Introduction

The brachial plexus is a network of nerves that extend from the cervical spine to the distal upper limb to provide motor and sensory innervation [1]. Nerve lesions caused by trauma, tumors, inflammation, radiotherapy, or thoracic outlet syndrome often lead to brachial plexopathy, a clinical condition involving upper limb sensory disorders and movement impairments. To characterize the cause and precise location of brachial plexus lesion, several magnetic resonance neurography (MRN) techniques have been applied in the diagnosis of brachial plexopathy [1,2,3,4,5]. Among them, 3D T2-weighted short inversion time inversion recovery (3D-T2-STIR) sequence is the most commonly used [6,7,8,9]. With multiplanar and curved planar reconstruction combined with maximum intensity projection (MIP), it can cover the long-extension range of brachial plexus and visualize the entire nerve course from its spinal origin to terminal segments [1, 5]. However, there is one limitation for the conventional 3D-T2-STIR MRN. In addition to nerve fibers showing long-T1 and long-T2 signal characteristics, neighboring tissues such as venules, lymphatic vessels and nodes exhibit similar MR signals, which interfere with the display of brachial plexus and affect diagnosis accuracy. In recent years, some new technologies, such as multi-echo 3D turbo spin-echo modified Dixon (mDixon) sequence and mildly diffusion-weighted sequence, have been developed for MRN to suppress vascular signals surrounding the nerves [10,11,12], but these methods are more suitable for 3 T systems. Currently, 1.5 T MR scanners are still the most commonly used systems worldwide. Due to the difference in hardware/software, 1.5 T MR scanners are less equipped with the aforementioned sequences. Therefore, for MRN imaging obtained at 1.5 T, reducing the interference from the neighboring tissues to improve the visualization of nerves is an important problem needing research attention. It has recently been reported that applying a contrast agent in 3D-T2-STIR MRN can improve the visibility of peripheral nerves [13,14,15].

On the other hand, the distribution of brachial plexus is spatially proximate to that of the blood vessel system. The nerves and blood vessels usually accompany each other as they travel through other tissues. Such a close apposition between the brachial plexus and vascular system makes blood vessel dysfunction, one of the major concomitant diseases of brachial plexus neuropathy [16,17,18]. It has been noted that brachial plexopathy can be associated with arterial/venous injury such as blood vessel rupture, occlusion and compression [19, 20]. Therefore, it is necessary to examine the brachial plexus and its adjacent arteries/veins simultaneously for a comprehensive diagnosis of the disease. In this regard, the conventional 3D-T2-STIR MRN is not able to visualize the neighboring vessels for assessing their involvement in brachial plexopathy.

Based on the above two considerations, we hypothesize that it would be beneficial to use a single injection of contrast agent to carry out both contrast-enhanced 3D-T2-STIR MRN, for better visualization of brachial plexus, and contrast-enhanced (CE) MR angiography (MRA), to reveal surrounding vessels simultaneously. This paper summarizes our studies over the past 3 years and is intended to present a systematic assessment of whether such a combined imaging approach can significantly improve the evaluation of brachial plexopathy with concomitant vascular diseases.

Materials and methods

All patients were informed about imaging procedure details, who signed a written informed consent before the examination. This study was approved by the Human Clinical Research and Ethics Review Board of The First People’s Hospital of Zhangjiagang City.

Subjects

This prospective study includes 27 patients with suspected brachial plexopathy (10 men [mean age, 42.3 years; range, 18–69 years] and 17 women [mean age, 45.9 years; range, 22–71 years]; overall age: mean, 44.5 years; range, 18–71 years), who were referred for MR imaging evaluation between December 2014 and December 2017. Clinical history indicated that 9 patients suffered from upper limb numbness, paresthesia, and weakness; 16 had neck and shoulder trauma; 2 had masses in the shoulder. Among all the 27 cases, clinical evaluation, which was based on clinical charts, electrophysiological analyses, and surgical findings, confirmed that 18 patients indeed had brachial plexopathy, while the remaining 9 patients did not show clear evidence for a definitive conclusion. These 18 confirmed cases, which comprised 14 cases of trauma, 2 of tumor and 2 of thoracic outlet syndrome, were all unilaterally affected, with 8 patients on the left side and ten on the right side. The clinical data and imaging findings of confirmed cases are summarised in Supplementary Table 1.

