Introduction

Zirconium is present at different stages of the nuclear fuel cycle, as it is a component of the PWR fuel rod cladding [1] and can also be found as a fission product [2]. In such case its isotope composition is not natural. In the recycling process, fuel rods are dissolved in concentrated nitric acid. Depending on chemical conditions, zirconium tends to precipitate through time in presence of molybdenum in nitric acid [3,4,5,6,7].

When wanting to study Zr precipitation kinetics in nitric acid, many samples are collected for analysis (at least seven, along the course of the reaction). Some analytical techniques are available directly in hot cells such as X-ray Fluorescence (XRF) [8], but in order to determine the zirconium isotope composition and concentration with uncertainties lower than 3%, it is necessary to use either thermo-ionisation mass spectrometry (TIMS) or multi collection inductively coupled plasma mass spectrometry (MC-ICP-MS) [9, 10]. The ATALANTE analyses laboratory (L2AT) only owns a nuclearized TIMS in a glove box, needing important dilution factors because of the high level of radiations the samples present.

The zirconium analysis method by TIMS for nuclear samples was presented by Quemet et al. [10], and necessitates to spike the samples in the shielded cell and perform separations before being able to analyse them as after the amount-of substance needed and the uncertainties required are not compatible with important dilution factors. Here, fourteen samples from two different kinetics experiments were supposed to be aliquoted. The average duration for a single sample analysis is 15 working days and even though it can be mutualised, it is difficult to use only this method to follow the Zr precipitation kinetics over its entire course. It was necessary to propose a new analytical method not necessitating TIMS analysis only.

This paper proposes a method using TIMS analyses of the first and final samples of the kinetics and an ICP-MS relative analyses for the kinetics core samples. Only four samples are going to be analysed using the TIMS method, all the other are going to be analysed using the ICP-MS method developed here. Optimisation experiment and development parameters are going to be explained and results obtained on two different kinetics experiments be presented.

Experimental

Reagents

Nitric (w t = 67%, Prolabo, USA) and hydrofluoric (w t = 67%, Prolabo, USA) acids were of Ultrapure Normatom grade. Oxalic acid was purchased from Fluka (USA). Lutetium, zirconium, molybdenum and ytterbium in nitric acid standard solutions (1000 µg mL−1) were purchased from SPEX® (USA). Deionised water was provided by a MilliQ system (resistivity: 18.2 MΩ cm, Millipore, Milford, MA, USA).

Initial solutions consisted of 6.6 mol L−1 nitric acid, prepared by diluting concentrated HNO3 in deionised water volumetrically. Stabilisation solution consisted of 5 mol L−1 nitric acid prepared the same way.

ICP-MS dilution solution consisted of (2%/0.1%) (w t/w t) HNO3/HF and were prepared by volumetric dilution in deionised water. Samples were spiked with 10 µg L−1 ytterbium. ICP-MS rinse solution consisted of (10%/0.1%) (w t/w t) HNO3/HF.

For TIMS analyses and upfront separation, 4 and 8 mol L−1 nitric acid solutions were prepared by diluting high purity nitric acid (Merck, Suprapur) with deionised water. For the separation protocol, an elution solution consisted of (4 mol L−1/ 0.1 mol L−1) HNO3/H2C2O4 mixture. It was prepared by diluting nitric acid with deionized water and by weighing dihydrated oxalic acid.

UTEVA resins (Triskem®, USA, 100–150 µm particle size in 2 mL prepackaged columns) were used in the separation protocol in order to obtain a pure zirconium fraction.

Precipitation experiment

Dissolution solutions of irradiated UO x fuel were obtained after heating for 30 min hulls at 85 °C in initial solution consisting of 6.6 mol L−1 nitric acid. Zr and Mo concentrations were approximatively 2 and 1 g L−1, respectively. The solutions were heated under reflux over a few hours to dissolve the fuel. Zr was then let precipitating as zirconium molybdate for about 40 h. Aliquots were sampled on a regular basis during the precipitation experiment. To stop precipitation, samples were diluted and acidified in 5 mol L−1 HNO3 gravimetrically right after being collected. Samples were stored in this matrix until being used. This experiment was performed on two different initial solutions (hereafter referred to as solution 1 and solution 2). Once the dissolution was complete, solutions are called dissolution solution.

