Introduction

Intraoral scanning systems are increasingly gaining importance in the CAD/CAM (computer-aided design/computer-aided manufacturing) production of high-resistance all-ceramic restorations, especially regarding single-tooth restorations and short span-bridges. The high acquisition costs are further hindering clinicians from purchasing intraoral scanning systems. Therefore, conventional impression techniques are still used for the majority of impressions performed in a clinical setting. The standard procedure for achieving the implementation of the CAD/CAM method in this setting is by scanning the plaster model resulting from the conventional impression by using a laboratory scanner [1]. However, mainly due to gypsum expansion, the model casting process is highly prone to error. The direct laboratory scan of the conventional impression, which is one work-step ahead in the conventional workflow, could be an approach to eliminate the entire model casting process and may be thus promising for reducing the overall error and therefore improving the fitting accuracy of restorations.

The marginal fit of the restoration is one of the main factors that determine the longevity of the restoration and its overall success [2]. In the oral environment, an inadequate marginal fit can lead to the dissolution of the cement film [3]. Subsequent plaque accumulation [4] can facilitate the development of periodontitis and [5] secondary caries [6], as well as microleakage and, as a consequence, endodontic inflammation of the tooth [7]. McLean and Fraunhofer stated that the marginal gap of a restoration should not exceed 120 μm [8]. Although there have been other investigations demanding values below 50 μm [9], the most recent studies mainly rely on the data of McLean and Fraunhofer [10].

A variety of factors can lead to an inadequate marginal fit of a restoration. Impressions that use an elastomeric material such as vinylpolysiloxane are a standard procedure in the conventional workflow, and the literature shows that a sufficient accuracy is achievable with conventional impression techniques [11,12,13]. However, a multitude of errors can occur in the intraoral phase and during laboratory procedures due to the high level of sensitivity of the technique and its dependence on material properties [14].

Recent investigations show a higher fitting accuracy in vivo with an all-digital workflow using direct intraoral digitization systems, such as Lava COS (3 M, St. Paul, USA) [15], CEREC (Sirona, Bensheim, Germany) [16], Cara TRIOS (Heraeus, Hanau, Germany) [17], and iTero (Align Technology, Amsterdam, Netherlands) [18]. One of the major drawbacks of these systems is that they are not able to record areas that are not fully optically accessible. However, this is also true for conventional impression techniques that also require a moisture-free and accessible preparation margin for successful impression. In these situations, further methods such as surgical crown lengthening may be necessary.

To our knowledge, the accuracy of the indirect approach by scanning the elastomeric impression has only been sparsely investigated. Thus far, this method of indirect digitization has shown acceptable outcomes in vitro [19,20,21,22] that are comparable to those of direct intraoral scanning, but no in vivo investigations have been published.

The following null hypothesis was tested: There is no significant difference between the marginal fit of restorations that are manufactured on the basis scans obtained from two established intraoral scanners and those that are manufactured on the basis of a laboratory scan of a conventional impression in vivo.

Materials and methods

This study was conducted at the Department of Prosthodontics of the Justus-Liebig-University in Giessen, Germany. A total of 63 teeth in 23 patients (12 females and 11 males; ages 50 to 84 years) were included in this study. The inclusion criteria included the indication for a crown restoration (crowns, bridges, and telescopic crowns), an asymptomatic tooth, and a visible finish line. Exclusion criteria included unstoppable bleeding, an inability to keep the area dry, and an indistinguishable finish line. Additionally, all of the subjects who were included gave their written consent after being informed of the aims and the procedure of the study. The study protocol was approved by the Ethics Committee of the Justus-Liebig-University in Giessen, Germany (267/13), and was registered in the German Clinical Trial Register (DRKS00009371).

Clinical procedures

All of the patients were given local anesthesia for the preliminary treatment of the abutment teeth, which included, if necessary, excavation of caries and adhesive build-ups (Optibond FL, Kerr, Rastatt, Germany; Rebilda LC blue, Voco, Cuxhaven, Germany). A chamfer finish line was prepared with diamond burs with rounded edges and polished with a fine diamond bur according to generally accepted guidelines [23]. Afterwards, the provisional restoration preparation (Luxatemp, DMG, Englewood, USA) was made and was temporarily cemented (Temp Bond, Kerr, Rastatt, Germany).

