1 Introduction

Atrioventricular (AV) nodal reentrant tachycardia (AVNRT) is the most common form of paroxysmal supraventricular tachycardia [1]. AVNRT represents reentry in the area of the AV node, but the exact circuit remains elusive. Considerable evidence points to the inferior nodal extensions providing the substrate for slow pathway (SP) conduction [1].

Catheter ablation of the SP is considered first-line therapy for AVNRT [2]. This treatment option is curative for patients with AVNRT, and has high acute and long-term success and low recurrence rates [2, 3]. Despite a high success rate, ablation of the SP can be challenging in certain cases, likely due to anatomical variations [4]. During SP ablation procedures, the positioning of the ablation catheter is guided anatomically with attention to intracardiac electrograms [5,6,7,8]. For anatomical orientation, fluoroscopy and/or three-dimensional (3D) electroanatomical mapping system (EAMS) are used. However, these methods do not allow direct visualization of intracardiac structures, therefore are less helpful for SP ablation in patients with atypical anatomy [9].

Intracardiac echocardiography (ICE) is a unique, real-time imaging technique to assist catheter ablations by visualizing intracardiac structures, ablation catheter position, stability, and lesion [10, 11].

We hypothesized that real-time, direct visualization of the SP region using ICE can identify anatomical variations and facilitate the ablation procedure. The aim of this study was to compare conventional, fluoroscopy-guided versus ICE-guided SP ablation in consecutive patients with AVNRT.

2 Methods

2.1 Study population

In our prospective singe-center study, 91 consecutive patients undergoing electrophysiologic study and SP radiofrequency (RF) ablation for AVNRT were randomized into fluoroscopy-guided or ICE-guided groups. We excluded patients referred for a second (redo) procedure, those in whom other arrhythmias were also identified, and patients under 18 years of age. The protocol of the study is in accordance with the Declaration of Helsinki and the study protocol was approved by the regional ethics committee (Approval no.: 4649). All patients provided written informed consent for the study protocol.

2.2 Study protocol

Antiarrhythmic drugs were discontinued for at least five half-lives prior to the procedure. The electrophysiologic (EP) study was performed under conscious sedation using midazolam ± fentanyl, in a fasting state. After local anesthesia, following femoral venous access, one decapolar steerable catheter (interelectrode spacing 2–5-2 mm, Dynamic Deca, Bard Electrophysiology, Lowell, MA, USA) was positioned in the coronary sinus (CS), one quadripolar electrode catheter in the right ventricular apex (RVA), and one mapping/ablation catheter was positioned in the His bundle region. Twelve lead electrocardiogram (ECG) and intracardiac electrograms were recorded and stored on a digital recording system (CardioLab, GE Healthcare, Chicago, IL, USA), using a band pass filter between 30 and 500 Hz. Atrioventricular nodal conduction and the induction of AVNRT were tested by using electrical stimulation techniques. The S2 coupling interval was shortened by 10 ms after each drive-train until tachycardia was induced, or AV conduction block occurred, or the atrial refractory period was reached. If the tachycardia was not inducible, isoprenaline infusion (1–2 μg/min) was administered to increase the heart rate by at least 20%, and the same stimulation protocol was repeated during infusion and during the washout phase.

The diagnosis of AVNRT was based on established electrophysiologic criteria and pacing maneuvers including an A-(H)-V response after ventricular overdrive pacing, with an SA-VA interval > 85 ms and corrected postpacing interval minus tachycardia cycle length > 110 ms [12].

After the diagnostic EP study, in patients randomized to the ICE-group, the RVA quadripolar electrode catheter was removed and an 8F ICE catheter (AcuNaV™ 90 cm, Siemens Medical Solutions, Mountain View, CA, USA) was inserted to guide mapping and SP ablation. The echo-transducer was positioned in the low right atrium at the 6 o’clock position and rotated clockwise towards the septum. The imaging plane allowed the visualization of anatomic landmarks such as the septal cavotricuspid isthmus, ostium of the coronary sinus, tricuspid valve, right ventricle, and aortic root. The RF ablation catheter (CELSIUS® Catheter, Biosense Webster, CA, USA) was then placed between the coronary sinus ostium and the tricuspid ring guided by ICE (Fig. 1). The proximity of the ablation catheter to the compact AV node was judged by the distance from the aortic valve, which invariably marks the recording site of a proximal His potential. RF energy was delivered in sinus rhythm starting just below the CS with a power output of 30 W and a preset temperature of 55 °C. Applications were considered effective when junctional rhythm appeared. Effective applications were continued for 30 to 60 s. RF application delivery was immediately discontinued in the case of catheter displacement, sudden impedance rise, PR interval prolongation, anterograde AV, or retrograde VA block. In case of ineffective RF delivery, the ablation catheter was moved slightly closer to the aortic valve (but never closer than 0.5 cm) and RF application was attempted again.

