Abstract
The widely applied reversed phase high-performance liquid chromatography (RP-HPLC) is an indispensable purification technique in drug discovery. During drug discovery, recovery was usually calculated based on the weight of the purified product after drying over the weight of the crude material multiplied by the assumed purity from HPLC/UV area percent of the product. Such a purity assumption can be off significantly when the crude material contains water, solvents, other UV-inactive impurities and inorganic salts. In this paper, we report a simple and efficient way to estimate recovery of preparative HPLC purification process. It is based on the ratio of the HPLC/UV peak area measured for the product in the crude solution and that in the final collected fraction with both accounted for their volumes. This approach eliminates not only the need for drying of the collected fraction to calculate recovery but also the inaccuracy associated with the true content in the crude sample using the traditional method. A systematic study was conducted to verify this method using caffeine mixed with various UV-active and -inactive impurities. The calculated recoveries using this approach were found to be consistent within 4% with the true recoveries based on dry weight estimation. The approach has been successfully applied for our in-house purifications. Furthermore, the approach was extended to library purifications, where in many cases heart-cutting the desired peaks is used to meet the purity requirements.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Reversed phase high-performance liquid chromatography (RP-HPLC) is a preferred purification technique in drug discovery. Although the value of HPLC purification is well acknowledged [1] and high-throughput purification is increasingly reported [2, 3], there is little discussion on fast recovery assessment during the HPLC purification in the literature. A reliable recovery assessment of HPLC purification process is important for developing and optimizing synthesis and purification of often precious target drug molecules. One of the earlier reported methods for recovery measurement makes use of commercially available drugs or drug-like compounds as standards and the estimated recovery is calculated by comparing the weight of the compound of interest before and after purification [4]. Although this can confirm the instrument and the method performance, the question still remains unanswered about the purification of a real sample where many other factors have to be considered. Usually, the weight of the compound of interest before purification is estimated by multiplying the purity of the component with the weight of the crude material. Thus, accurate recovery determination depends on the accurate assessment of the purity of compound of interest in the crude material. During drug discovery, since reference standard of a target drug molecule is rarely available, HPLC/UV peak area % is used for purity assessment of the crude material and final product. When crude material contains UV-inactive components such as inorganic salts, moisture and solvents, the purity of the crude could be off significantly. As a result, purification recovery could be underestimated and becomes a point of concern. To the best of our knowledge, there is no reported method for recovery assessment which is independent of the weight and purity of the crude material.
Zhang et al. reported mass spectroscopy (MS) directed automated HPLC purification [4, 5], which enhanced the reliability of fraction collection and automation. Nowadays, preparative HPLC/MS instruments have been widely adopted for purifying library compounds. However, MS can selectively detect the target component even with co eluting peaks, HPLC/MS does not help in purity and recovery determination.
There are a few methods reported for the use of HPLC with a chemiluminescent nitrogen detector (CLND) or ELSD as a means of estimating compound concentration in library purifications [6–10]. These methods were used as quality control check for the assessment of the purity and concentration of compound in the DMSO solutions for high-throughput screening after the purification. Although these methods are viable, these detectors are not as trouble-free as UV detectors especially for crude samples. CLND also has the limitations of using nitrogen-free mobile phase. ELSD can not respond well towards volatile and semi-volatile compounds and has limitation of non-linear response curve with gradient change in the mobile phase composition. Furthermore, none of these detectors offer a true universal response for a wide variety of compounds for purity determination. In order to determine recovery quickly for HPLC preparative purifications, we developed a simple and efficient HPLC/UV approach which is independent of the weight and true purity of the crude or the dry weight of the purified compound. At the same time, this approach will be independent of the presence of impurities with much higher or lower UV response in the crude sample relative to the product. The approach involves the HPLC injections of crude sample solution as well as the collected fraction after the purification. By comparing the peak area of the compound of interest accounted for the volumes of the crude sample solution and post-preparative HPLC fraction, one can calculate the amount in the crude sample versus the amount in the fraction solution which reflects the recovery of the HPLC purification process. Of course, the approach does not cover any loss or decomposition that might occur in the final drying. Therefore, characterization of the final dried compound is required especially when loss or decomposition is suspected. Herein, we demonstrate our approach using caffeine mixed with different amount of UV-inactive impurities, impurity with high UV response, water, solvent and inorganic salts. HPLC purification recovery by this method was compared with the true recovery calculated based on dry weights. This process has been implemented for the routine recovery assessment for all the compounds which are purified in our drug discovery facility.
Materials and Methods
Reagents and Chemicals
HPLC grade acetonitrile was purchased from Rankem (Avantor Performance Materials ltd, Ahmedabad, India). Water was purified and delivered by Millipore’s Milli-Q system (Bedford, MA, USA).
Sodium sulphate was purchased from Spectrochem (Milwaukee, WI, USA). HPLC-grade methylsulfoxide (DMSO), 6-amino benzothiazole and naphthalene were purchased from Sigma Chemical Co. (St. Louis, MO, USA). Caffeine was purchased from Aldrich (St. Louis, MO, USA). In-house compounds used for the recovery study were produced by the chemistry group.
