1 Introduction

The rapid increases in environmental metal pollution levels observed in recent decades have resulted in heightened concern over both public and environmental health. Traditionally, analytical measurements performed for environmental monitoring focused on the total content of each element of interest (e.g., toxic elements such as lead, mercury, and cadmium, as well as elements utilized in living organisms such as cobalt, selenium, and magnesium) in a sample (Kot and Namiesnik 2000). But, as knowledge advanced, it was soon realized that data on total elemental concentrations were insufficient. Biochemical and toxicological investigations have shown that the behavior and effects of a trace metal greatly depend upon its speciation (Adnívia et al. 2016; Florence 1986; Huang and Wang 2003). Therefore, it is essential to determine the relative abundances of the species of each trace element in varying environments.

Copper (Cu) is an important trace element that is required for the proper functioning of biological processes within aquatic organisms (Grosell and Wood 2002; de Polo and Scrimshaw 2012; Jacquot et al. 2014). However, Cu is toxic at increased concentrations as it interferes with ion transport (Tait et al. 2016; Whitby and van den Berg 2015). The toxicity of Cu is also dependent on its speciation (Tait et al. 2016; Guo et al. 2010). Cu is mostly found in inorganic and organic complexes in the environment, and this complexation decreases the fraction of Cu available to interact with organisms and to cause toxicity (Santore et al. 2012). The concentration of free Cu2+ ranges from 0.1 to 2 nM, which is toxic to a variety of ecologically important phytoplankton and zooplankton species (Byrne and Miller 1985). Research by Batley and Florence (1971) established that there are seven potential chemical and physical forms of metal ions in aquatic environments, such as free metal ions, labile metals weakly complexed with organic compounds, metals strongly bound in organic complexes; and so on. Among others, van den Berg proposed the concept of active ions and established a method for detecting them using electrochemical methods (Gledhill and van den Berg 1995). The concentration of labile Cu was determined by Huang and Wang using differential pulse anodic stripping voltammetry (Huang and Wang 2003). The correlation between the ASV-labile concentration of Cu and the toxicity derived from bioassays can be useful for predicting the toxicity of Cu.

The conditional stability constant reflects the strength of an interaction between a metal ion and a ligand. The conditional stability constant of an organic Cu2+ complex, K, is defined as K = [CuL]/[Cu2+][L′], where [CuL] is the concentration of Cu complexed by natural ligands, [Cu2+] is the Cu ion concentration, and [L′] is the concentration of the free ligand. Hirose divided the organic ligands in seawater into three classes: very strong (logK > 13), strong (logK ≈ 12), and weak (logK < 10) (Hirose 1994). In an attempt to further simplify speciation analysis processes, we decided to divide the total pool of metal speciation present in water samples into four speciation groups: electroactive metals, acid-dissolved metals, inert metals, and total dissolved metals. Electroactive metals were defined as ions that are directly detectable by electrochemical methods without any additional processing. While they are generally present in low concentrations, they are associated with high bioavailability and toxicity in marine environments. Metal ions have been found to complex rapidly with any organic matter present (Hirose 2006; Abdelraheem et al. 2017), and acid-dissolved metals were defined as metal ions that form complexes that are easily destroyed by acidic environments. Acid-dissolved metals have been found to possess relatively low activity but high levels of bioavailability and absorption by surrounding ecosystems (Guo et al. 2010; Huang and Wang 2003; Tait et al. 2016; Hirose 1994). Inert metals were defined as metal ions that strongly couple with an organic complex, leading to high levels of stability and low bioavailability. In terms of quantification, the concentration of the inert metal was considered to be equivalent to the total metal concentration minus the acid-dissolved metal and electroactive metal fractions. A significant benefit of the method proposed here is that water samples can be divided up and then easily prepared for electroanalysis using either filtration, acidification, or digestion, allowing different metal species to be readily identified and detected.

In the study reported in the present paper, after carrying out simple preparation processes, different Cu species in seawater samples were detected and quantified using a sensitive electrochemical detector. As well as being an efficient, straightforward, and inexpensive technique for speciation analysis, this technique was observed to be highly stable and reproducible when employed for the long-term determination of Cu species in marine environments.

