Abstract
Purpose
To determine between-method differences in corneal endothelial cell parameters using center and automated methods of non-contact specular microscopy (CellCheck software of Konan, Inc.) in glaucomatous eyes.
Methods
We analyzed the central corneal endothelial cell density (ECD) of 245 glaucomatous eyes using center (ECD-Ce) and automated methods (ECD-Au). Based on the ECD-Ce, we allocated subjects to Groups 1 to 10 (at 250 cells/mm2 intervals) and evaluated the ECD, coefficient of variation in cell area (CV), and percentage of hexagonal cells (HEX).
Results
There was a close correlation (r = 0.91) between the ECD values measured using both methods. However, ECD-Au were significantly higher than those measured by the center method when ECD-Ce was less than 2500 (in Groups 1 to 8; P < 0.001 to P = 0.006). The regression equation of (ECD-Au—ECD-Ce) = 1028–0.397*ECD-Ce shows greater deviation in eyes with lower ECD, and this difference became 0 when ECD -Ce was 2593 cells/mm2. None of the 44 subjects with an ECD-Ce of < 1000 cells/mm2 recorded an ECD-Au < 1000 cells/mm2. Compared with the center method, the automated method had higher and lower median CV and HEX values, respectively (P < 0.001). The between-method differences in both CV and HEX were negatively correlated with ECD-Ce (r = −0.49, P < 0.001 and r = −0.25, P < 0.001, respectively).
Conclusion
The automated method of the CellCheck software overestimates ECD in eyes with lower ECD values and may overlook risk of corneal decompensation.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Corneal endothelial cells have ion transport systems that counteract water imbibition by the corneal stroma, and therefore maintain corneal clarity. In humans, corneal endothelial cells do not proliferate. Severe endothelial cell loss can cause bullous keratopathy. Currently, there are multiple glaucoma treatment methods; among them, trabeculectomy and tube shunt surgery may affect the corneal endothelium [1]. Therefore, it is important to accurately assess corneal endothelial cell parameters to evaluate the risk of corneal decompensation and choose appropriate glaucoma treatment methods. Endothelial cell density (ECD) is an important parameter for assessing endothelial function and health. Polymegathism, determined by the coefficient of variation in cell area (CV), and pleomorphism, represented by the percentage of hexagonal cells (HEX), are other important parameters that indicate stress to the endothelium [2].
The Food and Drug Administration of USA recommends the center method of specular microscopy as the “gold standard,” and it is used by virtually every professional reading center [3, 4]. On the other hand, the automated method is widely used in daily clinical practice because of its simplicity. When we use ECD data utilizing the automated method (ECD-Au), we must be careful that the ECD-Au may overestimate the ECD in glaucomatous eyes [5], post-Descemet stripping endothelial keratoplasty eyes [6], and eyes with low corneal ECD [5, 7]. As far as glaucoma eyes are concerned, there is only one report on overestimation of ECD and concerns for overestimation of ECD are yet to be well recognized.
Assuming ECD values obtained using the center method (ECD-Ce) approximate the real data, it is possible to evaluate the reliability of the automated method. In this study, we compared endothelial parameters of ECD, CV, and HEX in glaucomatous eyes obtained using both methods. To the best of our knowledge, there has been no study on the specific ECD range where a significant between-method difference becomes apparent. Therefore, to determine the critical ECD level, we provided a scatterplot to study the correlation between the difference between ECD-Ce and ECD-Au, and we allocated subjects into groups based on the ECD-Ce (250 cells/mm2 intervals) and analyzed between-method differences.
Materials and methods
This was a retrospective comparative study. The primary outcome measures were the ECD, HEX, and CV. We enrolled patients with a clear central cornea, allowing the acquisition of high-quality images of the corneal endothelium. We analyzed the eyes with a minimum of 30 contiguous countable cells.
We excluded images of poor quality due to halation, defocus, and darkness. Moreover, we excluded subjects with active anterior uveitis with cells or flare in the anterior chamber, active infectious conjunctivitis, keratoplasty history, severe corneal erosion, epithelial defects, and parenchymal opacity. We included patients with eyes with pterygium and corneal neovascularization if there was no encroachment into the image area.