HPLC and MS Instrumentation
The purification system consisted of an Agilent (Santa Clara, CA, USA)-1200 series (German-1361A) preparative HPLC system, which included two solvent delivery pumps (G2260A), one DAD detector (G1315D) and one system controller (G1364D) to control the HPLC gradient. Sample injection and collection were controlled by Chemstation software.
Columns, Mobile Phases and Preparative HPLC Conditions
All preparative HPLC separations were carried out using Sunfire OBD C18, 30 mm × 250 mm, 5µ (waters) column with analytical method developments on Sunfire OBD C18, 4.6 mm × 250 mm, 5µ (waters) column. Milli-Q water and acetonitrile were used as mobile phase for all separations and purifications. Diluents for all sample preparations were 80:20 acetonitrile:water (V/V). The gradient used for analysis and preparative purifications are as shown in Table 1.
Sample Preparation
Following three compounds were used for the preparation of test mixture 2.
Eleven test mixtures were prepared for the recovery study to mimic the actual sample components which may interfere with recovery assessment in purification:
Mixture 1 97.5 mg of caffeine was dissolved in 1.5 mL of diluent.
Mixture 2 52.5, 53.6 and 52.3 mg of caffeine, 6-amino benzothiazole and naphthalene, respectively, were dissolved in 1.1 mL of diluent.
Mixture 3 98.5 mg of caffeine and 60.2 mg of Na2SO4 was dissolved in 4 mL of diluent.
Mixture 4 101.2 mg of caffeine and 50.1 mg of DMSO were dissolved in 1.5 mL of diluent.
Mixture 5 300.5 mg of caffeine was dissolved in 2 mL of ACN and 3 mL of water, and the sample was transferred to a 100 mL round bottom flask and the solvents were evaporated in such a way that the compound did not dry completely. 110.3 mg of this sample was dissolved in 1.2 mL diluent.
Mixture 6 50.2 mg of caffeine and 54.5 mg N-boc-trans-4-hydroxy-l-prolin methyl ester (UV inactive) were dissolved in 1.1 mL of diluent.
Mixture 7–11 30.9 mg ammonium chloride, 34.6 mg potassium chloride, 31.0 mg sodium chloride, 31.4 mg sodium bicarbonate, and 38.6 mg calcium chloride were mixed with caffeine separately in approximately 1:0.3 ratios (wt/wt) and dissolved in 1 mL diluents.
Mixture 12 10.2 mg of caffeine and 11.6 mg of an impurity with high UV response were dissolved in 1.1 mL of diluent. All mixture solutions were injected completely on the prep column in 1–5 injections (depending on the solution volume) with an injection volume of 1.0 mL. The peak corresponding to caffeine was collected and evaporated by lypholization to get the final solid which was weighed for actual wt/wt recovery. Crude sample solution and the final fraction (combined fraction from multiple injections) were injected on an analytical HPLC to get the peak area for prep recovery calculation. Crude samples were diluted ‘n’ folds as appropriate before injection to avoid the UV detector saturation (peak height <800 mAU) in the analysis, and n has been incorporated as a dilution factor in the recovery calculation formulae.
Results and Discussion
We have developed a simple method to assess HPLC purification recovery for each sample without the need to determine the exact weights of the component of interest before and after the purification. The recovery is calculated based on the peak areas of the component of interest accounting for the volumes in the crude sample solution and in the collected fraction (all fractions were pooled together at the completion of the purification and recovery estimation was done from this pooled solution.) using the following formulae:
where A f is the peak area in the final collected fraction; V f is the fraction volume; Injc is the injection volume of the crude sample; A c is the peak area in crude; V c is the crude volume; Injf is the injection volume of the fraction; n is the dilution factor of the crude sample solution for HPLC injection.
Based on the equation, the factors contributing to the accuracy of the recovery result include accuracy of the volume measurements (crude and fraction volume), performance of the analytical HPLC used for the analysis and proper dilutions to ensure the concentration of crude and collected samples are within the detector linear range. Properly controlling these experimental factors can result in accurate recovery estimation to the nearest 1%. Usually recovery in the actual sample is affected due to the presence of following components- UV active impurities, inorganic impurities, water, solvent and UV inactive impurities. To demonstrate our approach, we prepared eleven samples of caffeine mixed with various amount of UV-active and UV-inactive components including water, DMSO and salts. The mixtures were subjected to prep-HPLC purification and the caffeine peak was collected. The final collected fractions as well as the crude mixtures (after dilution) were injected on an analytical HPLC. The recovery was calculated for the HPLC purification of each mixture sample using the proposed formulae. The results are listed in Table 2. Representative chromatograms for mixture 2 are shown in Figs. 1, 2, 3. Figure 4 shows the caffeine mixture with an impurity of high UV response. Actual recovery (wt/wt) was also calculated for comparison using the following formulae:
Data in Table 2 clearly shows that the traditional approach yielded significantly larger error (from 4.65 to 48.98%) in the recovery assessment than the new proposed method (from 0.01 to 3.74%) especially for the purifications of the samples containing UV inactive impurities. This confirms that the proposed HPLC area-based recovery calculation method is not affected by the presence of these UV-active or inactive components in the crude samples. If any UV inactive impurity is co eluting with the peak of interest the present recovery estimation may still give the right recovery as it is solely based on the HPLC peak area but the recovery based on final dried weight of the compound may differ due to addition of this impurity in final weight. That may be verified by further method of development to separate the undesired impurity or with the help of other techniques like NMR or MS to cross check the purity of the sample.
