Abstract
In this study, an innovative analytical methodology capable of selectively identifying and quantifying mercury contamination by the association of solid-phase extraction using ion-imprinted polymers as a sorbent phase and differential pulse anodic stripping voltammetry is proposed. To this end, the ion-imprinted polymers were synthesized and characterized by infrared spectroscopy and atomic force microscopy. The sorption capacities and the selectivity of the ion-imprinted polymers were compared to the ones related to the non-imprinted ones. Next, the experimental parameters of this solid-phase extraction method (IIP-SPE) were evaluated univariately. The selectivity of this polymeric matrix against other cations (Cd II, Pb II, and Cu II) was also evaluated. Limits of detection (LOD) and quantification (LOQ) obtained for the here proposed methodology were 0.322 μg L−1 and 1.08 μg L−1, respectively. Also, the precision of 4.0% was achieved. The method was finally applied to three water samples from different sources: for the Piratininga and Itaipu Lagoon waters, Hg II concentrations were below the LOQ and for Vargem River waters a concentration equal to 1.35 ± 0.07 mg L−1 was determined. These results were confirmed by recovery tests, resulting in a recovery of 96.2 ± 4.0%, and by comparison with flame atomic absorption spectrometry, resulting in statistical conformity between the two methods at 95% confidence level.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Due to population growth and the development of the industrial sector, the supply and availability of potable water have become a global concern. Increased industrial activities lead to the production of contaminated effluents, rich for instance in toxic metals that are placed directly into the environment. In that sense, the toxicity, persistence, and high bioaccumulation potential of metals are the most dangerous points to deal with. Cadmium, lead, arsenic, and mercury are the metals that stand out based on their toxicity levels. Mercury is considered the most dangerous among them by the US Environmental Protection Agency (EPA - United States Environmental Protection 1985). It is a natural element that can be found in the four subsystems of planet Earth (geosphere, hydrosphere, biosphere, and atmosphere) in several possible forms. The organic forms, such as methyl mercury and ethyl mercury, and the elemental form are neurotoxic, and the inorganic forms, such as mercury chloride, are mainly nephrotoxic (Oliveira et al. 2007; UNEP, 2013).
Many methodologies have been developed for the determination of mercury in aqueous matrices. The techniques employed are the most diverse such as UV-Vis molecular absorption (Babu and Reddy 2012), molecular fluorimetry (Liang et al. 2018), atomic fluorescence spectrometry (AFS) (Schlathauer et al. 2017), inductively coupled plasma optical emission spectrometry (ICP-OES) (Wuilloud et al. 2002), inductively coupled plasma mass spectrometry (ICP-MS) (Allibone et al. 1999), atomic absorption spectrometry (Sobhi et al. 2017), or voltammetry (Martín-Yerga et al. 2013). However, the instrumentation necessary to many of these methodologies is expensive, making them inaccessible to some research groups. Also, for most of these techniques, except for ICP-MS, the quantification of mercury species in effluents is strongly affected by concomitants present in the matrix; thus, it is very common to make a previous step of sample preparation in order to promote the elimination of these interferences (Ghasemlou et al. 2013).
Solid-phase extraction (SPE) is a well-known method for isolating analytes from complex matrices. That approach makes use of the different interactions of a solid phase inside of a cartridge and the components of a liquid sample, allowing the separation of these components. It may offer the possibility of complex samples simplification, but many times the disadvantages of low selectivity and high cost of commercial cartridges are remarkable (He et al. 2007).
Obtaining new materials that might offer greater selectivity to the SPE sorbent phases when compared to the conventional commercial ones is a desired aspect in the development of new methodologies. In this context, many works in the field of molecularly and ion-imprinted polymers (MIP and IIP) have been developed by employing these ones as SPE sorbent phases (MIP-SPE/IIP-SPE), aiming the separation of interferents by the mechanism of analyte recognition (He et al. 2007; Moussa et al. 2017; Yusoff et al. 2017; Khoddami and Shemirani 2016).
Molecularly imprinted polymers (MIP) are synthesized in the presence of the template molecule, so that the growth of the polymer occurs around the substance of interest (Mazzotta et al. 2016). After polymerization, the template molecule is leached, leaving the binding sites available and resulting in a molecular memory by the polymer. This recognition mechanism is based on the selective mechanisms present in living beings such as enzyme-substrate and/or antigen-antibody (Han et al. 2016). When this model molecule is replaced by an ion, ion-imprinted polymers (IIP) are obtained (Shakerian et al. 2016; Luo et al. 2015).
