Abstract
The authors describe an electrochemical method for the determination of the blood clotting factor IX (FIX). A nanogapped dielectrode (with a < 100 nm junction) was modified with antibody against FIX, and the resulting system was characterized by both impedance spectroscopy and voltammetry. In order to attain the improved sensitivity, gold nanoparticles were electrostatically attached to FIX. The current to voltage (I-V) measurement was carried out from 1 V to 5 V, where the entire calibration plot with 5 V was taken. This results in a limit of detection as low of 1 pM, which is much lower compared to the real concentration of FIX (87 nM) in human blood serum. The analytical range extends from 1 pM to 0.1 μM. The electrode is highly specific over other serum proteins.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Healthy human plasma contain 3–5 mg.L−1 of FIX [1], required for normal coagulation process and a lower level of FIX leads to a bleeding deficiency [2,3,4]. FIX deficiency can be detected in human plasma using different methods such as coagulation screening test, coagulation factor assay, and bleeding test [4, 5]. Radioimmunoassay [6], ELISA [7], liquid chromatography [8], gel electrophoresis [9], Surface Plasmon Resonance [10], waveguide mode sensing and BioDVD platform [11] have also been used to quantify the level of FIX in the test sample. Generally antibody and aptamer are used as the probes against FIX in the above methods. To overcome the current challenges, such as usage of minimal sample, simplified system for point-of-care testing, the current study has been launched. For point-of-care testing, it is highly expected that the system better be feasible with both AC and DC currents. In this study, an analytical system has demonstrated to analyze FIX using antibody as the probe. For this analysis, nanogapped system between two electrodes was fabricated with the gap sized below 100 nm as stated before [12] and used for the determination and discrimination of FIX by antibody conjugated gold nanoparticle (GNPs).
GNP mediated sensing proved to enhance the detection of biological elements due to the higher surface area to accommodate a large number of molecules [13, 14]. Additionally, GNPs have good catalytic property, biocompatibility, conductivity and promote the limit of detection [15,16,17,18]. Lakshmipriya et al. [1] have detected FIX using aptamer by waveguide mode sensor and improved the detection using streptavidin-conjugated GNPs. Biomolecules can be easily immobilized on GNP through electrostatic interaction or physical adsorption and also by chemical functionalization. For the current study, the antigen (FIX) was immobilized on the GNPs to improve the detection on the FIX antibody immobilized polysilicon surface having dielectrode. Interactive analyses were performed on the nanogap created between the electrodes with an increment of current proportionally when biomolecules are bound to the surface, similar to other available electric sensors [19]. Our analytical system is suitable for both impedance and current versus voltage (IV) measurements, demonstrated here for its feasibility with the analysis of FIX and its antibody.
Materials and methods
Reagents and biomolecules
Aminopropyl-trimethoxysliane (APTES; http://www.sigmaaldrich.com/catalog/product/aldrich/281778?lang=en®ion=MY), Glutaraldehyde solution (50 wt.% in H2O; http://www.sigmaaldrich.com/catalog/product/sial/340855?lang=en®ion=MY&gclid=CPSokJDmo9QCFdOKaAodAQkPUg) and 10 mM Phosphate Buffered Saline (PBS; pH 7.4, at 25 °C) containing NaCl 138 mM; KCl - 2.7 mM (http://www.sigmaaldrich.com/catalog/product/sigma/p3813?lang=en®ion=MY&gclid=CNm1lqHmo9QCFUGVaAodlpkEeQ), were purchased from Sigma-Aldrich (USA). Ethanolamine was procured from Fisher Scientific, UK (https://www.fishersci.co.uk/shop/products/ethanolamine-extra-pure-slr-fisher-chemical/10508040). Human FIX was purchased from Abcam (Malaysia; http://www.abcam.com/natural-human-factor-ix-protein-ab62544.html). FIX antibody was purchased from were purchased from Sigma-Aldrich, USA (http://www.sigmaaldrich.com/catalog/product/sigma/f2645?lang=en®ion=MY), Gold nanoparticle was purchased from Sigma-Aldrich, USA (http://www.sigmaaldrich.com/catalog/product/aldrich/741957?lang=en®ion=MY). The purchased reagents and biomolecules were used without further purification. All chemical and biological reactions on the fabricated active polysilicon surface were done at ambient room temperature. Fabrication of nanogapped dielectric polysilicon surface was carried out as stated before [12].
