{\text{TiO}}}_{2}\) layer. The refractive index (np) of β-LG can be formulated employing a two term Cauchy Eq. (4) given below [20]:

$${n}_{{\text{p}}}=B+\frac{9.6\times 10{10}^{3}}{{\lambda }^{2}}$$
(4)

where the intercept B for β-LG has been taken 1.568.

For different concentration of β-LG in water, the dielectric constant of the bio-sample (ϵse) can be calculated employing Maxwell-Garnet effective medium theory with the help of Eq. (5) [20].

$$\sum\nolimits_{m}{\frac{{\in }_{m }-{\in }_{se}}{{\in }_{m }+2{\in }_{se}}v}_{m }=0$$
(5)

From this equation, \({v}_{m}\) and \({(\in }_{m }={n}_{m}^{2}\)) denote volume fraction of the mth component and the dielectric constant, respectively.

The unit cell of β-LG protein layer (restricted in the core area) forms a consistent cubic layer that contains spherical β-LG molecules along with water molecules. The dielectric molecule of this effective layer (\({\in }_{x}\)) can be determined employing superposition principle with the help of Eq. (6) [20].

$${\in }_{x}=\frac{{\in }_{p }{v}_{p}+{\varepsilon }_{{\text{water}}}{v}_{{\text{water}}}}{{v}_{t}}$$
(6)

From this equation, \({\varepsilon }_{{\text{water}}}\) and \({(\in }_{p }={n}_{p}^{2})\) denote water molecule and dielectric constants of β-LG. While \({v}_{p}\), \({v}_{t}\), and \({v}_{{\text{water}}}\) signify the volume employed by spherical β-LG molecule, total volume of the cubic unit cell, and volume employed by water molecules, respectively.

The molar concentration of β-LG per layer can be evaluated employing given Eq. (7) below [20]:

$$\text{Molarity}=\frac{\text{No}.\;\mathrm {of}\;\mathrm {moles}\;\mathrm {of}\;\beta\text{-}\mathrm{LG}}{\mathrm {Volume}\;\mathrm {of}\;\mathrm {solution}}$$
(7)
$${\text{No}}.\;\mathrm{of\;moles\;of}\;\beta\text{-}{\text{LG}}=\frac{\mathrm{Mass\;of}\;\beta\text{-}\mathrm{LG}}{\mathrm{Molar\;mass\;of}\;\beta\text{-}{\text{LG}}}$$
(8)

The molar mass and mass of a β-LG molecule is 18,400 gmol−1 and 1.88 g \(\times {10}^{-20}\), respectively. On supposing the β-LG molecules dilation in 1 ml of water, the molar concentration of β-LG monomer can be calculated as 0.75 fM. The molar concentration for a single layer including dimers will be doubled to 1.5 fM. The density of β-LG protein molecule has been used as 1.33 g/cc in the proposed work.

Alteration of molar concentration M in g/ml:

$${\text{Concentration}} ={\text{molar}}\;{\text{mass}} \times {\text{molar}}\;{\text{concentration}}\;({{\text{gml}}}^{-1})$$
(9)
$$\begin{aligned}1\;\mathrm M\;\mathrm{of}\;\beta\text{-}\mathrm{LG}&=1\;\mathrm{mole}\;\mathrm l^{-1}\;of\;\beta\text{-}\mathrm{LG}\;1\;\mathrm{mole}\;\mathrm l^{-1}\times18400\;\mathrm{gmole}^{-1}\\&=18400\;\mathrm{gl}^{-1}=18.4\;\mathrm{gml}^{-1}\end{aligned}$$

With the help of this Eq. (9), molar concentration can be simply transformed into g ml−1 concentration easily by multiplying the molar mass of the molecule.

Formulation and Theoretical Interpretation

Numerical simulation methods are valuable tools for designing and predicting the optical behavior of biosensors. These methods permit for a quick initial assessment of sensor performance and allow insights into the included optical event, all while lowering optimization and development. In reference to transmitting waves through intricate structures, the finite element method (FEM)-based COMSOL Multiphysics 5.4 is the preferred choice.

A key advantage of FEM is its exceptional adaptability. It permits a precise representation of all geometry and the seamless inclusion of materials with intricate, time-dispersive, and anisotropic characteristics. In comparison to conventional resolution techniques, FEM enables the handling of complex material properties and introduces novel opportunities for analysis of nonlinear properties [25].