These materials called IIP have attracted much attention as a highly selective sorbent for SPE in order to pre-concentrate and clean-up samples prior to analyses. IIPs show outstanding characteristics such as high selectivity, low cost, high surface area, durability, and reusability (Rajabi et al. 2017; Shamsipur et al. 2014). A variety of IIP have been reported for separation of copper (Rajabi et al. 2017), selenium (Khajeh et al. 2007), uranyl (Shamsipur et al. 2007), mercury (Ghanei-Motlagh et al., 2016; Alizadeh et al. 2011; Rajabi et al. 2013), cesium (Shamsipur and Rajabi 2013), zinc (Shamsipur et al. 2014), nickel (Vatanpour et al. 2011), and many other ions.
Three steps are involved in the ion-imprinting process: (i) complexation of templates (metal ions) to a suitable ligand, (ii) polymerization of this complex, and (iii) removal of the template after polymerization. After ion-imprinting polymerization, the imprinted metal ion is removed from the polymeric material by leaching with a mineral acid which generates available binding sites. These polymeric imprinted particles are complementary in shape and size to the metal ion and, for that reason, show an affinity for the template ion over other ones structurally related (Rajabi et al. 2017; Shamsipur et al. 2014).
In order to combine the sensitivity of the voltammetric instrumental technique and the selectivity of the IIPs as solid phases for IIP-SPE, a new analytical methodology capable of determining mercury in environmental water samples (river and lagoons waters) was proposed here. At first, an Hg+2-IIP was synthesized and characterized, then the IIP-SPE procedure was optimized and quantifications were performed by differential pulse anodic stripping voltammetry using gold electrode.
Table 1 shows a comparison of different methodologies for mercury determination in environmental water samples, some of them using IIP-SPE as sample preparation step and others that employ different strategies.
The results presented in this work show that the here developed methodology offers limits of detection (LOD) and quantification (LOQ) and recoveries as good as or better than others suggested in literature, except for AFS, ICP-OES, and ICP-MS. Though, it should be highlighted that this work proposes a more affordable alternative, one that can be accessible to many more research groups. Also, the results here exposed confirm that this present method offers better limits of detection and quantification than other works that synthesized Hg+2-IIP.
Experimental procedures
Reagents, solutions, and instrumentation
4-Vinylpyridine (4-VP), benzoyl peroxide (BPO), and ethylene glycol dimethacrylate (EGDMA) were purchased from Sigma-Aldrich (USA) and mercury II chloride from Qhemis (Brazil). All reagents were of the analytical grade used without further purification. Ultrapure water was obtained by Arium Pro Sartorius (Germany) purification system.
1.25 mL of a 1000 mg L−1 Hg II standard solution was transferred to a 250.0 mL volumetric flask and the final volume was completed to the mark with deionized water, resulting in a 5.00 mg L−1 intermediate solution. Working solutions were prepared daily by diluting the intermediate solution.
A 4.00 × 10−2 mol L−1 Britton Robinson buffer solution was prepared by dissolving 0.248 g boric acid, 1.433 g sodium phosphate, and 0.328 g sodium acetate in about 50.0 mL of deionized water, adjusting the pH with NaOH aqueous solution and completing the volume to the mark of 100.0 mL in a volumetric flask with deionized water.
The IR spectra were measured by a Spectrum Two FTIR Spectrometer (PerkinElmer, USA). The morphological structure and surface roughness of the polymers were investigated by a NanoSurf Atomic Force Microscope (Switzerland).
Hg II was detected by differential pulse anodic stripping voltammetry, using Metrohm 747 VA Stand (Switzerland) and a gold electrode as a working electrode, stainless steel as counter electrode and Ag/AgCl as a reference electrode. The supporting electrolyte solutions were prepared daily by dissolving 0.147 g of sodium chloride (VETEC, Brazil) and 2.70 mL of sulfuric acid 95% (Merck KGaA, Germany) in deionized water to the final volume of 200.0 mL in a volumetric flask.
The Ashcroft (Brazil) SPE manifold to perform IIP-SPE was coupled to a TBV5 Tander (Brazil) vacuum pump. The comparative AAS method for Hg II quantification was performed by using AA-6200 Flame Atomic Absorption Spectrometer (Shimadzu, Japan).
Electrochemical measurements
In order to improve the electrochemical signal of measurements, a conditioning procedure was performed daily on the gold electrode. It consisted of six voltammetric cycles from 0.100 to 0.550 V in NH4OH 2.5% m v−1 solution (METROHM 2018).
The electrochemical measurements were realized by transferring 25.00 mL of the supporting electrolyte to a 50.00 mL borosilicate cell, then adding 100 μL of the sample solution, followed by successive additions of 10 μL of a standard solution of Hg II.
