Introduction

Hydrogen peroxide (H2O2) is a common oxidant currently used in a wide range of industries, such as textile, paper, food, and pharmaceutical ones [1, 2]. In recent years, studies have shown that the excessive use of H2O2 in food industry is harmful to human health. Consequently, the Food and Agriculture Organization of the United States has limited the highest level of H2O2 in dairy products to 0.05% (14.7 mM). According to the Joint Expert Committee on Food Additives (JECFA, 2004), the maximum addition of H2O2 to food must be of 60 mg kg−1, whereas the Chinese National Standard (GB5009.226–016) has established this value to be less than 3 mg kg−1 [1]. In addition, studies have shown that high levels of H2O2 in living organisms affect profoundly the natural, physiological processes and can lead to cardiovascular and vascular diseases, as well as cancer, neurodegeneration, DNA damage, and so on [3, 4]. The level of H2O2 in human blood ranges from 1.0 to 5.0 μM, whereas in inflammatory diseases this value can reach up to 50 μM [1]. It is very important to determine the values of H2O2 in food and clinical diagnostics, particularly because oxidase enzymes also produce H2O2 as an important intermediate product. The rapid and accurate determination of the H2O2 is also of great importance in the development of effective biosensors [5]. Various techniques, such as chemiluminescence [6], spectrophotometry [7], titrimetry [8], chromatography [9], and electrochemistry [10,11,12,13,14,15] have been reported in the literature with regard to the determination of H2O2. Among these techniques, enzymatic biosensors are more attractive than others since they are inexpensive, simple, sensitive, and selective.

Carbon-based nanomaterials are widely used in biosensor studies due to their ability to provide large surface area, good mechanical and chemical stability, and high electrical conductivity [16]. Biosensors using carbon nanotubes (CNTs) as modification materials have been reported to exhibit high sensitivity, low detection limits, and wide working ranges [17, 18]. In biosensor fabrication, CNTs have been used with ionic liquids (ILs) to obtain composites with high sensitivity and good biocompatibility [19].

ILs belong to a special group of ionic electrolytes without molecular solvents [20]. ILs are used to improve the response of electrochemical biosensors due to their high ionic conductivity and biocompatibility [21]. Moreover, they are chemically and thermally stable as well as relatively ion-conductive, and display negligible vapor pressure, while possessing a wide electrochemical potential range [12]. Imidazolium and pyridinium salts are the main classes of ILs. They have attracted a significant degree of interest. The use of imidazolium ILs as physical films [20, 22,23,24,25] or in combination with carbon paste as bulk modified electrodes [26,27,28] is a major advancement in the field of electrochemistry.

Horseradish peroxidase (HRP) is a monomeric heme-containing phytoenzyme, which has attracted great interest in biosensor and biotechnology applications due to its high stability in the aqueous solution [29]. HRP contains a prosthetic heme group which catalyzes the oxidation of different substrates [5]. The enzymatic activity of HRP is due to the oxidation and reduction of the iron atom in the heme group. The immobilization step in enzyme-based biosensor fabrication determines the reliability and performance of the biosensor. Conventional immobilization methods include non-covalent (entrapment and adsorption) or covalent immobilization [30]. Each of these methods has both advantages and disadvantages. Various enzyme-based electrochemical biosensors have been reported as being used for H2O2 determination [13,14,15], but such studies involve complex enzyme immobilization procedures.

Poly-L-histidine (PLH) is a synthetic polyamino acid containing an imidazole group. PLH can be used to facilitate both the reduction and oxidation of myoglobin and cytochrome and to catalyze the reduction of the NAD+ [31]. PLH modified sensors and biosensors are used to determine various analytes [32, 33]. Several methods for polymerizing L-histidine on the electrode surface have been previously reported [5, 32, 33].

In this study, an amperometric biosensor has been fabricated by immobilizing HRP on PLH, multiwall carbon nanotube, and IL modified glassy carbon electrode (HRP/PLH/MWCNT-IL/GCE) for the sensitive quantification of H2O2. HRP was electrostatically immobilized on the PLH/MWCNT-IL/GCE. Following the optimization of the experimental conditions the analytical performance characteristics of the HRP/PLH/MWCNT-IL/GCE have been determined. The analytical application of the biosensor for the H2O2 determination in real samples was also studied.