Fig. 1
figure 1

Direct visualization of the ablation catheter in the slow pathway region by intracardiac echocardiography (ICE). Abbreviations: Abl, ablation catheter; Ao, aortic root; RA, right atrium; RV, right ventricle

In the conventional arm, catheter positioning was guided by fluoroscopic landmarks provided by the EP catheters and aiming at an atrial-to-ventricular electrogram amplitude ratio of 1:3–1:5. Crossover from the fluoroscopy-guided group to an ICE-guided procedure was allowed when after 8 RF applications, the ablation endpoint was not reached.

The procedural success of the ablation was obtained if the arrhythmia could not be induced after a 20-min waiting period and if there were no observations of > 1 echo beats, on and off isoprenaline.

Procedure time (in minutes) was measured from the time point of first femoral puncture until the withdrawal of the last venous sheath. Mapping plus ablation time was measured from the beginning of the mapping for the SP to the end of the last ablation attempt. Fluoroscopy time (in minutes), radiation dose (in mGy), and dose-area product (DAP in Gycm2) were automatically recorded by the fluoroscopy system. Ablation data (total number of the RF applications, sum of delivered RF energy expressed in Ws, and the total ablation time expressed in seconds) were calculated and stored by the EP recording system (CardioLab, GE Healthcare).

2.3 Statistical analysis

The distribution pattern of the data was evaluated using the Shapiro–Wilk tests. All tests were performed two-tailed with a significance level set to p < 0.05. Continuous data were presented as the mean ± SD or median (interquartile range, IQR), as appropriate while categorical variables are presented as absolute numbers and percentages. For comparisons, chi-square test and Mann–Whitney U test were used as appropriate. Statistical analyses were performed using SPSS 24 software (SPSS, Inc., Chicago, IL, USA).

3 Results

A total of 91 patients were randomized in 2 groups: 48 patients to the fluoroscopy-only and 43 patients to the ICE-guided group. There was no difference regarding the baseline characteristics of the study patients between the groups (male sex: 40.4% vs. 28.6%, p = 0.27; age: 53.4 ± 13.8 min vs. 51.0 ± 17.2 years, p = 0.49).

The procedural endpoint was achieved in all 91 cases; thus, acute success rate was 100%. We found no difference in puncture to mapping time between fluoroscopy-only vs. ICE-guided groups (43 ± 16.8 min vs. 42.8 ± 16.1 min, p = 0.95). Mapping plus ablation time (18.8 ± 16.1 min vs. 11.6 ± 15.0 min, p = 0.031), fluoroscopy time (4.9 (2.93–8.13) min vs. 1.8 (1.2–2.8) min, p < 0.001), and total ablation time (144 (104–196) s vs. 81 (60–159) s, p = 0.001) were significantly shorter in the ICE-guided group (Fig. 2). ICE-guidance was associated with reduced fluoroscopy dose (13.2 (8.2–13.4) mGy vs. 3.7 (1.5–5.8) mGy, p < 0.001) and radiation exposure (0.99 (0.56–1.57) Gycm2 vs. 0.22 (0.14–0.51) Gycm2, p < 0.001).

Fig. 2
figure 2

Box plots presenting mapping plus ablation time (A), fluoroscopy time (B), number of RF applications (C), and RFA time (D) in the fluoroscopy-only and the ICE-group. Abbreviations: ICE, intracardiac echocardiography; RF, radiofrequency; RFA, radiofrequency ablation

The sum of delivered RF energy (3866 (2786–5656) Ws vs. 2283 (1694–4284) Ws, p = 0.002) and number of RF applications (8 (4.25–12.75) vs. 4 (2–7), p = 0.001) were also lower with ICE-guidance (Table 1).

Table 1 Procedural parameters in the study population

Out of 48 patients randomized to the fluoroscopy-only group, twelve patients (25%) crossed over to the ICE-guided group. All were treated successfully thereafter and the number, time, and cumulative energy of RF applications delivered after crossover were similar to those recorded in the ICE-guided group (Table 2).