Differential pulse anodic stripping voltammetry (DPASV) was performed in the range from 0.200 to 0.750 V, using 5 s for resting time, 0.005 V for potential step, 0.200 V for time step, 0.050 V for pulse amplitude, 0.300 V for deposition potential, and 60 s for deposition time. Each analytical measurement was taken in three replicates. It is worth to mention that square wave anodic stripping (SWASV) voltammetry was also studied. Though SWASV is a faster technique, it showed less peak resolution and more noise; thus, the experiments were conducted in DPASV.
Hg+2-IIP synthesis
The imprinted material synthesis was an adaptation of the reaction described by Mergola and coworkers (Mergola et al. 2015), in which the concept of bulk polymerization is employed. 0.050 mmol of HgCl2 and 0.20 mmol of 4-VP were initially dissolved in 3.00 mL of a mixture containing acetonitrile and water (4:1, v/v−1). That solution was table stirred for 10 min to guarantee complete complexation.
Thereafter, 0.90 mmol EGDMA and 0.040 mmol benzoyl peroxide were added. After fully dispersing and deoxygenating the mixture by nitrogen gas purge, it was heated to 60 °C in a water bath for 24 h under stirring. The resulting polymers were crushed, ground, and sieved with a 0.20-mm sieve. Subsequently, the synthesized polymers were treated with 20.0 mL EDTA 0.100 mol L−1 for 2 h under stirring, followed by washing steps to remove the Hg II ion, firstly with 10.0 mL ethanol, then with 15.0 mL HNO3 (10% v/v−1) and finally with water enough to adjust the pH to neutral condition. The Hg2+-IIP was finally dried at 60 °C in a laboratorial oven. Non-imprinted polymers (NIP) were synthesized under the same conditions, including the washing steps, although not containing the ion template, in order to later evaluate the sorption characteristics of the imprinted material.
IIP-SPE
Forty milligrams of the polymer were transferred to empty commercial cartridges. The solid-phase column was conditioned with 5.00 mL of 4.00 × 10−2 mol L−1 Britton Robinson buffer at pH 6.0. Then, 20.0 mL of sample or standard solution, buffered at the same concentration and pH were loaded. The cartridge was later washed with 5.00 mL of deionized water, and finally, the adsorbed analyte was eluted with 10.0 mL HNO3 25% v/v−1.
These parameters were studied and optimized univariately. All extracts were analyzed in three replicates by differential pulse anodic stripping voltammetry.
Sample preparation
About 500 mL of water samples from the Vargem River, the Itaipu Lagoon, and the Piratininga Lagoon, all located in the State of Rio de Janeiro, Brazil, were superficially sampled in a sterilized glass bottle and stored under refrigeration at 5 °C. Before the IIP-SPE procedure, the samples were subjected to a sequence of filtration steps, using decreasing porosities of filters to eliminate the particles that were suspended in the samples. After this treatment, the samples’ pH was adjusted to pH 4.0 with a small volume of HCl 6.0 mol L−1 (Ghanei-Motlagh et al., 2016; Alizadeh et al. 2011; Rajabi et al. 2013).
Results and discussion
Hg+2-IIP preparation
As shown in Fig. 1, the IIP active sites were formed by the reaction between 4-VP and Hg II. This bifunctional ligand is capable of coordinating to the analyte by forming a stable complex and further has a polymerizable vinyl group which, through the action of the binding reagent (EDGMA) and the radical initiator (BPO), forms the polymer matrix (Shakerian et al. 2016).
For the complete template ion removal, three washing protocols were studied. First, 1.00 g of the synthesized IIP was put in contact with 20.0 mL of EDTA 0.100 mol L−1 for 2 h in a polypropylene tube and agitated on a horizontal stirring table, followed by washing and filtrating with 100.0 mL of deionized water. The second method was the same as the first in terms of EDTA complexation, though the washing step was performed with 100.0 mL of HNO3 10% v/v−1. The third method was identical to the first one, except for the contact time with EDTA, changed for 72 h.
To evaluate the efficacy of template ion removal by each of the three methodologies, aliquots of 5.00 mL of the washing solutions were collected after the filtration process. The analyte concentration in these aliquots was determined by anodic stripping voltammetry.
Both second and third procedures successfully corresponded to the same result of totally leaching the analyte. The one based on EDTA together with HNO3 was chosen by considering the best throughput for the entire method.
EDTA is a chelating agent widely used to complex with metal ions. Because of the presence of acetate and amino groups on its structure, the best EDTA performance is generally in basic pH. However, the Hg II ions present the tendency to precipitate in hydroxide form; therefore, the best solution, in this case, is to use a strongly acid medium (Skoog et al. 2006). This assesses the mercury ion removal efficiency performed by the second procedure.