Results and discussion

Optimization of HRP/PLH/MWCNT-IL/GCE composition

In order to improve the efficiency of HRP/PLH/MWCNT-IL/GCE, certain amounts of MWCNT, IL, and HRP used in the electrode composition were optimized. For this, the amount of the optimization material varied, whereas the amount of the other components was kept constant. In the optimization of the amount of MWCNT and IL, the level of HRP on the surface of the electrode was kept constant at 20.4 U.

The effects of the IL amount on the response of the HRP/PLH/MWCNT-IL/GCE were investigated by varying its amount in the MWCNT-IL-gelatin mixture as 2.5, 5.0, 7.5, and 10.0 mm3. The amount of MWCNT is of 1.5 mg cm−3. Figure 1A shows the H2O2 responses to HRP/PLH/MWCNT-IL/GCEs as fabricated with the studied IL amounts. As the amount of IL increased from 2.5 mm3 to 7.5 mm3, the current response improved and then decreased. The biosensor showed the best current response to be at 7.5 mm3 IL and this value was used as the optimal amount of IL for the fabrication of the modified electrode.

Fig. 1
figure 1

Effects of A IL, B MWCNT, C HRP amounts on the response of HRP/PLH/MWCNT-IL/GCE (in 0.05 M pH 7.0 PBS, Eapp = −0.1 V)

MWCNT-IL-gelatin mixture containing different amounts of MWCNT (1.0, 1.5, 2.0, and 2.5 mg cm−3) was investigated for its effect on the amperometric H2O2 response by four different HRP/PLH/MWCNT-IL/GCE. As shown in Fig. 1B, the sensitivity of the fabricated H2O2 biosensor increased with the amount of MWCNT up to 2.0 mg cm−3 and decreased afterwards. At high MWCNT amounts, the thickening of the electrode surface caused the shedding of the modified material over time. This is the reason why current responses with low repeatability were recorded. Consequently, the optimum amount of MWCNT was selected to be of 1.5 mg cm−3.

The effect of HRP amount on HRP/PLH/MWCNT-IL/GCE biosensor response was also investigated. The H2O2 sensitivity of the fabricated biosensors at different enzyme units is shown in Fig. 1C. The sensitivity increased with the HRP up to 13.9 U and decreased afterwards. The decreased biosensor response at higher HRP units is hypothesized to be the result of an increased electron transfer resistance [34]. Consequently, 13.9 U HRP was selected as the optimal amount.

Determination of the surface morphology of modified biosensor

SEM and EDX techniques were used to evaluate the surface morphology of the modified H2O2 biosensor. The SEM images of MWCNT/SPE, MWCNT-IL/SPE, PLH/MWCNT-IL/SPE, and HRP/PLH/MWCNT-IL/SPE are given in Figs. 2A-a and A-b, respectively. The SEM image of MWCNT-IL/SPE clearly shows the homogeneous distribution of the MWCNT-IL composite (Fig. 2A-b). A more porous morphology was observed with the addition of IL to MWCNT/SPE. Distinct plates and rods which were observed on the PLH/MWCNT- IL/SPE surface (Fig. 2A-c) indicate the well-coated PLH on the electrode surface [5]. The transformation of the rod and plate structures on the PLH/MWCNT-IL/SPE surface into a more ordered form confirms the immobilization of HRP on the electrode surface (Fig. 2A-d). The electrostatic immobilization of the negatively charged HRP on the positively charged PLH/MWCNT-IL/SPE surface can be explained by different surface morphology. The EDX spectrum of HRP/PLH/MWCNT-IL/SPE is also given in Fig. 2B. The C, O, and N peaks in the spectrum can be attributed to histidine on the electrode surface. The P and F peaks in the spectrum are due to the existence of IL.

Fig. 2
figure 2

A SEM images of (a) MWCNT/SPE, (b) MWCNT-IL/SPE, (c) PLH/MWCNT-IL/SPE, (d) HRP/PLH/MWCNT-IL/SPE; B EDX spectra of HRP/PLH/MWCNT-IL/SPE

Electrochemical properties of HRP/PLH/MWCNT-IL/GCE

The electrochemical properties of the electrode surface occurring after each modification step were investigated by CV and EIS methods. For this purpose, cyclic voltammograms and EIS of bare GCE, MWCNT-IL/GCE, PLH/MWCNT-IL/GCE, and HRP/PLH/MWCNT-IL/GCE were recorded in 5.0 mM K3Fe(CN)6/K4Fe(CN)6 (1:1) solution containing 0.1 M KCl (Fig. 3).

