Static-pressure head−flow characteristics were studied for six models of rotary blood pumps: Incor (BerlinHeart, Germany), HeartMate II (Thoratec Corp., USA), Sputnik and Sputnik 2 (JSC ZITC, Russia), Heartware HVAD (Heartware International, Inc., USA), and Duraheart (Terumo, Japan). Preload and afterload sensitivities were calculated for each pump model. Preload sensitivity was found to decrease in the maximum blood flow area. It was shown that preload sensitivity depends on the geometry of the pump rotor bundle rather than the type of the pump (centrifugal or axial-flow).
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Moazami N., Fukamachi K., Kobayashi M. et al., “Axial and centrifugal continuous-flow rotary pumps: A translation from pump mechanics to clinical practice,” J. Heart Lung Transplant., 32, No. 1, 1-11 (2013).
Salamonsen R. F., Mason D.G., and Ayre P. J., “Response of rotary blood pumps to changes in preload and afterload at a fixed speed setting are unphysiological when compared with the natural heart,” Artif. Organs, 35, No. 3, 47-53 (2011).
Fukamachi K., Shiose A., Massiello A. et al., “Preload sensitivity in cardiac assist devices,” Ann. Thorac. Surg., 95, No. 1, 373-380 (2013).
Gregory S. D., Stevens M., Timms D., and Pearcy M., “Replication of the Frank-Starling response in a mock circulation loop,” in: IEEE EMBS Conf. Proc., 6825 (2011).
Khalil H. A., Cohn W. E., Metcalfe R. W., and Frazier O. H., “Preload sensitivity of the Jarvik 2000 and HeartMate II left ventricular assist devices,” Am. Soc. Artif. Organs J., 54, No. 3, 245-248 (2008).
Petukhov D. S. and Telyshev D. V., “Analysis of the preload and afterload sensitivity of the Sputnik rotary blood pump,” Biomed. Eng., 49, No. 6, 362-365 (2015).
Camboni D., Zerdzitzki M., Hirt S. et al., “Reduction of INCOR® driveline infection rate with silicone at the driveline exit site,” Interact. Cardiovasc. Thorac. Surge. (2016) (in press).
Frazier O. H., Gemmato C., Myers T. J. et al., “Initial clinical experience with the HeartMate® II Axial-Flow left ventricular assist device,” Tex. Heart Inst. J., 34, 275-281 (2007).
Selishchev S. and Telyshev D., “Ventricular assist device Sputnik: Description, technical features and characteristics,” Trends Biomat. Artif. Organs, 29, No. 3, 207-210 (2015).
Selishchev S. V. and Telyshev D. V., “Optimisation of the Sputnik-VAD design,” Int. J. Artif. Organs, 39, No. 8, 407-414 (2016).
Larose J. A., Tamez D., Ashenuga M., and Reyes C., “Design Concepts and Principle of Operation of the HeartWare Ventricular Assist System,” Am. Soc. Artif. Organs J., 56, No. 4, 285-289 (2010).
Nojiri C., Kijima T., Maekawa J. et al., “Recent progress in the development of Terumo implantable left ventricular assist system,” Am. Soc. Artif. Organs J., 45, 199-203 (1999).
Salamonsen R. F., Mason D. G., and Ayre P. J., “Response of rotary blood pumps to changes in preload and afterload at a fixed speed setting are unphysiological when compared with the natural heart,” Artif. Organs., 35, No. 3, 47-53 (2011).
Fukamachi K., Shiose A., Massiello A. et al., “Preload sensitivity in cardiac assist devices,” Ann. Thorac. Surg., 95, No. 1, 373-380 (2013).
Moazami N., Fukamachi K., Kobayashi M. et al., “Axial and centrifugal continuous-flow rotary pumps: A translation from pump mechanics to clinical practice,” J. Heart Lung Transplant., 32, No. 1, 1-11 (2013).
Author information
Authors and Affiliations
Corresponding author
Additional information
Translated from Meditsinskaya Tekhnika, Vol. 50, No. 6, Nov.-Dec., 2016, pp. 49-52.
Rights and permissions
About this article
Cite this article
Telyshev, D.V. Load Sensitivity of Rotary Blood Pumps under Static Pressure Conditions. Biomed Eng 50, 425–428 (2017). https://doi.org/10.1007/s10527-017-9670-2
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10527-017-9670-2