Abstract
This study aims to investigate the potential safety hazards and provide reference for improving the medical waste disposal procedure in SARS-CoV-2 testing laboratory. Our SARS-CoV-2 testing group detected the RNA residue on the surface of medical waste with Droplet Digital PCR, and held a meeting to discuss the risks in the laboratory medical waste disposal process. After effective autoclaving, SARS-CoV-2 contaminated on the surface of medical waste bags was killed, but the average concentration of viral RNA residues was still 0.85 copies/cm2. It would not pose a health risk, but might contaminate the laboratory and affect the test results. When the sterilized medical waste bags were transferred directly by the operators without hand disinfection, re-contamination would happen, which might cause the virus to leak out of the laboratory. Furthermore, we found that sterilization effect monitoring and cooperation among operators were also very important. In summary, we investigated and analyzed the potential safety hazards during the medical waste disposal process in SARS-CoV-2 testing laboratory, and provided reasonable suggestions to ensure the safety of medical waste disposal.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
The COVID-19 epidemic has already spread around the world (Ali et al. 2020; Liu et al. 2020). Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) as the causative agent of COVID-19 is highly infectious and lethal (Harrison et al. 2020, Koff and Williams 2020, Phua et al. 2020). The continuous emergence of variants such as D614G, Spike Y839, and 20A.EU1 make it spread faster (Baric 2020, Borges et al. 2020, Hodcroft et al. 2020, Kirby 2021).
Laboratory testing of SARS-CoV-2 is allowed to be performed in the Biosafety Level 2(BSL-2) laboratory, which is generating a large amount of medical waste every day, including discarded samples, reagents, consumables, and personal protective equipment (Saadat et al. 2020). Proper disposal of medical waste is a key to ensure laboratory safety and test quality. According to the laboratory biosafety guidance related to COVID-19 (WHO 2020), waste generated in the BSL-2 laboratory must be autoclaved, then transferred out through the transfer window, and handed over to a professional medical waste recycling department for incineration and destruction. However, the guide does not mention the process of transferring decontaminated waste out of the laboratory (WHO 2020). There may be some potential safety hazards in the waste disposal of the SARS-Cov-2 test laboratory, but few studies have evaluated about them. In this study, we tried to detect the denaturation effect of autoclaving on SARS-Cov-2 RNA by Droplet Digital PCR (ddPCR), investigate the risk factors in the process of medical waste transfer, and provide reasonable suggestions for prevention.
Methods
Sample collection
The Taihe Hospital in Shiyan City, Hubei Province, China, was one of the first hospitals to carry out nucleic acid detection of SARS-CoV-2. The sampling of medical waste was conducted from March 2 to March 6, as shown in Fig. 1. On March 2, 3, and 5, two batches of nucleic acid test were carried out, and each type of samples was collected twice. On March 4 and 6, only one batch of nucleic acid test was conducted, and each type of sample was collected once. The sampling area of each medical waste bag was 100 cm2 including four areas 5 cm × 5 cm. For gloves, the entire surface of fingers was sampled, and the sampling area was estimated. The samples indicated by BG, AG, BB, AB, and AGB were collected from outer gloves of operator before nucleic acid testing, outer gloves of operator after nucleic acid testing, medical waste bags before autoclaving, medical waste bags after autoclaving, and sterilized medical waste bags transferred by testers without hand disinfection after nucleic acid testing, respectively (Fig. 1).
ddPCR detection
After sampling, all the samples were detected immediately in BSL-2 laboratory. Following the manufacturer’s instruction, viral RNA was extracted using a Viral Nucleic Acid Isolation Kit (Bioperfectus, Cat: SDK60102). Then, the ddPCR was used to amplify the specific target genes (ORF1ab and N) of SARS-CoV-2 on the Bio-rad QX200 system. The reaction mixtures and conditions were followed by a previous report (Lv et al. 2020). Finally, the output data was analyzed with the Quanta Soft TM analysis software, and the concentration calculation was calculated by Poisson distribution.