We conducted a retrospective medical record review of central corneal endothelial images obtained using a non-contact specular microscope (Noncon Robo FA-3809II ver.4.05; Konan Medical Inc. Hyogo, Japan), which did not have an option to select patterns of cell size. We evaluated one image from each of the 260 glaucomatous eyes of 191 consecutive outpatients who visited the glaucoma service of Sensho-kai between September 2019 and January 2020; photographs of the endothelium were obtained and stored in a memory of device independently by 11 technicians blinded to the study between September 2010 and December 2019. As a control, we studied 36 eyes of 36 cataract patients who visited our institute in February 2020. These patients did not have a history of intraocular surgery, corneal disease, or intraocular inflammation.
We outputted corneal endothelial images stored in the computer memory, and an ophthalmologist (MM) reviewed each image and analyzed the data using both methods. For images obtained using the center method, the maximum number of contiguous cells was counted within the largest viewable field by dotting each cell’s center.
We classified the subjects into small groups based on ECD-Ce (group intervals of 250 cells/mm2). Moreover, we analyzed the CV, HEX, number of analyzed cells (NUM), and central corneal thickness (CCT) obtained using specular microscopy. CV values were presented as (standard deviation of cell areas, μm2/mean cell area, μm2) × 100. Moreover, we recorded the following for each participant: age, gender, Goldman applanation tonometric intraocular pressure (IOP) on the measurement, and glaucoma classification.
This study was approved by the Institutional Review Board of the Sensho-kai Eye Institute and followed the tenets of the Declaration of Helsinki. Written informed consent was obtained from all the participants.
Statistical analysis
All statistical analyses were performed using Bell Curve Excel Tokei 2016 (SSRI, Tokyo, Japan), JMP (SAS Institute Inc., Cary, NC, USA), and StatView5.0 (SAS Institute, Cary, NC) software. We assessed between-method agreements in the ECD, CV, and HEX values using the Mann–Whitney U test. Moreover, we used the Bland–Altman method to assess the between-method agreements of the ECD values. The Spearman test was used to evaluate the correlation coefficients between the evaluated data. Statistical significance was set at P < 0.05.
Results
Figure 1 shows the flow diagram of the study design. Two hundred and sixty eyes of 191 patients were classified into 14 groups (Groups 0–13) based on ECD-Ce at 250 cells/mm2 intervals. We excluded 15 subjects from four groups due to the small sample sizes of the subgroups. Consequently, we included 245 eyes from 183 patients in the final analysis.
Intra-individual reproducibility of data was studied in six right eyes of six control subjects by repeating examinations five times, and the intraclass correlation coefficient (ICC) and 95% confidence interval (CI), and the coefficient of variation (%) were calculated and is listed in Table 1. The ICC (1,5) of the ECD, CV, and HEX by the automated method (0.980, 0.777, and 0.706, respectively) and ECD-Ce (0.973) were greater than 0.7, showing good reproducibility, while those for HEX (ICC (1,5) = 0.514) and CV (ICC (1,5) = 0.412) by the center method showed fair reproducibility (Table 1).
Table 2 presents the demographic characteristics of the glaucoma and control groups. The type of glaucoma based on guidelines [8] was 78 primary open-angle glaucoma, 16 primary angle-closure glaucoma, 49 neovascular glaucoma, 42 exfoliation glaucoma, and 8 childhood glaucoma. Two eyes with traumatic glaucoma underwent blunt trauma 12 months and 7 years before the measurement date, respectively.
Figure 2 shows the number of participants in each group. Corneal endothelial damage was graded according to the Japanese corneal society guidelines [9] as follows: normal, ≥ 2000 cells/mm2; grade 1, 1000–2000 cells/mm2; and grade 2, 500–1000 cells/mm2. When the ECD was low, ECD-Au was greater than ECD-Ce. None of the 245 eyes was grade 2 using the automated method (ECD-Au less than 1000 cells/mm2), while 44 eyes were classified as Grade 2 (ECD-Ce 500–1000 cells/mm2) using the center method. When difference in ECD-Ce and ECD-Au was plotted against ECD-Ce, a regression formula obtained was: ECD-Au—ECD-Ce = 1028.5–0.397*ECD-Ce, r = −0.815, and R2 = 0.664, P < 0.001 (Fig. 3). The difference was 0 when ECD-Ce = 2593 /mm2.