Fig. 3
figure 3

A Cyclic voltammogram (scan rate: 0.05 V s−1), B Nyquist plots (a) bare GCE, (b) MWCNT-IL/GCE, (c) PLH/MWCNT-IL/GCE, (d) HRP/PLH/MWCNT-IL/GCE in 5 mM [Fe(CN)6]3−/4− solution containing 0.1 M KCl

Cyclic voltammetry measurements are effective tools to elucidate the electrochemical properties of the modified electrodes after each modification step. In CV measurements, well-defined redox peaks at + 0.420 V and + 0.090 V were observed at bare GCE (Fig. 3A-a). There is a difference of 0.330 V between the anodic and cathodic peak potentials of [Fe(CN)6]3−/4− at bare GCE. Figure 3A-b shows that the peak currents are more intense in MWCNT-IL/GCE than at bare GCE. There is also a difference of 0.090 V between the peak potentials of the redox probe. This result can be attributed to the larger surface area of the MWCNT-IL composite as compared to the GCE electrode surface and the enhanced electron transfer between the redox probe and the electrode [12, 19, 35]. In the voltammogram obtained at PLH/MWCNT-IL/GCE, the current intensity of the redox peaks increased while the peak separation remained the same (Fig. 3A-c). This result can be attributed to the attraction between the positively charged electrode surface and the negatively charged redox probe due to intense electrostatic interaction [5]. Immobilization of HRP enzyme on the electrode surface decreased significantly the redox peak currents, where the difference between the peak potentials was of 0.120 V (Fig. 3A-d). The decrease in the peak currents and the increase in the peak potential difference can be explained by the thickened electrode surface, the barrier effect of Nafion, and the repulsion between the negatively charged HRP and the redox probe. This result constitutes the evidence for the successful immobilization of the HRP on the surface of the electrode.

EIS measurements were carried out in order to investigate more thoroughly the electrochemical properties of the modified electrodes. In EIS measurements, the Rct value of bare GCE was determined as 1089 Ω (Fig. 3B-a). After GCE was modified with MWCNT-IL composite, the Rct value decreased to 109.5 Ω (Fig. 3B-b). When the MWCNT-IL/GCE surface was coated with PLH, this value was determined as 17.01 Ω (Fig. 3B-c). The reason for the significant decrease in Rct value is the strong electrostatic interaction between the positive amino groups in PLH and the negatively charged redox probe. The immobilization of HRP on the PLH/MWCNT-IL/GCE surface increased the Rct value to 374.9 Ω (Fig. 3B-d). This increase can be attributed to the repulsion between the negatively charged HRP and the redox probe. This result is a consequence of the successful immobilization of HRP on PLH/MWCNT-IL/GCE.

Optimizing experimental conditions

The operating potential and pH have been optimized to improve the performance of the H2O2 biosensor. For this purpose, the effects of pH on HRP/PLH/MWCNT-IL/GCE performance were tested at pH values between 5.0 and 8.0. Figure 4A shows that the H2O2 sensitivity of HRP/PLH/MWCNT-IL/GCE increases from pH 5.0 to 6.5 and decreases from pH 6.5 to 8.0. The optimum pH was found to be of pH 6.5 with the highest current response.

Fig. 4
figure 4

The effects of A buffer pH (in 0.05 M PBS, Eapp = −0.1 V) and B applied potential (0.05 M pH 6.5 PBS) on the response of HRP/PLH/MWCNT-IL/GCE

The operating potential values were changed between −0.15 V and 0.0 V to determine the optimum potential (Fig. 4B). As the applied potential changed from −0.15 V to −0.1 V, the response of the biosensor increased and decreased at more positive potential values. The highest sensitivity was obtained at −0.1 V and this working potential was selected as the optimum potential.