Chemical functionalization on nanogapped dielectric polysilicon surface
For the molecular assembly on the nanogapped dielectric polysilicon surface, chemical functionalization was performed. Briefly, nanogapped chip was firstly cleaned by deionized water. To establish a self-assembled monolayer of amine terminated groups, 100 μL of 100 mM APTES was dropped on the nanogap and incubated for 1 h at room temperature. The unreacted APTES was removed by rinsing with 5 fold volume by absolute ethanol followed by deionized water. 100 μL of 2.5% Glutaraldehyde was added on amine-modified surface in order to generate a monolayer of aldehyde terminated group and incubated for 30 min. Under this condition, one aldehyde end of glutaraldehyde reacts with amino group of APTES and other end of glutaraldehyde remains available freely. The unreacted glutaraldehyde was removed by rinsing with 5 fold volume of 10 mM PBS buffer.
Biomolecular interactive analysis
After the above surface chemical functionalization, 5 μL of FIX antibody (100 nM), which could cover the active region was dropped on the nanogap having the aldehyde terminated surface and left for 30 min. Unreacted antibody was removed by washing with 10 fold volume of 10 mM PBS buffer (pH 7.4). Then, 10 μL of 1 M Ethanolamine as a masking agent was applied for blocking the aldehyde terminated groups which are not attached by antibodies. After masking, 5 μL of 100 nM of FIX was applied on the chip and kept at room temperature for 30 min and then washed with 10 fold volume of 10 mM PBS buffer (pH 7.4). Readings were taken by impedance and current versus voltage (IV) measurements.
High sensitive interaction of gold-FIX conjugates on nanogapped dielectric junction
To obtain high-sensitive detection, FIX was conjugated electrostatically on GNPs and then added on the FIX antibody bound surface as performed above. For that, the same surface modifications were carried out as in the above case. After immobilize FIX antibody on the glutaraldehyde modified surface, FIX-GNP conjugates were added with FIX concentrations, 1 pM, 10 pM, 100 pM, 1 nM, 10 nM and 100 nM. Samples of FIX-GNP conjugates were prepared by adding 5 μl of GNP into appropriate final concentration of FIX. To check the limit of detection, the resultant data were plotted using OriginPro 9.0. Incubation and washing steps were mandatorily followed as stated above.
Specific detection of FIX
To check the specific detection of FIX, after immobilizing the FIX antibody on glutaraldehyde modified surface, the higher concentration of globulin, albumin, other proteins (mixture of factors VIIa and XI) were independently passed on the antibody modified surfaces. These proteins were tested with the final concentration of 100 nM. Similarly, human blood serum was tested under the condition of spiking FIX (100 nM) and without spiking.
Impedance and IV measurements
All measurements were performed as stated before [12, 20]. Briefly, dielectric impedance spectroscopy measurements were performed using a Novocontrol alpha high-frequency analyzer (Hundsangen, Germany). Amperometric measurements of current to voltage (I-V) were carried out using Keithley 6487 Picoammeter, with 2 wired point-probing to characterize the dielectrode system on the polysilicon surface. The current to voltage (I-V) measurement was carried out from 1 V to 5 V, where the entire calibration plot with 5 V reading was taken. The impedance spectra of the real and imaginary parts of impedance, Zs’ and Zs” were received by sweeping the frequency of 1–100 MHz with the applied AC amplitude of 1 V RMS. The 100 Mhz data were selected for the sensitivity calibration plot. Dielectric properties altered by amine and glutaraldehyde functionalization, followed by interaction of antibody with different concentrations of FIX-GNP conjugates were studied by Nyquist plot. Each measurement was done at room temperature in triplicates.
Surface characterizations
The fabricated polysilicon surface was characterized using Scanning Electron Microscopy (SEM, JEOL JSM-6460LA) at 20 kV. Surface chemical functionalization was analyzed by Fourier transform infrared spectroscopy (FTIR, Spectrum 65, Perkin-Elmer). The analyses were performed for the surface with APTES, APTES/glutaraldehyde and APTES/glutaraldehyde/Antibody immobilization. All samples were examined at the wavelength ranges from 0 to 4500 cm−1.