Table 3 and Fig. 4 present the ECD data for each group. Compared with ECD-Ce, ECD-Au was significantly higher in Groups 1–8 (ECD-Ce: 500–2499 cells/mm2). The P value was < 0.001 for Groups 1–7 and P = 0.006 for Group 8. The respective between-method differences in the median ECD values were 860, 731, 558, 476, 275, 198, 176, and 127 cells/mm2, with the differences being larger in low cell density groups. When the ECD-Ce exceeded 2500 cells/mm2, there were no between-method differences in ECD. This suggests that the critical ECD value where the automated method becomes reliable is 2500 cells/mm2.
Figure 5 presents the Bland–Altman plots of the between-method comparisons of the ECD value, which showed that the ECD-Au overestimated the ECD by a mean difference of + 328 cells/mm2 (P < 0.001). Most of the values were not included within the limits of agreement. The differences greatly exceeded the upper limit of agreement for the low ECD groups. An inclination in the scatter of the dots is indicative of a proportional bias.
Figure 6a presents a scatter graph showing the correlations between ECD-Au and ECD-Ce in glaucomatous eyes. The correlation coefficient was r = 0.91, R2 = 0.8205, P < 0.001, and the linear regression equation was ECD-Au = 0.6033 × ECD-Ce + 1028.5 cells/mm2. Although there were significant between-method differences in the ECD values, there was a close correlation in the ECD values of both methods. When the correlation between the ECD-Ce and ECD-Au was studied in the 36 control eyes, the correlation coefficient was high again (r = 0.861, R2 = 0.742, P < 0.001). The linear regression equation was ECD-Au = 0.5871 × ECD-Ce + 1103.1 cells/mm2. This equation was nearly equal to that in the glaucomatous eyes (Fig. 6b). Thus, close correlation and overestimation of ECD in eyes with low ECD are a common finding in both glaucoma and control subjects.
Table 4 shows the age, IOP, CCT, CV, and HEX values in each group. The Mann–Whitney U test showed significant between-method differences in the CV and HEX values. In all groups, the automated method yielded higher CV values and lower HEX values than those obtained using the center method (P < 0.001).
Figure 7 shows an increasing tendency of the between-method CV deviation with decreasing ECD-Ce values. Figure 8 demonstrates the increased between-method differences in HEX with decreased ECD-Ce. Between-method differences in CV and HEX were negatively correlated with ECD-Ce (r = −0.49, P < 0.001 and r = −0.25, P < 0.001, respectively). This indicates that the differences in CV and HEX were larger at lower ECD values.
Figure 9 shows representative corneal endothelial images where significant between-method differences in the ECD values were observed. Images obtained from a 70-year-old-man in Group 2 revealed ECD values of 867 cells/mm2 and 1669 cells/mm2 and a CV of 28% and 51% using the center and automated methods, respectively. The automated method misidentified the cell border and erroneously divided one large cell into many small cells (Image A). Image B was obtained from a 65-year-old woman in Group 7 with ECD values of 2128 cells/mm2 and 2315 cells/mm2 by the center and automated methods, respectively. Even though the between-method deviation in the ECD decreased with increasing ECD-Ce, the between-method difference was still significant in Group 7 (P < 0.001).
Discussion
Few studies have reported differences in the endothelial parameters of glaucomatous eyes between the center and automated methods. Several studies have reported that the fully automated analysis significantly overestimates ECD values in individuals with low ECD [5,6,7]. Huang et al. compared corneal endothelial parameters between both methods in 106 glaucomatous eyes [5]. However, they did not classify the subjects into smaller ECD-based groups. To the best of our knowledge, there has been no study on the specific ECD level where a significant between-method difference emerges.
The difference between ECD-Ce and ECD-Au is 0 when ECD-Ce = 2593. However, there is a variation in data, and the compatibility between ECD-Ce and ECD-Au is slightly wider and is held if ECD-Ce is 2500 or higher (Figs. 3 and 4). When ECD-Ce was less than 2500 cells/mm2, we found significant between-method differences (Table 2, Fig. 3). Therefore, an ECD-Ce of 2500 cells/mm2 appears to be the critical ECD level for scientific evaluation using the automated method.