To enhance the performance of element quality and performance accuracy, extremely fine mesh element size with electromagnetic wave frequency domain (ewfd) has been used during simulation. The complete mesh consists of 151,616 domain elements and 17,592 boundary elements. The properties of mesh elements are defined here: to ensure better simulation accuracy, the complete mesh area consists of 397 μm2 and average element quality 0.2919. The number of vertex, triangular, and edge elements are 84, 17,432, and 150,416, respectively. To investigate the characteristics of confinement loss for distinct analytes can be formulated using the given Eq. (10) [26]:

$${\alpha }_{{\text{l}}}=8.686\times \frac{2\pi }{\lambda }\times {\text{Img}}\left({n}_{{\text{eff}}}\right)\times {10}^{4} \left(\frac{{\text{dB}}}{{\text{cm}}}\right)$$
(10)

Here, \(\lambda\) expresses the operating wavelength (μm) and \({n}_{{\text{eff}}}\) effective refractive index. It has been noticed that owing to the change in analyte’s RI (1.36 to 1.41), having 0.001 step size shows significant impact on the loss spectrum. For na = 1.36, a smaller loss spectrum,whereas for na = 1.41, higher loss spectrum has been achieved. The higher power switches from core mode to SPP mode when the loss spectra are high. In general, the SPP mode shows high sensitivity when varying the RI of analytes. Sensitivity can be calculated for a particular wavelength with amplitude interrogation technique from the given Eq. (11) [4]:

$${S}_{A}=-\frac{1}{\alpha \left(\lambda ,{n}_{a}\right)}\frac{\delta \left(\lambda ,{n}_{a}\right)}{\delta {n}_{a}}$$
(11)

Here, \(\alpha (\lambda ,{n}_{a})\) shows propagation loss, \(\delta (\lambda ,{n}_{a})\) the loss difference, and na dielectric RI.

To investigate the wavelength sensitivity employing wavelength interrogation method to improve the sensor performance, Eq. (12) can be employed [27]:

$${S}_{\lambda }=\frac{\Delta {\lambda }_{{\text{peak}}}}{\Delta {n}_{a}}$$
(12)

Here, \(\Delta \lambda\) and \(\Delta {n}_{a}\) represent the wavelength change of the peak resonance. The sensor resolution is an important parameter for detecting the small changes in analyte RIs. It shows the identification capability of the offered sensor. The realization of the resolution of the sensor is expressed using Eq. (13) [28]:

$${R}_{S }=\Delta {n}_{a}\times \frac{\Delta {\lambda }_{{\text{min}}}}{\Delta {\lambda }_{{\text{peak}}}} ({\text{RIU}})$$
(13)

Here, \(\Delta {\lambda }_{{\text{min}}}\) and \(\Delta {\lambda }_{{\text{peak}}}\) represent the small spectral resolution and resonant peak shift. Sensor resolution is evaluated with negligible instrumental noise and extra noise generated by external disturbances. Figure of Merit (FoM) is also a crucial parameter. It can be expressed to evaluate the detection limit of the sensor with the help of Eq. (14) [29]:

$${\text{FoM}}=\frac{{\text{Sensitivity}}}{{\text{FWHM}}} \left({{\text{RIU}}}^{-1}\right)$$
(14)

Here, FWHM expresses full width at half maximum. For better detection limit, FoM should be high. It can be noticed that if the sensitivity rises and FWHM lowers, high FoM is achieved from the equation.

Optimization of Sensor

To measure the performance of the sensor, y-polarized fundamental core mode is preferred due to sharp and high loss peak in this propagation direction. The interpretation of the offered sensor can be calculated with optimized different geometrical parameters inclusive of silver (tAg) and TiO2 layer thickness, pitch (p) size, and air hole diameters d1, d2, and d, respectively. To evaluate the impact of the performance change, a specific parameter has been changed while placing the other parameters constant.

Effect of Phase Matching and Distribution of EM Field

Photonic crystal fiber-based surface plasmon resonance works on the principle of guided evanescent field. Appropriately designed geometry of the core and cladding enhances the evanescent field propagation toward the metal surface. The free electrons get excited by the propagated wave that is incident on metallic film. When the wavelength or frequency of the guided core field is equal to the wavelength or frequency of the surface electrons, it generates SPW. The generated SPWs have been notably sensitive to the near RI. So, to identify the variation in RW, amplitude, local RIs, and spectral mode are monitored appropriately. For a specific analyte with a rise of the wavelength, the effective RI of the core mode decreases significantly. In Fig. 2, the characteristics of phase matching of core and SPP mode have been depicted for RI of na = 1.40. Hence, it is noticed that the effective RI of the core (real portion) and SPP mode index is equal and cut at wavelength 0.85 μm. So, 0.85 μm expresses a phase-matching wavelength, and resonance occurs at this position. Maximum power shifts from core mode to SPP mode at the resonance point can be seen as a sharp loss peak. With the rise of analyte RI, the resonance point has shifted red.