MR imaging

All imaging experiments were performed with a 1.5 T Avanto MR scanner (Siemens Medical System, Erlangen, Germany). A multi-channel head-neck joint coil was combined with a body matrix coil to cover the neck, shoulder and arms. MR images were acquired in the following order, with all acquisition parameters listed in Table 1. Step 1: T1-sequence coronal scanning of 3D volumetric interpolated breath-hold examination (VIBE) to cover the entire brachial plexus on both sides. Step 2: coronal scanning of non-enhanced 3D-T2-STIR used a 3D short inversion time inversion recovery imaging sampling perfection with application optimized contrasts using different flip angle evolutions (3D-STIR SPACE) sequence. Step 3: CE-MRA. A contrast agent, gadopentetic acid dimeglumine salt (Magnevist; Bayer Pharma AG, Berlin, Germany) at a dose of 0.15 mmol/kg, was intravenously injected at a rate of 3.0 ml/s. Right after the emerging of the aortic arch through CARE bolus real-time monitoring, a 3D-FLASH coronal sequence was acquired to reveal the arteries located in the neck and shoulder region. The 3D-FLASH sequence was repeated immediately after an arterial phase scan to obtain the venous images. Step 4: Contrast-enhanced 3D-T2-STIR scan was performed as in Step 2. Step 5: Contrast-enhanced 3D-T1-VIBE scan was conducted as in Step 1.

Table 1 Scanning parameters of the MR pulse sequences used in this study

Image post-processing and quantitative analysis

Coronal thin-slab MIP reformation was performed using a GE Advantage Workstation (Version 4.5; GE Healthcare, Buc, France), with a 10 mm slice thickness at 3 mm intervals. To compare between the non-enhanced and contrast-enhanced 3D-T2-STIR images, signal intensities of the brachial plexus C7 root and its surrounding tissues, including the C7 vertebra body, scalene muscle, paraspinal vein, and lymph node, were measured by a radiologic technologist (J.Z., with 6-year experience in image post-processing) in the source images of patients’ normal healthy sides (Fig. 1). These signal intensities were then compared for each tissue by calculating the ratio of post-versus pre-contrast values in percentage. Further, to evaluate the signal interference of various tissues surrounding the brachial plexus, the contrast ratios between the brachial plexus and other tissues were calculated as follows: CR = (Signalnerve – Signalsurrounding tissue) / (Signalnerve + Signalsurrounding tissue) [12, 13].

Fig. 1
figure 1

Representative regions of interest (ROIs) for measuring the signals of the brachial plexus. (1) C7 root, (2) scalene muscle, (3) deep lateral cervical lymph node, (4) paraspinal vein and (5) C7 vertebra body

Quality evaluation of MR images

To compare the image diagnostic quality of 3D-T2-STIR scans, the 2 groups of non-enhanced and contrast-enhanced images, including MIP and source formats, were mixed, yielding a random sample pool of 54 image sequences derived from 27 patients. These image sequences were then evaluated by two experienced radiologists (J.C. and T.Z., with 30 and 20 years of experience) in a blind and independent manner, without knowing the imaging acquisition method, electrophysiological finding, previous diagnosis conclusion as well as each other’s rating score.

To rate the image diagnostic quality, a three-point grading scale was applied according to the following criteria: 0-point stands for poor nerve visualization, serious interference from surrounding tissues and non-diagnostic; 1-point for moderate nerve visualization, some interference from surrounding tissues but diagnostic; 2-point for excellent nerve visualization, little interference from other tissues and completely diagnostic [10].

Combining contrast-enhanced MRN with MRA

Previous clinical studies have revealed that a significant proportion of brachial plexopathy cases involving concurrent blood-vessel disorders [18, 20,21,22], while the diagnostic value of MRN for vascular lesions is relatively limited. This prompted us to review and assess the source and MIP images of CE-MRA for all patients thereby evaluating the concomitant vascular diseases. We merged 3D images of MRA with MRN images using the 3D SynchroView software (GE Healthcare, Buc, France) on the AW4.5 workstation. The software automatically performed the coregistration of MRN and MRA images according to the spatial voxel positioning, then jointly demonstrated the brachial plexus and vascular images. Brachial plexopathy and relevant vascular lesions were observed from different views to reveal their spatial relationship.

