Abstract
Objective
To determine tinnitus prevalence and severity in a cohort of unselected first-time cochlear implant (CI) recipients whose primary motive for CI was sensorineural hearing loss (SNHL), and to evaluate the effect of CI on tinnitus after cochlear implantation.
Methods
Prospective longitudinal study of 45 adult CI recipients with moderate to profound SNHL. Patients completed the Danish version of the Tinnitus Handicap Inventory (THI) and a visual analogue scale (VAS) for tinnitus burden before implantation, 4 months after implantation and 14 months after implantation.
Results
The study included 45 patients, of which 29 (64%) had pre-implant tinnitus. Median THI score (IQR) significantly decreased from 20 (34) to 12 (24) at first follow-up (p < 0.05) and to 6 points (17) at second follow-up (p < 0.001). Median VAS (IQR) for tinnitus burden decreased from 33 (62) to 17 (40; p = 0.228) and 12 (27, p < 0.05) at the first and second follow-ups, respectively. Tinnitus was totally suppressed in 19% of patients, improved in 48%, remained unchanged in 19% and worsened in 6%. 2 patients reported new tinnitus. At the second follow-up, 74% of patients had slight or no tinnitus handicap, 16% had mild handicaps, 6% had moderate handicaps, and 3% had severe handicaps. High pre-implant THI and VAS scores correlated with greater decrease in THI scores over time.
Conclusion
64% of the patients with SNHL had pre-implant tinnitus, which was decreased 4 and 14 months after implantation. Overall, 68% of patients with tinnitus improved their tinnitus handicap after CI. Patients with higher THI and VAS scores had a larger decline and the highest benefits in terms of tinnitus handicap improvement. The study findings demonstrate that the majority of patients with moderate to profound SNHL eligible for cochlear implantation benefit from complete or partial tinnitus suppression and improved quality of life after implantation.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Tinnitus is frequently observed in patients with moderate to profound sensorineural hearing loss (SNHL) [1], and it causes difficulties in concentration, sleep, anxiety and depression [2,3,4], and overall quality of life [5,6,7,8]. The prevalence of tinnitus, which affects 65 million adults in Europe with 4 million experiencing severe symptoms, is influenced by patient age and the degree of hearing loss. [9,10,11] Patients with hearing loss are three times more likely to experience tinnitus than those with normal hearing [12].
Cochlear implants (CI) are the preferred treatment for moderate to profound SNHL when hearing aids (HA) are ineffective, with the potential to reduce tinnitus in both unilateral and bilateral SNHL patients [13,14,15]. In one study, 80% of candidates with bilateral profound SNHL experienced pre-implant tinnitus, and after implantation, 86% of patients reported a decrease in tinnitus. [14]. Despite various hypotheses such as auditory masking, surgical effects, central auditory plasticity [16], and intra-cochlear electrical stimulation, the mechanisms by which CI alleviate or suppress tinnitus remain unknown. [13, 14, 17, 18]. Earlier studies found that cochlear implants suppresses tinnitus, but the factors that influence this suppression differ across studies [9, 17, 19,20,21,22,23].
In this study, we aim to evaluate the prevalence and severity of tinnitus in patients with sensorineural hearing loss who are eligible for cochlear implantation and to investigate the impact of cochlear implantation on tinnitus. We use a tinnitus handicap questionnaire to investigate whether CI suppresses tinnitus handicap and how tinnitus suppression relates to speech reception and patient-reported quality of life.
Materials and methods
Adult patients with SNHL who meet the Danish national candidacy criteria for CI were included in this prospective longitudinal study, regardless of whether they had tinnitus [24]. Patients under the age of 18, those who were unable or unwilling to participate in the study, and those who had bilateral cochlear implantation were excluded. If a patient received a second cochlear implant later, the data collected up until that point was included in the analysis.
Cochlear implantation
Following mastoidectomy and posterior tympanotomy to reach cochlea, we used the round window surgical technique to implant the electrode. In Denmark, the brand of implant device is chosen in collaboration between the patient and the speech and language pathologist, but the implant electrode is generally chosen by the surgeon. Four different surgeons conducted CI, and they had all performed over 100 CIs. All surgery and post-operative procedures were conducted at Department of Otorhinolaryngology, Head and Neck Surgery, and Audiology, Rigshospitalet, Copenhagen, Denmark.
