Introduction

According to the World Health Organization (WHO), dengue has been identified as one of the ten threats to global health in 2019 [1]. This disease has become a main public health concern over last years, overcoming the status of neglected tropical disease [2, 3]. Dengue is transmitted by Aedes mosquitoes [4], infected by Flaviviridae virus (DENV) [5], and has four different serotypes (DENV-1, DENV-2, DENV-3, and DENV-4), which makes it difficult to create a tetravalent vaccine [2]. Over 3 million cases occurred in Brazil, only in 2022 [6]. To change this situation, early diagnosis is required with low cost, fast procedures, and high selectivity.

Dengue virus leads the infected cells to produce ten proteins: three structural proteins (capsid, membrane, and enveloped glycoprotein, or protein E) and seven non-structural proteins (NS1, NS2A, NS2B, NS3, NS4A, NS4B, and NS5) [4]. Between them, NS1 is the only glycoprotein, with approximately 46–50 kDa, secreted from all serotypes [7] that is released in the blood and plays several roles in viral replication and immune evasion [8, 9]. This unique feature enables the detection of NS1, making it a biomarker of dengue during acute infection. High levels of NS1 protein can be detected since the first symptoms up to 9 days of disease [10]. Alcon et al. reported NS1 levels in human serum from 0.04 to 2 µg/ml in the first and 0.01 to 2 µg/ml in the second infection [11].

Nowadays, the clinical tests are based on the detection of NS1 using IgM and IgG antibodies in enzyme-linked immunosorbent assay (ELISA), a current standard technique or nucleic acid amplification tests for RNA [8]. All these laboratory methodologies are efficient; however, they require expensive equipment and specific and technical expertise [9, 12].

Sensitive platforms using electrochemical techniques for biosensors have made great progress, mainly due to the efforts of different research fields. Electrochemical, and especially potentiometric, biosensors are one of the most promising classes with relatively simple devices, low fabrication costs, and easy miniaturization processes [9]. Electrochemical impedance spectroscopy (EIS) is widely used to investigate modified surfaces capable of detecting biomarkers. However, the need for a redox probe in solution to obtain the data reduces the possible applications. Cecchetto et al. reported a sensitive electrochemical capacitance spectroscopy (ECS) technique for dengue detection through antibodies [12]. One other promising possibility is the use of specific and artificial DNA sequences (aptamers) to interact with the biomarker [13]. Compared with immunoassay-based methods, which are strongly dependent on antibody quality, aptamers exhibit a high specificity, affinity, excellent stability, easy modification, and low cost to develop the biosensing platform [14].

In a previous study [15], the focus was investigating the aptamer interaction with two different NS1 serotypes using impedimetric detection. However, in this method, it was necessary to add a redox molecule to the solution to verify and quantify the interaction. In this study, gold electrodes were modified with DNA aptamers to detect NS1 protein using ECS. This device was based on a self-assembled monolayer (SAM) where the aptamers were co-immobilized with 6-(ferrocenyl)hexanethiol (FCH) and 6-mercapto-1-hexanol (MCH). The probe surface density was optimized, and the biosensor performance was tested in phosphate-buffered saline (PBS) and undiluted human serum.

Experimental

Materials

The Serotype 4 NS1 Protein (Abcam, ab181957) was used as a target. The DNA aptamer sequence (5′-HS(CH2)6–TTTTT ACTAGGTTGCAGGGGACTGCTCGGGATTGCGGATCAACCTAGTTGCTTCTCTCGTATGAT–3′), Trizma base (C4H11NO3, ≥ 99%), magnesium chloride (MgCl2, ≥ 98%), octaethylene glycol monododecyl ether (C12E8, ≥ 97%), 6-(ferrocenyl)hexanethiol (C16H22FeS, ≥ 97%), 6-mercapto-1-hexanol (HS(CH2)6OH, ≥ 97%), bovine serum albumin (BSA), sodium phosphate dibasic (Na2HPO4, ≥ 99%), potassium phosphate monobasic (KH2PO4, ≥ 99%), and ethylenediaminetetraacetic acid (C10H16N2O8, ≥ 99%) were purchased from Sigma-Aldrich. Sulfuric acid (H2SO4, ≥ 95%) and sodium chloride (NaCl, ≥ 99%) were purchased from Synth; sodium hydroxide (NaOH, ≥ 97%) was purchased from Dinamica and potassium chloride (KCl, ≥ 99%) was purchased from Cinetica. Ultra-pure water (18.2 MΩ.cm, Merck Millipore Direct-Q 5UV) was used to prepare all aqueous solutions. The Biopak Polisher filter (Merck Millipore, CDUFB-001) was used in Millipore equipment to remove all nuclease in ultra-pure water. Commercial human serum was purchased from Sigma-Aldrich (human male AB plasma, USA origin, sterile-filtered) and stored in the freezer (− 20 °C). All the chemicals used in this study were analytical or commercial grade.

