Abstract
Trace heavy metal ions are probable carcinogens with concentrations at ppb levels. Ion selective electrode (ISE) technology is a conventional and widely used method to detect various ions with the advantages as rapid measurement, low power consumption and low cost. However, the detection limit of a conventional bulky ISE device is only able to reach ppm level so far, not sensitive enough for practically detecting heavy metal ions. A micro ion selective electrode (μISE) array which can realize multiplex detection of heavy metal ions on one chip is reported in this paper. Its detection limit for Pb2+, Cd2+, AsO2−, and Hg2+ can reach 1, 3, 10, and 1 ppb, respectively, within the permissible limits of drinking water. The micrometer scale device arrays from microfabrication processes demonstrated enhanced sensitivity, uniformity, and reduced response time (18 s) compared to conventional ISE. Our devices showed superb selectivity, and the performance dependence on temperature and pH values was investigated as well.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
1 Introduction
TRACE ion elements of lead (Pb2+), cadmium (Cd2+), mercury (Hg2+), and arsenite (AsO2−) can cause numbers of health problems, and are probable carcinogens for human beings with concentrations at ppb levels (Wang et al. 2014). Therefore, it is essential to detect the concentration of these toxic ions in various samples including drinking water and food. The most common and reliable methods for the determination of heavy metal ions include atomic absorption spectroscopy (AAS) (Xie et al. 2008), X-ray fluorescence (XRF) (Ravisankar et al. 2015), inductively coupled plasma-mass spectroscopy (ICP-MS) (Rui and Hao 2012), and inductively coupled plasma-atomic emission spectroscopy (ICP-AES) (Cong and Cai 2010). Although these laboratory-based methods exhibit high performance, the large cost and complicated sample pretreatment prohibit them from wide sensing application.
Ion selective electrode (ISE) technology is a conventional and widely used method to detect various ions with the advantages as rapid measurement, low power consumption and low cost (Malinowska et al. 1994). They generally do not need complicated sample pretreatment and can be used for real time analysis. However, detection limit of a conventional ISE device can only reach ppm level so far (McGraw et al. 2007; Naushad et al. 2014; Gupta et al. 2003; Gupta and Agarwal 2005; Gholivand and Raheedayat 2004; Gupta et al. 2003), not sensitive enough for practically detecting trace heavy metal ions. I, a commercialized bulky ISE device is only able to detect one specific ion. As such, more than one device is needed for multiple heavy metal ions detection, which further limits its applications. Here we used microfabrication technique to largely reduce the geometry size and simplify the complex structure of conventional ISE (Li et al. 2015). Four or even more μISEs can be integrated on one chip, increasing the efficiency of both fabrication and detection. Its detection limit for Pb2+, Cd2+, AsO2−, and Hg2+ can reach 1, 3, 10, and 1 ppb, respectively, which is able to meet the permissible limits of the guideline for drinking water quality provided by the World Health Organization (WHO). The micrometer scale device arrays derived from microfabrication processes demonstrated enhanced sensitivity, uniformity, and reduced response time compared to conventional ISE.
2 Experimental
The μISE arrays were fabricated by highly uniform microfabrication technique (Fischer et al. 2015). Metal electrodes were patterned by lift-off process on a glass substrate (Fig. 1a–d). Firstly, negative photoresist NR9-1500PY was spun coated on glass substrate at 3000 rpm for 60 s, followed by alignment, UV light exposure, and development to define the electrode arrays. Then sputtering process was applied to deposit metal layer (10 nm Cr/100 nm Au) on top. The metal electrode arrays were formed on glass after removing the photoresist together with the redundant Cr/Au by acetone. In order to investigate the dependence of the performances on device geometry size, metal electrodes with three different pad sizes (500, 800, and 1200 μm) were fabricated. Finally, glass wafer was diced into small chips with 4 or more electrodes on each one (Fig. 1d). As such, multiplex detection of heavy metal ions on one chip can be realized.