Analytical characterization of biosensor

The response of HRP/PLH/MWCNT-IL/GCE to H2O2 was investigated after optimizing the experimental conditions. The amperometric response of HRP/PLH/MWCNT-IL/GCE was measured by successive addition of H2O2 in 0.05 M pH 6.5 PBS at −0.1 V. Figure 5A shows that a stable, rapid, and well-defined amperometric response is observed after each addition of H2O2. This indicates the efficacy of HRP/PLH/MWCNT-IL/GCE for the biorecognition event. The corresponding calibration curve plotted from the i-t data is shown in Fig. 5B. The response of HRP/PLH/MWCNT-IL/GCE to H2O2 showed two different linear operating ranges from 2.0 × 10–2 µM to 9.1 × 10–1 µM and from 1.8 µM to 180 µM. The limit of detection (LOD) and the limit of quantitation (LOQ) of HRP/PLH/MWCNT-IL/GCE were calculated as 2.9 × 10–3 µM and 9.5 × 10–3 µM from Eqs. 3 sb/m and 10 sb/m, respectively. In the given equation, m is the slope of the calibration curve, sb is the standard deviation of the current differences obtained by different measurements (n = 10) of H2O2 solution at constant concentration (2.0 × 10–2 µM). HRP/PLH/MWCNT-IL/GCE reached 95% of the steady-state current in approximatively 15 s.

Fig. 5
figure 5

A Current–time response of HRP/PLH/MWCNT-IL/GCE to the successive additions of H2O2 and B calibration curves of the HRP/PLH/MWCNT-IL/GCE (a) first linear range (b) second linear range (in 0.05 M pH 6.5 PBS, Eapp = −0.1 V)

The reproducibility of the proposed biosensor was assessed by plotting the H2O2 calibration curves of five HRP/PLH/MWCNT-IL/GCEs fabricated through the same process. The slope of the calibration curve was used to calculate the reproducibility of the HRP/PLH/MWCNT-IL/GCE in terms of relative standard deviation (RSD) and it was found to be of 0.71%. This result shows that the presented (bio)sensor has a very good reproducibility.

The response of HRP/PLH/MWCNT-IL/GCE to 0.16 µM H2O2 solution was recorded for ten times in order to prove its working stability. After these 10 measurements, only a 2.52% loss in activity was observed, which confirms the good operating stability of the H2O2 biosensor. The long-term stability of HRP/PLH/MWCNT-IL/GCE was investigated. The H2O2 response of the biosensor was recorded for three weeks. Modified GCEs were stored at + 4 °C during the experiments. HRP/PLH/MWCNT-IL/GCE showed 93.6%, 82.5%, and 80.1% of the amperometric response at the end of days 5, 10, and 21, respectively. Based on this result, it can be confirmed that the biosensor possesses a good long-term stability. The performance parameters of the HRP/PLH/MWCNT-IL/GCE biosensor is given in Table 1.

Table 1 The analytical performance parameters of the developed biosensor

The selectivity of the HRP/PLH/MWCNT-IL/GCE biosensor was investigated in the presence of some possible interferents, such as sodium chloride, glutamic acid, oxalic acid, and glucose. Amperometric measurements of the HRP/PLH/MWCNT-IL/GCE biosensor to successive additions of 0.09 mM H2O2 and 0.09 mM of each interfering substance to a stirred 0.05 M PBS solution were recorded. The results indicated that sodium chloride, glutamic acid, oxalic acid, and glucose provide 3.2%, 0.4%, 0.2%, and 1.7% interference on the H2O2 response of HRP/PLH/MWCNT-IL/GCE, respectively. The low operating potential and the blocking role of Nafion are responsible for the good selectivity.

Determination of H2O2 in oxygenated water

The amount of H2O2 in commercial disinfectant samples was determined by the standard addition method to verify the potential applications of HRP/PLH/MWCNT-IL/GCE. Table 2 shows the amounts of H2O2 in the disinfectant sample determined by the developed (bio)sensor. The recoveries were found to be between 97.2% and 101.7%. These results show that the fabricated H2O2 biosensor is accurate and precise to an acceptable level.

Table 2 H2O2 determination in oxygenated water sample

Table 3 compares the analytical performance of the H2O2 biosensor fabricated by us with some other biosensors documented in the literature. In the table, it can be seen that HRP/PLH/MWCNT-IL/GCE offers high sensitivity, wide operating range, and low detection limit as compared to most of the previous H2O2 biosensors. This improved efficiency is due to the strong immobilization of PLH and HRP, the synergistic effects of the modifying substrates, and the low working potential.