2 Experimental

2.1 Instruments

An electrochemical workstation (CHI 660D, Chenhua Instruments, Shanghai, China) was used for all electrochemical experiments. The electrochemical sensor contained a conventional three-electrode system consisting of a Au-film-modified glassy carbon working electrode (Au/GCE, 3 mm in diameter), a platinum foil auxiliary electrode, and a silver chloride reference electrode (Ag/AgCl, 3 M KCl). All potentials were measured against the Ag/AgCl reference electrode. The electrochemical UV digestion apparatus employed in this work was supplied by the Marine Electrochemistry Group (Department of Earth and Ocean Sciences, School of Environmental Science, University of Liverpool, UK). UV light was generated by a high-pressure mercury vapor lamp with a power output of 125 W, which peaked at around 250 nm. The sample collector (Taiheng Plastics Corporation, Henan, China) was filled with 0.1 M HCl (analytical grade) for at least one day and then thoroughly washed with deionized water before use. A portable pH/temperature measurement apparatus (pH-100B, Bangxi Scientific Instruments Corporation, Shanghai, China) was used for in situ measurements of water characteristics and parameters during sampling. Inductively coupled plasma mass spectrometry (ICP-MS, ELAN DRC, PerkinElmer, Waltham, MA, USA) was carried out to quantitatively analyze Cu species, and the resulting dataset was compared with the datasets obtained using other methods.

2.2 Chemicals and materials

The stock standard solution of 1000 mg L−1 Cu was supplied by the National Research Center for Certified Reference Materials (CRMs; Beijing, China). Separate standard solutions were prepared by diluting the stock solution with 10% v/v HCl. HAuCl4 stock liquid was purchased from Sinopharm Chemical Reagent Co. (Shanghai, China), and a 1 mM HAuCl4 solution was prepared in 0.2 M Na2SO4 for electrodeposition. All other reagents were of analytical grade and used without further purification unless otherwise stated. Deionized water (ρ = 18.2 MΩ cm) was obtained using a Pall (Port Washington, NY, USA) Cascada laboratory water system and was used in all the experiments. Environmental copper standard solutions (BY400031, BY40033) supplied by the Tanmo Quality Inspection Technology Corporation (Beijing, China) and calibrated by inductively coupled plasma optical emission spectrometry (ICP-OES) were used to validate and evaluate the new analytical method. The Chinese seawater standard GBW(E)130011 with a salinity of 34.99 was purchased from the National Standard Material Network. Polyethylene bottles (Xiangyun Plastic Products Factory, Hebei, China) that were used to hold water samples were filled with 0.1 M HCl (analytical grade) for at least 1 day and then thoroughly washed with deionized water before use. Cellulose acetate membranes that were used to filter water samples were supplied by Shanghai Xingya Purification Material Factory (Shanghai, China).

2.3 Sample collection

Seawater samples were all collected from the same location, a coastal point in the Huanghai Sea (121.47 E, 37.49 N), which is significantly impacted by tidal and anthropogenic activity. Surface seawater samples (from depths of between 0 and  1 m) were collected using pretreated 2.5 L sample collectors twice per day (at 08:00 a.m. and 14:00 p.m.). Once collected, the samples were immediately analyzed for temperature (T) and pH and then transferred into individual 500 mL acid-cleaned polyethylene bottles.