Results and discussion
Hemophilia is the disease, also called ‘a Royal disease’ caused by the defective level of FIX in human blood, leads to a clotting deficiency. In the human blood clotting cascade, FIX involves both extrinsic and intrinsic clotting pathways and with the deficiency in these pathways clotting is not proceed successfully [4]. A threshold quantitative detection of FIX helps to identify the disease for the proper treatment. In the past, quantification of FIX has been demonstrated with different analytical and non-analytical sensing systems, which include Surface Plasmon Resonance, waveguide mode sensing, BioDVD platform [1, 4, 5, 21]. Herein, additional evidence was introduced to support the detection of FIX by its antibody with the help of nanogapped dielectric polysilicon substrate, suitable to operate with DC and AC currents. To attain a high-performance with this analytical system, antibody-GNP conjugates were used and specificity analysis was done with proteins other than FIX, reside in human blood serum.
Surface characterizations
Figure 1 a, b displays the surface image of the nanogapped dielectric system under scanning electron microscopy (SEM) observation. The measurement with SEM shows the gap between two electrodes to be <100 nm. Due to the lesser nanogap, it has already been proven its efficient detection of biomolecules with this chip [12]. This nanogapped polysilicon surface was functionalized chemically by APTES followed by glutaraldehyde in order to capture FIX antibody. The steps involved in the complete surface chemical modifications and biomolecular assembly processes are indicated in the Fig. 1 c-g. These modifications are confirmed by Fourier transform infrared spectroscopy (FTIR) spectral profiles. FTIR results clearly displayed the changes in the spectral arrangements for APTES, APTES-glutaraldehyde and APTES-glutaraldehyde-antibody modified surfaces.
Impedance measurements for molecular assembly
The molecular assembly/interaction during the experiments steps were measured by both impedance and IV. Figure 2 infers that there is a change on the surface with the assembly of molecules. It can be obviously seen from the semicircle that indicate the conformational changes on the surface upon step by step process with increasing overall charge transfer resistance (Rct) of the analytical system. The bare polysilicon showed the lowest impedance and after introduce APTES on the chip; the resistance was increased to the value of 1.8 × 105 Zs’ ohms. This result confirms the proper modification of amine on the nanogapped dielectrode polysilicon surface. Then with glutaraldehyde immobilization, there was an increment with the Rct to 3.7 × 105 Zs’ ohms. Antibody and ethanolamine attachments were confirmed by increasing the Rct to 4.6 × 105 and 4.8 × 105 Zs’ ohms, respectively. Since antibody covered most of the aldehyde tethered surface, ethanolamine shows only a small variation in the changes. Finally, FIX was added on the FIX antibody modified surface with the concentration of 100 nM and the spectrum shows the clear increase in the Rct to 5.9 × 105 Zs’ ohms.
Current vs Voltage (IV) measurements for molecular assembly
The above-mentioned immobilization and assembly processes were also monitored by a Picoammeter, with 2 wire point-probing systems IV measurement. As shown in Fig. 3, the bare surface without any surface modification shows the current flow at the lower level and after attaches the APTES, current flow was increased to 1.0 × 10−4 V, it confirms the proper modification of the surface with amine. With further modifications, the current flows were decreased by each attachment. After attaching glutaraldehyde, the current decrement was to 8.0 × 10−5 V. Then, the current was changed to 5.0 × 10−5 V after immobilizing FIX antibody. It clearly confirms that the proper antibody binding on the glutaraldehyde modified surface. After the antibody attachment, the remaining places of glutaraldehyde were blocked by ethanolamine, it lead to the current flow with small variation due to the higher attachment of FIX antibody bound to the aldehyde-modified surface. This step is exactly imitating the above impedance measurement, where also the higher immobilization of antibody was evidenced. Finally, FIX protein was passed on the FIX antibody modified surface and it was found that the current flow has a great difference with decrement upon the addition of FIX, and the changes were from 5.0 × 10−5 to 2.3 × 10−5 V.
