Introduction

Because osseointegration immediate and long-term success has been extensively reported, the need to expand knowledge on implant-supported prosthodontics success concerning mechanisms involved in the fatigue and failure process has been strongly suggested in a recent systematic review [1]. For single implants, one of the most commonly reported mechanical problems is screw joint instability potentially resulting in loosening and/or fracture of the abutment screw. The second most common failure is loosening or fracture of the abutment with few reports of implant fracture [2]. With variations in implant design and diameter, complication rates should be well understood especially in cases where a narrow implant is under loading, given that stresses will be born and distributed to a significantly smaller structure compared to a standard diameter implant (approximately 4 mm) [3].

The use of narrow implants, when correctly indicated and placed, has gained significant attention in implant dentistry since it may allow the restoration to be inserted in areas that would require grafting procedures to create enough volume to insert a standard diameter implant. That could increase not only treatment time but also costs and treatment morbidity. Other situations where they may also be indicated are limited prosthetic spaces, as in the maxillary and mandibular lateral incisors [4]. However, little is known about the mechanical performance and expected failure mechanisms of different implant-abutment connection designs of narrow diameter implants subjected to fatigue [3, 5]. The use of narrow-neck and reduced-diameter implants resulted in an overall increase in stress and strain magnitudes around supporting implants in comparison with support from two standard solid-screw implants [6]. Some manufacturers attempted to decrease the potential failures by increasing the implant diameter and modifying the abutment connection design [4, 5, 7].

Narrow implants are either presented as one piece, where its coronal portion is prepared to receive the crown, or as two pieces when an abutment is selected to be torqued and to receive the prosthetic crown. Clinical pros and cons aside, the variety of implant-abutment connection designs warrants further investigation so potential differences between systems can be acknowledged. In addition, two-piece implants may present abutments and their screws fabricated as a monolithic piece, which may improve the performance of the system given that spaces for micromotion are eliminated, potentially resulting in increased survival when higher loads operate under function.

Long-term randomized clinical trials are always highly desired to obtain information on the longevity and complication modes of new implant system’s design. However, due to high cost and the time consuming nature of this kind of research, laboratory testing involving fatigue is an important step in the hierarchical approach for rapidly understanding material’s overall performance before randomized trials can be performed [8]. Fatigue testing is an important tool to understand the probability of survival and failure modes of prosthetic components and implant [9, 10]. Fatigue testing has demonstrated to be relevant to simulate the mechanical response of prosthesis systems under function [11, 12]. Fatigue and its related failures are a time-dependent phenomenon with a proportional increase in the incidence of implants and prosthetic component failures observed as time elapses with the system under load [9, 10, 13].

This study sought to evaluate the probability of survival and failure modes of five narrow implant systems, each comprised by a different implant-abutment connection design and restored with maxillary central incisor crowns when subjected to step stress-accelerated life testing (SSALT) in water. The postulated null hypothesis was that two-piece narrow implants from several manufacturers would not present different probability of survival when used for anterior single-unit replacements.

Materials and methods

Experimental design

Five different narrow implant systems were selected and divided (n = 18 each) as follows: Astra (ASC) OsseoSpeed TX Small implant with 3.5 mm diameter (Astra Tech, Waltham, MA, USA) with a standard pre-fabricated titanium connection/TiDesign (Astra Tech, Waltham, MA, USA); BioHorizons (BSC) narrow implant of 3.5 mm diameter (PYR35105, Birmingham, AL, USA) with standard pre-fabricated titanium connection/PYNEA (BioHorizons, Birmingham, AL, USA); Straumann Roxolid (SNC) narrow neck CrossFit® connection with 3.3 mm diameter (Straumann, Basel, Switzerland) using a standard pre-fabricated abutment (NC anatomic abutment, Straumann, Basel, Switzerland); Intra-Lock (IMC) narrow implant (3.4 mm diameter) (SQ Connection, Intra-Lock International, Boca Raton, FL, USA) with a standard multilobular connection; and Intra-Lock narrow neck implant (ILO) (3.4 mm diameter) with a one-piece abutment and screw (One Piece Connection, Intra-Lock International, Boca Raton, FL, USA) with a modified square connection (Table 1).