Hg+2-IIPCharacterization
The IR spectra of NIP, Hg+2-IIP, and IIP (after Hg II extraction) are shown in Fig. 2. When analyzing the spectra corresponding to NIP and Hg+2-IIP, it is possible to observe great similarity, as expected, since the same reagents were used in both syntheses, being distinguished only by the template ion. Around 2990 cm−1 region, there is a doublet in Hg+2-IIP spectrum, that is assigned to quaternized nitrogen atoms belonging to 4-VP. In this same spectral region, NIP shows a single band that can be attributed to the axial deformation of the N-H bond of aromatic amines. These observations demonstrate the formation of the complex with mercury ion (Valle et al. 2011). When comparing the IR spectra of the washed and complexed IIP, it can be noticed that the doublet absent for the first one, attesting that the washing process is efficient.
Atomic force microscopy (AFM) was used to analyze morphological differences between the Hg+2-IIP and NIP particles. The images by AFM are shown in Fig. 3. On the one hand, the Hg+2-IIP analysis (a) made it possible to observe a more uniform distribution of particles and a higher average roughness (Rms = 51,45 nm), which is of extreme importance in sorption studies, since it facilitates mass transfer from a liquid phase, containing the analyte, to the synthesized solid phase (Herrmann et al. 1997). It was also possible to conclude that the Hg+2-IIP particles are products of smaller particles aggregation, thus conferring on them greater porosity. On the other hand, NIP (b) showed larger particles, lower average roughness (Rms = 33.43 nm) and a heterogeneous surface. This behavior suggests advantages of inserting the printing practice in the synthesis, to promote the formation of selective cavities on the solid material (Huang et al. 2019; Liu et al. 2018; Liu et al. 2016).
Optimization of IIP-SPE
Study of loading pH
The effect of loading pH over the percentage of Hg II removal using IIP as sorbent phase was evaluated in the range from 3.0 to 8.0. The experiments were conducted by conditioning the cartridges with 5.00 mL of 4.00 × 10−2 mol L−1 Britton Robinson buffer at the same pH that 10.0 mL of 15.0 mg L−1 Hg II solution prepared in this buffer were loaded. Forty milligrams of IIP were employed as solid phase, using 5.00 mL of deionized water at the washing step and 10.0 mL of HNO3 25% v/v−1 for elution.
This study was restricted to the pH range from 3.0 to 8.0, since the polymer is unstable in extremely acid environments and in very basic medium mercury ions tend to form hydroxides (Mergola et al. 2015).
It is possible to observe in Fig. 4 that the analytical signal increases with increasing pH to 6.0, after decreasing in more alkaline medium. The IIP sorption process depends both of electrostatic interactions between the species of opposite charges available in the systems and of the availability for entrapment in the cavities (Rao et al. 2004). Since 4-VP presents pKa = 5.2, below pH 5.2, the nitrogen atoms present in the major portion of these molecules are protonated, that is, unavailable for extracting Hg II ions. This fact justifies the lower recovery of mercury ions when the loading medium is strongly acid (Batlokwa et al. 2012).
At pH 5.5, 4-VP pKa region, about 50% of the total amine groups remain protonated and theoretically unavailable to form complexes with mercury ions, in agreement with the experimental result shown in Fig. 4.
As expected, it was experimentally observed that the increase of loading pH reduced the percentage of protonated 4-VP nitrogen atoms, then increasing the interaction between them and Hg II ions and consequently the recovery efficiency. At loading pH 6.0 and 6.5, the recovery percentages were between 85 and 97%.
In more alkaline conditions, however, the recovery percentages tend to lower, since mercury hydroxides are formed, decreasing the amount of free Hg II ions to be adsorbed on the polymer cavities (Batlokwa et al. 2012).
Analyzing this study, higher percentages of analyte retention were achieved at pH 6.0, so that this was the sample loading pH employed in further experiments.
Study of IIP and NIP sorption capacities
The maximum sorption capacity for IIP and NIP cartridges were calculated by building breakthrough curves, shown in Fig. 5. This study consisted in monitoring the analyte concentration after passing through the polymeric solid phase, and then plotting the relation between the sorbed analyte mass per one gram of solid material as a function of the volume of analyte solution loaded. Naturally, the effects of analyte dispersion and mass transfer until the columns achieve total saturation have to be considered (Knox et al. 2016; El-Sayed et al. 2016).
As shown in Fig. 5, the sorption capacities of both polymers initially increased when increasing the volume of loaded analyte solution (Hg II 15 mg L−1 buffered at pH 6.0). After loading more than 100.0 mL of the analyte solution, NIP reached exhaustion, so that its sorption capacity remained constant in 19.8 ± 0.2 mg g−1. IIP reached exhaustion after loading a greater volume of the same solution (more than 400 mL); thus, its sorption capacity was calculated as 54.9 ± 0.6 mg g−1, being greater than the NIP sorption capacity.