2.4 Sample pretreatment and determining the Cu speciation

Water samples were immediately processed following collection according to the methodology summarized in Fig. 1. Water sample pretreatment was divided into three steps: filtration, acidification, and UV digestion. During filtration, the water samples were filtered through 0.45 μm cellulose acetate membranes. Acidification involved the addition of 1 mL HCl per 100 mL water in order to drop the pH value of the sample below 2.0. The sample was then maintained at 4 °C for 24 h. An acidification period of 24 h to ensure that all of the acid-dissolved Cu had reached equilibrium was chosen based on the results of a previous experiment (Laglera et al. 2013). Acidification was performed to break down complexes of Cu with organic matter, releasing the Cu. The procedure also ensured that there was no metal adsorbed on the container walls during storage (Mart 1979; Patterson et al. 1976). In the UV digestion step, the acidified samples were transferred to quartz tubes and digested for 45 min under irradiation from a high-pressure mercury vapor lamp, which destroys all complexes and compounds and thus allows the total dissolved Cu concentration to be determined (Capodaglio et al. 1994, 1998). Samples that only underwent filtration as pretreatment were then immediately analyzed by anodic stripping voltammetry to avoid metal adsorption; direct electrochemical analysis was employed without pH adjustment to establish the concentrations of electroactive Cu species. Samples pretreated with filtration and acidification steps were analyzed by the same electrochemical method described above, but the overall concentration of acid-dissolved Cu and electroactive Cu in each marine sample was obtained. The concentration of acid-dissolved Cu was then derived by subtracting the previously obtained concentration of electroactive Cu species from the overall concentration of acid-dissolved Cu and electroactive Cu. Finally, the total dissolved Cu concentration was determined for samples pretreated with filtration, acidification, and UV digestion. The concentration of inert Cu was then derived by subtracting the acid-dissolved Cu and electroactive Cu concentrations from the total dissolved Cu concentration.

Fig. 1
figure 1

Diagram showing the process used to analyze the speciation of Cu

2.5 Electrochemical analysis

Following the preprocessing steps, the samples were analyzed by electrochemical anodic stripping voltammetry using the electrochemical sensor. Before using it, a bare glassy carbon electrode (GCE) was polished sequentially with a 0.3 and 0.05 μm alumina slurry, washed thoroughly with deionized water, and then ultrasonicated for 1 min. The GCE was then subjected to potential cycling (− 0.2 to 0.8 V, 10 mV s−1) in 0.5 mol L−1 H2SO4 until stable and reproducible cyclic voltammograms were obtained. Finally, the GCE was immersed in a 0.2 mol L−1 Na2SO4 solution containing 1 mmol L−1 HAuCl4 to electrodeposit a Au film onto the surface of the GCE, thus generating the working Au/GCE. The deposition parameters used were as follows: constant potential, − 0.2 V; deposition time, 60 s.

Cu speciation analysis was performed directly without further pH adjustment in 10 mL pretreated marine samples by anodic stripping voltammetry using the electrochemical sensor. This involved two main steps: accumulation and stripping out. First, Cu2+ was reduced to Cu on the surface of the Au/GCE under a potential of − 0.4 V for 120 s with continual agitation. After an equilibration period of 10 s, square-wave stripping voltammetry (SWV) was performed from 0.2 to 0.7 V in increments of 0.004 V. It was found that the Au/GCE could be reused multiple times as long as it was electrolytically cleaned at a constant potential of 0.5 V for 10 s with slow stirring after each use. The concentrations of different Cu species were determined using the standard addition method.

3 Results and discussion

3.1 Speciation of Cu in seawater

Different Cu species were detected in the seawater samples, and relevant SWV current densities are presented in Fig. 2. Cu species concentrations obtained via the standard addition method are reported in the Electronic supplementary material (ESM). In Fig. 2, curve (a) was obtained for a sample pretreated with filtration only, so it shows the oxidation peak of electroactive Cu. A large amount of organic matter was present in the sample, and this matter interfered with the electrochemical detection and distorted the shape of the peak. Conversely, the shapes of the acid-dissolved Cu curve (b) and the total dissolved Cu curve (c) are less distorted because additional acidification or digestion pretreatment steps were applied to the relevant samples. In addition to the peak from Cu, another peak can be seen in curve (c). This can be attributed to the oxide species (bromine and chlorine) that are present in the seawater after UV irradiation. Acidification or UV irradiation destroys organic matter, releasing bound metal and ensuring that any adverse effects from the presence of organic matter on the voltammetric determination are eliminated (Mart et al. 1980; Annibaldi et al. 2011). The peak potentials of different species of Cu varied to a minor extent depending on the pretreatment method(s) applied. This can be attributed to the influence of the organic matter present in the samples on the rate of Cu oxide dissolution. It should be noted that the peak currents of the three Cu speciation groups differed significantly due to the varying Cu concentrations in the samples. The peak currents of different Cu species are closely related to both the content of organic matter and the physicochemical parameters of the natural water sample; Cu ions may react differently in different aquatic environments (Palma et al. 2016; van den Berg, 1982).