Table 1 Characteristics of groups included in the study

Sample preparation

Each abutment was connected to the respective implant and vertically embedded in polymethyl-methacrylate resin (Orthodontic Resin, Dentsply Caulk, Philadelphia, PA, USA) leaving 1 mm of the implant-abutment finishing line exposed above the potting surface. The resin block with the embedded implant-abutment assembly was used as a matrix for fabrication of all samples within the same group.

All groups were restored with standardized central incisor metallic crowns obtained by milling and produced from one single .stl file which fabricated a series of identical samples (Co-Cr alloy, Wirobond 280; Bego, Bremen, Germany). The abutments were tightened with a torque gauge (Tohnichi BTG150CN-S, Tohnichi America, Northbrook IL, USA) according to the manufacturer’s instructions, and crowns cemented (Rely X Unicem, 3M ESPE, St. Paul, MN, USA) on the abutments.

Mechanical testing and reliability analysis

Mechanical testing was conducted with specimens positioned at a 30° axial inclination, as per ISO 14801:2007, to provide a bending component during fatigue loading which was performed under water at 9 Hz with a servo-all-electric system (TestResources 800L, Shakopee, MN, USA). Step-stress accelerated life testing mild, moderate, and aggressive profiles were used to fatigue samples [11, 1418]. Distribution of the 18 samples of each group across the three step-stress profiles was as follows: nine to the mild, six to the moderate, and three to the aggressive profiles. These profiles refer to the gradual increase in load as fatigue cycles elapse until specimen failure or survival. Therefore, samples assigned to the aggressive profile run less cycles to reach the same load levels of others ran in the mild profile [19]. The rationale for such profile design and distribution is based on the need to distribute failures across different step loads allowing improved statistical prediction and accelerated life testing [20, 21]. The samples were tested until failure (fracture or bending of the fixation screw and/or of the abutment and/or implant) or survival after the end of step-stress profiles.

Based upon the step-stress distribution of the failures, use level probability Weibull curves (probability of failure versus cycles) with use stress of 130 and 180 N at 90 % two-sided confidence intervals were calculated and plotted (Synthesis Alta 9, ReliaSoft, Tucson, AZ, USA) using a power law relationship for damage accumulation. For the mission reliability and β parameters calculated in the present study, the 90 % confidence interval ranges were calculated as follows:

$$ IC=E(G)\pm {Z}_{\alpha}\mathrm{sqrt}\left(Var(G)\right) $$
(1)

where IC is the confidence bound (CB), E(G) is the mean estimated reliability for the mission calculated from Weibull statistics, Z α is the Z value concerning the given IC level of significance, and Var(G) is the value calculated by the Fisher information matrix [20, 22].

If the use level probability Weibull calculated beta was <1 for any group (which indicates the failure rate over time), then a Probability Weibull Contour plot (Weibull modulus vs. characteristic strength) was plotted (Synthesis Weibull ++9, ReliaSoft, Tucson, AZ, USA) using final load magnitude to failure or survival of all groups. The Weibull modulus (m) and characteristic strength Eta (η) (63.2 % of the specimens would fail up to the calculated “η”) were identified for examining differences between groups.

Failure analysis

The failed samples were inspected in a polarized light stereomicroscope (MZ-APO stereomicroscope, Carl Zeiss MicroIang, Thornwood, NY, USA) and classified according to the failure criteria for comparisons between groups. In order to identify fractographic marks and characterize the failure origin and the direction of crack propagation, the most representative failed samples of each group were inspected under a scanning electron microscope (SEM) (S-3500N; Hitachi, Osaka, Japan) [16, 23].