After using O2 plasma to clean the metal electrodes for 10 min to further remove the photoresist and other organic residue, we spun coated POT (Poly (3-octylthiophene-2, 5-diyl), Sigma-Aldrich) conductive polymer solution (2.5 g/ml in chloroform;, 3-5 μL) on each working electrode with speed of 1000 rpm for 15 s before ionophore coating. POT layer was applied since it is reported to be able to increase the stability of ISE inasmuch as it avoids the interference from O2 which may result in the potential shift of the device (Sutter et al. 2004). Next, tetrahydrofuran (THF) solvent with different types of ionophore specific for Pb2+, Cd2+, AsO2−, and Hg2+ were spun coated on different electrodes, respectively (Fig. 1d), with speed 4000 rpm for 40 s. The sample was kept in air at room temperature for 5–10 min to let the solvent, THF, evaporate and form a solid ionophore film. Theoretically, for a certain type of ionophore film, only one kind of ions could be selectively permeate through the film as shown in Fig. 2a. Therefore, ionophore film works as a filter which can block other ions, and increases the selectivity of the sensor.
After introduction of sodium acetate buffer solution (0.1 M, pH 4.6) with various heavy metal ion concentrations at room temperature, potentials between an Ag/AgCl reference electrode and corresponding μISE electrodes were measured using an Agilent digital multimeter 34461A (Fig. 2b). According to the principle of Nernst effect (Gulbault 1981), the potential of reference electrode keeps constant, while the potential of μISE changes with different concentrations of ion detected.
3 Results and discussion
The performance ion selective electrode should follow modified Nernst (Nikolsky) Equation (Gulbault 1981):
where E is experimentally observed potential of ISE, E0 is the potential under standard conditions, R is the gas constant (8.31 J K−1mol−1), T is the thermodynamic temperature (in K), n is an integer with sign and magnitude corresponding to the charge of the principal ion, F is the Faraday constant (9.65 × 104 C mol−1), γis the activity coefficient of the ion, N is the concentration of the ion. According to Eq. 1, the potential of ISE should increase with higher concentrations of ion detected. In our experiment, as shown in Fig. 3a, when the Pb2+ concentration increased from 10 ppb to 100,000 ppb, a clear potential increase of μISE was observed, indicating that modified Nernst (Nikolsky) Equation can be applied to μISE as well. This result also suggested that the Pb2+ can be effectively detected over wide concentration range.
Spin coating, a standard and highly uniform microfabrication technique, was chosen in our experiment to form ionophore layer on metal electrodes. To investigate the influence of ionophore coating method, we tested devices based on spin coating and drop casting (a conventional method for bulky ISE devices), respectively. All the devices tested had identical geometry size. The thicknesses of drop-casted ionophore were typically in the orders of millimeters, while spin-coated ionophore film were much thinner (in the orders of micrometers), so ions detected can pass through it more easily. As a result, devices with spin-coated ionophore demonstrated much shorter response time than drop-casted ionophore shown in Fig. 3b. We fitted the curve with equation (Li et al. 2015):
where V is the instantaneous potential difference between ISE and reference electrode, V0 is the initial potential difference after Pb2+ introduction, Vf is the final potential difference when the output signal is stable, t is the time, and τ is time constant. The extracted time constants for ISE devices based on spin-coated and drop-casted ionophore were 18 s and 145 s (for 1000 ppb Pb2+), respectively (Fig. 3b). Notably, larger response time constants were observed with smaller Pb2+ concentration (Fig. 3c). Additionally, the thickness of drop-casted ionophore film varies from device to device, while spin-coated ionophore has highly uniform thickness, which is responsible for the improved uniformity of the sensors as shown in Fig. 3d. Therefore, spin-coating method can not only enhance the response rate but also improve the uniformity of the sensor.