A conventional impression and two intraoral scans were taken 7–10 days after the preparation to allow for possible injuries of the gingiva caused by the preparation to heal. To ensure a sufficient retraction of the gingiva and in order to prevent bleeding, intrapapillary anesthesia (Ultracain, Sanofi-Aventis, Frankfurt am Main, Germany) was administered and retraction cords, which were soaked in 25% alumina sulfate (Orbat sensitive, lege artis, Dettenhausen, Germany), were placed in the sulcus prior to each impression (Ultrapak, Ultradent, South Jordan, USA). The double cord technique was employed, meaning that one smaller cord was plugged deep into the sulcus beneath the finish line and one larger cord was plugged right above it. The latter was then removed for the scanning procedure, while the previous cord remained in the sulcus in order to prevent the gingiva from collapsing during the procedure. Dry tips and disposable soft tissue retractors for cheeks and lips (Optragate, Ivoclar Vivadent, Schaan, Liechtenstein) were used to keep the scanning area dry at all times. Two intraoral scans were conducted using different powder-free intraoral scanners (CEREC AC Omnicam, Sirona, Bensheim, Germany; Cara TRIOS, Heraeus Kulzer, Hanau, Germany) in a randomized order. The manufacturer’s scanning protocols were abided by at all times. Before proceeding to the conventional impression, the data that were captured were checked for artifacts. The resulting files were exported from the intraoral scanners and, if necessary, converted into stl format (CrossManager, Datakit, Lyon, France). All scans and impressions were performed by the same calibrated dentists (P.B.). Prior to the study, the operator was trained by the respective manufactures and supervised by the senior authors (P.R. and B.W.). Before starting the study it was ensured that the operator had performed a minimum of 30 scans per system.

The conventional impression was always taken after the intraoral scanning procedure was completed so that the retraction measures would not be compromised. Perforated metal stock trays were used. The handles had to be shortened by approximately 3 cm beforehand in order for the trays to fit in the laboratory scanner without disrupting the scanning process. After varnishing the metal stock trays with a tray adhesive, a two-step putty-wash technique was implemented using a scannable VPS (Flexitime Fast & Scan Putty and Light Flow, Heareus Kulzer, Hanau, Germany). The remaining retraction cord was removed just before applying the washing material so that the cord would not get impacted in the impression. The impression material was given 3:30 min to set, as recommended by the manufacturer. All of the impressions were disinfected and then sent to a dental technician laboratory.

For each tooth included in the study, three zirconia copings were manufactured, resulting in a total of 189 copings.

Laboratory procedures

In the dental technician laboratory, all of the undercuts of the conventional impressions were trimmed without harming the prepared teeth. Undercuts directly below the preparation margin are in general small feather-edges of less than a millimeter. These were not removed in order not to destroy the impression.

Thereafter, the conventional impression was digitized by a trained dental technician using an automated laboratory scanner (D700, 3Shape, Copenhagen, Denmark). The resulting files were sent back to the department, where they were converted into stl file format (CrossManager, Datakit, Lyon, France).

The conventional impression and the intraoral scans each resulted in one stl-dataset per tooth. These were sent to a dental laboratory (Dentallabor Kappert, Essen, Germany) for design and manufacturing of the zirconia copings. The design parameters were set at 65 μm for the cement gap, 600 μm for the overall layer thickness and 200 μm for the edge reinforcement (ExoCAD, Darmstadt, Germany). These parameters were identified in a laboratory conducted series of systematic pretests with different values for the cement gap, the overall layer thickness, and the edge reinforcement to deliver optimal results for the marginal fit of each impression system.

All of the copings were milled from pre-sintered zirconia blocks (Copran Zr, White Peaks, Wesel, Germany) by means of a 5-axis milling machine (DWX 50, Roland, Irvine, USA). After sintering at 1500 °C, the copings were sent back to the dental operatory for try-in.

Thereafter the conventional impression was poured with type IV plaster and further used for the production of the intended restoration for the patient. For scientific clearness, we only compared the three zirconia copings and left out the final restoration.

Replica technique

The marginal fit of the copings was determined by means of the replica technique. After removing the provisional crown, the abutment teeth were thoroughly cleaned. Before implementing the replica technique, the initial fit of the coping was checked visually. If it was not possible to seat the coping on the tooth, a low viscosity silicone (Fit Test C&B, Voco, Cuxhaven, Germany) was used as fit-checker and up to 3 adjustments using a fine diamond bur (25 μm grain size) were made. For the replica technique, the copings were filled with a low-viscosity silicone (Fit Test C&B, Voco, Cuxhaven, Germany), seated on the preparation and held in place with maximum finger pressure for 3 s, simulating the clinical cementation of the crown. After 2 min of setting time, the coping was removed from the tooth. For this purpose, it was crucial that the resulting thin layer of silicone remained adhered to the inner surface of the coping and did not tear when removed. If this was not achieved, the procedure was repeated. Afterwards, the lumen of the coping was filled with another low viscosity silicone, stabilizing the first layer and mimicking the tooth. After the silicone had set, the replica was carefully removed from the coping and sectioned. Using a sharp scalpel, one cut was made mesio-distally, and one was made buco-orally, resulting in four sections. The samples were embedded in putty-mass (Giroform Putty, Amann Girrbach, Pforzheim, Germany) on a microscope slide and adjusted horizontally with a parallel-press. The measurements were conducted using a digital microscope (Smartzoom5, Zeiss, Jena, Germany) at × 200 magnification. Image analysis software was used to measure the marginal discrepancy (the shortest distance from the margin of the coping to the abutment surface closest to the finish line), which was performed in accordance with the procedure described by Holmes et al. (Fig. 1). All of the replicas were prepared and all of the measurements were made using the same calibrated operator. The measurements were recorded in microns and exported to a spreadsheet.