Enhancing detection by FIX-GNP conjugates
To enhance the level of FIX detection and to demonstrate the high-performance detection, herein analyzed the interaction of FIX-GNPs against FIX antibody. GNP can binds with the protein through electrostatic interaction or ionic bonding or chemical modification [12, 22, 23]. Smaller sized biomolecule can be easily modified by chemical means using thiol-group to attach on GNP, but the bigger sized molecules (protein or antibody) with a larger molecular weight can conjugate GNP without a specific chemical modification due to more attraction between amines on the protein and GNPs [23]. In this study, FIX protein with 55 kDa was bound to the GNPs by mixing. Previously, Gopinath et al. [23] have demonstrated the stable binding of FIX on the GNP surface electrostatically even under the high salt concentration. Other biomolecules, such as nucleic acid is also found to interact with GNP electrostatically [24]. In this study, prepared the FIX-GNP conjugates and titrated from 1 pM to 1 nM of FIX and detected on the FIX antibody modified surfaces. Figure 4 shows clear changes in charge transfer resistance with the concomitant increase with FIX concentrations. The high concentration (1 nM) of FIX yielded the Rct to the level of 3.7 × 106 Zs’ ohms. With the lowest concentration of 1 pM, the Rct level shows the background level and considered as not significant. FIX with 10 and 100 pM gives the Rct of 0.9 × 106 and 1.5 × 106 Zs’ ohms, respectively. With 1 nM of FIX the current flow was noticed to be with the Rct of 1.8 × 106 Zs’ ohms. The higher concentrations of FIX, 10 and 100 nM displayed with 2.9 × 106 and 3.7 × 106 Zs’ ohms, respectively. These results indicated clearly that the lower detection of FIX (1 pM) was achieved. An overview on the reported nanomaterial-based methods for the determination of factor IX is displayed in the Table 1. Previously, Lakshmipriya et al. [1] have detected the FIX by its aptamer on the waveguide surface with the limit of detection of 100 f. as the highest sensitivity achieved with the assistance of polymer conjugated GNPs, and they also detected the FIX protein using the sandwich pattern with aptamer and antibody until the limit of 100 pM [10]. The current study with the sensitivity level of 1 pM may consider being the better after the methods designed by Lakshmipriya et al. [1, 10]. Other applied methods stated in the Table 1 with different substrates, the sensitivities were attained in the range of picomolar [4, 5, 21, 25]. Since healthy human plasma contains FIX around 3–5 μg/ml, the limit of detection attained in the present study is also well suited to detect the FIX in human plasma. On the other hand, the signal enhancement with FIX-GNP conjugates was ~100 folds higher compared to the level without GNPs, brought the sensitivity to 1 pM. The main appealing characteristics of the current work compared to other established methods are relatively cheaper, convenient, consumes lesser sample and with battery operating feasibility for point-of-care testing.
High-analytical performance by nanogapped dielectrode system
The sensitivity was calculated based on the experimental values and a linear correlation of the differences in the charge transfer resistance, \( \varDelta \mathrm{Rct}={Rct}_{FIXantibody}-{Rct}_{FIXantigen} \) with respect to the logarithm of FIX concentrations are shown in Fig. 5 a. The sensitivity of the reported analytical assay was calculated using the equation below,
The slope, m = 7.5 × 10−7 μA μM − 1 was got from the Fig. 5 a and area; A is the maximum detection spot on the active surface (0.00262cm2). Therefore, the sensitivity of the system calculated was 286.26 μA μM−1 cm−2. It was observed that ∆Rct increases linearly with increasing FIX concentrations from 1 pM to 0.1 μM. The value (∆Rct) difference between Rct on antibody immobilized surface and antigen interacted surface was found to be well proportional to the natural logarithm of FIX concentration with a linear equation of ∆Rct = 1.534E6 + 1.986E7x, (R2 = 0.98906). The limit of detection was estimated to be 1 pM using the signal to noise ratio of more than 3σ (where σ is the standard deviation of the blank solution, n = 5). The detection limit is lower than previous report using another probe (aptamer) interacted with FIX, the dissociation constant (Kd) determined by Rusconi et al. [10] was 580 pM using radioisotope labeled aptamer. Similarly, Gopinath et al. [3] reached the Kd value of 365 and 418 pM with aptamer against FIXa and FIX, respectively, determined by surface plasmon resonance biosensor.