Table 2 Procedural parameters in the crossover group

There was no complication including AV block during the study. No recurrence occurred during the 14.8 ± 2.9-month follow-up.

4 Discussion

We found in this randomized comparison of fluoroscopy- versus ICE-guided ablation of the SP for AVNRT that ICE use is beneficial in reducing mapping plus ablation time, radiation exposure, and unnecessary RF delivery. It also facilitates treatment of challenging cases.

Supported by its low complication and a high success rate, catheter ablation is the treatment of choice for paroxysmal supraventricular tachyarrhythmias including AVNRTs [2, 3]. Katritsis et al. have shown that ablation at the anatomical area of the slow pathway (SP) is the therapy of choice for symptomatic AVNRT, regardless of whether the typical or atypical form is present [13]. The contemporary anatomical approach for SP ablation targets the lower below the CS ostium [14, 15], traditionally using fluoroscopy-guidance, often with prolonged X-ray and ablation times [9, 14]. This exposes both patients and medical staff to a potentially dangerous amount of ionizing radiation and results in unnecessary RF delivery to the heart in proximity to the conduction system, with a possible association with late AV nodal conduction disturbances [16].

Koch’s triangle varies in terms of anatomy and electrogram distribution (Supplementary videos 1 and 2) [4, 7]. ICE allows direct real-time visualization of the relevant anatomical structures resulting in more precise and targeted mapping and ablation. Bencsik et al. previously demonstrated that ICE-guided ablation of the cavotricuspid isthmus significantly shortens the procedure and fluoroscopy time and time spent for ablation in comparison with fluoroscopy-only procedures without an EAMS [17]. Previously studies reported on the advantage of mechanical radial ICE-imaging for SP ablation [18, 19]. Most recently, data from a retrospective analysis of 47 patients who underwent ICE vs. fluoroscopy-guided SP cryoablation showed that the use of ICE shortens the cryo-application duration; however, time needed for catheter placement is longer, when compared with conventional fluoroscopic guidance [20].

Recently the 3D mapping system-guided approach for catheter ablation of simple arrhythmias like AVNRT has become widespread [21,22,23]. However, while EAMSs can significantly reduce fluoroscopy time and dose, they may be less helpful in detecting variations in anatomy due to the lack of direct visualization of anatomical structures. Therefore, they may not provide benefit in terms of the number and cumulative time of RF ablations. This was shown in a meta-analysis published in 2016 comparing randomized controlled trials of zero/near zero-fluoroscopy versus conventionally performed procedures, and there was no difference in either total ablation time or procedure time between the groups [24]. Subsequent trials also did not show difference either regarding ablation time or application number [25,26,27]. One study showed increased procedure time with the use of an EAMS in AVNRT without benefit in success or complications [9].

These studies, however, did not use contact force sensing technology that could be similar to ICE in detecting contact between the ablation catheter and atrial tissue.

Compared to the above-mentioned studies on EAMS in AVNRT ablation, our study has shown that using ICE-guidance during RF ablation of the SP not only reduces radiation exposure, but also minimizes unnecessary RF delivery by direct visualization of both the ablation catheter and the SP region. Reduced RF energy delivery may impact the development of late conduction disturbances after SP ablation [28, 29].

Remarkably, in challenging fluoroscopy-guided procedures, after crossover to ICE-guidance similar number, time and cumulative energy of RF applications were needed as in the original ICE-guided group. This finding suggests that SP ablation in case of unusual anatomy is just as straightforward when the target area and catheter are directly visualized by ICE.

The only drawback of an ICE-guided strategy to SP ablation can be the extra cost of the catheter. Compared to the extra cost of using an EAMS, this expenditure is similar [30], or even lower if using reprocessed ICE catheters [31]. Compared to an EAMS, ICE offers the potential added advantage of a better targeted ablation with less energy delivered and facilitation of treatment of more challenging cases.

4.1 Limitation

Our series is a single center study, which may limit its applicability in other centers. Although the study was randomized, it could not be blinded. The level of ICE experience varied among the investigators. As discussed above, the extra cost of using an ICE catheter cannot be neglected. It also requires a larger venous sheath; however, no increase in access site complications occurred due to this. Multicenter trials are required to assess the cost-effectiveness and ability to improve clinical outcomes with this technique.

5 Conclusion

Compared to conventional fluoroscopy-guided procedures, ICE-guidance for anatomic SP ablation significantly reduces mapping and ablation time, radiation exposure, and RF delivery. When conventional ablation is ineffective after a fair effort, switching to ICE-guided ablation can be recommended.