Fig. 2a–c
figure 2

Voltammetric curves of electroactive Cu (a), acid-dissolved Cu (b), and total dissolved Cu (c) in real seawater

3.2 Validation of the analytical method

The method created in this work was used to analyze real seawater directly. A good stripping peak for Cu was obtained, demonstrating that this method is not influenced by the salinity of real seawater. Since this particular experiment was performed in seawater directly, it was impossible to explore the blank level of this method. In order to investigate the detection limit of this method, experiments were carried out with a seawater standard, and the detection limits of the method for electroactive Cu, acid-dissolved Cu, and total dissolved Cu were found to be 1.1, 0.8, and 0.5 nM, respectively.

To further validate the proposed analytical procedure, the method was applied to determine the total dissolved Cu in analytical standard reference materials and real environmental seawater samples. The concentrations of Cu in the environmental seawater samples determined using the electrochemical method were compared to the corresponding results obtained using ICP-MS (Table 1), and good agreement was observed. Before the ICP-MS determination, 0.5 mL of 10 mg mL−1 Ga3+ were added to each 10 mL seawater sample. Then, by adjusting the pH of the solution to 9.0 with ammonia hydroxide, interference from the salt in the water was eliminated through coprecipitation. After that, the solution was examined by ICP-MS using argon as a carrier gas, and the Cu in the solution was quantified by a standardized curve method. The measured concentrations of total dissolved Cu in the environmental standard reference materials were found to be in good agreement with the certified values within experimental error, showing that the measurements were accurate.

Table 1 Analytical results (n = 3) for the determination of Cu (nM) in certified environmental standard reference materials and an environmental sample

These findings demonstrate the applicability of this method to the direct determination of the total dissolved Cu in real seawater samples. The currently favored analytical method, ICP-MS, detects and quantifies the total dissolved Cu concentration against standard reference materials, and therefore does not quantify electroactive Cu or acid-dissolved Cu. Since the same method was used to separately determine the electroactive Cu, the acid-dissolved Cu, and the total dissolved Cu, and this method was validated, we can infer that the electroactive Cu and acid-dissolved Cu concentrations determined using the method were accurate.

3.3 Reproducibility and repeatability

If it is to be used for environmental monitoring applications, the sensor must be stable. Therefore, the reproducibility of the signal from the electrochemical sensor was checked by performing Cu detection in environmental seawater samples twenty consecutive times. Real seawater that contained 200 nM Cu was used, and the acid-dissolved Cu was determined as described above. After each determination, the Au/GCE was electrolytically cleaned and renewed by oxidizing any Cu that remained on the electrode. The observed fluctuation in the response of the sensor across the twenty consecutive analyses is shown in Fig. 3. The horizontal dashed line in Fig. 3 is the average of the twenty determinations. The relative standard deviation (RSD) of the peak current was 7.05%. This plot illustrates that the electrochemical sensor can be used for continuous analysis, as it shows high stability and reproducibility while detecting various Cu speciation groups.

Fig. 3
figure 3

Variance and reproducibility of the peak current from the electrochemical sensor upon continuous repeated use in real seawater containing 200 nM Cu ion. Dashed line is the average of all peak current values

The repeatability of the electrochemical sensor was investigated by determining the total dissolved Cu in real seawater samples containing 200 nM Cu each day across a seven-day period. The electrode was used twice each day (once in the morning and once in the afternoon). After each determination, the electrode was renewed through potential (electrolytic) cleaning and then immersed in deionized water until the next determination. All of the observed current signals are shown in Fig. 4, where the dashed line is the average of all the current values. The current produced by the electrode fluctuated more dramatically across the seven days than in the reproducibility experiment. This is because all of the measurements were performed in the same environment in the reproducibility experiment, so the variation in the current signal observed in that experiment is an almost direct reflection of the stability and repeatability of the electrode. However, the current fluctuations observed during the repeatability experiment were caused by both electrode instability and variations in the analytical matrix or experimental environment over the seven-day period. The RSD of the peak current during the seven days was calculated to be 11.71%; further optimization should allow the sensor to meet the stability requirements for effective continuous monitoring.