To check the specificity of antibody used in this study to detect FIX, different proteins which are abundant in the human serum were evaluated by passing them independently on FIX antibody immobilized surface. The normal level of albumin in the human serum was reported to be 45 g.L−1, similarly the level of other factors FVIIa is 3.6 mg.L−1 and FXI is 3–6 g.L−1 [1]. For the specificity analyses, different samples were used, namely albumin, globulin, mixture of factors (factors VIIa, FXI). In which, the current analytical system displays ~5 folds higher detection by FIX antibody against FIX compared to other tested proteins from the serum (Fig. 5 b). We also analyzed with real human serum alone, 100 nM of FIX-spiked human serum and mixture of FIX (100 nM) and the above negative proteins. There were no significant differences between FIX alone and FIX-spiked samples. However, with only human serum there was a reduction in the signal compared to the FIX alone. But, FIX spiked human serum has shown an increment, this might be due to abundance of FIX in the human serum.
Conclusions
Detection and quantification of FIX are the mandatory events to know the level of disease and help to treat. Here nanogapped dielectric analytical system was used to detect FIX (conjugated to GNPs) against its antibody by impedance and IV measurements. With these analyses, could achieve the picomolar level of detection with a higher specificity by discriminating other proteins and the human serum. This detection level is suitable for detecting the FIX abundance in real human serum and comparable with other currently available methods, even though the voltammetry measurement shown here is not the best. This method is favorable to diagnose clotting deficiency, feasible with AC and DC currents for point-of-care testing. Compared to the currently demonstrated sensing systems for detecting factor IX, this analytical system has positive characteristics such as, convenient to handle without prior experiences and consumes the lower volume of samples with the feasibility for bed-side analysis.
References
Lakshmipriya T, Fujimaki M, Gopinath SCB, Awazu K, Horiguchi Y, Nagasaki Y (2013) A high-performance waveguide-mode biosensor for detection of factor IX using PEG-based blocking agents to suppress non-specific binding and improve sensitivity. Analyst 138:2863–2870
Mehta JB, Mehta S (1990) Acquired haemophilia. J Assoc Physicians India 38:895–896
Gopinath SCB (2008) Anti-coagulant aptamers. Thromb Res 122:838–847
Gopinath SCB, Shikamoto Y, Mizuno H, Kumar PKR (2006) A potent anti-coagulant RNA aptamer inhibits blood coagulation by specifically blocking the extrinsic clotting pathway. Thromb Haemost 95:767–771
Rusconi CP, Scardino E, Layzer J, Pitoc GA, Ortel TL, Monroe D, Sullenger BA (2002) RNA aptamers as reversible antagonists of coagulation factor IXa. Nature 419:90–94
Richard B, Chung K, The S (1980) Detection of Factor Calcium. October 56:608–614
Takamiya O, Kinoshita S (1997) A simple method for detection of human factor IX inhibitor using ELISA. Scand J Clin Lab Invest 57:683–688
Lin PC, Su YN, Liao YM, Chang T-T, Tsai S-P, Shu H-L, Chiou S-S (2014) Efficient detection of factor IX mutations by denaturing high-performance liquid chromatography in Taiwanese hemophilia B patients, and the identification of two novel mutations. Kaohsiung J Med Sci 30:187–193
Satoh C, Takahashi N, Asakawa J, Hiyama K, Kodaira M (1993) Variations among Japanese of the factor IX gene (F9) detected by PCR-denaturing gradient gel electrophoresis. Am J Hum Genet 52:167–175
Lakshmipriya T, Horiguchi Y, Nagasaki Y (2014) Co-immobilized poly(ethylene glycol)-block-polyamines promote sensitivity and restrict biofouling on gold sensor surface for detecting factor IX in human plasma. Analyst 139:3977–3985
