Introduction

In recent years, the number of patients who need an orthopedic implant has considerably increased. Consequently, the total hip arthroplasty has been developed, becoming one of the most successful surgical procedures. The materials used for orthopedic implants play a very important role in implants fixation. The choice of material strongly influences the rigidity, the corrosion characteristics, the biocompatibility, and the tissue receptivity of the implant [1,2,3,4].

The most used materials in orthopedics applications are surgical grade pure titanium (Ti), Ti alloys, cobalt-chromium alloys, and stainless steels [5,6,7]. Among these alloys, stainless steel has less interesting corrosion and mechanical properties for a permanent implant application [8]. Before the use of titanium, the cobalt-based alloys had largely replaced stainless steel as materials for permanent implants due to their higher wear resistance [6]. These properties justify the choice of the Co28Cr6Mo alloy for the fabrication of the head of the hip implant. Indeed, this part of the implant exhibits high strength and ductility.

Despite the high mechanical performances of the CoCrMo alloy, clinical research has shown the presence of cobalt and chromium ions in the urine of patients with a total hip replacement in CoCrMo. The metal ion release inside the human body during the corrosion process can cause adverse tissue reactions and the premature failure after a total hip arthroplasty [9]. Pains and risks of cancer are also possible for the patient. For these reasons, the corrosion behavior of implanted materials remains a persistent challenge in orthopedic surgery.

However, some studies show that the optimization of material microstructure and mechanical properties by means of surface modification techniques, such as surface mechanical attrition treatment (SMAT) [10], ultrasonic shot peening [11], high energy shot peening (HESP) [12], extrusion [13], and cryogenic burnishing [14], can improve the corrosion behavior of materials [15]. For example, the SMAT introduces localized plastic deformation, allowing the grains refinement without any change of the chemical composition. This mechanical treatment is known to modify the surface of materials by inducing changes of the mechanical properties and of the crystallographic structure.

Until today, the SMAT has been successfully performed on some materials such as 304-316-409 stainless steel [16,17,18], titanium alloy [10, 19], and cobalt-based alloy [20, 21]. The results show that the effect of the SMAT on the corrosion resistance depends on the nature of the alloy and on the mechanical treatment parameters.

In the present investigation, a low carbon CoCrMo alloy was treated by the SMAT. The objective of this work was to investigate the effect of the SMAT on the corrosion behavior of CoCrMo alloy in Ringer’s solution. On the other hand, metallic articular bearing surfaces are systematically polished to produce low coefficients of friction [22]. Therefore, in the present work, the influence of a biomedical polishing was also investigated.

The material surface was characterized by using X-ray diffraction (XRD) and scanning electron microscopy (SEM).

Experimental

Materials

The material used for the present study is a low carbon Co28Cr6Mo alloy (CRITT-MDTS, Charleville-Mézières, France). This material is suitable for medical applications, and its chemical composition complies with the international standard ISO 5832-12 (64.5 wt% Co, 27.6 wt% Cr, 5.5 wt% Mo, 0.8 wt% Mn, 0.6 wt% Si, 0.45 wt% Fe, 0.3 wt% Ni, 0.16 wt% N, 0.05 wt% C, 0.02 wt% W, 0.01 wt% Cu, 0.003 wt% P, and 0.0005 wt% S). For all experiments, the CoCrMo alloy was machined into disc specimen with a 50-mm diameter and 6-mm thickness.

Mechanical treatment and sample preparation

Before each mechanical treatment, the CoCrMo alloy was prepared using silicon carbide (SiC) papers of 800 and 1200 grit followed by an ultrasonication with acetone and high-purity water. Two mechanical treatments were applied: the SMAT and the biomedical polishing.

SMAT

The CoCrMo alloy was subjected to SMAT during 15 min with 3-mm diameter balls in 100Cr6 alloy.

