Abstract
Background
Infusion tests, which measure resistance to outflow (Rout), are used in selecting patients suspected for idiopathic normal pressure hydrocephalus (iNPH) for shunt surgery. Infusion tests can be performed through an external ventricular drain (EVD). A 24-hour time gap from EVD insertion to an infusion test is a routine practice at our department due to concerns that the surgical procedure might influence the test results in the immediate postoperative period. The objective of the study was to investigate if timing of an intraventricular infusion test influences the results of the test in patients suspected for iNPH.
Methods
Ten patients scheduled for an intraventricular infusion test were included. Measurements of baseline intracranial pressure (ICP) and plateau ICP were obtained during constant rate intraventricular infusion test performed at two time points (1 and 24 h after EVD insertion) and Rout was calculated from these measures and compared within patients.
Results
Eight patients completed both infusion tests. In one of the 18 infusion tests performed, it was not possible to define an ICP plateau and this infusion test was excluded, leaving 7 paired infusion tests. Median Rout was 12.9 mmHg/ml/min (range 7.0–22.0) 1 h after EVD insertion and 11.3 mmHg/ml/min (range 7.8–18.1) after 24 h. Overall, there were no statistically significant differences in Rout (P = 0.83), baseline ICP (P = 0.70), or plateau ICP (P = 0.81) between the recordings performed 1 h and 24 h after EVD insertion. For two of the seven patients with paired infusion tests, there was poor agreement between Rout values at 1 and 24 h.
Conclusion
Overall, Rout estimates do not change significantly between 1 and 24 h after EVD insertion. We therefore propose that infusion tests can be performed shortly after surgery to reduce the period of indwelling EVD and duration of hospitalization.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Patients with idiopathic normal pressure hydrocephalus (iNPH) often present with gait disturbance, dementia, and urinary incontinence, and have ventriculomegaly on CT and MR imaging [14]. A ventriculoperitoneal (VP) shunt can be inserted to relieve these symptoms [5, 13]. Infusion tests are performed to improve selection of patients who will benefit from shunt surgery [8, 9, 25]. Based on the infusion test, it is possible to estimate the absorption of cerebrospinal fluid (CSF) measured as CSF resistance to outflow (Rout), as deficit in the absorption of CSF is assumed to be part of the pathology of iNPH [7]. Because of a positive predictive value as high as 86% for Rout > 12 mmHg/ml/min regarding effect of VP shunt treatment [25], infusion tests are widely accepted in the diagnosis of iNPH [6, 16, 19]. There are different methods for performing infusion tests; the most reliable being outflow resistance obtained with steady state, e.g., constant rate infusion [24]. Infusion test can be performed both via lumbar puncture and via an intraventricular drain [4]. In contrast to the lumbar test, which is performed within minutes after needle insertion, an intraventricular infusion test can be performed sooner or later after drain insertion. This raises the question if timing of the infusion test influences the measurement.
The present study was conducted to investigate if infusion tests performed shortly after surgery provide the same Rout as tests performed after 24 h in the same patient.
Methods
The study was approved by the Danish Data Protection Agency (ID 2012-58-0004) and the National Committee on Health Research Ethics for the Capital Region of Denmark (ID H-15016032). Written informed consent was obtained from all patients included in the study.
Study population
We included 10 patients scheduled for an intraventricular infusion test at the Department of Neurosurgery, Rigshospitalet, between January 2016 and September 2017. All patients had hydrocephalus on CT/MRI, symptoms of normal pressure hydrocephalus (NPH), and were scheduled for intraventricular infusion test at our department as part of their clinical evaluation. Patients were not included if the hydrocephalus nurse was not available for both infusion tests. Our hydrocephalus service includes a clinical nurse specialist allocated specifically to patients with hydrocephalus and other CSF disorders.
Infusion test protocol
Practical aspects of infusion tests are center-specific with no published guideline for the timing of the infusion test after external ventricular drain (EVD) insertion. In our center, we have routinely used a 24-hour time gap between EVD placement and the infusion test due to a concern that intracranial pressure (ICP) is affected by surgery in the immediate postoperative period, and that CSF may egress along the drain until brain tissue glues to the drain yielding falsely low Rout values.
