The efficiency of a newly developed single-use blood pump was evaluated in comparison with the Safira single-use blood pump manufactured by Braile Biomedica (Brazil) used in modern clinical practice. The performance of prototype pumps was compared by evaluating the normalized index of hemolysis (NIH) calculated from data obtained during hemolysis tests. These studies showed that the level of hemolysis during perfusion using prototypes at this stage of development was not statistically different from that of single-use pumps used in clinical practice.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Bokeriya, L. A., Milievskaya, E. B., and Skopin, A. I., “Cardiovascular surgery. Cardiopulmonary bypass in the Russian Federation, 2018,” Grud. Serdech. Sosud. Khirurg., 4, No. 61, 255-264 (2019).
Averina, T. B. and Samuilova, D. Sh., “What do cardiologists need to know about cardiopulmonary bypass?,” Kreativ. Kardiol., No. 1-2, 102-118 (2007).
Blackshear, P. L., in: Biomechanics: Its Foundations and Objectives, Y. C. Fung, N. Perrone, and M. Anliker, (ed.), Prentice-Hall (1972).
Yasuda, T., Funakubo, A., and Miyawaki, F., “Influence of static pressure and shear rate on hemolysis of red blood cells,” ASAIO J., No. 47, 351-353 (2001).
Apel, J., Paul, R., and Klaus, S., “Assessment of hemolysis related quantities in a microaxial blood pump by computational fluid dynamics,” Artif. Organs, No. 25, 341-347 (2001).
Khubulava, G. G., Romanovskii, D. Yu., Volkov, A. M., Biryukov, A. V., Skibro, I. R., Butuzov, A. G., Lyubimov, A. I., Sizenko, V. V., and Didenko, A. V., “Effects of cardiopulmonary bypass on the morphological characteristics of erythrocytes and the gas transport function of blood,” Vestn. Ross. Voenno-Med. Akad., 66, No. 2, 7-12 (2019).
Lipunova, E. A. and Skorkina, M. Yu., The Red Blood System: Comparative Physiology. A Monograph [in Russian], Belgorod State University, Belgorod (2004).
ASTM F1841–97 Standard Practice for Assessment of Hemolysis in Continuous Flow Blood Pumps (2005).
Bomba Safira - Safira Centrifugal Pump Code 608568, https://www.Braile.Com.br/en/admin/uploads/arquivos/1588079369_72.pdf (accessed April 29, 2022).
Paul, R., “Recent findings on flow induced blood damage: Critical shear stresses and exposure times obtained with a high shear stress Couette system,” Artif. Organs, No. 23, 680 (1999).
Author information
Authors and Affiliations
Corresponding author
Additional information
Translated from Meditsinskaya Tekhnika, Vol. 56, No. 4, Jul.-Aug., 2022, pp. 23-26.
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Semenikhin, P.V., Skvortsov, A.E., Filipenko, K.V. et al. Assessment of the Efficiency of a Single-Use Blood Pump in Hemolysis Tests. Biomed Eng 56, 251–255 (2022). https://doi.org/10.1007/s10527-022-10213-3
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10527-022-10213-3