Analysis of risk factors in the medical waste transfer
During the COVID-19 outbreak, multiple groups of operators took turn to carry out nucleic acid detection every day. Meanwhile, the medical waste generated from each batch need to be safely and properly disposed. Based on the actual operating experience, a nucleic acid testing team consisting of 11 people in this hospital held a meeting to discuss the risks in the process of sterilization, transfer, and handover of medical waste.
Results
Denaturation effect of autoclaving on SARS-CoV-2 RNA
The test results of SARS-CoV-2 RNA residues on the surface of medical waste bags before and after autoclaving were shown in Table 1. Before autoclaving, all the surface of medical waste bags were positive for SARS-CoV-2 RNA and the concentration ranged from 16.80 to 37.80 copies/cm2, with an average of 22.84 copies/cm2. After autoclaving at 103.4 mPa, 121.3 °C for 28 min, 5 out of 8 samples were positive, and the average concentration of SARS-CoV-2 RNA was 0.85 copies/cm2.
The autoclave procedure was recommended by the instrument manufacturers. The sterilization effect biological indicator (AttestTM 1262) used in this autoclaving did not change the color after incubating at 56 °C for 48 h, which showed that the sterilization was successful and all microorganisms had been killed. However, the outer surface of some medical waste bags were still positive for SARS-CoV-2 RNA, which indicated that autoclaving was designed to kill the virus, but it could not completely degrade the viral nucleic acid.
Re-contamination of sterilized medical waste
Before the nucleic acid test, 8 samples of tester’s outer gloves were negative for SARS-CoV-2 RNA. In contrast, after the nucleic acid test, all of them had been contaminated by the SARS-CoV-2 RNA, and the average concentration of SARS-CoV-2 RNA was 19.54 copies/cm2. When the medical waste bags were transferred after autoclaving by operators without hand disinfection, the surface of all the sterilized medical waste bags were positive for SARS-CoV-2 RNA again. The concentration ranged from 0.84 to 5.78 copies/cm2, and the average concentration of SARS-CoV-2 RNA increased from 0.85 to 3.36 copies/cm2. It indicated that the sterilized medical waste was re-contaminated by the operator’s gloves, as shown in Fig. 2. Considering that there might be live virus remaining on the surface of the tester’s gloves, it might cause the virus to leak out of the laboratory in the medical waste transfer process.
Risk factors during medical waste handover process
After group discussion, we found that there were several risks in the process of medical waste disposal in SARS-CoV-2 testing laboratory.
-
1.
In the BSL-2 laboratory, non-strict sterilization effect monitoring may increase the risk of virus leakage through medical waste transfer.
-
2.
During the COVID-19 outbreak, multiple groups of testers have to perform nucleic acid test in batches every day to screen large cluster of patients. Poor communication and cooperation among operators in medical waste disposal may be also a risk.
-
3.
In negative pressure laboratory, physical discomfort caused by wearing personal protective equipment and high-intensity labor may increase the error probability in the handover of medical waste. However, the post-processing personnel outside the laboratory usually do not care about the changes in the appearance of the medical waste after autoclaving (Fig. 3). So, mistakes in the transfer of medical waste cannot be discovered in time.
Discussion
Approximately 10% of typical medical waste is infectious (Chartier et al. 2014). Incineration, chemical disinfection, and physical disinfection are commonly used for hospital waste disinfection (Ilyas et al. 2020; Wang et al. 2020). Improper medical waste management increases the potential for COVID-19 spread, especially in developing countries (Nzediegwu and Chang 2020). Safe and effective medical waste management is one of China’s experiences in successfully controlling the COVID-19 epidemic (Ma et al. 2020, Singh et al. 2020). All the medical waste generated in SARS-CoV-2 testing laboratory are considered infectious; this study exclusively focused on the risks of laboratory medical waste disposal.