Fig. 2
figure 2

a The phase matching effects and bd EM field distributions for na = 1.40

Effect of Silver and TiO2 Thickness Variation

The thickness of the silver layer has a crucial role in the loss and sensitivity as the thickness of the metal layer has a remarkable effect on the RW change. Sensitivity and CL peak depend on metal film thickness. Hence, the effect of change of Ag thickness on the complete sensing performance has been evaluated. With variation in Ag thickness for different analyte refractive index na = 1.39, 1.40, 1.41, confinement loss peak has been investigated, respectively. From Fig. 3a, CL peak expresses a red shift with a rise of tAg varying from 45 to 55 nm. The maximum peak has been observed for thin layer of silver which results in strong coupling. From Fig. 3b it has been observed that for the value of tAg = 55 nm, the AS peak is maximum with the fixed thickness of TiO2 layer tt = 5 nm. So, the optimized thickness of the silver film is 50 nm. The effect of variation in the TiO2 (tt) film thickness on CL spectra for na = 1.39, 1.40, 1.41 is displayed in Fig. 3c. From Fig. 3d, the maximum AS peak is observed on the tt = 5 nm film thickness. So, 5 nm is the optimized thickness of the tt.

Fig. 3
figure 3

a, c Confinement loss spectrum. b, d Amplitude sensitivity with the change in thickness of silver (tAg) and TiO2 (tt) layers

Effect of Pitch Variation

The impact of different pitch size on the loss peak and amplitude sensitivity is shown in Fig. 4. Hence, the loss spectrum reduces with an increase in pitch size because it raises the RI difference between core and cladding. Furthermore, increasing size of pitch results in a blue shift and reduces CL depth. CL spectra has been plotted and observed for different pitch changing from 2.9, 3, 3.1 μm for na = 1.39, 1.40, 1.41, respectively. The maximum CL spectra have been obtained for the pitch value of 2.9 μm as shown in Fig. 4a. At the value of pitch (p) = 3.1 μm, the maximum AS peak has been observed as shown in Fig. 4b. On the contrary, the sensing region has been restricted from 1.33 to 1.41. So, the pitch (p) = 3.0 μm is optimized pitch of the sensor.

Fig. 4
figure 4

a CL spectra. b Amplitude sensitivity with variation of pitch (p) size

Effect of Air Hole Diameter Variation

The maximum AS has been attained for dc = 0.145p because of strong coupling of core and SPP mode. So, dc is optimized center diameter. The effect of regular air hole diameter d on CL spectra is expressed in Fig. 5a. It is clear that with increase of d from 0.5p, 0.6p, and 0.7p, the CL spectra has been increased for 0.6p and reduced for 0.7p, although no change has taken place in RW. From Fig. 5b, the highest AS peak has been observed for na = 1.39, 1.40, and 1.41, respectively. The impact of variation in the air hole diameter d1 is shown in Fig. 5c. It has been noticed that the CL spectra has been decreased, when diameter d1 has been scaled up from 0.7p to 0.8p which results in a redshift and conversely results in scaling up of d1 from 0.8p to 0.9p. To maintain the proper spaces between the external ring and the metal surface, optimized air hole diameters are d = 0.6p and d1 = 0.8p. From Fig. 5d, it has been observed that the highest AS peak occurs with 0.7p, 0.8p, and 0.9p for na = 1.39, 1.40, and 1.41, respectively. Apart from this, the effect of air hole diameter d2 contrast in the CL spectrum has been shown in Fig. 5e. It has been noticed that unchanged scaled-down air holes show inconsistency in the CL spectrum. Increasing the value of d2 = 0.05p, 0.1p, and 0.15p shows a blue shift. Figure 5f shows that for d2 = 0.1p, maximum AS has been achieved. So, d2 = 0. 1p is the optimized diameter.

Fig. 5
figure 5

a, c, and e CL spectra. b, d, and f The spectrum of AS with variation of d, d1, and d2

Sensing Performance with Variation of Different Analyte RIs

The CL and AS of the offered sensor employing optimized model parameters have been illustrated for various na with Eqs. (4) and (6) as shown in Fig. 6. The significant increase of na results in a small change of RI difference between core guided and plasmonic mode. Therefore, most evanescent waves travel through the cladding area rather than the core portion. Consequently, light the coupling improves through the metal surface and effectively interacts with the dielectric medium. So, the CL spectrum increases slowly and shifts toward red. As the changing of analyte refractive index from na = 1.36 to 1.41, CL spectrum shifts toward maximum wavelength and significantly increases the RW peak as shown in Fig. 6.