Electrode preparation

All procedures were developed by your group and published in previous works [15]. Briefly, gold electrodes with a 2 mm diameter (Metrohm) were desorbed using cyclic voltammetry in NaOH (0.5 M) from − 2.0 to − 0.8 mV (vs. Ag/AgCl) for 30 cycles. Thereafter, the surface was polished with different alumina sizes (1 and 0.3 µm), followed by sonication in water for 5 min between each polish. The electrochemical cleaning was performed in H2SO4 (0.5 M) by scanning the potential from − 0.5 to 1.1 V (vs. Ag/AgCl) for 25 cycles [16]. The effective surface was calculated by the reduction peak area and used to normalize all data [17, 18]. Clean electrodes were dried with N2.

Electrode functionalization

DNA aptamer was incubated at 94 °C in a water bath for 10 min and immediately diluted to 1 µg/mL in Selex solution (20 mM Trizma pH 7.5 + 150 mM NaCl + 1 mM MgCl2 + 2.7 mM KCl + 0.005% C12E8) at room temperature (24 °C) for 40 min [13, 15]. The Selex solution with aptamer was mixed with 1 mM 6-(ferrocenyl)hexanethiol (FCH) (10% ethanol) in different molar fractions. The fractions are presented in this work as (Apt / (total thiol)), where total thiol corresponds to the sum between aptamer and FHC. The electrodes were co-immobilized with aptamer and FHC in 150 µl of Selex solution for 18 h, in a humidity chamber, at 4 °C [19].

Modified gold electrodes were rinsed with three wash solutions (200 mM phosphate buffer (PB); 10 mM PB + 100 mM ethylenediaminetetraacetic acid (EDTA); 10 mM PB) for 5 min each. Following this procedure, the electrodes were backfilled with 1 mM 6-mercapto-1-hexanol (MCH, 10% ethanol) for 1 h to block the remaining spaces on the surface. The electrodes were rinsed with ultra-pure water (Milli-Q) and immediately placed in PBS (137 mM NaCl + 2.7 mM KCl + 1 mM Na2HPO4 + 0.18 mM KH2PO4, pH 7.4) for 2 h to stabilize [16]. Finally, the gold electrodes were exposed to BSA (0.1%) or serum for 30 min. Figure 1a presents an illustration of SAM architecture. The capacitance responses from blocked electrodes with BSA, or serum, were used as a baseline for all NS1 responses. Figure 1b exhibits the capacitive response after the BSA blocking and interaction with 1 µg/mL NS1 protein.

Fig. 1
figure 1

Schematic illustration of self-assemble monolayer in a. The illustration is out of scale and does not represent the real aptamer folding. In b, the capacitive signal after the BSA blocking. The signal decreases upon the interaction with 1 µg/mL of NS1 protein. In detail, the decrease in capacitance curve

Measurement and apparatus

All measurements were obtained using the potentiostat (Metrohm/Autolab PGSTAT302) with a three-electrode cell configuration: Ag/AgCl (FACTE) as the reference electrode, platinum wire (Metrohm) as the counter electrode, and gold electrode as the working electrode. All procedures were carried out in a 15-ml electrochemical cell with PBS buffer. The ECE measurement was performed in the 100-kHz to 100-mHz frequency range with a 10 mV a.c. potential difference superimposed on a d.c. bias of the open circuit potential (OCP) configuration in PBS buffer. Autolab software (Nova 2.1) was used. The OCP measurement can be used to detect the potential variation due to the changes in biolayer capacitance [19]. To analyze the charge on the surface, the OCP values obtained for SAM/BSA response were fixed during all procedures for each electrode. All data were originally obtained as complex impedance signals (Z*(\(\omega )\)), Eq. 1. The capacitance complex signals (C*(\(\omega\))) were obtained by deriving the last equation in two new equations (Eqs. 2 and 3):

$${Z}^{*}\left(\omega \right)= \frac{1}{i\omega {C}^{*}\left(\omega \right)}$$
(1)
$${C}^{{^\prime}{^\prime}}\left(\omega \right)= \frac{{Z}{^\prime}}{\omega {\left|{Z}^{*}\right|}^{2}}$$
(2)
$$C{^\prime}\left(\omega \right)= \frac{{-Z}^{{^\prime}{^\prime}}}{\omega {\left|{Z}^{*}\right|}^{2}}$$
(3)

where C′ is the real capacitance; C″ is the imaginary capacitance [20]. These data generated a capacitive spectrum (Fig. 1b).