In case of close eluting peaks with difficult separations during the library purifications, it is a normal practice to heart-cut the fractions by removing the peak fronting or tailing which leads to a lower recovery. To check the applicability of this method for heart-cut fraction purifications, we collected the caffeine peak in three fractions and checked the recovery of each fraction separately using the proposed area-based method and wt/wt method. Figure 5 shows the chromatogram of caffeine with heart-cut fraction collection.
As shown in Table 3, the separate recovery from each fraction as well as overall recovery including all fractions is very close to the dry weight recovery assessments. This also confirms that around 75% material could be recovered if only the middle fraction is taken for high purity requirement which could be helpful to make a decision for the collection of fractions.
This area-based approach requires the measurement of the fraction volume for the recovery calculation which would add additional time and efforts for each sample. To further simplify the procedure, the fraction volume could also be calculated directly from the chromatogram using the following formulae:
where Rtend is the retention time at the end of peak collection, and Rtstart is the retention time at the start of peak collection.
However, recovery calculation using this formula may not be accurate if there is any instrumental error in terms of the retention volume during multiple injections.
Conclusion
We report a simple and efficient approach to assess the recovery of a prep-HPLC purification based on the peak area of the component of interest in the crude and in the collected fraction solutions. This approach does not require the determination of weight and % purity of the crude sample which is often inaccurate especially when the sample contains UV-inactive impurities. The method was demonstrated to be reliable using caffeine mixed with various UV-active and inactive components. This process has been implemented for the routine recovery assessment for all the compounds which are purified in our drug discovery facility. In addition, it also worked well for the assessment of recovery for different fractions of the peak collected so it can be used successfully for library purifications as well.
References
Yan B, Fang L, Irving M, Zhang S, Boldi AM et al (2003) Quality control in combinatorial chemistry: determination of the quantity, purity, and quantitative purity of compounds in combinatorial libraries. J Comb Chem 5:547–559
Leister W, Strauss K, Wisnoski D, Zhao Z, Lindsley C (2003) Development of a custom high-throughput preparative liquid chromatography/mass spectrometer platform for the preparative purification and analytical analysis of compound libraries. J Comb Chem 5:322–329
Isbell J, Xu R, Cai Z, Kassel DB (2002) Realities of high-throughput liquid chromatography/mass spectrometry purification of large combinatorial libraries: a report on overall sample throughput using parallel purification. J Comb Chem 4:600–611
Zhang X, Picariello W, Hosein N, Towle M, Goetzinger W (2006) A systematic investigation of recovery in preparative reverse phase high performance liquid chromatography/mass spectrometry. J Chromatogr A 1119:147–155
Goetzinger W, Zhang X, Bi G, Towle M, Cherrak D et al (2004) High throughput HPLC/MS purification in support of drug discovery. Int J Mass Spectrom 238:153–162
Shah N, Gao M, Tsutsui K, Lu A, Davis J et al (2000) A novel approach to high-throughput quality control of parallel synthesis libraries. J Comb Chem 2:453–460
Popa-Bruke LG, Issakova O, Arroway JD, Bernasconi P, Chen M et al (2004) Streamlined system for purifying and quantifying a diverse library of compounds and the effect of compound concentration measurements on the accurate interpretation of biological assay results. Anal Chem 76:7278–7287
Corens D, Carpentier M, Schroven M, Meerpoel L (2004) Liquid chromatography–mass spectrometry with chemiluminescent nitrogen detection for on-line quantitative analysis of compound collections: advantages and limitations. J Chromatogr A 1056:67–75
Mathews BT, Higginson PD, Lyons R, Mitchell JC, Sach NW et al (2004) Improving quantitative measurements for the evaporative light scattering detector. Chromatographia 60:625–633
Lane S, Boughtflower B, Mutton I, Paterson C, Farrant D et al (2005) Toward single-calibrant quantification in HPLC. A comparison of three detection trategies: evaporative light scattering, chemiluminescent. nitrogen, and proton NMR. Anal Chem 77:4354–4365
Acknowledgements
We would like to thank Ramakanth Sarabu for his encouraging support to do this work. We would also thank DAS purification team members, Nancharaiah, Vijayalakshmi Sathiyamoorthy, Vinay Kempegowda and Sudam Mane for their support in optimizing this procedure by sharing their valuable feedbacks. Our sincere thanks to Syngene Intl Ltd. for providing all resources to accomplish this work.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Rights and permissions
About this article
Cite this article
Bajpai, L., Asokan, K., Samy, S. et al. A Simple and Efficient Approach for Estimating Recovery of a Preparative Reversed Phase HPLC Purification. Chromatographia 79, 1577–1583 (2016). https://doi.org/10.1007/s10337-016-3191-0
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10337-016-3191-0