Autoclaving is the most efficient way to sterilize the medical waste. In this article, the spore population of AttestTM 1262 used for sterilization effect monitoring is 4.0 × 105. After autoclaving, they could not be revived by culture, which indirectly indicates that all viruses contaminated on medical waste had been inactivated. The genome of a virus is DNA or RNA, which is more susceptible to be damaged under hydrated conditions than in dry conditions (Choi et al. 2014). Autoclaving may take 2 h to effectively eliminate nanogram quantities of contaminating nucleic acid (Gefrides et al. 2010). However, the maintenance time of autoclaving at 121 °C is generally about 30 min, which is not enough to completely degrade the nucleic acid of pathogenic microorganisms. Studies have shown that after autoclaving, shorter DNA or RNA fragments produced by incomplete degradation of viral nucleic acid may be recovered by molecular amplification techniques (Choi et al. 2014; Unnithan et al. 2014). Our results also showed that autoclaving could not completely degrade the viral nucleic acid. After autoclaving, the SARS-CoV-2 was killed and did not pose a health risk. However, the viral RNA residue might contaminate the laboratory environment and affect subsequent test results. The real-time quantitative PCR testing of SARS-CoV-2 targeted one or more genes, such as ORF1ab and N (Mathuria and Yadav 2020). If there was a target gene sequence in the nucleic acid residue, a false-positive result would be obtained. Therefore, after sterilization, medical waste should not be accumulated in the laboratory, and must be removed from the laboratory as soon as possible to prevent the residual nucleic acid from contaminating the laboratory environment and affecting the test results.
Disinfection or replacement outer gloves before transferring sterilized medical waste are just a small detail. In this study, if the operator did not comply, sterilized medical waste was re-contaminated. There might be live virus remaining on the surface of the tester’s gloves, which might cause the virus to leak out of the laboratory. Therefore, we recommend that in biosafety laboratory, operators should disinfect or replace outer gloves before transferring sterilized medical waste to prevent re-contamination. Laboratory managers can put a warning sign on the lid of the autoclave to remind the operators to avoid the re-contamination of medical waste as much as possible.
Sterilization effect monitoring is very important for ensuring laboratory biosafety. Especially in developing countries, there is a high proportion of sterilization failure (Panta et al. 2019a). In addition to qualified sterilization facilities and strict compliance with operating procedures, a proper sterilization effect monitoring method is essential. Medical waste can only be transferred out of the laboratory after passing the sterilization effect monitoring. There are biological indicators and chemical indicators can be used to monitor the sterilization effect of autoclaving. Biological indicator monitoring is the safest and most reliable method for measuring the effectiveness of autoclaving (Panta et al. 2019b), but it cannot show the sterilization effect at the end of autoclaving (Garibaldi et al. 2017). The conventional biological indicators AttestTM 1262, Proof Plus, Assert, and Biosign usually need to be incubated at 56 °C for 48 h to get the monitoring result (Skaug and Berube 1983). The rapid Readout biological indicator AttestTM 1292 is equivalent to conventional biological indicators and needs to be placed in a fluorimetric auto-reader for 3 h to detect fluorescence (Rutala et al. 1996). Chemical indicators mainly monitor the process of autoclaving through the color changes (Jabbari et al. 2012). Six classes of chemical indicators have been defined by the ANSI/AAMI/ISO 11140-1:2005 Standard (Puttaiah et al. 2014). Class 1 and class 2 chemical indicators only indicate whether the waste in the autoclave has gone through a sterilization cycle by color changes, but cannot assess the effectiveness of autoclaving. Class 3 and class 4 chemical indicators only measure one and two parametric variable (such as temperature, pressure, and maintenance time) in autoclaving process. Class 5 and class 6 chemical indicators are integrators and theoretically expected to be equivalent to biological indicators in terms of assessing effectiveness of autoclaving.
The best sterilization effect monitoring program is to use biological indicators for regular and periodic monitoring of autoclaves, and use chemical indicators for monitoring each sterilization cycle (Puttaiah et al. 2014). However, the quality and quantity of sterilization effect indicators, using regulations and enforcement, vary in different countries, which lead to different levels of sterilization effect monitoring failure. SARS-Cov-2 is too contagious; any negligence in the process of medical waste disposal in the testing laboratory can cause virus leakage or even personnel infection. Therefore, we recommend that the SARS-CoV-2 testing laboratory develop a strict sterilization effect monitoring program, use qualified biological indicators and chemical indicators, and implement them strongly. The autoclaves should be monitored with biological indicators at least once a week, or even daily when conditions permit. For each autoclaving cycle, a class 1 chemical indicator (autoclave indicator tape) and a class 5 chemical indicator can be used simultaneously. Medical waste will be moved from autoclaves to transfer window only when class 5 chemical indicator showed “ACCEPT.” Autoclave indicator tapes, pasted on the surface of medical waste bags, can help subsequent processing personnel outside the BSL-2 laboratory to determine that the medical waste has gone through autoclaving cycle.