Infusion tests were performed with constant rate of 60 ml/hour intraventricular infusion of artificial cerebrospinal fluid (Ringer Lactate) through an EVD while measuring ICP with the patient in a supine position. The Raumedic Neurovent was inserted as EVD. The device has a tip transducer integrated into the drain and can be used both to measure ICP and perform CSF drainage. The first infusion test was performed 1 h after insertion of the EVD, and repeated with the same set-up 24 h after EVD insertion.
Baseline ICP was measured for 10 min before the infusion was started. When ICP reached a plateau (determined visually by the hydrocephalus nurse), infusion was stopped [10, 12]. After the test was completed, ICP monitoring was continued until it reached baseline level. Infusion was discontinued if the patient experienced headache or nausea, or if ICP increased above 40 mmHg.
ICP was monitored and analyzed with the software package ICM+ (Cambridge Enterprises Ltd, Cambridge, UK) [21].
The resistance to CSF outflow (Rout) was calculated by the formula:
All infusion tests were performed by a specialized hydrocephalus nurse (authors DC and SDJ), which ensured consistency in practical procedure routine.
Study variables
For all patients, we collected baseline data on age, sex, symptomology (gait disturbance, incontinence, and signs of dementia), and duration of symptoms. Results extracted from infusion tests were: Rout, plateau ICP, baseline ICP, duration of test, and discomfort during test. Baseline and plateau ICP was assessed through blinded evaluation of ICP curves (by consensus among authors THA, ALC, AVH, and MJ) (Fig. 1). We also noted treatment consequences of the infusion test as well as clinical status at 3-month follow-up. As the predictive value of Rout is not the focus of this investigation, clinical outcome after surgical treatment was assessed retrospectively by reviewing medical records for the 3-month follow-up with self-reported clinical status.
Statistics
The main study endpoint was the reproducibility of Rout and Rout values from infusion tests after 1 and 24 h and was compared using Wilcoxon signed rank test and a Bland-Altman plot. A P value < 0.05 was considered the statistical level of significance. The statistical software package “R” version 3.2.0 (The R Foundation for Statistical Computing, Vienna, Austria) was used for analyzing data and visualization of the collected data. Sample size calculation was performed for paired sample significance testing with power 0.9 and alpha 0.05, and the results from a previous publication by Tans et al. [24] reporting a standard deviation of 1.2 mmHg/ml/min between repeated measurements of Rout.
Results
Patients
Ten patients (6 men and 4 women) scheduled for an intraventricular infusion test at the Department of Neurosurgery at Copenhagen University Hospital, Rigshospitalet, were included in this study. Median and mean age was 75 years (range 64–81). All patients had signs of communicating hydrocephalus on head CT scan, and had symptoms consistent with NPH (Table 1). Eight of the ten patients had memory difficulties, all ten patients had gait and/or balance disturbance, and all ten patients had urinary incontinence. Of the 10 patients included in the investigation, 8 patients completed both infusion tests. Patients 4 and 5 were the two patients not completing the infusion test planned 24 h after EVD insertion (Table 1). One patient was excluded as an ICP plateau was not clearly visible during the infusion test performed 1 h after EVD placement. The final study population thus comprised 7 patients.
Infusion tests
Median Rout one hour after EVD insertion was 12.9 mmHg/ml/min (range 7.0–22.0), and after 24 h median Rout was 11.3 mmHg/ml/min (range 7.8–18.1). This difference in Rout was not statistically significant (P = 0.83). In 4 out of the 7 patients, the difference in Rout was < 1 mmHg/ml/min. The remaining three patients had the following difference in Rout values between the test after 1 and 24 h + 2.4, − 4.7, and + 5.2 mmHg/ml/min (Table 2). Differences in Rout between repeated infusion tests can be illustrated by a Bland-Altman plot (Fig. 2).