The biosensor surface was characterized via atomic force microscopy (AFM) using a Shimadzu SPM-9600 Scanning Probe Microscopy (Shimadzu Corporation, Japan). The analysis was obtained in air at 25 °C using the same gold electrode.

Results and discussion

Electrochemical capacitance spectroscopy analysis

The EIS is composed of two processes: Faradaic and non-Faradaic. The Faradaic process needs a redox specie (e.g., the redox couple [Fe(CN)6]4−/[Fe(CN)6]3−) to investigate the interactions on the electrode surface [16]. The non-Faradaic measurements do not require any external redox probe, indicating the existence of dielectric (associated with a double layer) contribution and making this technique able to analyze small changes in the interface charge [21]. In this case, the technique uses a redox probe (FCH) co-immobilized with the biological receptive layer. The capacitance values are related to this molecule’s coverage and concentration. The interaction between the aptasensor and target changes the charge over the surface, limits the redox process, and reduces the capacitive response [12].

To analyze the capacitive data, it is important to find the highest phase angle by plotting the dose-dependent response from the capacitive signal change over frequency [22]. The phase angle is presented in Fig. 2a and indicates the maximum capacitive response at ~ 1.7 Hz, which corresponds to 86°. Fixing the analysis in 1.7 Hz, Fig. 2b shows the real capacitance (C′) reduction when the NS1 concentration increases from 1 pg/mL to 1 µg/mL, indicating a successful binding between protein and aptamer. Figure 2c exhibits the alteration in dielectric properties on the biolayer capacitance for different protein concentrations. The interface SAM/BSA indicates 1.19 µF as capacitance response, and this value decreases to 1.15 µF with the addition of 1 µg/mL NS1 protein, which corresponds to 40-nF variance in real capacitance (− ΔC′).

Fig. 2
figure 2

Electrochemical capacitance spectroscopy measurements in PBS buffer. In a phase response as a function of frequency, b real capacitance response (C′) from different NS1 concentrations with a highlight at 1.7 Hz in c

Surface functionalization

The surface modification for ECS measurements was tested with two different protocols. First, the surface was co-immobilized with DNA aptamer and FCH. After 18 h, the electrodes were backfilled with MCH. This result is presented in Fig. 3 and corresponds to aptamer + FCH/MCH. In the second protocol, aptamer and MCH were co-immobilization and backfilled with FCH (aptamer + MCH/FCH).

Fig. 3
figure 3

Change in real capacitance (− ΔC′) as a function of NS1 concentrations for two different protocols for co-immobilization and backfilling: Selex + FCH with MCH in backfilling and Selex + MCH with FCH in backfilling

Figure 3 shows the change in real capacitance [− ΔC′ = (CNS1 − CSAM)] over NS1 concentrations from 10 pg/mL to 1 \(\upmu\) g/mL. When the surface is co-immobilized with aptamer + FCH, the NS1 increase changes the capacitance response. However, the aptamer + MCH co-immobilization with FCH just added in the backfilling does not show any increase in capacitance response. This result is related to the presence of FCH on the SAM; once it takes more time to interact with the gold surface and organize as part of the SAM [23], the backfilling is not appropriate.

When FCH is correctly immobilized on the gold surface, it can accumulate charge in the interface, increasing the capacitive value and generating a high electron transfer rate between interface and surfac [24, 25], which can be observed with the change in − ΔC′ when NS1 interacts with aptamer.

Surface characterization and protein interaction were also characterized using atomic force microscopy (AFM). The images are shown in Figure S1 (Supplementary information), where the clean bare electrode (Figure S1(a)) exhibited an 83.53-nm peak. After surface immobilization, backfilling, and BSA blocking, the peak value increased to 145.10 nm, indicating that the immobilization of biomolecules on the gold surface increased the surface roughness due to the different sizes of the molecules (Figure S1(b)). After 1 µg/mL NS1 interaction, the peak value decreased to 101.34 nm. This change indicated that the protein interaction on the surface occupies the remaining spaces and brings the aptamers close to the electrode surface, resulting in a smoother surface (Figure S1 (c)).