Statistical evaluation

Statistical analysis was performed using SPSS 21.0 software (IBM, Chicago, IL, USA). A paired t-test was used to compare various tissues’ signal intensities and contrast ratios. Analysis of variance of randomized block design was used to compare the percentage of signals acquired with contrast agent versus without. In case a significant difference was found, a paired t-test with Bonferroni correction was performed for multiple testing. Wilcoxon signed-rank test was applied to compare the three-point grading scores of the non-enhanced and contrast-enhanced images. P < 0.05 was considered statistically significant. Kappa inter-observer variability was used to measure the degree of agreement between the two radiologists. Kappa ≤ 0.4, 0.4 < Kappa < 0.8, 0.8 ≤ Kappa were considered, respectively, as poor, good and excellent agreement.

Results

Quantitative analyses of the effects of contrast enhancement

Both contrast-enhanced 3D-T2-STIR MRN and contrast-enhanced MRA were successfully performed in all subjects with a single injection of contrast agent. The brachial plexus was usually not clearly visible in the pre-contrast images, excluding the possibility of tracing the entire nerve course (Fig. 2a). This was largely due to strong background-tissue signals, which severely interfered with the display of brachial plexus. Applying the contrast agent, however, suppressed such background and improved the visibility of brachial plexus dramatically (Fig. 2b).

Fig. 2
figure 2

The effects of applying the contrast agent on 3D-T2-STIR display of the brachial plexus. a The pre-contrast volume rendering (VR) image showed poor visualization of brachial plexus. b Applying the contrast agent suppressed the signals of the background-tissue, a clear display of brachial plexus was obtained in the post-contrast VR image. This 29-year-old female appeared to have a normal brachial plexus

In the pre-contrast images, the signal intensities for brachial plexus, scalene muscle, lymph node, small vein, and vertebra body were 92.0 ± 20.3, 49.6 ± 8.8, 141.0 ± 27.9, 148.4 ± 39.4, 67.3 ± 16.9, respectively (Fig. 3a, black columns). Applying the contrast agent suppressed the signal levels of all the five tissues (Fig. 3a, grey columns), with their signal intensities reduced to 84.8 ± 18.7, 29.6 ± 6.7, 39.4 ± 16.8, 26.1 ± 10.1, 32.5 ± 9.3, respectively. Comparing these pre- and post-contrast signal intensities revealed statistically significant differences, with P ˂ 0.001 for all the five tissues. The signals of small veins and lymph nodes were found to be suppressed the most dramatically, with only 15.3% and 26.7% of their pre-contrast signal remained (Fig. 3b). The vertebra and muscle retained 46.7% and 60.5% of pre-contrast signals. On the other hand, the suppressing effect on the brachial plexus was quite small, with 92.9% of signal remained. Therefore, the signal suppression of brachial plexus was significantly lower than those of the other four tissues (P < 0.001). As a result, introducing the contrast agent significantly enhanced the contrast ratios between the brachial plexus and its four surrounding tissues (Fig. 3c). The enhancement was from 0.29 ± 0.10 to 0.48 ± 0.11 for scalene muscle, from − 0.21 ± 0.13 to 0.39 ± 0.15 for lymph node, from − 0.22 ± 0.15 to 0.53 ± 0.14 for small vein, and from 0.15 ± 0.16 to 0.44 ± 0.14 for vertebra (P < 0.001 for all four comparisons). It should be noted that the nerve-to-vein and nerve-to-lymph contrast ratios exhibited an even more dramatic increase, as revealed by the much steeper slope between each patient’s pre- and post-contrast data points in Fig. 3c.

Fig. 3
figure 3

Quantitative analysis of 3D-T2-STIR images acquired with and without contrast agent. a Signal intensity levels of the five tissues before and after injection of the contrast agent, with P < 0.001 for all five pairs of comparison. b Percentage of signals with versus without contrast agent. The contrast enhancement-induced suppression of the brachial plexus was much less obvious than the suppression of its four surrounding tissues (P < 0.001), with the strongest suppression occurring to small vein and lymph node. c Changes of contrast ratios between the brachial plexus and the other four tissues in 27 patients. P < 0.001 for all four pairs of pre- versus post-contrast comparison

Image diagnostic quality evaluation

The results of the image diagnostic quality analysis for all patients are shown in Table 2. The diagnostic quality scores in images with contrast agent (the median of the score was 2 in both readers) were significantly (P < 0.01) higher than those in images without contrast agent (the median of the score was 1 in both readers), as reported by both readers. By introducing a contrast agent, the image diagnostic quality scores from both radiologists significantly improved. For inter-observer agreement, Kappa values for images with and without contrast agent were 0.514 and 0.465, respectively. Applying the contrast agent revealed brachial plexus and its lesions more clearly in the 3D-T2-STIR MRN (Figs. 4 and 5).