Solutions and samples preparation

Initial solutions were prepared gravimetrically in a gauged flask, in order to validate density and consisted of 6.6 mol L−1 HNO3. In order to reduce uncertainties and follow dilution factors for ICP-MS measurements, it was spiked with 100 mg L−1 lutetium (hereafter referred to as process internal standard). Indeed, it is time consuming to weigh samples in a shielded cell, thus it was decided to follow a process internal standard and perform volumetric dilutions. It was verified lutetium has no influence on zirconium separation. Results for ICP-MS would be given relatively to Lu.

For ICP-MS, an approximate (as Lu is present as process internal standard) 5000-fold dilution was performed in a shielded facility to reduce radiation level. Afterwards, a 100-fold dilution was performed in a glove box leading to a 500,000-fold dilution.

Tests were performed on non-interfered isotopes using natural elements standard solutions beforehand: 90Zr and 98Mo.

However, for TIMS, such a dilution could not be performed as the amount-of-substance would not have been sufficient, thus a separation was necessary to remove isobars as well as emitting elements. For TIMS analysis, a separation step is required to obtain a pure Zr fraction in order to eliminate all interfering elements (such as Mo). It also helps decreasing the radiation level mainly due to cesium and actinides present in the sample to obtain a radiation level of the solution compatible with glove box work. This separation was performed in shielded cell. It was described entirely by Quemet et al. [10]. The sample, previously diluted with 8 mol L−1 nitric acid and deionized water in order to obtain a 4 mol L−1 nitric acid matrix, was deposited onto the column. The column was then washed with 4 mol L−1 HNO3 in order to eliminate all interfering elements (the Zr is extracted on the column). Zirconium was eluted with a HNO3/H2C2O4 mixture. After the separation, the elution solution was transferred to the TIMS analysis laboratory by pneumatic transfer. The elution solution was then evaporated and re-dissolved with nitric acid in order to obtain a Zr concentration around 5 µg µL−1. Then, 1 µL of this solution was deposited onto the filament for the isotope analysis.

For the sample isotope analysis, one separation was performed in the shielded cell. Then, 3 independent isotope analyses were performed by TIMS. Zirconium concentration determination was performed by isotope dilution. For each concentration determination, 3 sample-tracer mixtures were prepared gravimetrically with 200 µL of tracer (natural Zr SPEX solution at 1000 µg L−1) and 250 µL of sample. The separation step was performed on the 3 mixtures before the isotope analysis.

Instruments

The ICP-MS measurements were performed on an Elan DRCe (PerkinElmer, USA) equipped with a glove box. Plasma conditions were verified 30 min after start up and optimised if needed. A Meinhard© concentric nebuliser was used with a low volume baffled cyclonic spray chamber. To avoid any measurement delays or carry over, uptake and rinse parameters were optimized (Table 1). In order to obtain the lowest RSD possible and limit sample consumption 60 replicates were performed. Samples were analysed using peak hopping and integration time was set to 50 ms amu−1. Ytterbium was used as an analytical internal standard to monitor plasma variations. As Zr was a fission product, its occurring isotopes are: 90, 91, 92, 93, 94, 95 and 96. 90Zr being interfered by 90Sr and 93Zr by 93Nb, isotope 92 and 94 were measured for the kinetics experiments to avoid isobar interferences.

Table 1 Sampling parameters

Description of filament loading techniques and details on the TIMS Thermo Fischer Triton instrument (Waltham, MA, United States) are given by Quemet et al. [10, 11]. Concentration analysis by isotope dilution was performed with this developed method [10] while the isotope composition was determined using the classical method.

Zirconium concentration determination by isotope dilution

The zirconium concentration determination was performed using isotope dilution. The equation for the sample concentration determination is defined by Eq. 1.