Results

Mechanical testing

The mean beta and associated upper and lower bounds β value (confidence interval range) derived from use level probability Weibull calculation (probability of failure versus number of cycles) were 2.10 (1.24–3.55) for ASC, 1.34 (0.75–2.37) for BSC, 0.40 (0.23–0.70) for SNC, 3.19 (2.22–4.57) for IMC, and 1.05 (0.61–1.81) for ILO (Table 2). The beta value for SNC indicated that strength was the main factor dictating its failure behavior, while failures for the remaining groups were influenced by fatigue and damage accumulation. Note that the beta value (or Weibull shape factor) describes failure rate changes over time where beta <1, failure rate is decreasing over time, commonly associated with “early failures” or failures that occur due to egregious flaws; beta ∼1, failure rate that does not vary over time, associated with failures of a random nature; and beta >1, failure rate is increasing over time, associated with failures related to damage accumulation [24].

Table 2 Reliability values for a mission of 100,000 cycles show that at 130 N, systems are not different (upper and lower bounds overlap). However, at 180 N, a significant decrease was observed in the reliability of ASC, SNC, and IMC, whereas the reliability of BSC and ILO was maintained. The beta values at the bottom of the table indicate failure rate changes over time.

The calculated Weibull modulus (m) and characteristic strength (η, in Newtons) are numerically presented for each group in Table 3 and graphically in the contour plot (Fig. 1). The characteristic strength values for SNC (220.38 N) and ASC (237.37 N) were not significantly different from each other, but only the distribution of values of ASC was significantly lower compared to IMC (270 N) and BSC (286 N) (Fig. 1). The latter two groups (IMC and BSC) were not significantly different between each other, but both presented characteristic strength significantly lower than the ILO (288 N) which was significantly higher than all others. Differences between groups are depicted in the contour plot (Fig. 1) which represents the values of the combination of both parameters (Weibull modulus and characteristic strength), and differences in specimen population are detected if contour overlap between groups does not exist [25].

Table 3 Weibull modulus and characteristic strength values for groups
Fig. 1
figure 1

a Use level probability Weibull plot shows the probability of failure as a function of elapsed cycles. b Contour plot which is the summary statistics of the failed fatigue samples and their load at fracture during fatigue, or characteristic strength, combined with the information of Weibull modulus. Non-overlap between groups indicates that they are significantly different

The calculated reliability with 90 % confidence intervals for a mission of 100,000 cycles at 130 N showed that cumulative damage from loads would lead to ∼93 % implant-supported restoration survival in group ASC, ∼98 % in BSC, ∼85 % in SNC, ∼99 % in IMC, and ∼100 % in ILO. At 180 N and same cycle mission, the probability of survival was <0.1 % ASC, ∼77 % for BSC, ∼26 % for SNC, ∼93 % for IMC and ∼100 % for ILO (Table 2). Reliability was not significantly different between groups at 130 N. However, at a 180-N load, a significant decrease in reliability was observed for all groups, except for ILO where an overlap was still present between the upper and lower bounds (Table 2). When comparing the reliability between groups at 180 N, only IMC was not significantly different from ILO, whereas the remaining had significantly decreased. Also, the reliability for group BSC was not statistically different from IMC (Table 2).

Failure modes

All specimens failed after SSALT. When component failures were evaluated together, failures comprised the combination of abutment screw bending or fracture, and abutment fracture (Fig. 2). Failure modes are described in detail in Table 4. For all groups, failure predominantly involved the abutment screw fracture, and for the ILO group, the abutment fracture. Observation of the polarized light and SEM micrographs of the fractured surface of the abutment screws allowed the identification of fractographic markings, such as compression curl and the identification of the fracture origin at the opposite side along with the direction of crack propagation (Fig. 2).

Fig. 2
figure 2

SEM micrographs of representative fractures during SSALT showing (a, b) failure of the one-piece abutment screw system. a A lingual view, where fracture initiated and b an occlusal view with a compression curl located at the top which indicates that the bottom opposing side is the tensile, where fracture initiated. For the remaining figures, which presented abutments and screws as separate parts (cj), a similar failure mode was observed among them with failure confined to the abutment screw and occasionally involving the abutment (as in i, j)