Median baseline ICP one hour after surgery was 11.0 mmHg (range 3.7–16.6), and 24 h after surgery it was 12.6 mmHg (range 4.2–29.8) mmHg (P = 0.70). Baseline ICP changed considerably between the two tests in three patients (patient 4, 5, and 7, see Table 3). In patient 7, this change in baseline ICP from 9.5 to 16.0 mmHg did not translate to a large difference in Rout values, since the plateau ICP changed in parallel from 18.7 to 24.7 mmHg. Patients 4 and 5 had higher baseline ICP during the test performed after 24 h, and subsequent infusion tests were not performed.
The median plateau ICP 1 h after surgery was 25.7 mmHg (range 17.2–36.1), and 24 h after surgery the median plateau ICP was 23.5 mmHg (range 12.3–31.1, P = 0.81, Table 4).
Clinical consequences of infusion tests
Using a cutoff at Rout > 12 mmHg/ml/min as the threshold for pathological Rout, five patients had increased Rout in both their infusion tests and were selected for surgical treatment. Four patients underwent shunt surgery and one patient with an infusion test indicating obstructive hydrocephalus and symptoms of NPH underwent ETV (endoscopic third ventriculostomy). The remaining five patients were not considered eligible for surgery due to Rout below the threshold.
Three patients reported clinical improvement at the 3-month follow-up visit after shunt surgery. One of these reported possible effect on memory but no effect on urinary incontinence or gait, and two reported effect on memory and cognition. The fourth patient with a VP-shunt had no clinical effect after shunt insertion and head CT indicated overdrainage through the shunt system. The patient had no symptoms of overdrainage, and a subsequent head CT showed regression of overdrainage. The patient who underwent ETV experienced short-term effect on gait, and the patient was seen again after 6 months and reported effect on all NPH symptoms (gait, memory, and urinary incontinence). One patient had a complication of ventriculitis and was treated with intrathecal antibiotics.
Discussion
The diagnosis of iNPH relies on clinical presentation, head CT/MRI scans, and additional invasive diagnostic tests, e.g., infusion tests and tap tests [12]. In the present investigation, we tested the hypothesis that Rout differs according to time after EVD insertion and found no significant difference in Rout calculated from ICP measurements during infusion tests performed 1 and 24 h after EVD insertion. For 4 out of 7 patients, the difference in Rout was < 1 mmHg/ml/min between infusion tests performed 1 and 24 h after EVD insertion. The remaining 3 patients had a difference in Rout of more than 2 mmHg/ml/min.
Reproducibility of Rout
We routinely do not perform intraventricular infusion tests earlier than 24 h after EVD insertion due to concerns that surgery might affect testing by introducing biological changes to the CSF system. Theoretically, a CSF spill during the EVD insertion, CSF leakage along the catheter, or ICP changes induced by recent surgery might affect testing. In our experience, the CSF spill is minimal, since the surgeon plugs the EVD as soon as the ventricular system has been reached. In lumbar infusion tests, a falsely high increase in pressure can be measured if there is blockage of tissue at the orifice of the needle [12]. Our study does not support these theoretical concerns. With a reliable result 1 h after EVD insertion, the infusion test can be performed earlier than we do in our current clinical practice. The EVD can then be explanted a few hours after surgery and the patient discharged on the same day. This will shorten length of stay for each patient and reduce infection risk. This study therefore offers a faster expedition of diagnosis, and presumably a shorter waiting period until shunt surgery, which improves the result of shunt surgery for patients with iNPH [3].