Audiological and patient-reported outcome measures
At baseline (T0), all patients were tested with pure-tone audiometry thresholds for six frequencies, ranging from 250 to 8000 Hz (PTA6) and speech reception score (SRS). Two speech recognition tests, Dantale I and Hearing in Noise Test (HINT), and three patient-reported outcome measures (PROMs)—Nijmegen Cochlear Implant Questionnaire (NCIQ), Speech, Spatial, and Qualities of Hearing Scale (SSQ-12), and Tinnitus Handicap Inventory (THI)—were conducted at baseline (T0) and two post-implantation follow-ups (T1 and T2). T0, T1, and T2 coincided with routine clinical follow-up visits. From auditory performance and patient-reported quality of life outcomes (NCIQ and SSQ-12) from an earlier study [25], we will compare these findings to those related to tinnitus.
Tinnitus outcome measures
The Danish 25-item Tinnitus Handicap Inventory (THI) is a patient-reported outcome measure for tinnitus burden [26, 27]. The patients can answer each item on a scale of “yes” (4 points), “sometimes” (2 points) and “no” (0 points), resulting in a total score of 0–100 points. Patients rated their subjective tinnitus burden on a visual analog scale (VAS) from “No symptoms” to “Worst imagined symptoms” [28]. The VAS score was converted into a scale of 0—100 points. We assessed tinnitus handicap (THI total score) and subjective burden (VAS) before implantation (T0), and at 4 months (T1) and 14 months (T2) after implantation.
The tinnitus severity grading developed by McCombe et al. grades the tinnitus handicap into five severity levels: (1) slight or no handicap (0–16 points), (2) mild handicap (18–36 points), (3) moderate handicap (38–56 points), (4) severe handicap (58–76), and (5) catastrophic handicap (78–100)[29]. The minimal clinically significant change score for tinnitus handicap is seven points, as proposed by to Zeman et al. [30].
Statistical analysis
All analyses and graphics were conducted in RStudio [31]. Demographic data were examined using means, medians, ranges, and percentages. We analysed and compared pre- and post-operative questionnaire scores to detect significant differences in tinnitus and associated patient characteristics before and after CI. Since THI and VAS are non-normally distributed ordinal scale data, we used non-parametric analysis and reported median and interquartile range (IQR). The Friedman test was used to compare T0, T1 and T2 changes over time. Post hoc Wilcoxon signed-rank test with Bonferroni-adjusted p values was used to do pairwise comparisons. We used Spearman correlation coefficients to examine the strength and direction of correlations between speech reception scores, PROMs, and the change in tinnitus after implantation.
Results
Patient characteristics
49 patients were enrolled; however, four were lost to follow-up. Among the 45 included patients, 18 (40%) were female and 27 (60%) were male. All patients were given an ID number that followed them during the study. Patients’ age ranged from 26 to 88 years. Pure-tone audiometry in the implanted ears had a mean score (SD) of 89 (16) dB HL, while contralateral ears had a mean score of 79 (17) dB HL. Mean speech reception score (SD) was 24% (22) and 45% (28) for implanted and contralatereal ears. On the implanted ears, 20 patients (44%) had profound hearing loss and 20 patients (44%) had severe hearing loss. On the contralateral ear, 7 patients (16%) had profound hearing loss and 24 patients (53%) had severe hearing loss. The patients had used hearing aids on average for 19 years (range 0–57). Table 1 divides the demographic characteristics into two groups: those who had tinnitus (n = 31) during the study and those who did not (n = 14). Table 2 shows the individual patient data from the tinnitus group.
The Nucleus Cochlear CI522 (n = 26, 58%) or Nucleus Cochlear CI622 (n = 6, 13%) electrodes was implanted in most patients. The remaining 21 patients received Oticon Medical Zti EVO (n = 5, 11%), Advanced Bionics ULTRA 3D Midscale (n = 5, 11%), and MED-EL Flex 28 Synchrony (n = 2, 4%). One surgeon implanted 26 CIs (58%), while three other surgeons inserted 10 (22%), 6 (13%), and 3 (7%) CIs. All of the cochlear implant electrode arrays were fully inserted. 39 (87%) patients had normal impedances measured. Impedances were not measured in 6 (13%) of the patients. In 33 (73%), normal eCAP responses were measured. Six patients (13%) had no eCAP response measured. 3 patients had one absent electrode eCAP response, 2 (4%) had two absent electrode contacts, and 1 (2%) had more than three absent electrode contacts.
Non-congenital hereditary SNHL was the etiology of hearing loss in 10 patients (22%), congenital hearing loss in 4 patients (9%) and unknown etiology in 16 (36%) patients. Other causes of hearing loss were found in 15 (33%) patients, including Ménière’s disease, otosclerosis, meningitis, and syndromic hearing loss.