Table 3 Comparison of the analytical performance parameters of different amperometric H2O2 biosensors earlier reported in the literature

Conclusion

The present study proves that the modified electrode surface fabricated by electrochemical polymerization of L-histidine on MWCNT-IL composite provides a suitable environment for electrostatic immobilization of HRP. In order to develop a biosensor with high sensitivity and accuracy, the amount of MWCNT, IL, and enzyme was optimized and suitable operating conditions were determined. The designed biosensor shows wide operating range, high sensitivity, low detection limit, good stability, and good storage stability. These results indicate that the limited biosensor application involving electrostatic immobilization of HRP should be more widely used in the future.

Experimental

HRP (from horseradish type VI with a specific activity of 273 units/mg solid), hydrogen peroxide, sodium monohydrogen phosphate, sodium dihydrogen phosphate, sodium hydroxide, hydrochloric acid, 1-butyl-3-methylimidazolium tetrafluorophosphate, potassium hexacyanoferrate(III), potassium hexacyanoferrate(II), Nafion (5% by weight in lower aliphatic alcohols), oxalic acid, bovine skin gelatin (type B), and L-histidine were purchased from Sigma-Aldrich and used as provided. Potassium chloride, glutamic acid, and sodium chloride were obtained from Merck, and glucose was obtained from Fluka. MWCNTs (O.D. < 8 nm; length 10–30 µm) were obtained from Cheap Tubes Inc. (Brattleboro, USA). Double distilled water was used for all experiments. Commercial oxygenated water samples for the sampling practices were obtained from a local pharmacy in the city of Tekirdağ, Türkiye.

Electrochemical measurements and electrochemical impedance measurements (EIS) were performed by Microstat 400 (Dropsens, Spain) and Gamry Instrument (Reference 300) bipotentiostat/galvanostat instruments, respectively. A three-electrode cell system consisting of a modified GCE (BASi MF 2012, diameter 3 mm) as working electrode, a platinum wire (BASi MW1032, Bioanalytical Systems, Inc., USA) as auxiliary electrode and an Ag/AgCl electrode (BASi MF 2052) as reference electrode were also used. SEM and EDX studies were performed using a Field Emission Scanning Electron Microscope (FE-SEM, QUANTA FEG-250). SPCE (Dropsens C110), containing a carbon working electrode, a silver pseudo reference electrode, and a carbon counter electrode, was used to perform the SEM and EDX measurements.

Electrochemical measurements

In the optimization studies, the amperometric response of the biosensor to H2O2 concentrations between 0.03 mM and 0.2 mM (0.03, 0.05, 0.07, 0.09, 0.1, and 0.2 mM) was measured, and their sensitivities (slopes of the calibration graphs) were compared. CV and EIS studies were performed in 5.0 mM K3Fe(CN)6/K4Fe(CN)6 (1:1) with 0.1 M KCl from − 0.8 V to + 1.0 V at a scan rate of 50 mV s−1. The figures are shown with error bars. The error bars represent the standard deviation of the three measurements.

Fabrication of HRP/PLH/MWCNT-IL/GCE

The GCE surface was treated with Al2O3 paste (0.05 μm) for 10 min. Then the GCE surface was sonicated with distilled water and ethanol-pure water (1:1) mixture for 5 min in an ultrasonic bath. 1 mg of gelatin was suspended in 1 cm3 distilled water and kept on a magnetic stirrer at 35 °C−40 °C for 10 min. A mixture of 1.5 mg MWCNT and 7.5 mm3 IL was added to 1.0 cm3 of gelatin solution and dispersed homogeneously in an ultrasonic bath for 1 h. 5 mm3 of MWCNT-IL-gelatin mixture was dropped onto GCE. The mixture was allowed to dry at room temperature. Cyclic voltammograms of MWCNT-IL/GCE versus Ag/AgCl in solution containing 5 mM L-His (0.05 M pH = 5.0 PBS) were recorded at 50 mV s−1 with 30 cycles in the potential range of 0.0 to – (− 1.5 V). PLH/MWCNT-IL/GCE was cleaned with distilled water and allowed to dry at room temperature. 2.93 mg of HRP were added to 200 mm3 of 0.05 M PBS (pH 7). 3.5 mm3 of the enzyme dilution (4 U/mm3) was pipetted onto PLH/MWCNT-IL/GCE. Finally, 5.0 mm3 0.1% Nafion was added. The HRP/PLH/MWCNT-IL/GCE was stored at + 4 °C overnight (Fig. 6).

Fig. 6
figure 6

Fabrication of HRP/PLH/MWCNT-IL/GCE