During the COVID-19 outbreak, multiple groups of testers have to perform nucleic acid test in batches every day to screen large clusters of patients. The latter batch of testers often needs to help the former batch of testers transfer the autoclaved medical waste, or even complete the autoclaving. Personal protective equipment can protect the operators from infection (Cook 2020), but the air permeability is poor. Especially in the BSL-2 laboratory with negative pressure, operators often feel stuffy and breathless, hoping to complete the test and leave the laboratory as soon as possible, which will increase the risk of errors in the medical waste disposal process. To reduce the risk, we recommend that the medical waste transfer process in the BSL-2 laboratory operation guide needs to be improved, so that the operators can cooperate closely in medical waste treatment. In addition, the post-processing personnel outside the BSL-2 laboratory should carefully observe the appearance of medical waste before handing over and receiving. Once unsterilized medical waste is found, the transfer should be stopped immediately.
This article also has some limitations. During the COVID-19 outbreak in this city, all our energies were focused on carrying out nucleic acid testing to screen new confirmed cases from fever patients, suspicious cases, and close contacts. Since March 1, no new confirmed cases had been found, and we mainly conducted discharge testing of confirmed cases and environmental monitoring in public places. Then, we took the time to investigate the potential safety hazards during medical waste disposal in SARS-CoV-2 testing laboratory. The sampling time range and the sample size were small, and no statistical reasoning had been provided. Furthermore, in some developing countries with poor economic conditions, there is a lack of standard BSL-2 laboratories, qualified autoclave facilities, even personal protective equipment in primary hospitals. For them, the SARS-CoV-2 testing laboratory faces more risks in the medical waste treatment process, and further research is needed.
Conclusions
Proper disposal of medical waste is a key to the laboratory safety. In this study, we found that there were some potential safety hazards during medical waste disposal in the SARS-CoV-2 testing laboratory. Autoclaving cannot completely degrade the viral genome, so sterilized medical waste should be transfer out of the laboratory as soon as possible to prevent nucleic acid contamination. Operators must disinfect or replace outer gloves before transferring sterilized medical waste to prevent re-contamination. The SARS-CoV-2 testing laboratory need to develop a strict sterilization effect monitoring program, use qualified biological indicators and chemical indicators, and implement them strongly. In the process of medical waste transfer and handover, close cooperation between operators and careful observation of follow-up processing personnel outside the laboratory can reduce the risk of virus leakage due to mistakes.
The above findings and recommendations can alert the health authorities to pay more attention to laboratory biosafety management, especially to improve the laboratory medical waste treatment process. Eliminating potential safety hazards is able to protect the internal and external environment of the laboratory from contamination, which will ensure the safety and health of laboratory operators and surrounding personnel.