Surface density

Auletta et al. report that the proportion between FCH and aptamers influences the interaction between target and probe [26]. Electrodes were exposed to different molar fractions of FCH and aptamers (1:10, 1:50, 1: 100, 1: 150, 1: 200) in immobilization solution and tested with 100 pg/mL NS1 diluted in buffer solution (PBS + BSA (0.1%), pH 7.4). Figure 4 shows the influence of molar fraction in capacitance response as an important step for the biosensor performance, once FCH concentration affects the SAM organization. The surface coverage does not need a large amount of aptamer (1:10) or FCH (1:200) because the SAM organization depends on the alkanethiol chain and van der Walls interactions, making an intermolecular film in the gold surface [27]. The ratio 1:10 showed the lowest capacitance signal (2 nF), which indicates an excess of aptamer and high repulsion between DNA oligonucleotide aptamer chains, generating an unstable monolayer. On the other hand, the ratio 1:200 presented a low capacitance signal (4 nF) due to the low amount of aptamer. The most stable biolayer was performed by 1: 150 molar fraction (with 11 nF) and was used for biosensor development.

Fig. 4
figure 4

Change in real capacitance (-ΔC’) for aptamer/FCH molar ratio (aptamer: total thiol) with 100 pg/mL NS1 in PBS (pH 7.4) + BSA (0.1%) solution. Error bars show the standard deviation for at least three measurements from different electrodes

Analytical curve and negative control

The optimized ratio was used to test the system and obtain a calibration plot. The performance of this biosensing platform was tested from 10 pg to 1 µg/mL in buffer (Fig. 5a) and commercial undiluted human serum (Fig. 5b). For each measurement, the SAM was incubated with 100 µl of NS1 concentrations for 30 min, followed by washing with PBS to remove unbound NS1.

Fig. 5
figure 5

Analytical curve for NS1 diluted with a linear response in a buffer; b commercial undiluted human serum; c change in real capacitance (− ΔC′) for each NS1 concentration in buffer (open black square) and undiluted serum (open red circle). Error bars show the standard deviation from at least three individual electrodes. d Negative control shows the change in real capacitance (− ΔC′) for 1 ng/mL of NS1 and envelope protein (protein E) in serum

Figure 5c shows the calibration plot as − ΔC′ increases for NS1 concentrations in buffer and undiluted serum. The platform has a linear behavior from 10 pg/mL to 1 µg/mL (buffer) and from 10 pg/mL to 100 ng/mL (undiluted serum). Serum has a high amount of protein that can saturate an electrode surface, reducing the linear range.

The limit of detection (LOD) was calculated as 3.3 × σ/b, where σ represents standard deviation, and b represents the slope of the analytical curve fitting on a linear range [28]. In buffer, the biosensor presented a LOD of 30.9 fg/mL with 3-nF sensitivity per decade. In undiluted serum, the platform reached 41.8 fg/mL and a sensitivity of 4 nF per decade. The relative standard deviation (RSD) was calculated from both calibration curves (0.09% and 0.1% from PBS and serum, respectively). These results exhibited a high sensibility in comparison with other techniques reported in the literature. Table S1 (supplementary material) compares the results from the last 5 years of different platforms to detect NS1 in the clinical range for first (0.04 to 2 µg/mL) and second (0.01 to 2 µg/mL) infection.

The negative control was performed with 1 ng/mL E protein in undiluted serum. Envelope protein (E protein) is a dengue protein expressed in infected cells and is responsible for virus protection and target cell binding [29]. Figure 5d shows the difference between the NS1 response with 14 nF in comparison to the E protein (8 nF) with a 43% decrease in the signal response. This difference validates the selectivity characteristic of this aptamer to detect Dengue NS1 protein.

It is crucial to acknowledge the current limitations of this platform, including its non-disposable electrodes, and the need for improved NS1 detection in real human samples. All the measurements were performed with at least six different electrodes and performed at two different times. The results present a 5% difference between batches. The aptamer platform storage lifetime was previously tested using other sequences, in the same conditions, and presented a 40% signal reduction after 2 weeks. Despite these limitations, the proposed method for capacitive biosensors enables dengue and other disease detection using a label-free system.

Conclusion

A new platform that selectively quantifies NS1 protein using DNA aptamer on the gold electrode was developed. As a highlight, it was demonstrated that electrochemical capacitive spectroscopy, using a simple approach with FCH on the SAM, was able to detect NS1 in PBS and in serum biological matrix presenting a linear detection range from 0.01 to 1000 ng/mL in buffer and 0.01 to 100 ng/mL in undiluted serum. The biosensor achieved low limits of detection of 30.9 fg/mL and 41.8 fg/mL in buffer and undiluted serum, respectively. The main advantage of this platform is that there is no need to add a redox molecule to the solution, which is particularly useful to reduce patient sample manipulation, increasing applications. Moreover, the ECS platform exhibited a rapid response time, easy operation, and cost-effectiveness.