Fig. 1
figure 1

Replica at × 200 magnification, (1: tooth, 2: coping, 3: cement gap, 4: marginal discrepancy)

Statistical procedures

Based on the data of Boeddinghaus et al. [24], a sample size of approximately 60 specimens was calculated for achieving a power of 80%. The statistical analysis was carried out using SPSS (22, IBM Statistics, Armonk, USA). The level of significance was set at α = 0.5%. Because of dependencies within the variables, a random-intercept model was applied. The mixed procedure was calculated with logarithmic values in order to determine estimations with approximately normally distributed residuals. The Levene test was applied to check for variance homogeneity. In the case of a significant F-test when testing the four-staged factor “type of tooth,” pairwise comparisons of the estimated margins, which were corrected for multiple testing (Sidak), were interpreted. The mean values and standard deviations were reported.

Results

In total, 63 teeth from 23 patients were examined. Thirty-three teeth were situated in the upper jaw, and 29 in the lower jaw. Altogether, 16 incisors, 22 canines, 16 premolars, and 9 M were analyzed. Only six copings initially needed adjustments. However, no visually acceptable fit could be achieved and therefore, these six copings could not be tested: one based on the data of CEREC and five based on the data of the impression scan. The mean values of all of the methods were tested and compared. The marginal discrepancies were 86.09 μm (± 64.46 μm) for CEREC AC Omnicam, 88.95 μm (± 54.46 μm) for Cara TRIOS, and 143.29 μm (± 100.71 μm) for the impression scan (Fig. 2). Significant differences could be established between the laboratory scan and the two intraoral scanners (p < 0.05). The comparison of the different types of teeth (Table 1) showed no significant differences within each acquisition method (p > 0.05). Therefore, the null hypothesis was rejected.

Fig. 2
figure 2

Boxplot diagram of the impression methods

Table 1 Marginal fit of the different types of teeth. There was no significant difference in between the four types of teeth within each group (p > 0.05)

Discussion

This study aimed at measuring and comparing the marginal fit of all-ceramic restorations fabricated from three different impression systems, two direct and one indirect. The focus of this investigation was on the indirect laboratory scan of a conventional impression. Due to various results from studies over the last few years [25, 26], which showed a greater accuracy for single-tooth restorations that were manufactured on the basis of direct intraoral scans in comparison to the conventional workflow, we decided to utilize two well-established and validated intraoral scanners (CEREC AC Omnicam; Cara TRIOS) as references. As this study analyzed the entire workflow, we optimized the production parameters prior to the study to insure an optimal result for the clinically most relevant parameter “marginal fit” for the different methods compared. We ensured that the marginal fit was not effected by axial, incisal, or occlusal discrepancies. Therefore, we omitted the further analysis of the internal fit of the copings.

All types of teeth (molars, premolars, and anteriors) were included in this study, although not in equal numbers. This is not suspected to have any influence on the differences found between the acquisition methods, as the same conditions were provided for every technique and reproduction of every type of tooth. There was no limitation concerning the position of the preparation margin, as long as it was visible. This was particularly necessary for the intraoral scanners, which operate solely with optical acquisition systems, yet it is also necessary for the application of VPS for the conventional impression to the preparation margin. The absolute marginal discrepancy according to Holmes et al. [27] was selected as the primary outcome variable as this measurement does imply the full extent of the misfit, and includes under-extension and over-extension of the restoration margins. As also most authors measure the marginal discrepancy [10], so we chose to do the same to increase the comparability of the study. Compared to an in vitro setup, where teeth and cemented restoration can easily be sectioned, the in vivo assessment of the marginal gap is rather difficult. Therefore, the well-established replica technique was applied, which allows for a non-invasive and easily performed means for assessing the adaptation of a restoration to the tooth surface [28]. The concordance of the thin outer silicone layer in the replica to the cement-space of a cemented restoration has been validated multiple times [29, 30]. Known limitations of the technique, such as its slice preparation and deep preparation margins, were excluded from the beginning and therefore did not have an effect [31].