Data availability
Not applicable
References
Ali SA, Baloch M, Ahmed N, Ali AA, Iqbal A (2020) The outbreak of Coronavirus Disease 2019 (COVID-19)—an emerging global health threat. J Infect Public Health 13:644–646. https://doi.org/10.1016/j.jiph.2020.02.033
Baric RS (2020) Emergence of a highly fit SARS-CoV-2 variant. N Engl J Med 383:2684–2686. https://doi.org/10.1056/NEJMcibr2032888
Borges V, Isidro J, Cortes-Martins H, Duarte S, Gomes JP (2020) Massive dissemination of a SARS-CoV-2 Spike Y839 variant in Portugal. Emerg Microbes Infect 9:2488–2496. https://doi.org/10.1080/22221751.2020.1844552
Chartier Y, Emmanuel J, Pieper U, Pruss A, Rushbrook P, Stringer R, Townend W, Wilburn S, Zghondi R (2014) Safe management of wastes from healthcare activities. WHO. https://www.who.int/water_sanitation_health/publications/wastemanag/en/. Accessed June 2014
Choi WS, Rodríguez RA, Sobsey MD (2014) Persistence of viral genomes after autoclaving. J Virol Methods 198:37–40. https://doi.org/10.1016/j.jviromet.2013.12.021
Cook T (2020) Personal protective equipment during the coronavirus disease (COVID) 2019 pandemic–a narrative review. Anaesthesia 75:920–927. https://doi.org/10.1111/anae.15071
Garibaldi BT, Reimers M, Ernst N, Bova G, Nowakowski E, Bukowski J, Ellis BC, Smith C, Sauer L, Dionne K (2017) Validation of autoclave protocols for successful decontamination of category a medical waste generated from care of patients with serious communicable diseases. J Clin Microbiol 55:545–551. https://doi.org/10.1128/JCM.02161-16
Gefrides LA, Powell MC, Donley MA, Kahn R (2010) UV irradiation and autoclave treatment for elimination of contaminating DNA from laboratory consumables. Forensic Sci Int Genet 4:89–94. https://doi.org/10.1016/j.fsigen.2009.06.008
Harrison AG, Lin T, Wang P (2020) Mechanisms of SARS-CoV-2 transmission and pathogenesis. Trends Immunol 41:1100–1115. https://doi.org/10.1016/j.it.2020.10.004
Hodcroft EB, Zuber M, Nadeau S, Comas I, Candelas FG, Stadler T, Neher RA (2020) Emergence and spread of a SARS-CoV-2 variant through Europe in the summer of 2020. MedRxiv. https://doi.org/10.1101/2020.10.25.20219063
Ilyas S, Srivastava RR, Kim H (2020) Disinfection technology and strategies for COVID-19 hospital and bio-medical waste management. Sci Total Environ 749:141652. https://doi.org/10.1016/j.scitotenv.2020.141652
Jabbari H, Alikhah H, Alamdari NS, Behzad MN, Mehrabi E, Borzui L, Bakhshian F (2012) Developing the use of quality indicators in sterilization practices. Iran J Public Health 41:64–69 http://dspace.tbzmed.ac.ir:8080/xmlui/handle/123456789/49684
Kirby T (2021) New variant of SARS-CoV-2 in UK causes surge of COVID-19. Lancet Respir Med 9(2):e20–e21. https://doi.org/10.1016/S2213-2600(21)00005-9
Koff WC, Williams MA (2020) Covid-19 and immunity in aging populations—a new research agenda. N Engl J Med 383:804–805. https://doi.org/10.1056/nejmp2006761
Liu Y, Gayle AA, Wilder-Smith A, Rocklöv J (2020) The reproductive number of COVID-19 is higher compared to SARS coronavirus. J Travel Med 27:taaa021. https://doi.org/10.1093/jtm/taaa021
Lv J, Yang J, Xue J, Zhu P, Liu L, Li S (2020) Detection of SARS-CoV-2 RNA residue on object surfaces in nucleic acid testing laboratory using droplet digital PCR. Sci Total Environ 742:140370. https://doi.org/10.1016/j.scitotenv.2020.140370
Ma Y, Lin X, Wu A, Huang Q, Li X, Yan J (2020) Suggested guidelines for emergency treatment of medical waste during COVID-19: Chinese experience. Waste Dispos Sustain Energy 1–4. https://doi.org/10.1007/s42768-020-00039-8
Mathuria JP, Yadav R (2020) Laboratory diagnosis of SARS-CoV-2-A review of current methods. J Infect Public Health 13:901–905. https://doi.org/10.1016/j.jiph.2020.06.005
Nzediegwu C, Chang SX (2020) Improper solid waste management increases potential for COVID-19 spread in developing countries. Resour Conserv Recycl 161:104947. https://doi.org/10.1016/j.resconrec.2020.104947