ECD overestimation in damaged corneas may be clinically problematic. Assuming a 0–30% cell loss after intraocular surgery, patients should have an ECD of at least 1000–1200 cells/mm2 for safety purposes to reduce the risk of postoperative corneal edema [2, 10]. The true ECD of eyes with pre-surgical ECD-Au of 1500 cells/mm2, which is estimated by the regression equation (Fig. 6a), is close to 782 cells/mm2. This finding suggests that the eyes with ECD of 1500 by the automated method have a high risk of corneal decompensation and may not tolerate glaucoma surgery.
ECD is an important parameter for assessing endothelial function and health. However, there is a need to assess changes in other parameters. Theoretically, polymegathism determined by CV and pleomorphism (cell shape variations represented by HEX) have been suggested as a more sensitive measure of the endothelium under stress [2]. Corneas with a CV > 40% or the presence of < 50% HEX should be considered abnormal and have an increased risk of postoperative edema [2].
In this study, there were further concerns regarding the reliability of CV values obtained using the automated method since they exceeded 40% in all 10 groups. Moreover, these values were higher than those measured using the center method, especially with low ECD values. Furthermore, HEX values obtained using the automated method were < 50% in all 10 groups, which is considered abnormal. In our study, the between-method deviation in the HEX exceeded 14% in the subgroup eyes with ECD > 2000 cells/mm2.
As shown in Image A (Fig. 9), the automated method often misrecognized intracellular structures or endothelial cell nuclei as cellular borders and misidentified large cells as multiple small cells. This caused an overestimation of ECD and CV by the automated method in the low ECD group. This kind of CV overestimation by the automated method tends to occur in individuals with low ECD with damage-induced polymegathism.
Concerning the system of fully automated methods, there is another problem that needs to be addressed. The previous models of the Konan specular microscope had the option of selecting patterns of cell size (S, M, L and XL). However, recent models, including the machine we have used in this study, have no such option and have a default setting of S cell size. As described in previous studies [4, 5], the setting of S cell size is not appropriate in corneas with low ECDs, because this would cause misidentifying the cells as being smaller than they really were and overestimating the cell density. This may be caused by the inherent assumption of the automated method that all endothelial cells in the frame are of uniform cell size. When polymegathism is present, this assumption may lead to incorrect counting. Price et al. reported that even with appropriate use of the sizing feature, the Konan Robo automated software significantly overestimated ECD in the Descemet stripping endothelial keratoplasty eyes [6]. However, in the latest model of the Konan specular microscope, the automated analysis algorithm has been refurbished and may have higher precision in the analysis results. The precision of the refurbished automated analysis will be a subject for future studies. Recently, several studies have reported precise and accurate results using deep learning automatic segmentations of corneal endothelial cell images [10,11,12,13]. We expect the development of new software with precise automated analysis in the near future.
There have been various recent programs for corneal endothelial cell analysis in specular microscopy, with each having its advantages and disadvantages. The center method has been approved by the FDA for use in clinical trials [1, 3]. However, the center method excludes the outermost digitized cells of a contiguous group, which leads to a decrease in the number of analyzed cells and an increase in CV. The flex-center method allows the inclusion of the outermost cells and increases the NUM (the number of analyzed cells) and may be recommended for cases with small visible cell numbers [4]. However, the flex-center method requires manual boundary tracing of the contiguous cell in addition to cell-center dotting and may be more time consuming than the traditional center method. Doughty et al. studied the appropriate cell count required to obtain accurate ECD, and they recommended a cell count > 75; however, they only assessed the normal cornea (average ECD = 3519 cells/mm2) [14]. Individuals with low ECD may have only a few countable cells. Huang and others reported that in corneas with guttae, the center method yields an accurate ECD value when there are ≥ 30 contiguous countable cells in an image. If the number of countable cells is less than 30, the flex-center method may be more suitable [15], and a minimum of 30 cells may be appropriate to obtain an accurate ECD value using the center method.
In conclusion, the fully automated method for glaucomatous eyes significantly overestimates the cell count in eyes with low ECD. An ECD of 2500 cells/mm2 is a critical cell count where overestimation occurs in the automated method. The ECD of 1500 using the automated method does not warrant that the eye can tolerate glaucoma surgery.
Limitation of this study
Although we carefully analyzed images and studied intra-individual deviations, the center method could vary across examiners. We did not determine inter-individual variations, and it is a subject for future studies. This study was conducted using only one type of non-contact specular microscope; thus, the results may not apply to other machines.