As there are several work steps between the capturing of the intraoral situation and the completion of the actual zirconia coping, not only the acquisition method but also the whole workflow with all of its variables has to be taken into account. As all three of the systems produce digital data, from this point on, the same standardized workflow was conducted for all of the methods.

Considering the data obtained in this clinical investigation, the null hypothesis was rejected because significant differences were found for the marginal discrepancies of the copings fabricated from the direct and indirect digital impression methods. The mean marginal gap was 86.09 μm for CEREC AC Omnicam, 88.95 μm for Cara TRIOS, and 143.29 μm for the impression scan. To our knowledge, there are no in vivo investigations available in the literature that compare intraoral scanning systems with the indirect laboratory scan of a conventional impression. However, there are some studies comparing intraoral scanning systems with the laboratory scan of a gypsum cast [15, 17, 18, 24, 32] or with the entire conventional workflow [25, 26].

Berrendero et al. [17] assessed the marginal gap of full-ceramic crowns in vivo, comparing the CaraTRIOS intraoral scanner to a laboratory scan of a plaster model resulting from a conventional impression. The mean marginal discrepancy was 106.6 μm for the intraoral scan and 119.9 μm for the scan of the gypsum cast. Boeddinghaus et al. [24] investigated both intraoral scanners that were utilized in this investigation and measured a mean marginal gap of 112 μm for CaraTRIOS and a mean marginal gap of 149 μm for CEREC AC Omnicam. Additionally, they tested another intraoral scanning system, Lava True Definition, and observed a mean marginal gap of 88 μm. The study of Brawek et al. [16] compared the marginal gaps of crowns fabricated from the data captured with the Lava COS intraoral scanner to those fabricated from the data of CEREC AC in vivo and measured 51 μm for Lava COS and 83 μm for CEREC AC. Other authors also found smaller gaps with Lava COS in vivo [15, 32]. However, both 3 M scanning systems, Lava COS and Lava True Definition, operate only with the application of a scanning powder and therefore cannot be compared directly to powderless scanning systems. The slightly lower values for Cara TRIOS and CEREC AC Omnicam in the present study could have been due to the use of more recent software versions in both intraoral scanners and lower settings for the cement space. Regarding the laboratory scan of a conventional impression, in vitro investigations show considerably smaller marginal gaps compared to the in vivo outcomes of the present study [20, 33]. Lee et al. compared two indirect laboratory procedures and found mean marginal discrepancies ranging from 54 μm to 75 μm for the scan of the plaster model and from 69 μm to 118 μm for the scan of the conventional impression [21]. Shembesh et al. investigated all three methods of digitizing the intraoral situation. The best outcome was shown by the intraoral scanning system Lava True Definition, which measured a mean marginal gap of 27 μm. The laboratory scan of the gypsum model achieved a mean value of 50 μm, while the laboratory scan of a silicone impression reached the highest value of 81 μm, which still is well within the clinical parameters of acceptance [22]. It is common for studies to show higher accuracies in vitro than in vivo [34]. Clinical factors, such as limited space, patient movement, saliva, bleeding, and sulcus-fluid do not have any influence on the outcome in in vitro settings, whereas they foster errors when working in vivo. Here, digital systems appear to have an advantage over conventional impressions because they allow the user to check for inaccuracies on a magnified model on a screen after finishing the scanning procedure. Therefore, errors can be detected more easily.

In this study, the mean marginal gaps for the different types of teeth within each acquisition method were also examined. Although none of them displayed significant differences, a tendency is recognizable within the data of the laboratory scan of the conventional impression (Table 1). Incisors and canines, which are teeth that are typically longer and more angulated when prepared for a crown, exhibit greater mean marginal gaps and greater standard deviations compared to premolars and molars. The latter show a stronger similarity to the shape of model teeth that are typically used in in-vitro settings. The major disadvantage when scanning conventional impressions appears to be the insufficient ability to capture long and deep hollow spaces and inside contours, as found in the impression of prepared incisors and canines [35, 36]. Figure 3 shows a schematic representation why the capturing process might fail to complete. The angle α remains the same for both rays of light shown. The green ray can be reflected to the sensor whereas the red ray is hindered by the border of the impression resulting in an incomplete acquisition of the same tooth. Therefore the limitations do not seem too dependent on the scanner, but on the geometry of the tooth.

Fig. 3
figure 3

Incomplete acquisition of a tooth in the scanning process of a conventional impression

Conclusion

Within the limitations of the present study, it can be concluded that the laboratory scan of a conventional impression is not comparable to that obtained using intraoral scanners regarding the marginal fit of zirconia copings manufactured on the basis of the resulting data. Only the copings that were obtained from the intraoral scanners CEREC AC Omnicam and Cara TRIOS displayed marginal gaps within the limits of clinical acceptability.