Panta G, Richardson AK, Shaw IC (2019a) Effectiveness of autoclaving in sterilizing reusable medical devices in healthcare facilities. J Infect Dev Ctries 13:858–864. https://doi.org/10.3855/jidc.11433
Panta G, Richardson AK, Shaw IC, Chambers S, Coope PA (2019b) Effectiveness of steam sterilization of reusable medical devices in primary and secondary care public hospitals in Nepal and factors associated with ineffective sterilization: A nation-wide cross-sectional study. PLoS One 14:e0225595. https://doi.org/10.1371/journal.pone.0225595
Phua J, Weng L, Ling L, Egi M, Lim C-M, Divatia JV, Shrestha BR, Arabi YM, Ng J, Gomersall CD (2020) Intensive care management of coronavirus disease 2019 (COVID-19): challenges and recommendations. Lancet Respir Med 8:506–517. https://doi.org/10.1016/S2213-2600(20)30161-2
Puttaiah R, Griggs J, D’Onofrio M (2014) A preliminary evaluation of reusable digital sterilization indicator prototype. J Contemp Dent Pract 15:626–635. https://doi.org/10.5005/jp-journals-10024-1591
Rutala WA, Jones SM, Weber DJ (1996) Comparison of a rapid readout biological indicator for steam sterilization with four conventional biological indicators and five chemical indicators. Infect Control Hosp Epidemiol 17:423–428. https://doi.org/10.1086/647333
Saadat S, Rawtani D, Hussain CM (2020) Environmental perspective of COVID-19. Sci Total Environ 728:138870. https://doi.org/10.1016/j.scitotenv.2020.138870
Singh N, Tang Y, Zhang Z, Zheng C (2020) COVID-19 waste management: effective and successful measures in Wuhan, China. Resour Conserv Recycl 163:105071. https://doi.org/10.1016/j.resconrec.2020.105071
Skaug N, Berube R (1983) Comparative thermoresistance of two biological indicators for monitoring steam autoclaves. 3. Comparison performed at 121 degrees C in a hospital prevacuum steam sterilizer. Acta Pathol Microbiol Immunol Scand B 94:107–112. https://doi.org/10.1111/j.1699-0463.1986.tb03028.x
Unnithan VV, Unc A, Joe V, Smith GB (2014) Short RNA indicator sequences are not completely degraded by autoclaving. Sci Rep 12:4070. https://doi.org/10.1038/srep04070
Wang J, Shen J, Ye D, Yan X, Zhang Y, Yang W, Li X, Wang J, Zhang L, Pan L (2020) Disinfection technology of hospital wastes and wastewater: suggestions for disinfection strategy during coronavirus Disease 2019 (COVID-19) pandemic in China. Environ Pollut 262:114665. https://doi.org/10.1016/j.envpol.2020.114665
WHO ( 2020) Laboratory biosafety guidance related to coronavirus disease ( COVID-19) : interim guidance. World Health Organization. https://apps.who.int/iris/handle/10665/332076. Accessed 13 May 2020
Acknowledgements
We thank all members of the SARS-CoV-2 testing team in Taihe Hospital for their solidarity and cooperation.
Funding
This research was funded by the Fundamental Research Funds for the Central Universities 2662017PY011, 2662018PY028, 2662019YJ014, talent funding RCQD002 from Taihe hospital, and Foundation of Health and Family Planning Commission of Hubei Province (WJ2018H260).
Author information
Authors and Affiliations
Contributions
Conceptualization, LJ and LS; methodology, LJ and YJ; investigation, LJ, ZP, XJ, and LLF; resources, YJ and LLF; data curation, LJ and YJ; funding acquisition, LS; writing - original draft preparation, LJ, YJ, XJ, and ZP; and writing - review and editing, LJ and LS. All authors have read and agreed to the published version of the manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
Not applicable
Consent to publication
Not applicable
Competing interests
The authors declare no competing interests.
Additional information
Responsible Editor: Lotfi Aleya
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Lv, J., Yang, J., Xue, J. et al. Investigation of potential safety hazards during medical waste disposal in SARS-CoV-2 testing laboratory. Environ Sci Pollut Res 28, 35822–35829 (2021). https://doi.org/10.1007/s11356-021-13247-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11356-021-13247-4