$$\left[ {\text{Zr}} \right]_{\text{S}} \,=\, \left[ {\text{Zr}} \right]_{\text{T}} \times \frac{{m_{\text{T}} }}{{m_{\text{S}} }} \times \frac{{M_{\text{S}} }}{{M_{\text{T}} }} \times \frac{{(90)_{\text{T}} }}{{(91)_{\text{S}} }} \times \frac{{\left( {{\raise0.7ex\hbox{${90}$} \!\mathord{\left/ {\vphantom {{90} {91}}}\right.\kern-0pt} \!\lower0.7ex\hbox{${91}$}}} \right)_{\text{Mix}} - \left( {{\raise0.7ex\hbox{${91}$} \!\mathord{\left/ {\vphantom {{91} {90}}}\right.\kern-0pt} \!\lower0.7ex\hbox{${90}$}}} \right)_{\text{T}} }}{{1 - \left( {{\raise0.7ex\hbox{${91}$} \!\mathord{\left/ {\vphantom {{91} {90}}}\right.\kern-0pt} \!\lower0.7ex\hbox{${90}$}}} \right)_{\text{Mix}} \times \left( {{\raise0.7ex\hbox{${90}$} \!\mathord{\left/ {\vphantom {{90} {91}}}\right.\kern-0pt} \!\lower0.7ex\hbox{${91}$}}} \right)_{\text{S}} }}$$
(1)

where T refers to the tracer used with the major isotope 90, Mix refers to the sample-tracer mixture and S refers to the sample; m are the masses involved in the mixture; M are the atomic weights; (90) and (91) are the isotope abundances of each major isotope; 90/91 and 91/90 are the isotope ratios.

For ICP-MS analysis, the most important contribution to uncertainties is the dilution factors. Here, the experiments were performed in a shielded line and the analysis in a glove box. Thus, the average dilution factor was 500,000-fold. That is why the initial solution (~6 M HNO3) was spiked with lutetium, in order to be able to compute a ratio between intensities (Eq. 2). It would allow following the evolution of concentration as Zr precipitates between t 0 and t f.

$$R = \frac{Zr(cps)}{Lu(cps)}$$
(2)

With Zr(cps): Zr intensity in count s−1; Lu(cps): lutetium intensity in count s−1; R ratio

Uncertainties associated with such a ratio depend only on the relative standard deviation (RSD) (Eq. 3):

$$U_{\text{R}} = 2 \times \frac{{{\text{RSD}} (R)}}{{\sqrt {{\text{number}}\; {\text{of}}\;{\text{replicates}}} }}$$
(3)

With R ratio; \({\text{RSD}} = \frac{\sigma \left( X \right)}{{\bar{X}}}\) \(\bar{X}\) average intensity; UR expanded uncertainties

The kinetics evolution Δ is expressed by Eq. 4:

$$\Delta = \frac{{X_{{{\text{t}}_{0} }} - X_{{{\text{t}}_{\text{f}} }} }}{{X_{{{\text{t}}_{0} }} }}$$
(4)

With Δ Zr evolution; X concentration for TIMS and ratio for ICP-MS; t 0 starting time; t f end time.

Uncertainties estimation for the evolution (Δ) follows Eq. 5 for ICP-MS measurements:

$$U_{\Delta } = 2 \times \sqrt {\frac{{R_{{t_{\text{f}} }}^{2} }}{{R_{{t_{0} }}^{4} }} \cdot u^{2} \left( {R_{{t_{0} }} } \right) + \frac{1}{{R_{{t_{0} }}^{2} }} \cdot u^{2} \left( {R_{{t_{\text{f}} }} } \right)}$$
(5)

where U expanded uncertainty; u uncertainties; R ratio

Results and discussion

TIMS analysis in such a set up allows for low uncertainties, however it is time consuming and not adapted for quick kinetics analysis. Thus, only the first (t 0) and final (t f) samples were analysed this way. The kinetics core samples were analysed by ICP-MS relatively to the process internal standard put into the initial solution (Lu).

Optimisation of ICP-MS analytical parameters

For ICP-MS analysis of intermediary samples, the dilution matrix is usually 2% HNO3 (w t). Natural Mo and Zr were mixed at a concentration of 10 µg L−1. However, zirconium shows lagging tendencies in presence of other elements in such medium (Fig. 1a). Solutions often contain hydrofluoric acid or oxalic acid for ICP analysis [12,13,14]. Oxalic acid cannot be used where it can be avoided as it is quite difficult to deal with as an analytical effluent in the nuclear fuel cycle, so HF was chosen. Its concentration being limited as well a 2%/0.01% HNO3/HF mixture was used and found efficient to stabilize Zr (Fig. 1b). HNO3 concentration was increased up to 10% in the rinse solution to avoid any carryover.