These differences in behavior between IIP and NIP suggest that the complexation between Hg II ions and 4-VP has played an important role in terms of sorption capacity of the analyte. The IIP morphological characteristics significantly favor this parameter, since there are more active sites available to interact with the analyte.
Study of the elution conditions
Choosing the appropriate eluent for the IIP-SPE process was considered fundamental for achieving satisfactory recovery percentages of the analyte. Thus, 10.0 mL of different solutions were evaluated: ethanol, EDTA 0.100 mol L−1, the supporting electrolyte (H2SO4 0.270 mol L−1, NaCl 0.0126 mol L−1) and HNO3 10% v/v−1. Table 2 shows the analyte recovery obtained when each eluent was employed.
The best results were obtained in the strongly acid medium, in agreement to the observations discussed in item 3.1. Even though the supporting electrolyte is also composed by sulfuric acid, this solution presents a higher pH than the HNO3 one, so that the recoveries are not satisfactory.
In order to improve the recovery percentages, different concentrations of HNO3 (10, 25 and 50% v/v−1) were studied and these results are shown in Fig. 6. The same volume (10.0 mL) of eluent was loaded in all cases.
It was expected that increasing HNO3 concentration the analyte recoveries would also increase, since the pH of eluent would be lower. However, in the case of 50% v/v−1 HNO3 solution, this was not observed. This behavior can be attributed to the fact that the polymeric matrix is unstable in such acidic environments, and part of it may be leached or oxidized by the eluent. That decreases the analyte recoveries and interferes in the voltammetric quantification, contributing to the instability of the analytical signal, resulting in a greater standard deviation in measurements. In regard to that study, HNO3 25% v/v−1 was used as eluent in further experiments.
To evaluate the pre-concentration capacity of the IIP-SPE column, the elution step was performed by adding different volumes of HNO3 25% v/v−1. The results of this study are shown in Fig. 7.
There was no gain in terms of analyte recovery when volumes greater than 10.0 mL were used. In order to achieve a higher pre-concentration factor for the methodology, and, as a consequence, lower limits of detection and quantification, 10.0 mL HNO3 25% v/v−1was employed as eluent in the IIP-SPE protocol.
When analyzing samples, 20.0 mL of them were loaded in the cartridges and submitted to the optimized IIP-SPE protocol, assigning to the method a pre-concentration factor of two times.
Validation of methodology
After optimizing the IIP-SPE steps, in order to ensure the reliability of the method, several parameters were evaluated as follows.
Linear range, typical analytical curve, and coefficient of determination
The typical analytical curve for the voltammetric method was Y = 1.969 × 10−9X + 8788 × 10−9, in the linear range from 0.400 to 25.3 μg L−1. The coefficient of determination was 0.9977.
Repeatability
The repeatability of the method was evaluated by submitting ten different Hg II 1.00 mg L−1 standard solutions to the optimized IIP-SPE protocol and determining the final concentration by differential pulse anodic stripping voltammetry. The coefficient of variation of these measurements was 4.0%, value employed to represent the repeatability of the proposed method.
Limits of detection and quantification (LOD and LOQ)
The limits of detection and quantification for the proposed method were calculated based on \( \raisebox{1ex}{$3\delta $}\!\left/ \!\raisebox{-1ex}{$S$}\right. \) and \( \raisebox{1ex}{$10\delta $}\!\left/ \!\raisebox{-1ex}{$S$}\right. \) criterion, respectively. Instrumental LOD was 0.644 μgL−1 and LOQ 2.15 μgL−1. These parameters were recalculated for the proposed method by dividing these values by the pre concentration factor obtained after the IIP-SPE optimization. Then, the LOD and LOQ for the method were equal to 0.322 μg L−1 and 1.08 μg L−1, respectively.
It is emphasized that these parameters for the proposed IIP-SPE voltammetric method were better than some others shown in Table 1, particularly ones that involve the synthesis of Hg+2-IIP. As expected, AFS, ICP-OES, and ICP-MS methods generate lower LOD and LOQ, but this work proposes a more affordable alternative.
Accuracy
Application to laboratorial sample
To test the accuracy of the method three aliquots of Hg II 1.00 mg L−1 standard solution were submitted to the proposed IIP-SPE protocol, assessing the accuracy of 98.6 ± 1.5%.
Application to environmental samples
The IIP-SPE methodology was applied to three different environmental water samples, Piratininga and Itaipu Lagoons, located in Niterói-RJ/Brazil, and Vargem River, located in Itaboraí, RJ/Brazil. These waters were sampled superficially and filtered as mentioned in item 2.5.