Table 2 Assessment of image diagnostic quality by two independent radiologists
Fig. 4
figure 4

A 44-year-old male with a right-neck trauma 10 days before imaging. a The swollen right-side scalene muscle (red arrowheads), small veins (blue arrowheads), displayed high signals similar to the nerves in the pre-contrast source image. b The high signals from the scalene muscle and veins were completely suppressed in the post-contrast source image, and the middle trunk of the right brachial plexus (pink arrow) could be distinguished clearly. c In the pre-contrast MIP image, the strong background signals obscured the display of brachial plexus. d Contrast enhancement suppressed these signals and significantly improved nerve delineation in the post-contrast MIP image, revealing the swelling and inhomogeneity of the upper and middle trunks and cords of the right brachial plexus. The small black spot of brachial plexus root displayed in (b) is the spinal ganglion (yellow arrow)

Fig. 5
figure 5

A 60-year-old female with a history of breast cancer resection suffered from the numbness of her right upper limb and limited mobility for 4 months. a The lesions in the right upper trunk of the brachial plexus were not shown clearly in the pre-contrast source image. b The post-contrast source image displayed a decreased signal in the local area of the upper trunk (red arrowhead). c The pre-contrast MIP image had a poor visualization of the brachial plexus and its lesions. d The post-contrast MIP image, however, demonstrated a filling defect area in the right upper trunk (red arrowhead), which was presumably caused by a damage in the blood-nerve barrier due to tumor invasion. This was confirmed by postoperative pathology revealing that metastatic adenocarcinoma invaded the brachial plexus

Combining contrast-enhanced MRN with MRA

The presence of vascular dysfunction was frequently observed in patients with traumatic brachial plexus injury. Of the 14 traumatic cases we examined, 6 patients showed abnormality either in the subclavian or axillary artery in CE-MRA images. Moreover, brachial plexus-related tumors potentially affected peripheral blood vessels. In one patient, we revealed the subclavian vein invasion and occlusion by metastatic tumors. These visible vascular lesions in CE-MRA images were all clinically confirmed in our study. Combining MRN with MRA provided a more complete picture of the relationship between nerve lesions and vascular diseases (Figs. 6 and 7), which was essential for guiding subsequent treatments.

Fig. 6
figure 6

A 25-year-old male with brachial-plexus nerve root avulsion and subclavian artery damage in the left shoulder. The images were taken 14 days after a motorcycle accident. Contrast-enhanced 3D-T2-STIR neurograph revealed neurotmesis in the C6-T1 roots of the left brachial plexus as well as formation of traumatic pseudomeningocele (as indicated by a green arrowhead in upper panel). There was wide-spread soft tissue edema over the left shoulder. Two red arrows in the contrast-enhanced MR angiograph (middle panel) indicate the damage and occlusion of the left subclavian artery. The display of the downstream axillary artery, which appeared to be much thinner than the axillary artery on the opposite side, was due to contrast agent entering through collateral circulation. The neurograph and angiograph were combined in the lower panel to show their spatial relationship

Fig. 7
figure 7

A 53-year-old male with brachial-plexus neurotmesis who had an axillary artery-bypass surgery about 1 year before imaging. Upper panel, a contrast-enhanced 3D-T2-STIR neurograph demonstrated neurotmesis in the upper, middle and lower trunks of the right brachial plexus, with small neuromata (yellow arrowheads) developed locally in these trunks. Middle panel, a contrast-enhanced MR angiograph revealed that an artery-bypass graft was made to the right axillary artery between the locations as indicated by two red arrows. Lower panel, the neurograph and angiograph were merged to visualize the spatial relationship between the graft vessel and the sites of brachial plexus injury

Overall, of the 18 clinical confirmed cases of brachial plexopathy, 7 patients showed concomitant subclavian and axillary vessel abnormalities, reaching an incidence rate of 39%. This highlights the necessity of combining MRN and MRA with a single injection of the contrast agent for the diagnosis of brachial plexopathy.

Discussion

In this study, we demonstrate that contrast-enhanced images of the brachial plexus with background suppression and images of adjacent vessels can be obtained simultaneously after a single injection of contrast agent at a 1.5 T system. Moreover, brachial plexopathy and vascular injury could be evaluated in parallel.