Other investigations of repeated infusion tests have also found a high reproducibility of Rout [1, 2, 22,23,24]. Studies are mainly on lumbar infusion tests or test both lumbar and intraventricular but does not perform subanalysis regarding location [23, 24]. Reproducibility of Rout was high in studies repeating infusion tests only 5 min apart [2] and up to 102 days between repeated tests [23]. Tans et al. [24] performed both intraventricular and lumbar infusion tests and performed repeated test in 10 patients. They found a mean difference in Rout of 1.7 ± 1.2 mmHg/ml/min, but the location (intraventricular or lumbar) for the 10 patients and the time gap between repeated infusion tests were not provided. Juniewich et al. [15] support the notion that the infusion test introduces changes to the CSF system. They measured resting ICP before and after infusion, performing both lumbar and intraventricular constant rate infusions tests, and found the differences between post and pre infusion resting ICP (∆ICP) higher than 1 mmHg in 20 infusion tests out of 27. This was considered significant. The mean ∆ICP for the 20 infusion tests was 3.0 ± 0.7 mmHg. However, decreasing ICP was only recorded for about 10 min after infusion and not necessarily until ICP returned to resting ICP. In our study, 3 patients exhibited an increase in baseline ICP before the second infusion test, and 2 exhibited decrease in baseline ICP, but the difference in baseline ICP for the entire group was not statistically significant. We also found no significant difference for plateau ICP between infusions tests. We can however not exclude the possibility that the infusion tests performed 24 h after EVD insertion are affected by the previous infusion test. Therefore, we assessed the agreement between replicate measurements by a Bland-Altman plot (Fig. 2). For patients 1 and 2, the agreement between paired infusion tests was poor, though the plot illustrates good agreement for most of the patients. Our data thus support that the infusion test does not need to be delayed beyond 1 h after drain insertion.
Treatment and pathological Rout
The diagnosis of iNPH can be a difficult, making infusion tests an important tool in selecting patients who would benefit from shunt surgery. Malm et al. [18] performed lumbar infusion tests in normal individuals aged 60–82 years, and found a mean Rout of 11 mmHg/ml/min. In a study with intraventricular infusion tests, a Rout below 10 mmHg min/ml was considered physiological, Rout 10–13 mmHg min/ml borderline and Rout above 13 mmHg/min/ml pathological [20].
Eide et al. [11] performed constant rate lumbar infusion tests while measuring ICP in both the lumbar cerebrospinal space and in the brain parenchyma. They reported a higher lumbar Rout than ventricular Rout with a mean difference of 2.7 mmHg/ml/min. This might be explained by differences in compliance of the two compartments. However, it suggests that differentiated thresholds for abnormal Rout should be adapted for intraventricular and lumbar infusion tests.
There is no complete consensus regarding the treatment threshold of Rout. Previous publications have concluded that shunt surgery based on Rout values above 18 mmHg/ml/min improves outcome [6, 7]. However, other investigators recommend a lower threshold at 12 or 13 mmHg/ml/min [11, 15, 17, 20]. Setting the limit at 18 mmHg/ml/min yields a lower sensitivity, possibly excluding patients who would benefit from a shunt [17]. The infusion tests for patients 1 and 2 had a difference in Rout of + 5.7 and − 4.7 mmHg/ml/min, respectively. This difference seems large enough to potentially change the clinical consequence drawn from the tests.
We performed power analysis to detect a difference of 2 mmHg/ml/min, as we believe a significant difference should be of at least 2–3 mmHg/ml/min. We have not chosen a specific value for a significant difference since a difference in Rout matters more when in the area of 15–18 mmHg/ml/min where a pathological Rout is unsure. A difference in Rout above 18 mmHg/ml/min or below 12–15 mmHg/ml/min is less essential since the clinical conclusion on pathological or not does change. For the patients with a large difference in Rout , it can certainly be debated that though the statistical value is not influenced, the clinical value might be. Swallow et al. [23] commented that the parameters estimated with infusion testing have to be considered as a range in which the actual value of the patient lies rather than as a fixed value. We concur with this statement, since analysis of the ICP curve partly depends on clinical assessment, and the plateau is not always easily identified. The risk of finding different results seems to be related to technical variability and inter-observer differences in analyzing the ICP curves.
Limitations
It was not possible to repeat infusion tests in all patients. Two patients had infusion tests with a pattern of obstructive hydrocephalus with a steep increase in ICP, and no ICP plateau. This study was intended for patients with normal pressure hydrocephalus only.
We believe that inter-observer variation was an important limitation in the analysis of ICP curves.
Conclusion
We compared Rout estimated through infusion tests performed 1 and 24 h after EVD insertion. For two of the seven patients, there was poor agreement between paired infusion tests, though overall there was no significant difference in Rout between paired tests. We therefore propose that infusion tests can be performed shortly after EVD insertion, though further studies seem needed to confirm this.