Figure 1 presents the experimental setup of the SMAT process. The parameters controlling the process are the balls’ diameter, the vibration frequency, the processing time, and the vibration amplitude. The vibration was generated by a sonotrode with a vibration frequency of 20 kHz. The distance between the sample and the sonotrode was fixed at 17 mm. When the balls are in resonance, the sample surface is impacted by a large number of flying balls during a short period of time. The flying directions of the balls inside the vibration chamber are random. Each ball induces several repeated multidirectional plastic deformations in the surface layer of the sample. In this condition, the plastic deformation mechanism of the CoCrMo alloy impacts about 10 μm in depth [23, 24].

Fig. 1
figure 1

Sketch of the SMAT process

Biomedical polishing

Before and after the SMAT, the samples were subjected to a biomedical polishing to reduce the surface roughness following the biomedical standard (arithmetic roughness R a < 0.5 μm). The biomedical polishing based on brush polishing method was performed by MARLE® orthopedics manufacturing company. For reasons of confidentiality and intellectual property, the steps of this treatment could not be detailed.

In this work, four CoCrMo samples were used for the corrosion tests and labeled according to their surface treatment. The sample noted (W) was obtained by cutting the CoCrMo alloy followed by a polishing step using silicon carbide (SiC) papers of 120, 600, and 1200 grit. For the second sample noted (W-P), a biomedical polishing (P) was realized on the (W) surface. The samples named (S15) are (W) discs subjected to the SMAT for 15 min, and those noted (S15-P) are (S15) discs with biomedical polishing (P).

X-ray diffraction

The phase’s transformations induced during the treatments were determined by X-ray diffraction (XRD) measurements using an X-ray diffractometer (Bruker D8 Advance). The X-ray pattern data was collected from  = 35 to 90° using a monochromatic Cu radiation with a step of 0.06° every 10 s. The Powder Diffraction Files (PDF) of the International Center for Diffraction Data (ICDD) were used for the phase’s identification.

Scanning electron microscopy

A LaB6 scanning electron microscope (JEOL JSM-6460 LV) operating at 0–30 kV was used to observe the surface topography of the untreated and treated CoCrMo samples.

Electrochemical corrosion testing

The corrosion behavior of the untreated CoCrMo alloy and that of those subjected to a mechanical treatment was evaluated by potentiodynamic polarization and electrochemical impedance spectroscopy (EIS). The potentiodynamic polarization tests were performed by scanning the applied potential from − 1.5 V vs SCE (saturated calomel electrode) to 0.8 V vs SCE with a scan rate of 1 mVs−1. Before testing, the samples were allowed to reach the open circuit potential E OCP for 15 min. Each test was repeated four times in order to reduce the margins of error and to verify the repeatability of results. The impedance measurements were carried out after 15 min at open circuit potential (OCP) using a PGZ potentiostat RADIOMETER 100. The frequency ranges from 10 kHz to 10 mHz with five points per decade and an amplitude of 10 mV. The impedance data was modeled using the ZView software.

The electrolyte used for the electrochemical corrosion measurements was Ringer’s solution purchased from Merck (9.000 g L−1 NaCl, 0.200 g L−1 NaHCO3, 0.420 g L−1 KCl, 0.480 g L−1 CaCl2). The physiological liquid is a saline solution (pH = 7.00 ± 0.02 at 25 °C) without vitamins and amino acids [25]. The temperature of the electrolyte was maintained at 37 °C during the experiments.

Results and discussion

XRD analysis

The X-ray diffraction patterns of the CoCrMo alloy before and after the mechanical treatments are presented in Fig. 2. The W and W-P samples reveal the presence of two phases: the austenite (γ-fcc) and the martensite (ɛ-hcp). These two diffraction patterns are quite similar which clearly indicates that the biomedical polishing does not generate any crystallographic transformation of the alloy.

Fig. 2
figure 2

X-ray diffraction patterns of the four CoCrMo alloys

The SMAT of the CoCrMo alloy (S15) has resulted in a decrease in the intensity of the (111 γ) and (200 γ) peaks and an increase in the intensity of the (101 ɛ) peak. This result clearly shows that the martensitic phase (ɛ-hcp) is generated at the surface of the CoCrMo alloy during the SMAT. Moreover, the biomedical polishing does not affect the martensitic transformation since the (101 ɛ) peak intensity in the case of the S15-P sample is greater than that of the W-P sample.