Compared with 1.5 T, 3 T systems are highly preferred for MRN imaging. A few newer types of sequences have been developed for MRN at 3 T systems recently. Multi-echo 3D turbo spin-echo mDixon sequence, which integrates multi-echo chemical shift acquisitions with a modified 3D TSE flip angle train, provides robust fat and blood suppression [10]. 3D nerve-sheath signal increased with inked rest-tissue rapid acquisition of relaxation imaging (3D SHINKEI) is also a new sequence that suppresses signals from blood vessels, muscles, and fat tissue using improved motion-sensitized driven equilibrium (iMSDE) and spectral attenuated inversion recovery [23, 24]. These novel methods can effectively suppress background-tissue signals, in particular, to reduce the interference of neighboring veins during nerve observation. Therefore, the new 3 T MR scanner is the top-notch equipment for MRN imaging currently.

However, The 1.5 T MR scanners are still the most widely used MR systems worldwide. Due to hardware and software limitations, the sequences mentioned above rarely can be applied to the 1.5 T MR scanner. The conventional MRN yields low image quality on the 1.5 T MR scanner compared with the 3 T, but the 1.5 T MR scanner has the advantages of lower specific absorption rate and fewer artifacts. In this study, we used the 3D-T2-STIR sequence for MRN after intravenous injection of contrast agent to the 1.5 T system. The contrast-enhanced 3D-T2-STIR MRN images provide better contrast ratios between nerve and surrounding tissues compared with non-enhanced images. As an essential quantitative imaging parameter, it was reported that the comparisons based on contrast ratios concur well with subjective visual evaluation, and the increased contrast ratios could improve the diagnostic ability [14]. Our subjective evaluation also confirmed that the quality of post-contrast images was better than that of pre-contrast images. Besides, benefiting from the 3D acquisition, the axial, coronal, sagittal, and oblique images can be reconstructed from source images of the contrast-enhanced 3D-T2-STIR sequence, facilitating the evaluation of brachial plexopathy.

The brachial plexus and some of its surrounding tissues, like small veins, venules, lymphatic vessels, and nodes, are characterized by a long T1 and long T2 relaxation time and thus display high signals in the conventional non-enhanced 3D-T2-STIR images. Small veins and lymph nodes have an even much higher signal level than the brachial plexus, and this explains why the conventional 3D-T2-STIR MRN usually has a strong background that obscures the nerves display. Our direct comparison between the non-enhanced and contrast-enhanced 3D-T2-STIR images reveals that the paramagnetic contrast agent can significantly improve the display of peripheral nerves by suppressing background-tissue signals, which is consistent with previous studies [13,14,15]. Among several factors, the STIR inversion pulse of the 3D-T2-STIR sequence appears to be one major contributor to this improvement. STIR sequence is originally designed for fat suppression by taking advantage of fat tissue’s short T1 relaxation time. When other non-fat background tissues (such as small veins and lymph nodes) absorb the contrast agent, their T1 relaxation time is shortened to be similar to that of fat tissue [25]. With an inversion time of 180 ms used in our 3D-T2-STIR sequence, the STIR pulse can non-selectively suppress the signals from these background tissues [26]. Meanwhile, the contrast agents also shorten the T2 relaxation time of the above tissues [25], while the TE of the 3D-T2-STIR sequence is long, with image characteristics of the T2-weighted sequence; thus, background signals could be further decreased.

The nerve tissue, on the contrary, absorbs little contrast agent [27] and thus has a very limited T1 and T2 shortening effect. Consequently, the brachial plexus can retain relatively high signal of its pre-contrast level, whereas the signals for small veins and lymph nodes are sharply reduced. Such a differential shortening of relaxation time presents a mechanism for increasing the contrast ratios between nerves and surrounding tissues to improve the image’s diagnostic quality. In the case of trauma, the injury-induced swollen muscles also absorb more contrast agent, and its high signal would be suppressed in the post-contrast images to avoid interference to the nerves.

One exception for peripheral nerves is the spinal ganglion, where there is abundant blood circulation. The relatively high absorption of contrast agent turns a ganglion into a dark breaking point along the otherwise continuous bright nerve trajectory [13]. On the other hand, if a contrast-enhanced 3D-T2-STIR image reveals a nerve-darkening area at a non-ganglion location, it is very likely that the nerve at this location is injured. The damaged blood-nerve barrier allows the entry of the contrast agent [27, 28], which induce signals suppression from the injured nerve. This may presumably be called “The Negative Contrast Enhancement Effect”. In this regard, an unusual darkening area along the nerve course can be an indicator of nerve lesion.