Initially, the three samples were directly analyzed by the voltammetric method, without employing the proposed IIP-SPE pretreatment, in order to investigate whether this procedure would be fundamental to the quantification or not. By doing so, no characteristic peak of mercury ions was obtained, even when spiking these samples with Hg II 12.5 μg L−1. That observation made it possible to affirm that other matrix components were interfering in the voltammetric analysis.
When the IIP-SPE procedure was applied to the lagoon samples also no analytical signal was observed, but when spiking these samples with Hg II 12.5 μg L−1 it was. In the case of spiking, a recovery of 90.0 ± 1.3% was achieved. These results show that the Hg II concentration in the lagoon samples is below the LOQ of the proposed methodology.
In opposition to the lagoon water samples, the Hg II concentration in Vargem River was high and determined 1.35 ± 0.07 mg L−1 after IIP-SPE, resulting in a recovery of 96.2 ± 4.0% after spiking with Hg II 0.250 mg L−1.
The accuracy of the method was also evaluated by comparing these results to a flame atomic absorption spectrometry (FAAS) reference method (Table 3). The IIP-SPE pretreatment was also necessary to accomplish the FAAS quantification, as other matrix components interfered in the analysis by strongly increasing the background signal.
By the application of Student’s t test at 95% confidence level to the results of the quantifications in Vargem River, it was possible to affirm that there was no statistical difference between the results obtained by the two techniques.
Study of possible interferents
In order to study the influence of other cations over the Hg II analytical signal for the proposed methodology, a laboratorial sample was simulated by considering the limits permitted by the Brazilian legislation CONAMA/357, which deals with river and lagoon waters quality (CONAMA, Conselho Nacional do Meio Ambiente 2005).
Thus, an aqueous solution containing Hg II 2.00 μg L−1, Cd II 10.0 μg L−1, Pb II 100 μg L−1, and Cu II 90.0 μg L −1 was submitted to the IIP-SPE method followed by differential pulse anodic stripping voltammetry quantification. It was possible to assess that in these conditions the studied cations did not influence the Hg II analytical signal, as the concentration of the analyte was determined 2.36 ± 0.42 μg L−1.
Conclusions
Both imprinted and non-imprinted polymers were successfully synthesized by the methodology here proposed. These two polymers were differentiated in terms of sorption capacity, as IIP presented a higher capacity (31.0 mg g−1) than NIP (20.0 mg g−1). This differentiation was corroborated by physical characterizations, by means of infrared spectrometry and atomic force microscopy.
After the optimum condition for IIP-SPE was settled, the imprinted polymer presented sufficient selectivity to Hg II against other cations (Cd II, Pb II, and Cu II), considering the expected concentration of those in river and lagoon waters according to the Brazilian legislation.
The extraction procedure using the ion-imprinted polymer as the solid phase was essential to perform mercury II quantification in the evaluated water samples, as it was responsible for eliminating the influence of other matrix components that caused interferences. The accuracy of the proposed methodology was 96.2 ± 4.0% calculated by means of recovery tests using river water samples. Also, the results of Hg II quantification in these samples using the IIP-SPE procedure and differential pulse anodic stripping voltammetry were compared to those obtained by using flame atomic absorption spectrometry for the analyte detection. Comparing these results by Student’s t test at 95% confidence level, no statistical difference was observed.
References
Allibone J, Fatemian E, Walker PJ (1999) Determination of mercury in potable water by ICP-MS using gold as a stabilising agent. J Anal At Spectrom 14(2):235–239
Alizadeh T, Ganjali MR, Zare M (2011) Application of an Hg2+ selective imprinted polymer as a new modifying agent for the preparation of a novel highly selective and sensitive electrochemical sensor for the determination of ultratrace mercury ions. Anal Chim Acta 689(1):52–59
Babu SV, Reddy KH (2012) Direct spectrophotometric determination of mercury (II) using 2-acetylpyridine thiosemicarbazone in environmental samples. Indian J Adv Chem Sci 1:65–72