Fig. 1
figure 1

a Zr measurement in presence of Mo in 2% HNO3 matrix, b Zr measurement in presence of Mo in 10%/0.01% HNO3/HF matrix—[Zr] = [Mo] = 10 µg L−1

Figure 2 shows the evolution of the ratio uncertainty \(\left( {U_{\text{R}} } \right)\) as the number of replicates increases for ICP-MS measurements. Zr at 10 µg L−1 was mixed with Mo at 10 µg L−1 in a HNO3/HF solution. It leads to a lower uncertainty as the number of replicates increases. It seems that increasing the number of replicates further would lead to a better uncertainty as the RSD would be lower but the sample consumption would increase dramatically. Thus, a total number of replicates was set to 60.

Fig. 2
figure 2

Evolution of the ratio expanded uncertainty with the number of replicates. [Zr] = 10 µg L−1 and [Mo] = 10 µg L−1 in the dilution matrix

These optimised parameters were used to analyse the aliquots sampled during the kinetics.

Kinetics

Two different dissolution solutions were analysed through time. TIMS results and associated uncertainties are presented in Table 2. Concentration evolution (Δ) is about 20% in both cases. Uncertainties are lower than 3% on each sample except for t f on solution 2 (3.4%). However, they are higher than the one presented in the previous paper [10]. Indeed, the first study was performed on an idealistic case. Here, the sample was spiked gravimetrically on a scale in a shielded cell which is less accurate than the one the glove box where the TIMS analysis is performed. It strongly affected the uncertainties estimations.

Table 2 Initial and final sample concentrations by TIMS

For solution 2, uncertainties for TIMS analyses are slightly more important than the ones for solution 1 as uncertainties are more important for this solution analyses. Indeed, these solutions were stored a while longer than solution 1. Despite the stabilisation step they may have not been completely stable as it has been noticed in previous unpublished studies.

As Zr isotope composition was not natural, 92 and 94 isotopes were measured by ICP-MS. Figure 3 shows the ICP-MS results obtained. Uncertainties on each ratio is too small to be seen on this plot (lower than 1%). Time has been randomised for confidentiality purposes. Uncertainties are below 2%. It is enough to consider each sample is different from the one before it, making the differences of kinetics shape significant.

Fig. 3
figure 3

Ratio (Eq. 2) evolution through time for a solution 1 and b solution 2

ICP-MS results are in good agreement with the TIMS ones regarding evolution (Δ). Zr concentration is reduced by about \(\Delta_{{94_{\text{R}} }} = (20.26 \pm 0.40) \%\) and \(\Delta_{{92_{\text{R}} }} = \left( {20.33 \pm 0.41} \right) \%\) between the starting and the final kinetics sample for solution 1, and \(\Delta_{{94_{\text{R}} }} = \left( {15.98 \pm 0.14} \right) \%\) and \(\Delta_{{92_{\text{R}} }} = \left( {16.32 \pm 0.21} \right) \%\) for solution 2.

Uncertainties obtained by ICP-MS for evolutions are 2.0, 2.0, 0.85 and 1.26% respectively. They are significantly lower than the one obtained by TIMS. This phenomenon can be explained by the fact gravimetric spiking does not have to be taken into account for ICP-MS measurements.

Analytical management

It was chosen to differentiate the two analyses (i.e. initial and final sample by TIMS, and evolution by ICP-MS), not only because of time consumption but also because of effluent and waste management. Table 3 shows the practical parameters obtained and the interest of not analysing all samples by TIMS.

Table 3 Practical parameters

The method presented here is more cost-effective than the one using only TIMS. Indeed, high level effluent generation was reduced by three quarter, some medium level effluents were generated instead and working hours, in shielded line especially, were reduced by half.

It is a quite important factor as high level effluents and wastes are difficult to deal with, in regard of storage and evacuation, work conditions in shielded line are more strenuous than in glove boxes and performing experiments in the latter should always be preferred whenever possible.

Conclusions

Spiking sample at the beginning of the kinetics experiment and analysing the sample either by TIMS or by ICP-MS allows for low uncertainties and rapid analysis of zirconium. As no other information than the measured isotope intensity is necessary, computing a ratio by ICP-MS allows for uncertainties lower than the one obtained by TIMS for concentration evolution. Thus combining TIMS and ICP-MS allows answering quantitatively when kinetics shapes are required, giving the initial and final concentration and the kinetics evolution with uncertainties lower than 3%.