Tinnitus outcomes
Postoperative THI responses, T1 and T2, were performed at 4 months (mean 120 days, range 66–220) and 14 months (mean 410 days, range 290–530) after implantation. During the study, tinnitus was experienced by a total of 31 patients (69%). 29 (64%) had tinnitus before implantation, while two patients (4%) developed tinnitus after implantation (Fig. 1). The first patient had no tinnitus at T2, while the other reported persistent severe tinnitus after T2. Tinnitus was totally suppressed in 6 patients (19%) and improved in 15 (48%) patients after CI. Two patients (6%) reported an increase in tinnitus. Tinnitus remained unchanged in 6 (19%) patients (THI change ≤ 7) (Table 3).
The Friedman test showed a statistically significant decrease in THI total scores with time, χ2(2) = 24, p < 0.001. Median (IQR) THI total score was 20 (34) at T0, 12 (24) at T1 and 6 (17) at T2. THI total scores decreased significantly between T0 and T1 (Z = 284, p < 0.05), between T0 and T2 (Z = 356, p < 0.001), but not between T1 and T2 (Z = 226, p = 0.091) (Fig. 2a).
The median VAS score also decreased significantly with time, χ2(2) = 12, p < 0.01. Post hoc test revealed that the median (IQR) VAS scores for T0, T1, and T2 were 33 (62), 17 (40), and 12 (27), respectively. There was a significant decrease in VAS between T0 and T2 (Z = 269, p < 0.01), and T1 and T2 (Z = 190, p < 0.05), but not between T0 and T1 (Z = 300, p = 0.228) (Fig. 2b).
Before the implant, 14 patients (45%) had slight or no handicap, 7 (23%) had mild handicap, 4 (13%) had moderate handicap, 4 (13%) had severe handicap, and 2 (5%) had catastrophic handicap (Table 4). At T2, 23 patients (74%) had slight or no handicap, 5 (16%) had mild handicap, 2 (6%) had moderate handicap, and 1 (3%) had severe handicap. No patients had catastrophic handicap at T2 (Table 4).
Correlation analyses
The decrease in THI between T0 and T2 was strongly correlated with pre-implant THI scores (\({\widehat{\uprho }}_{\mathrm{s}}\) = 0.95; p < 0.001 (Fig. 3). The decrease in THI between T0 and T2 was likewise strongly correlated with pre-implantation VAS (\({\widehat{\rho }}_{s}\) = 0.89; p < 0.001). None of the other factors we analysed showed any correlation to the decrease in THI scores (Table 5).
Discussion
In the present study, we found that 64% of this cohort of CI recipients with moderate to profound SNHL had tinnitus prior to implantation. Two patients reported newly developed tinnitus; the first had no tinnitus at the second follow-up, while the latter had persistent severe tinnitus. Out of all the patients with tinnitus, complete suppression of tinnitus was observed in 19%, while 48% showed improvement in tinnitus. As a result, 68% of the patients showed an improvement in tinnitus handicap. Two patients (6%) reported worse tinnitus after the second follow-up. Tinnitus in 6 patients (19%) remained unchanged. Tinnitus was reduced in most of the CI recipients within the first 4 months, but we also observed individual reductions after 14 months.
The median THI total score decreased significantly after implantation at both post-implant follow-ups compared to the baseline. Between pre-implant and post-implant follow-ups, the median VAS score was likewise significantly lower, indicating that the recipients were less burdened by their tinnitus following implantation. The decrease in THI scores correlated with pre-implant THI scores and pre-implant VAS, meaning that those patients with higher pre-implant THI and VAS experienced the largest drop in scores after implantation. In summary, we found that following implantation, there was a sustained significant reduction in tinnitus and few incidences of new or worsened tinnitus.
Comparison to other studies
The pre-implantation tinnitus prevalence (64%) in this cohort is slightly lower than the estimated tinnitus prevalence range of 67–100% found in a review by Baguley et al. [14]. Overall, 68% of patients had their tinnitus handicap reduced, which corresponds to previous studies that ranged between 64 and 100% in unilaterally implanted patients [13, 20, 22, 32, 33]. Previous studies have also reported on tinnitus worsening in a few patients following CI. Amoodi et al. found that tinnitus handicap worsened in 5% of patients [32], while Kompis et al. found that tinnitus worsened in 7–9% of patients after implantation [22]. Kompis et al. also found that 11% (n = 5/47) of patients experienced new tinnitus and found an association between the patients who had tinnitus induced after implantation and poor speech reception [22]. In a study with 35 patients, Kim et al. found no worsening or new tinnitus after CI [20], but in a 2016 study, Kim et al. found newly developed tinnitus in 20% (n = 5/20) of patients, which resolved later [17]. In our study, two patients (6%) developed tinnitus after implantation, but only one patient (3%) had persistent tinnitus and experienced severe handicap as a result.