The brachial plexus is divided into five parts from proximal to distal end, including the roots, trunks, divisions, cords, and branches. Among them, the division and cord nerves run closely with the subclavian and axillary vessels [29]. Any factors causing brachial plexopathy, such as trauma and tumors, would also likely lead to concurrent blood-vessel disorders [21, 22, 30, 31]. Under such a situation, the diagnosis of the vascular lesion can be easily overlooked by clinical examination due to collateral circulation compensation. However, such a vascular lesion alters blood supply in the affected upper limb, which can have adverse effects on the recovery and regeneration of injured brachial-plexus nerves. Furthermore, if not treated timely, the site of vascular damage can develop pseudoaneurysm, which would compress the brachial plexus and cause additional pathological changes [32]. Therefore, in the case of brachial plexopathy involving trauma, tumor, and thoracic outlet syndrome, it is essential to use CE-MRA to detect the injury, stenosis, occlusion, and compression of adjacent blood vessels [16, 18, 33, 34]. The 3D volume-rendering display of merged vascular and neural images can reveal the anatomical relationship between brachial plexopathy and adjacent vascular lesion, which may provide critical information for selecting appropriate treatment regimens; and guiding surgical operation and determining the prognosis of the disease.

Kappa values represent the degree of agreement between observers. When the Kappa values approach to 1, the agreement between the two observers is considered better. In this study, we obtained Kappa values for images with and without contrast agent as 0.514 and 0.465, respectively. Such Kappa values were considered good rather than excellent, and this may be ascribed to a small overall sample size. When the sample size is small, even a few cases of inconsistency in the scores graded by the two readers decrease the Kappa values significantly. In this study, the number of cases in which two readers graded different scores to the images with and without contrast agent was 4 and 6, respectively. In the cases with different scores, the difference-value between the scores of the two readers was 1 point, and none of the cases had a difference-value of 2 points, implying a relatively small difference in scores. Besides, the gap between the two readers' work experience was 10 years, and the difference in diagnosis experience might explain the discrepancy in scores between them.

One potential drawback for the contrast-enhanced 3D-T2-STIR MRN is related to the neoplastic lesion of brachial plexus. Tumor growth is often associated with angiogenesis, which brings in a rich supply of blood. This may lead to the suppression of tumor signals and thereby affect tumor detection. Therefore, we included the 3D-T1-VIBE sequence in our study to reveal the enhancement of tumor efficiently. Another problem is that 3D fusion of angiograph and neurograph requires a patient to remain in the same body position without movement throughout the entire imaging session. Otherwise, anatomical dislocation and mismatch can occur in the merged image. The five-step imaging procedure employed in this study generally took about 20 min, which can be difficult for those patients experiencing pain. The major time-consuming scans are the two 3D-T2-STIR sequences taken with and without contrast agent, with each scan lasting over 6 min. To shorten the imaging session, the non-enhanced 3D-T2-STIR scan can be omitted, especially for non-tumor patients. Due to the small sample size, we only included confirmed cases of brachial plexopathy with trauma, tumor, and thoracic outlet syndrome. Contrast-enhanced 3D-T2-STIR MRN might be most suitable for these lesions. A limitation of this study is the lack of evaluation for other types of brachial plexopathy, such as radiation-induced changes, idiopathic brachial neuritis, and chronic inflammatory demyelinating polyneuropathy. Therefore, follow-up studies should be directed to these types of lesions. The safety issues associated with gadolinium-based contrast agents such as nephrogenic systemic fibrosis, gadolinium deposition, also require in-depth attention. In the future, we intend to use safer macrocyclic gadolinium-based contrast agents such as gadobutrol instead of Magnevist (a linear gadolinium-based contrast agent) for imaging.

In summary, the contrast-enhanced 3D-T2-STIR sequence can selectively suppress background-tissue signals, and thereby significantly improve an image’s diagnostic quality. Combining contrast-enhanced neurography with angiography at a 1.5 T MR scanner for simultaneous display of the brachial plexus and adjacent vessels is a viable procedure for the diagnosis and assessment of both brachial plexus lesions and any surrounding vascular injury. With further optimization of scan parameters, this combined imaging approach will have broad applications in the clinical examination of peripheral neuropathy.