Abbreviations
- CSF:
-
Cerebrospinal fluid
- ETV:
-
Endoscopic third ventriculostomy
- EVD:
-
External ventricular drain
- ICP:
-
Intracranial pressure
- ∆ICP:
-
Differences between post and pre infusion resting ICP
- iNPH:
-
Idiopathic normal pressure hydrocephalus
- Rout :
-
Resistance to outflow
- VP-shunt:
-
Ventriculoperitoneal shunt
References
Andersson N, Malm J, Backlund T, Eklund A (2005) Assessment of cerebrospinal fluid outflow conductance using constant-pressure infusion-a method with real time estimation of reliability. Physiol Meas 26:1137–1148. https://doi.org/10.1088/0967-3334/26/6/022
Andersson K, Manchester IR, Laurell K, Giuliana Cesarini K, Malm J, Eklund A (2013) Measurement of CSF dynamics with oscillating pressure infusion. Acta Neurol Scand 128:17–23. https://doi.org/10.1111/ane.12063
Andren K, Wikkelso C, Tisell M, Hellstrom P (2014) Natural course of idiopathic normal pressure hydrocephalus. J Neurol Neurosurg Psychiatry 85:806–810. https://doi.org/10.1136/jnnp-2013-306117
Bech-Azeddine R, Gjerris F, Waldemar G, Czosnyka M, Juhler M (2005) Intraventricular or lumbar infusion test in adult communicating hydrocephalus? Practical consequences and clinical outcome of shunt operation. Acta Neurochir 147:1027–1035; discussion 1035-1026. https://doi.org/10.1007/s00701-005-0589-0
Bergsneider M, Black PM, Klinge P, Marmarou A, Relkin N (2005) Surgical management of idiopathic normal-pressure hydrocephalus. Neurosurgery 57:S29–S39 discussion ii-v
Boon AJ, Tans JT, Delwel EJ, Egeler-Peerdeman SM, Hanlo PW, Wurzer HA, Avezaat CJ, de Jong DA, Gooskens RH, Hermans J (1997) Dutch normal-pressure hydrocephalus study: prediction of outcome after shunting by resistance to outflow of cerebrospinal fluid. J Neurosurg 87:687–693. https://doi.org/10.3171/jns.1997.87.5.0687
Boon AJ, Tans JT, Delwel EJ, Egeler-Peerdeman SM, Hanlo PW, Wurzer HA, Hermans J (2000) The Dutch normal-pressure hydrocephalus study. How to select patients for shunting? An analysis of four diagnostic criteria. Surg Neurol 53:201–207
Borgesen SE (1984) Conductance to outflow of CSF in normal pressure hydrocephalus. Acta Neurochir 71:1–45
Borgesen SE, Gjerris F (1982) The predictive value of conductance to outflow of CSF in normal pressure hydrocephalus. Brain 105:65–86. https://doi.org/10.1093/brain/105.1.65
Czosnyka M, Whitehouse H, Smielewski P, Simac S, Pickard JD (1996) Testing of cerebrospinal compensatory reserve in shunted and non-shunted patients: a guide to interpretation based on an observational study. J Neurol Neurosurg Psychiatry 60:549–558. https://doi.org/10.1136/jnnp.60.5.549
Eide PK, Fremming AD, Sorteberg A (2003) Lack of relationship between resistance to cerebrospinal fluid outflow and intracranial pressure in normal pressure hydrocephalus. Acta Neurol Scand 108:381–388
Eklund A, Smielewski P, Chambers I, Alperin N, Malm J, Czosnyka M, Marmarou A (2007) Assessment of cerebrospinal fluid outflow resistance. Med Biol Eng Comput 45:719–735. https://doi.org/10.1007/s11517-007-0199-5
Ghosh S, Lippa C (2014) Diagnosis and prognosis in idiopathic normal pressure hydrocephalus. Am J Alzheimers Dis Other Dement 29:583–589. https://doi.org/10.1177/1533317514523485
Hellstrom P, Klinge P, Tans J, Wikkelso C (2012) A new scale for assessment of severity and outcome in iNPH. Acta Neurol Scand 126:229–237. https://doi.org/10.1111/j.1600-0404.2012.01677.x