In order to investigate the dependence of the performances on device geometry size, metal electrodes with three different pad sizes (500, 800, and 1200 μm) were measured (Fig. 3e), and their sensitivities extracted from linear region were 8.63, 7.21, and 6.04 mV/decade, respectively, which indicated a trend that reducing the size of ISE can enhance its sensitivity. We assume that the ion concentration gradient after equilibrium is different at the edge area of ionophore film from that at the center. When the size of ISE devices reduces to micrometer scale, the edge effect becomes prominent, and it can affect the entire output of the sensor (increase sensitivity). This is one of our major motivations of exploring μISE.
Detection limit of a sensor is generally defined as three times the standard deviation of its noise (defined by international union of pure and applied chemistry (IUPAC)) (Currie 1995). Micro ISE’s noise we measured (22 µV, Fig. 4a) was much smaller than that of regular ISE reported (0.6-0.8 mV). We assume the simplicity of our device structure avoids some interfaces which may introduce noise. Because of the low noise and enhanced sensitivity, we were able to achieve ultrasensitive detection of heavy metal ions. When we switched the solution from sodium acetate buffer to 1 ppb Pb2+, a potential shift of 1.1 mV was observed as shown in Fig. 4b. This value was large than three times the standard deviation of μISE’s noise. As such, μISE is capable of detecting Pb2+ at least down to 1 ppb. Additionally, the μISE’s detection limits for Cd2+, AsO2−, and Hg2+ can reach 3, 10, and 1 ppb, respectively (Fig. 4c, d, e). Selectivity is critical for ISE to identify each type of ions correctly and reduce the chance of misleading (Li et al. 2015). Micro ISE array demonstrated excellent ion selectivity. Lead working electrode’s response to Pb2+ was much larger than that to Cd2+, AsO2−, and Hg2+, indicating it is only sensitive to Pb2+ (Fig. 5a). Similar results were observed for working electrodes coated with cadmium, arsenic, and mercury ionophore (Fig. 5b, c, d). The results displayed in Fig. 5 demonstrate the good selectivity of μISE. It also proves that the potential change of our sensor was caused by the ion interaction instead of other factors in environment (Li et al. 2015).
The environment’s influence on the sensor was also investigated. Similar to regular ISE, µISE array’s output signals were sensitive to temperature variation. According to Eq. 1, higher temperature can result in larger potential value, and higher temperature can also increase the activity coefficient of the ion, which further increases the potential. Our experimental result in Fig. 6(a) is consistent with this theory. The results indicate that the device is able to work under a wide range of temperature value after calibration. The performance of µISE can also be influenced by H+ or OH−. Thus, pH dependent of device was tested over the pH range 1–7 for 1 ppb heavy metal ions. Figure depicts that the pH dependence of the potentials is insignificant in the pH range of 3–5 for devices based on Cd2+, AsO2−, and Hg2+ ionophore, and insignificant in the pH range of 4–7 for devices based on Pb2+ ionophore. Therefore, sodium acetate buffer solution (0.1 M, pH 4.6) we used can minimize the influence of pH value variation, and it is suitable for µISE to detect heavy metal ions, as shown in Fig. 6(b).
4 Conclusion
We investigated a new μISE array realizing multiplex detection of heavy metal ions on one chip. Micromanufacture processes including photolithography and lift-off were used to fabricate the micro chip with a high potential for integration. The micro device array using microfabrication processes demonstrates enhanced sensitivity, uniformity, and reduced response time, compared to conventional ISE. Its detection limit can reach ppb level, within the permissible limits of drinking water. The device demonstrates superb selectivity, and the performance dependence on temperatures and pH values was also investigated.