Batlokwa BS, Chimuka L, Tshentu Z, Cukrowska E, Torto N (2012) An ion-imprinted polymer for the selective extraction of mercury (II) ions in aqueous media. Water SA 38(2):255–260
CONAMA, Conselho Nacional do Meio Ambiente (2005) Resolução 357/. Available at: <http://www.mma.gov.br/port/conama/res/res05/res35705.pdf>. 17 Oct 2018
Dal Borgo S, Jovanovski V, Hocevar SB (2013) Antimony film electrode for stripping voltammetric measurement of Hg(II) in the presence of Cu(II). Eletrochim Acta 88:713–717
Dos Santos JS, De La Guárdia M, Pastor A, Dos Santos MLP (2009) Determination of organic and inorganic mercury species in water and sediment samples by HPLC on-line coupled with ICP-MS. Talanta 80(1):207–211
El-Sayed M, Ramzi M, Hosney R, Fathy M, Moghny TA (2016) Breakthrough curves of oil adsorption on novel amorphous carbon thin film. Water Sci Technol 73(10):2361–2369
EPA - United States Environmental Protection (1985) Ambient Water Quality Criteria for Mercury, United States. 8 p
Ghanei-Motlagh M, Taher MA, Heydari A, Ghanei-Motlagh R, Gupta VK (2016) A novel voltammetric sensor for sensitive detection of mercury(II) ions using glassy carbon electrode modified with graphene-based ion imprinted polymer. Mater. Sci. Eng. C Mater. Biol. Appl. 63:367-75,
Ghasemlou S, Aghaie H, Monajjemi M (2013) Thermodynamic study of Hg (II) ion adsorption onto nano hydroxyapatite from aqueous solution. JPTC 10(2):125–136
Han Q, Shen X, Zhu W, Zhu C, Zhou Z, Jiang H (2016) Magnetic sensing film based on Fe3O4@Au-GSH molecularly imprinted polymers for the electrochemical detection of estradiol. Biosens Bioelectron 79:180–186
Hande PE, Samui AB, Kulkarni PS (2017) Selective nanomolar detection of mercury using coumarin based fluorescent Hg(II)-ion imprinted polymer. Sensors Actuators B 246:597–605
He C, Long Y, Pan J, Li K, Liu F (2007) Application of molecularly imprinted polymers to solid-phase extraction of analytes from real samples. J Biochem Biophys Methods 70(2):133–153
Herrmann PSP, da Silva MAP, Bernardes RF, Job AE, Colnago LA, Frommer JE, Mattoso LHC (1997) Microscopia de varredura por força atômica: uma ferramenta poderosa no estudo de polímeros. Polímeros 97:51–61
Huang R, Shao N, Hou L, Zhu X (2019) Fabrication of an efficient surface ion-imprinted polymer based on sandwich-like graphene oxide composite materials for fast and selective removal of lead ions. Colloids Surf A 566:218–228
Khajeh M, Yamini Y, Ghasemi E, Fasihi J, Shamsipur M (2007) Imprinted polymer particles for selenium uptake: synthesis, characterization and analytical applications. Anal Chim Acta 581:208–213
Khoddami N, Shemirani F (2016) A new magnetic ion-imprinted polymer as a highly selective sorbent for determination of cobalt in biological and environmental samples. Talanta 146:244–252
Knox JC, Edner AD, Levan MD, Coker RF, Ritter JA (2016) Limitations of breakthrough curve analysis in fixed-bed adsorption. Ind Eng Chem Res 55(16):4734–4748
Liang X, Chen S, Gao J, Zhang H, Wang Y, Wang J, Feng L (2018) A versatile fluorimetric chemosensor for mercury (II) assay based on carbon nanodots. Sensors Actuators B Chem 265:293–301
Liu Y, Hu X, Meng M, Liu Z, Ni L, Meng X, Qiu J (2016) RAFT-mediated microemulsion polymerization to synthesize a novel high-performance graphene oxide-based cadmium imprinted polymer. Chem Eng J 302:609–618
Liu Z, Wu W, Liu Y, Hu X (2018) Preparation of surface anion imprinted polymer by developing a La(III)-coordinated 3-methacryloxyethyl-propyl bi-functionalized graphene oxide for phosphate removal. J Taiwan Inst Chem Eng 85:282–290
Luo F, Huang S, Xiong X, Lai X (2015) Synthesis and characterization of Hg(II)- ion- imprinted polymer and its application for the determination of mercury in water samples. RSC Adv 5:67365–67371
Martín-Yerga D, González-García MB, Costa-García A (2013) Electrochemical determination of mercury: a review. Talanta 116:1091–1104
Mazzotta E, Turco A, Chianella I, Guerreiro A, Piletsky SA, Malitesta C (2016) Solid-phase synthesis of electroactive nanoparticles of molecularly imprinted polymers. A novel platform for indirect electrochemical sensing applications. Sensors Actuators B: Chem 229:174–180
Mergola L, Scorrano S, Bloise E, Bello MPD, Catalano M, Vasapollo G, Sole RD (2015) Novel polymeric sorbents based on imprinted Hg (II)- diphenylcarbazone complexes for mercury removal from drinking water. Polym J 48:73–79