Gopinath SCB, Kumar PKR, Tominaga J (2011) A BioDVD Media with Multilayered Structure Is Suitable for Analyzing Biomolecular Interactions. J Nanosci Nanotechnol 11:5682–5688
Balakrishnan SR, Hashim U, Gopinath SCB, Poopalan P, Ramayya HR, Liu WW, Ruslinda AR, Haarindraprasad R, Perumal V (2015) Polysilicon Nanogap Lab-on-Chip Facilitates Multiplex Analyses with Single Analyte. Biosens Bioelectron 84:44–52
Gopinath SCB, Awazu K, Fujimaki M, Shimizu K, Shima T (2013) Observations of Immuno-Gold Conjugates on Influenza Viruses Using Waveguide-Mode Sensors. PLoS One 8:1–10
Azizah N, Hashim U, Gopinath SCB, Nadzirah S (2016) Gold nanoparticle-mediated method for spatially resolved deposition of DNA on nano-gapped interdigitated electrodes, and its application to the detection of the human Papillomavirus. Microchim Acta 183:3119–3126
Li Y, Schluesener HJ, Xu S (2010) Gold nanoparticle-based biosensors. Gold Bull 43:29–41
Jin NZ, Anniebell S, Gopinath SCB, Chen Y (2016) Variations in Spontaneous Assembly and Disassembly of Molecules on Unmodified Gold Nanoparticles. Nanoscale Res Lett 11:399
MacKay S, Hermansen P, Wishart D, Chen J (2015) Simulations of interdigitated electrode interactions with gold nanoparticles for impedance-based biosensing applications. Sensors (Switzerland) 15:22192–22208
Liopo AV, Conjusteau A, Oraevsky AA (2012) PEG-coated gold nanorod monoclonal antibody conjugates in preclinical research with optoacoustic tomography, photothermal therapy and sensing. Proc SPIE 8223:1–10
Kyu S, Cho H, Park H-J, Kwon D, Min LJ, Hyun CB (2009) Nanogap biosensors for electrical and label-free detection of biomolecular interactions. Nanotechnology 20:455502
Balakrishnan SR, Hashim U, Gopinath SCB, Poobalan P, Hariram R, Omar I, Dhahi TS, Haarindraprasad R, Veeradasan P (2015) A Point-of-Care Immunosensor for Human Chorionic Gonadotropin in Clinical Urine Samples Using a Cuneated Polysilicon Nanogap Lab-on-Chip. PLoS One 10:e0137891
Gopinath SCB, Awazu K, Tominaga J, Kumar PKR (2008) Monitoring biomolecular interactions on a digital versatile disk: A BioDVD platform technology. ACS Nano 2:1885–1895
Lakshmipriya T, Hashim U, Gopinath SCB, Azizah N (2016) Microfluidic-based biosensor: signal enhancement by gold nanoparticle. Microsyst Technol 22:2389–2395
Gopinath SCB, Lakshmipriya T, Awazu K (2014) Colorimetric detection of controlled assembly and disassembly of aptamers on unmodified gold nanoparticles. Biosens Bioelectron 51:115–123
He Y, Cheng F, Pang D-W, Tang H-W (2017) Colorimetric and visual determination of DNase I activity using gold nanoparticles as an indicator. Microchim Acta 184:101–106
Cheen OC, Gopinath SCB, Perumal V, Md Arshad MK, Lakshmipriya T, Chen Y, Haarindraprasad R, Balakrishnan SR, Hashim U, Pandian K (2017) Aptamer-based impedimetric determination of the human blood clotting factor IX in serum using an interdigitated electrode modified with a ZnO nanolayer. Microchim Acta 184:117–125
Gopinath SCB, Awazu K, Fujimaki M, Sugimoto K, Ohki Y, Komatsubara T, Tominaga J, Gupta KC, Kumar PKR (2008) Influence of nanometric holes on the sensitivity of a waveguide-mode sensor: Label-free nanosensor for the analysis of RNA aptamer-ligand interactions. Anal Chem 80:6602–6609
Acknowledgements
Dr. Yeng Chen would like to thank University of Malaya Research Grants (UMRG) RG454-12HTM.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
The author(s) declare that they have no competing interests.
Rights and permissions
About this article
Cite this article
Gopinath, S.C.B., Perumal, V., Rao, B.S. et al. Voltammetric immunoassay for the human blood clotting factor IX by using nanogapped dielectrode junctions modified with gold nanoparticle-conjugated antibody. Microchim Acta 184, 3739–3745 (2017). https://doi.org/10.1007/s00604-017-2389-7