The design of a KGK-2 γ-ray dose rate detector is described. The basic characteristics of the detector are presented. The readings of a KGK-2 gas-filled ionization chamber detector and VED-2 secondary-electron emission detector during power startup of BR-K1 and BR-1M reactors are compared. It is shown that by using KGK-2 and VED-2 detectors in combination the γ-ray dose rate can be measured in the interval 1 μGy/sec to 10 GGy/sec with the linear sections of the regulation curves overlapping to at least four decimal orders.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
V. F. Kolesov, Electronuclear Installations and Problems of Atomic Energy, RFYaTS – VNIIEF, Sarov (2013).
S. V. Chuklyaev, Yu. M. Odintsov, Yu. N. Pepelyshev, and A. S. Koshelev, “Secondary-emission detector VED-2 for measuring γ-radiation dose rate,” At. Enrg., 75, No. 4, 304–312 (1993).
A. S. Koshelev, L. E. Dovbysh, V. Kh. Khoruzhii, and S. V. Chuklyaev, “KGK-2 type gamma radiation power detector for diagnosing radiation fields of nuclear reactors in the range from 1 μGy/sec to 100 Gy/sec,” Vopr. At. Nauki Tekhn. Ser. Fiz. Yad. Reakt., No. 3, 43–49 (2012).
S. V. Chuklyaev, M. Ya. Grudskii, and V. A. Artem’ev, Secondary-Emission Detectors of Ionizing Radiation, Energoatomizdat, Moscow (1995).
V. V. Generalova, A. P. Zhanzhora, A. S. Koshelev, and A. T. Narozhnii, “A specialized source of gamma radiation spatially combined with a modeling neutron reference field MOS-K2,” Izmer. Tekhn, No. 1, 24–27 (2010).
A. S. Koshelev, V. Kh. Khoruzhii, M. A. Ovchinnikov, and G. N. Pikulina, “Possibilities of and prospects for the development of a comparator option for calibrating gamma radiation power detectors in the field of a specialized gamma source of the BR-K1 reactor,” Vopr. At. Nauki Tekh. Ser. Fiz. Yad. Reakt., No. 3, 10–18 (2011).
V. F. Kolesov, M. I. Kuvshinov, S. V. Vorontsov, et al., “Critical stands and pulsed reactors RFYaTs – VNIIEF,” in: 65th Anniv. VNIIEF. Physics and Technology of High Energy Densities, RFYaTs – VNIIEF, Sarov (2011), 2nd ed., Iss. 1, pp. 136–164.
B. A. Briskman, V. V. Generalova, E. A. Kramer-Ageev, and V. S. Troshin, In-Reactor Dosimetry: Practical Guide, Energoatomizdat, Moscow (1985).
S. V. Chuklyaev, Patent No. 2683786 RF, “Detector block for measuring photon radiation,” subm. 03.21.17, publ. 04.02.19, Byull. Izobret. Polezn. Modeli, No. 10 (2019).
Author information
Authors and Affiliations
Additional information
Translated from Atomnaya Énergiya, Vol. 128, No. 3, pp. 154–158, March, 2020.
Rights and permissions
About this article
Cite this article
Chuklyaev, S.V., Koshelev, A.S. Gas-Filled Ionization Chamber and Secondary-Emission Detector for Measuring γ-Ray Dose Rate. At Energy 128, 166–170 (2020). https://doi.org/10.1007/s10512-020-00669-5
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10512-020-00669-5