Table 4 Failure modes after mechanical testing (SSALT) according

Discussion

It is important to acknowledge that no consensus seems to exist in the terminology for narrow or small diameter implants, but as controversies may arise to differentiate them from “mini dental implants” [26] used in orthodontic treatment, it has been suggested that narrow implants have a diameter equal to or greater than 3 mm [27]. Since narrow diameter implants have been introduced in attempt to fulfill the challenging clinical situation of thin alveolar crests or limited proximal clearance, the possibility of restoring single-unit prosthesis, as simulated in this study, has raised concerns regarding its mechanical integrity under fatigue loading. As implant diameter decreases, wall thickness at the cervical area will also decrease which seems to be critical in implants with internal connections, as recently shown [13]. Therefore, manufacturers have given substantial attention to more complex engineering designs that would better dissipate stresses improving the overall mechanical performance of narrow implant systems.

Although the probability of survival for loads averaging 130 N was high and not different between systems, a slight increase to 180 N resulted in a substantial decrease for all groups, except for the one-piece abutment/screw system (ILO), and for a very narrow difference in confidence interval also for group BSC. Whereas such result may be expected for the one-piece abutment/screw system, it showed that the presence of an implant-abutment connection and its connecting parts may hamper the performance of most narrow implant systems under more challenging load scenarios. Conversely, the reliability of a system with less connecting parts may be predominantly influenced by the implant and integrated abutment/screw’s raw material (i.e., titanium grades or alloys) [28, 29]. The observed differences between the two-piece systems were not expected from an implant diameter perspective alone, given that all four groups were of similar dimensions. Comparing the two-piece implant/abutment groups, only the BSC presented statistically different from the one-piece group at 180 N. Potential reasons for this finding could be attributed to its ability to shield the abutment and its screw preventing them from failing at load levels that lead the remaining groups to failure.

A remarkable aspect observed when comparing the groups for characteristic strength was the non-overlap between the one-piece and all other two-piece systems meaning that the characteristic strength of the former was higher than all others. In addition to this finding, the one-piece system resulted in the highest Weibull modulus, which was expressed in the contour plot by its narrow aspect and indicates that loads required for failure under fatigue occurred under a very defined range for this group. Among the two-piece systems, the only group not intersecting both IMC and BSC, due to its significantly lower characteristic strength, was the ASC system. Considering that implant diameter was identical between BSC and ASC and even lower for IMC, differences in design and fit between connections have likely accounted for the lower characteristic strength observed for the ASC group. The characteristic strength of the IMC was not different from groups BSC and SNC, in spite of its smaller diameter. Possibly its modified connection geometric design (IMC) which presents a four-sided internal configuration provides improved fit, decreasing micromovement between parts [30]. This aspect is of importance as the moment of inertia in bending is proportional to the inverse of the part diameter cube and as such even small decreases or increases in fit will result in exponential variation in the systems’ bending resistance. Future studies evaluating the fit of these systems are warranted.

The lack of long-term studies reporting on prosthetic survival rates on narrow implants has recently been pointed out [27], which indicates the need of clinical trials that supports previous in vitro evaluation of their performance considering that an important failure mechanism in metals is fatigue [31]. The step-stress accelerated life testing method utilized in the present study consists in testing the samples at higher stress levels in order to accelerate failures, although every sample is subjected to progressively increasing loads which start at low levels before reaching the above mentioned higher levels [21, 30, 3235]. These failures are distributed over three different profiles allowing statistical predictions for a given load and cycle mission.

From a fracture evaluation standpoint, abutment screw fracture was the main failure mode for systems other than the monolithic abutment and screw system (ILO). In the latter case, failure was still confined to the abutment/screw system and never involved the implant fractures, which allows subsequent restoration replacement in a clinical scenario. Therefore, regardless of the region where the fracture took place, repair was possible for all groups which is surprising considering that implant wall thickness is engineered to its minimum dimensions.

Conclusion

The postulated null hypothesis which stated that narrow implant systems varying the connection design within similar diameter would not present different probability of survival when used for anterior single-unit replacements was partially accepted. Although the probability of survival was not significantly different between systems at a 130-N load, an increase in load to 180 N significantly decreased the probability of survival of all two-piece narrow systems, but the BSC and the one-piece abutment/screw ILO.