These results have been observed previously in literature in the case of AISI 304 stainless steel [16] and cobalt [26]. The mechanism of phase transformation was attributed to the increase in stacking-fault energy generated by the deformations. Wu et al. [26] have studied the grain refinement and the strain accommodation in cobalt (a mixture of hexagonal close packed (hcp) and face-centered cubic (fcc) phases) subjected to the SMAT. They observed a strain accommodation in the fcc phase started by the slip of dislocations at low strain, forming intersecting planar arrays of dislocations. With increasing strain, the γ → ɛ transformation occurred and contributed to the grain refinement.

SEM observations

Figure 3 shows the surface morphology of the four CoCrMo samples. The surface of W sample (Fig. 3a) shows stripes generated by the SiC polishing. These stripes are still slightly visible despite the biomedical polishing (Fig. 3b). After the SMAT, the surface of the S15 sample (Fig. 3c) is chaotic and rough but the biomedical polishing (S15-P) has significantly reduced its surface roughness and only residual stripes and ball impacts remain visible (Fig. 3d). The roughness (R a) was measured using a contact roughness tester. Its value has increased from about 0.02 to 0.64 μm after the SMAT (S15 sample). After the biomedical polishing, it has decreased to around 0.30 μm (S15-P sample).

Fig. 3
figure 3

SEM surface morphology of the CoCrMo alloys. a W. b W-P. c S15. d S15-P

Electrochemical corrosion testing

Polarization curves

The potentiodynamic polarization is used to compare the dynamic electrochemical responses of the four CoCrMo alloy surfaces. This characterization aims at determining the influence of each treatment step on the corrosion behavior of the CoCrMo alloy. Figure 4 shows typical Tafel polarization curves for the four CoCrMo samples in Ringer’s solution. These curves were obtained after stabilization of the open circuit potential during 15 min (results not shown).

Fig. 4
figure 4

Tafel curves of the four CoCrMo alloys in Ringer’s solution (scan rate 1 mV s−1)

In the cathodic branch, the hydrogen evolution reaction (HER) due to the electrolysis of water appears from − 1.5 V vs SCE to − 0.9 V vs SCE. Moreover, the oxygen reduction reaction (ORR) also occurs since the electrochemical experiment is carried out at open room atmosphere in a non-deaereted physiological solution. The anodic branch of the curves describes a typical behavior of metals that undergo a passivation when they are exposed to an appropriate electric current density. The formation of a passive oxide film on the sample surface inhibits the corrosion evolution, resulting in the formation of a plateau on the Tafel curves. For example, the polarization curve of the W sample shows a wide plateau from − 0.75 V vs SCE to 0.50 V vs SCE with a current density of about 10−1 mA cm−2. This plateau corresponds to the passivation of the alloy surface. Thus, the corrosion reactions are reduced and both active and passive behaviors occur. Next, the transpassive region is reached (no more passivation) where the current density increases with the applied potential. The corrosion current densities (i corr) can be determinate by extrapolating the linear part of the cathodic and anodic curves to the open circuit potential (E OCP vertical axis) [27]. The corrosion parameters deduced from these curves are reported in Table 1. We note that the corrosion current densities (i corr) and the corrosion potentials (E corr) for W and W-P samples are almost the same. Thus, the corrosion behavior of the CoCrMo alloy was not improved by the biomedical polishing.