References
Cong Q, Cai Y (2010) Determination of heavy metals in vegetables by microwave digestion/inductively coupled plasma atomic emission spectroscopy. Food Sci 31:290–292
Currie IA (1995) Nomenclature in evaluation of analytical methods including detection and quantification capabilities (IUPAC recommendations 1995). Pure Appl Chem 67:1699–1723
Fischer AC, Forserg F, Lapisa M et al (2015) Integrating MEMS and ICs. Microsyst Nanoeng 1:15005
Gholivand MB, Raheedayat F (2004) Chromium (III) ion selective electrode based on oxalic acid bis (cyclohexylidene hydrazide). Electroanalysis 16:1130–1135
Gulbault GG (1981) Recommendations for publishing manuscripts on ion-selective electrodes. Pure Appl Chem 53:1907–1981
Gupta VK, Agarwal S (2005) PVC based 5, 10, 15, 20-tetrakis (4-methoxyphenyl) porphyrinatocobalt (II) membrane potentiometric sensor for arsenite”. Talanta 65:730–734
Gupta VK, Chandra S, Agarwal S et al (2003a) Mercury selective electrochemical sensor based on a double armed crown ether as ionophore. Indian J Chem 42:813–818
Gupta VK, Chandra S, Agarwal S (2003b) Mercury selective electrochemical sensor based on a double armed crown ether as ionophore. Indian J Chem 42A:813–818
Li P, Zhang D, Liu J et al (2015a) Air-stable black phosphorus devices for ion sensing. ACS Appl Mater Interfaces 7:24396–24402
Li P, Zhang B, Cui T (2015b) TiO2 and shrink induced tunable nano self-assembled graphene composites for label free biosensors. Sens Actuat B Chem 216:337–342
Li P, Zhang B, Cui T (2015c) Towards intrinsic graphene biosensor: a label-free, suspended single crystalline graphene sensor for multiple lung cancer tumor markers detection. Bioesens Bioelectron 72:168–174
Li P, You R, Jing G, Cui T, Ultrasensitive micro ion selective arrays for multiplex heavy metal ions detection, MicroTAS, 2015
Malinowska E, Brzozka Z, Kasiura K et al (1994) Lead selective electrodes based on thioamide functionalized calyx [4] arenes as ionophores. Anal Chim Acta 298:253–258
McGraw CM, Radu T, Radu A et al (2007) Evaluation of liquid- and solid-contact, Pb2+-selective polymer membrane electrodes for soil analysis. Electro Anal 20:340–346
Naushad M, Inamuddin, Rangreez TA et al (2014) A mercury ion selective electrode based on poly-o-toluidine Zr (IV) tungstate composite membrane. J Electro Anal Chem 713:125–130
Ravisankar R, Chandrasekaran S, Chandrasekaran A et al (2015) Statistical assessment of heavy metal pollution in sediments of east coast of Tamilnadu using energy dispersive X-ray fluorescence spectroscopy (EDXRF). Appl Radiat Isot 102:42–47
Rui Y, Hao J (2012) Determination of nine heavy metals by inductively coupled plasma mass spectroscopy in groundwater from northeast rural of China. Asian J Chem 24:2825–2826
Sutter J, Radu A, Peper S et al (2004) Solid-contact polymeric membrane electrodes with etection limit in the subnanomolar range. Anal Chim Acta 523:53–59
Wang Z, Wang H, Zhang Z et al (2014) Sensitive electrochemical determination of trace cadmium on a stannum film/poly (p-aminobenzene sulfonic acid)/electrochemically reduced graphene composite modified electrode. Electrochim Acta 120:140–146
Xie F, Lin X, Wu X, Xie Z (2008) Solid phase extraction of lead (II), copper (II), cadmium (II) and nickel (II) using gallic acid-modified silica gel prior to determination by flame atomic absorption spectrometry. Talanta 74:836–843
Author information
Authors and Affiliations
Corresponding author
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
You, R., Li, P., Jing, G. et al. Ultrasensitive micro ion selective sensor arrays for multiplex heavy metal ions detection. Microsyst Technol 25, 845–849 (2019). https://doi.org/10.1007/s00542-018-4067-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00542-018-4067-z