Fig. 6
figure 6

Performance characteristics of the sensor. a CL spectra w.r.t the wavelength. b Amplitude sensitivity w.r.t the wavelength. c Wavelength sensitivity w.r.t the variation of na

Performance Analysis of the Sensor

After optimization of the sensor performance with the analyte medium refractive index, the proposed sensor has been employed to identify the bio-samples. For this purpose, different concentrations of the β-LG have been considered for the further analysis. The CL and AS can be noticed for the proposed sensor with variation of concentration of β-LG in Fig. 7. The spectral transition in resonance wavelength with variation in β-LG concentration (%) has been illustrated. In this instance, protein molecules have been filled throughout the height of the core area. In this situation, the spectral shift experiences a nearly exponential growth as the concentration of protein molecules rises.

Fig. 7
figure 7

a CL spectra. b Amplitude sensitivity. c Spectral shift in the resonance wavelength for distinct concentration with changing adlayer thickness in the core

Fabrication Feasibility and Applicability

The manufacturing process for the PCF sensor is notably intricate in comparison to traditional optical fibers. Various well-established fabrication methods, such as sol–gel, injection modelling, chemical vapor deposition (CVD), capillary stacking, atomic layer deposition (ALD), stack-and-drilling, stack-and-draw, 3D printing, wheel polishing method (WPM), and extrusion techniques, are widely recognized [30]. A computer-controlled mill employing the drilling technique has been utilized to create air holes in the solid rod and solid tube, drawing them to form inside and outside air holes fibers, respectively [31]. Silver has been coated on the external layer of the microfiber employing the chemical vapor deposition (CVD) technique, while TiO2 has been coated on the inner layer of the fiber. To identify the unspecified sample with the effect of the metal surface a thickness of (0.5p) sensing layer has been positioned on the outer part of the silver layer. A thickness of 0.5p (10% of PCF’s diameter) of a circular PML layer prevents the unapproved reflection of the light and absorbs the approaching evanescent wave from the PCF scattering boundary condition (SBC) which has been carried out.

For instance, an experiment briefly describes an Ag–TiO2-coated nanostructure in [32], changing the size and parameters of the layer. Furthermore, in the realm of fabrication, several surface plasmon resonance-based optical fiber sensors with Au-deposited [33] and MgF2/TiO2-coated optical filters have been experimentally demonstrated [34].

In summary, the proposed sensor’s extended operating wavelength period enhances its appeal for both telecommunications and sensing applications due to its short loss profile, concise, fabrication facility, and minimal back-reflection. For the application purpose, PCF-based SPR sensors find utility in bioimaging, organic chemical sensing, gas sensing, liquid sensing, tuberculosis detection, disease detection, glucose monitoring in urine, pregnancy testing, medical diagnostics, and environmental monitoring [35]. The suggested sensor, with an operating analyte ranging from 1.36 to 1.41, is well-suited for application as a biosensor. Many biological analyte’s refractive indexes fall within this range, including tuberculosis cell sensing ranging from 1.345 to 1.349 [36], pregnancy testing ranging from 1.335 to 1.343 [37], cancer cell detection ranging from 1.36 to 1.38 [38], alcohol sensing ranging from 1.333 to 1.3611 [39], and various blood component sensing (1.33–1.40) [24]. Consequently, the broad span of analytes and high sensing performance position the suggested sensor as a promising profile across numerous SPR-based biosensor application areas. For different range of RI and concentration of the analyte, the performance parameters are investigated in Tables 1 and 2, respectively. The performance parameter comparison in Table 3 has been incorporated with the recently reported articles.

Table 1 Performance parameters of the proposed sensor for various analyte RI
Table 2 Performance parameters of the proposed sensor for various concentration of β-LG
Table 3 Performance comparison of the proposed sensor with existing sensors

Conclusion

The present paper offers a dual-core PCF-based SPR biosensor to identify the different concentration of in β-LG in the solution FEM-based numerical simulation methods which are a valuable tool for designing and predicting the optical behavior of biosensors. These approaches enable a swift prime evaluation of sensor performance and offer insights into the optical phenomena included and also lower the optimization and development expenses. In this regard of transmitting waves through intricate structures, the finite element method is the preferred choice. The comparison in Table 3 shows highest wavelength sensitivity of the proposed sensor with existing sensors. The proposed sensor gives maximum 98.67764 RIU−1AS and 14,090 nm/RIU WS in the analyte sensing RI spectra from 1.36 to 1.41. It offers high FOM of 169.94669 RIU−1. It also offers maximum AS of 99.14327 RIU−1 and WS of 5970.73171 nm/RIU with changing the concentration of β-LG.