Fig. 4
figure 4

Variance in and reproducibility of the peak current of the electrochemical sensor upon its continuous use in real seawater containing 200 nM Cu ion over a seven-day period. Dashed line is the average of all peak current values

3.4 Application to the determination of different Cu species

Electrochemical stripping voltammetry was used to continuously monitor different Cu species in environmental seawater samples. Monitoring lasted for a 15-day period (1st March 2017 to 15th March 2017). Samples were collected at 08:00 a.m. and 14:00 p.m. each day, and the mean of at least three determinations was used for analyses. Relevant seawater characteristics and parameters as well as the concentrations and relative abundances of different Cu species are shown in Table S1 of the ESM, while a comprehensive summary of the concentration fluctuations is provided in Fig. 5. During days 1–5, the electroactive Cu concentration ranged from 5.91 to 14.87 nM, with an average concentration of 9.96 nM. The acid-dissolved Cu concentration ranged from 27.56 to 48.80 nM, with an average concentration of 36.45 nM, while the total dissolved Cu concentration ranged from 45.03 to 78.28 nM, with an average concentration of 60.05 nM. There was no substantial change in the concentration of each Cu species between the first and second sampling event, although the largest Cu concentrations were measured on day 11 and over the following few days. This phenomenon may be ascribed to strong rainfall in the Yantai area on day 11, resulting in river runoff. This runoff washed Cu and dissolved organic materials into the seawater, causing significant increases in the concentrations of electroactive Cu, acid-dissolved Cu, and total dissolved Cu species. Also, the disturbance of the sediment by the rainwater would have led to the resuspension of organic matter, causing major fluctuations in the levels of different Cu species. The maximum concentrations of electroactive Cu, acid-dissolved Cu, and total dissolved Cu were observed to be 27.23, 73.62, and 122.85 nM, respectively. However, there was no obvious increase in the concentration of inert Cu. As reported previously (Plavšić et al. 2008), the dissolved organic ligands present in rainwater during spring contain a particularly high number of active sites for Cu complexation, indicating that many of the organic ligands that complex with Cu in rainwater samples are not stable and can be destroyed by acid-based pretreatment methods. Three days after the rainfall, the concentrations of the different Cu species had generally returned to normal. Additionally, daily variations (i.e., fluctuations from a.m. to p.m.) in Cu concentrations were likely to have been due to tidal patterns, although specific patterns were unclear.

Fig. 5
figure 5

Concentrations of different species of Cu detected in real seawater samples during the experimental period. Along the abscissa, bar “a” represents the morning and bar “b” represents the afternoon

To further explore the correlations between the concentrations of different Cu species and the physicochemical parameters of seawater, a Pearson’s correlation analysis (Wang et al. 2014) was performed, and the results are reported in Table 2. Significant positive relationships were found between the electroactive Cu concentration and the acid-dissolved Cu level (r = 0.83, p < 0.01) as well as the total dissolved Cu level (r = 0.92, p < 0.01). The correlations of the inert Cu concentration with the electroactive Cu level (r = 0.53, p < 0.01), the acid-dissolved Cu level (r = 0.35, p < 0.05), and the total dissolved Cu level (r = 0.65, p < 0.01) were relatively poor. This may be because there was conversion of electroactive Cu into acid-dissolved Cu and vice versa under certain conditions, and both of these forms of Cu were directly related to the total dissolved Cu. However, the inert Cu was very stable. Seawater is a typical multicomponent system that contains a large number of metal, inorganic, and organic ligands. There would have been competitive reactions involving the different species of Cu in seawater, which would have eventually reached equilibrium. Inert Cu was strongly bound to organic complexes, so the concentration of inert Cu would have been mainly determined by the types of inorganic and organic ligands present in the seawater. Positive relationships were also observed between the pH and the various Cu species concentrations because the correlation coefficients (r) were all greater than 0.3 (p < 0.05). There were no significant correlations between T and the concentrations of the various Cu species.