METROHM (2018) Application Bulletin 422/2. Determination of mercury in water with the scTRACE Gold. Available at: <https://partners.metrohm.com/GetDocument?action=get_dms_document&docid=2743090>
Moussa M, Ndiaye MM, Pinta T, Pichon V, Vercouter T, Delaunay N (2017) Selective solid phase extraction of lanthanides from tap and river waters with ion imprinted polymers. Anal Chim Acta 963:44–52
Oliveira LMJ, Vidal-Torrado P, Otero XL, Ferreira JR (2007) Mercúrio total em solos de manguezais da Baixada Santista e Ilha do Cardoso. Quím Nova 30(3):519–524
Rajabi HR, Roushani M, Shamsipur M (2013) Development of a highly selective voltammetric sensor for nanomolar detection of mercury ions using glassy carbon electrode modified with a novel ion imprinted polymeric nanobeads and multi-wall carbon nanotubes. J Electroanal Chem 693:16–22
Rajabi HR, Zarezadeh A, Karimipour G (2017) Porphyrin based nano-sized imprinted polymer as an efficient modifier for the design of a potentiometric copper carbon paste electrode. RSC Advances 7:14923–14931
Rao TP, Daniel S, Gladis JM (2004) Tailored materials for preconcentration or separation of metals by ion-imprinted polymers for solid-phase extraction (IIP-SPE). TrAC Trends Anal Chem 23(1):28–35
Schlathauer M, Reitsam V, Schierl R, Leopold K (2017) A new method for quasi-reagent-free biomonitoring of mercury in human urine. Anal Chim Acta 965:63–71
Shakerian F, Kim K, Know E, Szulejko JE, Dasfarnia S, Shabani AMH (2016) Advanced polymeric materials: synthesis and analytical application of ion imprinted polymers as selective sorbents for solid phase extraction of metal ions. Trends Anal Chem 83:55–69
Shamsipur M, Rajabi HR (2013) Flame photometric determination of cesium ion after its preconcentration with nanoparticles imprinted with the cesiumdibenzo-24-crown-8 complex. Microchim Acta 180:243–252
Shamsipur M, Fasihi J, Ashtari K (2007) Grafting of ion-imprinted polymers on the surface of silica gel particles through covalently surface-bound initiators: a selective sorbent for uranyl ion. Anal Chem 79:7116–7123
Shansipur M, Rajabi HR, Pourmortazavi SM, Roushani M (2014) Ion imprinted polymeric nanoparticles for selective separation and sensitive determination of zinc ions in different matrices. Spectrochim. Acta A Mol. Biomol. Spectrosc. 117:24–33
Skoog DA, West DM, Holler FJ, Crouch SR (2006) Fundamentals of analytical chemistry. Boston, Cengage Learning, pp 435–455
Sobhi HR, Ghambarian M, Esrafili A, Behbahani M (2017) A nanomagnetic and 3-mercaptopropyl-functionalized silica powder for dispersive solid phase extraction of Hg (II) prior to its determination by continuous-flow cold vapor AAS. Microchim Acta 184(7):2317–2323
UNEP (2013) United Nations Environment Program. Global Mercury Assessment Sources, Emissions, Releases and Environmental Transport. Geneva
Valle ASS, Costa LC, Marques MRC, Silva CLP, Maria LCS (2011) Preparação de copolímeros à base de 2-vinilpiridina com propriedades bactericidas. Quím Nova 34(4):577–583
Vatanpour V, Madaeni SS, Zinadini S, Rajabi HR (2011) Development of ion imprinted technique for designing nickel ion selective membrane. J Membr Sci 373:36–42
Wuilloud JCA, Wuilloud RG, Silva MF, Olsina RA, Martinez LD (2002) Sensitive determination of mercury in tap water by cloud point extraction pre-concentration and flow injection - cold vapor – inductively coupled plasma optical emission spectrometry. Spectrochim Acta B At Spectrosc 57(2):365–374
Yusoff MM, Mostapa NR, Sarkar MS, Biswas TK, Rahman ML, Arshad SE, Sarjadi MS, Kulkarni AD (2017) Synthesis of ion imprinted polymers for selective recognition and separation of rare earth metals. J Rare Earths 35(2):177–186
Acknowledgments
The authors are grateful to the Brazilian institutions CAPES (Coordenação de Aperfeiçoamento de Pessoal de Nível Superior), CNPq (Conselho Nacional de Desenvolvimento Científico e Tecnológico), and FAPERJ (Fundação Carlos Chagas de Amparo à Pesquisa do Estado do Rio de Janeiro) for grants, scholarships, and financial support.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Additional information
Responsible editor: Severine Le Faucheur
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Francisco, J.E., Feiteira, F.N., da Silva, W.A. et al. Synthesis and application of ion-imprinted polymer for the determination of mercury II in water samples. Environ Sci Pollut Res 26, 19588–19597 (2019). https://doi.org/10.1007/s11356-019-05178-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11356-019-05178-y