Table 1 Corrosion parameters extracted from the polarization curves

For the S15 sample, the corrosion potential (E corr) increases from − 0.920 V vs SCE to − 0.737 V vs SCE and the corrosion current density (i corr) is divided by about 2.5 in comparison with the W sample. These results indicate that the passive film formed on the CoCrMo alloy surface after the SMAT is more protective. The improved surface passivity can be attributed to the surface changes generated by the SMAT. Indeed, the chemical composition of the passive layer plays an important role in the protection against corrosion. This layer is generally composed of metal oxides. For example, Hao et al. have shown that the passive layer rich in chromium oxides formed after SMAT of 316L stainless steel improves the resistance to the attack of corrosive chloride anions [17]. Balusamy et al. have studied the effect of SMAT on AISI 409 stainless steel and explain that it improves the corrosion resistance through the formation of a passive film associated with surface nano-crystallization [18]. Jelliti et al. studied the effect of SMAT on the TA6V alloy and showed that the improvement of the corrosion resistance in Ringer’s solution is due to the formation of a passive layer rich in titanium oxides [19]. Concerning the CoCrMo alloy, Petrov et al. have clearly shown that a surface treatment similar to SMAT leads to the formation of a passive layer consisting essentially of cobalt oxides (determined by XPS analysis) [20]. Combined with the martensitic transformation, this layer allows an improvement of the corrosion resistance in a medium similar to Ringer’s solution. In our case, we can consider that the same mechanisms take place. According to Petrov et al., the severe plastic deformation of the surface alloy promotes an intense penetration of oxygen atoms into the crystal lattices leading to the formation of the oxide passive film. At the same time, the formation of interfacial boundaries associated with high fraction of grains with (111)γ and (0002)ε planes parallel to the surface of the alloy promotes enhanced corrosion resistance.

Moreover, after the biomedical polishing, the E corr of S15-P slightly increases (from − 0.962 V vs SCE to − 0.858 V vs SCE) and the i corr slightly decreases (from 25.2 to 21.3 μA cm−2) in comparison with the W-P sample. This indicated that the biomedical polishing does not alter the advantages of the surface passivation generated by the SMAT.

Electrochemical impedance spectroscopy

The EIS is used to get information on the characteristics of the passive films and on the nature of the electrochemical processes occurring at the surface of the alloy in the physiological medium.

The Nyquist diagrams obtained for the four CoCrMo samples are shown in Fig. 5. It is observed the presence of an incomplete semi-circle in the high-frequency region with a larger diameter for the S15 sample. After biomedical polishing (S15-P), the diameter of the semi-circle remains greater than that of the W and W-P samples. The incomplete semi-circle observed on the Nyquist diagram highlights the charge transfer reactions between the sample surface and the electrolyte [21]. The increase of the semi-circle diameter indicates the enhancement of the passive film stability. The largest semi-circle diameter in the case of S15 and S15-P samples shows the noblest electrochemical behavior induced by the SMAT of the CoCrMo alloy.

Fig. 5
figure 5

Nyquist plot of the four CoCrMo alloys immersed in Ringer’s solution

The Bode impedance plots (Fig. 6) can be divided in three regions; each one reveals specific information on the electrochemical process at the surface of the CoCrMo alloy. In the first region (high frequencies (1000 Hz < f < 10,000 Hz), we observe that log |Z| forms a plateau with a weak slope. There is also a sharp fall of the phase values in this area (Fig. 7). These results describe a typical response of a resistive behavior of the electrolyte which is more highlighted in the case of S15 and S15-P samples.

Fig. 6
figure 6

Bode impedance plot of the four CoCrMo alloys immersed in Ringer’s solution

Fig. 7
figure 7

Bode phase plot of the four CoCrMo alloys immersed in Ringer’s solution

The second region (low frequencies 10−2 Hz < f < 10 Hz) generally provides information about the passive behavior of the material. In this area, we can observe that log |Z| shows a linear tendency which describes the character of a passive film on the surface of the material [18]. At 10−2 Hz, the log |Z| of the S15 sample reaches a higher value (~ 7000 Ω cm2) than the W sample (~ 6000 Ω cm2). These results show that the SMAT can generate a more efficient passive layer, improving the CoCrMo alloy behavior against corrosion in physiological medium.

In the third region (middle frequencies 10 Hz < f < 103 Hz), the maximum peak observed on the phase plot (Fig. 7) may indicate the formation of a passive film or its evolution [28, 29]. This peak is maximum for the S15 sample, confirming that the SMAT promotes the CoCrMo alloy passivation.