Juniewicz H, Kasprowicz M, Czosnyka M, Czosnyka Z, Gizewski S, Dzik M, Pickard JD (2005) Analysis of intracranial pressure during and after the infusion test in patients with communicating hydrocephalus. Physiol Meas 26:1039–1048. https://doi.org/10.1088/0967-3334/26/6/013
Kahlon B, Sundbarg G, Rehncrona S (2002) Comparison between the lumbar infusion and CSF tap tests to predict outcome after shunt surgery in suspected normal pressure hydrocephalus. J Neurol Neurosurg Psychiatry 73:721–726
Kim DJ, Kim H, Kim YT, Yoon BC, Czosnyka Z, Park KW, Czosnyka M (2015) Thresholds of resistance to CSF outflow in predicting shunt responsiveness. Neurol Res 37:332–340. https://doi.org/10.1179/1743132814y.0000000454
Malm J, Jacobsson J, Birgander R, Eklund A (2011) Reference values for CSF outflow resistance and intracranial pressure in healthy elderly. Neurology 76:903–909. https://doi.org/10.1212/WNL.0b013e31820f2dd0
Marmarou A, Bergsneider M, Klinge P, Relkin N, Black PM (2005) The value of supplemental prognostic tests for the preoperative assessment of idiopathic normal-pressure hydrocephalus. Neurosurgery 57:S17–S28 discussion ii-v
Meier U, Bartels P (2002) The importance of the intrathecal infusion test in the diagnosis of normal pressure hydrocephalus. J Clin Neurosci 9:260–267
Smielewski P, Czosnyka Z, Kasprowicz M, Pickard JD, Czosnyka M (2012) ICM+: a versatile software for assessment of CSF dynamics. Acta Neurochir Suppl 114:75–79. https://doi.org/10.1007/978-3-7091-0956-4_13
Sundstrom N, Andersson K, Marmarou A, Malm J, Eklund A (2010) Comparison between 3 infusion methods to measure cerebrospinal fluid outflow conductance. J Neurosurg 113:1294–1303. https://doi.org/10.3171/2010.8.jns10157
Swallow DM, Fellner N, Varsos GV, Czosnyka M, Smielewski P, Pickard JD, Czosnyka Z (2014) Repeatability of cerebrospinal fluid constant rate infusion study. Acta Neurol Scand 130:131–138. https://doi.org/10.1111/ane.12246
Tans JT, Poortvliet DC (1985) CSF outflow resistance and pressure-volume index determined by steady-state and bolus infusions. Clin Neurol Neurosurg 87:159–165
Wikkelso C, Hellstrom P, Klinge PM, Tans JT (2013) The European iNPH multicentre study on the predictive values of resistance to CSF outflow and the CSF tap test in patients with idiopathic normal pressure hydrocephalus. J Neurol Neurosurg Psychiatry 84:562–568. https://doi.org/10.1136/jnnp-2012-303314
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.
Informed consent
Informed consent was obtained from all individual participants included in the study.
Additional information
Comments
LP infusion test may be performed in all cases. It is unsafe when on MRI lack of communication between cranial and lumbar CSF space is seen. Otherwise LP infusion test is safe (no serious complications in our hospital, where around 3000 LP infusion tests were performed, infection rate <1%). It gives information on whole CSF dynamics, including Rout, elasticity, baseline pressure, compensatory reserve at baseline, slope of amplitude-pressure line, etc. LP test is much less invasive than the test through EVD and much cheaper- information is almost the same.
Marek Czosnyka
Cambridge, UK
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
This article is part of the Topical Collection on CSF Circulation
Rights and permissions
About this article
Cite this article
Andreasen, T.H., Lilja-Cyron, A., Holst, A.V. et al. Timing of intraventricular infusion test for diagnosing idiopathic normal pressure hydrocephalus. Acta Neurochir 162, 1011–1017 (2020). https://doi.org/10.1007/s00701-019-04168-w
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00701-019-04168-w