Data availability
Available on request.
References
Janson BJ, Alward WL, Kwon YH et al (2018) Glaucoma-associated corneal endothelial cell damage: a review. Surv Ophthalmol 63:500–506
Mannis MJ, Holland EJ (eds) (2017) Cornea, 4th edn. Rotterdam, Elsevier
McCarey BE, Edelhauser HF, Lynn MJ (2008) Review of corneal endothelial specular microscopy for FDA clinical trials of refractive procedures, surgical devices, and new intraocular drugs and solutions. Cornea 27:1–16
Konan Inc. KONAN MEDICAL [CellCheck Overview Video]. Available at: https://konanmedical.com/cellchek. Accessed February 22, 2020.
Huang J, Maram J, Tepelus TC et al (2018) Comparison of manual & automated analysis methods for corneal endothelial cell density measurements by specular microscopy. J Optom 11:182–191
Price MO, Fairchild KM, Price FW Jr (2013) Comparison of manual and automated endothelial cell density analysis in normal eyes and DSEK eyes. Cornea 32:567–573
Maruoka S, Nakamura S, Matsuo N et al (2018) Comparison of semi-automated center-dot and fully automated endothelial cell analysis from specular microscopy images. Int Ophthalmol 38:2495–2507
The Japan Glaucoma Society guidelines for glaucoma (4th Edition). (2018) J Jpn Ophthalmol soc 122:5–53.
Kinoshita S, Amano S, Inoue K et al (2014) Grading for corneal endothelial damage. J Jpn Ophthalmol soc 118:81–83
Fabijańska A (2018) Segmentation of corneal endothelium images using a U-Net-based convolutional neural network. Artif Intell Med. 88:1–13
Kolluru C, Benetz BA, Joseph N et al (2019) Machine learning for segmenting cells in corneal endothelium images. Proc SPIE Int Soc Opt Eng. https://doi.org/10.1117/12.2513580
Heinzelmann S, Daniel MC, Maier PC et al (2019) Automated cell counting using “deep learning” in donor corneas from organ culture achieves high precision and accuracy. Klin Monbl Augenheilkd 236:1407–1412
Joseph N, Kolluru C, Benetz BAM, et al. (2020) Quantitative and qualitative evaluation of deep learning automatic segmentations of corneal endothelial cell images of reduced image quality obtained following cornea transplant. J Med Imaging (Bellingham) [serial online]. doi: https://doi.org/10.1117/1.JMI.7.1.014503.
Doughty MJ, Muller A, Zaman ML (2000) Assessment of the reliability of human corneal endothelial cell-density estimates using a noncontact specular microscope. Cornea 19:148–158
Huang J, Liu X, Tepelus TC et al (2017) Comparison of the center and flex-center methods of corneal endothelial cell analysis in the presence of guttae. Cornea 36:1514–1520
Acknowledgements
We would like to thank technicians Ms. Morigaki T, Yamasaki S, Hashimoto N, Nishimura R, Suhara C, Kinoshita C, Matsushita A, Oku Y, Furuhashi T, Fujino R, and Mr. Machiba N for their technical assistance in obtaining endothelial images. We have not received any support from other parties to be acknowledged for this work.
Funding
We have not received any fund for this work.
Author information
Authors and Affiliations
Contributions
Mayumi Minami involved in conceptualization; methodology; formal analysis and investigation; writing—original draft preparation; Etsuo Chihara took part in review and editing, and supervision. E Chihara served as corresponding author and is responsible to accuracy and integrity of the work.
Corresponding author
Ethics declarations
Conflict of interest
The authors have no relevant financial or non-financial interests to disclose.
Ethics approval
This manuscript follows the COPE guidelines. This study was approved by the internal review board at Sensho-kai, and it certifies that the study was performed in accordance with the ethical standards as laid down in the 1964 Declaration of Helsinki and its later amendments or comparable ethical standards.
Consent to participate
Informed consent was obtained from all patients participated.
Consent to publication
Images used do not identify any patients.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Minami, M., Chihara, E. Overestimation of corneal endothelial cell density by automated method in glaucomatous eyes with impaired corneal endothelial cells. Int Ophthalmol 42, 133–145 (2022). https://doi.org/10.1007/s10792-021-02008-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10792-021-02008-4