The Bode phase diagrams in Fig. 7 show two time constants: one at low frequency (in the first region) and the second at the interface second region/third region corresponding to relaxation times. These results suggest that the passive film formed on the CoCrMo surface is made of two layers: the inner dense passive layer and the outer porous layer [28, 30].

Furthermore, the phase angles for all the samples are lower than 90°, indicating that the CoCrMo alloy behavior does not correspond to that of an ideal capacitor. To explain the distribution of relaxation times resulting from surface heterogeneities, the formation of passivation films and the adsorption of species, a model of the sample/electrolyte, is necessary. The equivalent electrical circuit proposed for the different samples is given in Fig. 8. R s corresponds to the electrolyte resistance; C dl and R ct represent the double layer capacitance and the charge transfer resistance at the CoCrMo sample/electrolyte interface. Particularly, R ct quantifies the impact of the experimental conditions on the kinetics of the oxidation reaction during the corrosion process given by

$$ M\to {M}^{z+}+z\ {e}^{-} $$
Fig. 8
figure 8

Equivalent circuit used to model the EIS data of the CoCrMo alloys

M is a metallic atom of the alloy surface.

C f and R f are the capacitance and the resistance of the passive film. The global values of the polarization resistance R p are calculated from the values of R f and R ct by the following relationship: R p = R f + R ct.

As an illustration, Figs. 9 and 10 present the comparison between the experimental values and the calculated data obtained from this model in the case of S15 and S15-P samples. A good agreement is obtained, and the electrochemical characteristics identified from the equivalent electrical circuit for all the samples are summarized in Table 2.

Fig. 9
figure 9

Experimental and simulated Nyquist diagrams of the CoCrMo alloys. a S15. b S15-P

Fig. 10
figure 10

Experimental and simulated Bode impedance and phase diagrams of the CoCrMo alloys. a S15. b S15-P

Table 2 Corrosion parameters of the four samples obtained from the proposed model

All of these corrosion results indicate that the chloride ions in the Ringer’s solution induce a competition between the local formation and the destruction of the passive film formed on the surface of the different CoCrMo samples. This corrosion mechanism leads to the charge transfer resistance R ct decrease and to the double layer capacitance C dl increase during immersion in Ringer’s solution. It can be noticed that for the S15 sample, the R ct value remains higher than that of the W sample. R f is the electrical resistance of ion transfer through the passive film pores, its value is maximum for the S15 and S15-P samples. The R f value evolution indicates that the electrolyte diffusion into the passive film pores is maximum in the case of W and W-P samples while it is strongly reduced after the SMAT because of the surface grains refinement.

On the other hand, the capacitance of the passive film C dl formed on the S15 surface is approximately twice lower than that of the W sample. This result shows that the SMAT reduces the active surface area and increases the thickness of the passive film. Compared with the W sample, the formation of this passive film in the case of S15 sample leads to a better protection of the CoCrMo alloy in Ringer’s solution, as shown by the EIS diagrams [12, 26]. Indeed, the SMAT process has produced many severe plastic deformations in the CoCrMo alloy that caused a large number of dislocations and stacking faults [29]. These structural changes have induced numerous diffusion pathways for oxygen that quickly increased the sub-surface oxygen content. Consequently, the SMAT process has promoted the formation of a dense passive layer at the surface of the CoCrMo alloy that improves the protection of the metallic implant against corrosion in an aggressive physiological solution.

Conclusion

In this study, the corrosion behavior of untreated and mechanically treated CoCrMo alloys was evaluated by potentiodynamic measurements and electrochemical impedance spectroscopy. These electrochemical studies reveal that the surface mechanical attrition treatment (SMAT) of the CoCrMo alloy by using 3-mm balls for 15 min promotes the passive film formation and increases the corrosion resistance during immersion in Ringer’s solution. Indeed, after SMAT, the CoCrMo alloy shows a more positive corrosion potential and a lower corrosion current density. It is also shown that the specific biomedical polishing coupled with the SMAT process does not alter the advantages of the surface passivation generated by the SMAT.