Table 2 Pearson’s correlation matrix for the concentrations of different Cu species and the physical and chemical parameters of seawater samples

The relative abundances of the different Cu species comprising the total dissolved Cu were investigated. In complex environmental matrices, the abundances of different Cu species may be significantly driven by the types and concentrations of the organic species and matter present (Plavšić 2003; Schroeder et al. 2017). Electroactive Cu includes partially free Cu ions, which are believed to be the strongest influence on the bioavailability and toxicity of Cu (Guo et al. 2010; Brown and Markich 2000). Dissolved inorganic and organic ligands are abundant in aquatic environments. Cu exists mostly in inorganic and organic complexes, which decreases the fraction of Cu available to interact with organisms and to cause toxicity (Tait et al. 2016). Upon considering the type of dissolved organic material present (Plavšić et al. 2006), if the dissolved organic material forms stable complexes with Cu and those complexes are difficult to destroy, the percentage of inert Cu increases and the Cu becomes less ecotoxic. Additionally, the Cu in complexes destroyed by acid digestion shows low activity but can be transformed and absorbed by living organisms in certain cases, leading to a high risk of trophic transfer. Based on the findings shown in Figs. 5 and 6, the dominant form of Cu in this region was acid-dissolved Cu, which accounted for around 60% of the total dissolved Cu, while the relative abundances of electroactive Cu and inert Cu were smaller (about 20% each). This implies that, in the studied area, the Cu generally formed weak complexes with dissolved organic materials, and acidification destroyed most of the dissolved organic material, thus releasing Cu. The variations observed in the levels of the Cu speciation groups (especially inert Cu) were not substantial, which may be because all of the samples were collected from the same sampling location, thereby limiting potential variations due to changes in water quality conditions or organic matter content. The electroactive Cu and the acid-dissolved Cu levels fluctuated more sharply than that of the inert Cu, further supporting the idea that the electroactive Cu and the acid-dissolved Cu levels were more unstable and more readily influenced by the external environment than the level of inert Cu. It is worth noting that while heavy rainfall was found to influence the total dissolved Cu concentration, no significant change in the ratio of the abundances of the Cu speciation groups was observed.

Fig. 6
figure 6

Variations in the relative abundances of the different Cu species during the experimental period. Along the abscissa, bar “a” represents the morning and bar “b” represents the afternoon

4 Conclusions

Electroanalytical techniques have been found to be useful and convenient for rapid metal speciation analysis. The present study effectively used a stripping voltammetric method to determine different Cu species during the direct analysis of seawater samples using an electrochemical sensor. The speciation analysis performed for this method does not require a complicated pretreatment process or separation operation, and the stability and reliability of this method have been effectively verified. Critically, the study has shown that this method can be used for the accurate long-term determination of various Cu species in complex seawater matrices, as the salinity of the matrix did not lead to any interference.

Based on the fixed-point monitoring results, it was concluded that without any changes in the environmental conditions, the relative abundances of the different Cu species would remain constant, and that the ratio of abundances was directly correlated with seawater quality and humus content. Rainfall influenced the ratio of the abundances of different Cu species because of the external input and the resuspension of sediment. Positive relationships were found between pH and the different Cu species concentrations, and no significant correlations were found between T and the levels of the different Cu species. The strengths of the correlations between different Cu species varied greatly, although the concentration of electroactive Cu was found to be strongly positively correlated with the levels of other Cu species. The relationships of these concentrations to environmental and seasonal variations require further study.

Considering the importance of effective speciation analysis in natural water monitoring, this morphological classification and analysis method shows great promise, as it could provide a better understanding of the role of metal speciation in the marine environment.