The mechanism by which CI can suppress tinnitus is unknown, although a multimodal mechanism including habituation, auditory masking [3], direct cochlear stimulation, and cortical reorganization is postulated. [13]. Punte et al. argue that electrical stimulation of the auditory nerve reverse the plastic reorganization of central auditory cortex associated with peripheral deafferentation that may be the cause of tinnitus [34]. Mallen et al. showed better tinnitus suppression 2 months after CI activation compared to the time of CI activation, indicating that the initial effect of auditory masking can reduce tinnitus in the short term, while neural plasticity suppress tinnitus in the long term [35]. Tinnitus suppression may also be aided by increased awareness of surrounding noises [34]. Bovo et al. reported that CI was able to reduce tinnitus in some patients immediately after implantation, after 6 months and after 18 months of CI use [13]. This finding is consistent with our findings. Although we do not have post-implant outcomes immediately after implantation, we do demonstrate tinnitus reduction after 4 and 14 months of CI use. This study reports longitudinal suppression of tinnitus and continued improvement beyond 14 months of CI use, and this may be due to combination of masking and neural plasticity in the central auditory cortex.
Following CI, 2 of 31 (6%) patients developed new tinnitus, while 2 of 31 (6%) patients with pre-implant tinnitus experienced post-operative tinnitus worsening. These results are comparable to prior findings [14, 19, 22, 36]. The first recipient with newly developed tinnitus had tinnitus between T0 and T1, but it resolved by T2. The second recipient did not report any tinnitus at T0 but reported a severe handicap at T1 and T2 (THI total score of 58 and 66). It has previously been shown that patients who experienced tinnitus induced by CI had poor speech reception scores [22]. In our case, the patient did have low PTA6 and SRS at baseline, but this is a topic that needs further research.
Correlation analyses
Numerous studies have been conducted to assess predictors of speech reception outcomes after CI [37]. In general, research on tinnitus reduction after CI is limited and information on predicting factors for tinnitus burden outcomes is even more limited. Several predictors for tinnitus recovery were reported in two recent investigations, the first by Ramakers et al. [38] and the second by Dixon et al. [39]. Lower speech recognition scores (CVC) at baseline, unilateral tinnitus and loss of residual hearing at 250 Hz were shown to predict tinnitus improvement in the model of Ramakers et al. Dixon et al. likewise demonstrated that poorer residual hearing predict tinnitus recovery, albeit they suggested that, contrary to Ramakers et al., better baseline hearing could result in tinnitus improvement. Furthermore, they found that higher baseline THI scores were associated with a greater likelihood of tinnitus improvement. We also observed that higher THI and VAS scores before implantation was correlated with better tinnitus improvement. We found no effect of age, duration of HA prior to implantation, hearing level or speech reception score in our population. We also found no evidence that improved auditory performance and overall quality of life following implantation will result in less tinnitus.
Strengths and imitations
The study's strengths include a prospective longitudinal study design assessing benefits before and up to 14 months after implantation. Tinnitus changes are monitored at three time points. According to the findings of this research, tinnitus suppression by cochlear implants is stable, and tinnitus does not reappear over time.
One limitation of the study was that we did not record whether the tinnitus was exclusively in the CI ear, only in the non-CI ear, or bilaterally before surgery. Tinnitus in the contralateral ear may mask a change in tinnitus in the implanted ear that is either decreased or intensified. The patients completed the THI while their CI was switched on. The effects on tinnitus were not examined when the CI was switched off. Quaranta et al. found that 22% of patients had tinnitus when the CI was switched off, but it disappeared when it was switched on. [36]
Tinnitus is often present in specific hearing loss aetiologies, such as Ménière's disease, vestibular schwannomas, and otosclerosis. The etiologic groups in this investigation were too small to provide for meaningful statistical analysis.
Clinical implications
In addition to the significant audiological improvement observed in the adult CI group [25], there was also a decrease in tinnitus following cochlear implantation. Positive effects on tinnitus have been demonstrated earlier; however, the evidence from the Nordic countries is limited, and to our knowledge, Rodvik et al. from Norway is the only other group to describe tinnitus improvement after CI [40]. Tinnitus should not be considered a disqualifying factor for candidacy, but rather, tinnitus should also be taken into account along with audiological considerations while selecting a patient for CI. Our results are consistent with previous studies and support the use of cochlear implants in borderline cases of hearing loss with concomitant tinnitus [24, 41].
This study adds to the findings from previous studies on tinnitus improvement after CI and documents a positive benefit from CI in addition to the audiological benefits. Patients who receive CI based only on auditory criteria, regardless of tinnitus, may experience tinnitus suppression as a positive side effect of the implantation. Not only is tinnitus reduced, but numerous studies demonstrate a significant improvement in patient quality of life following implantation because of tinnitus reduction, indicating that CI has important clinical benefits in addition to speech perception improvements [8, 9, 32, 33]. Further studies should consider the effect of CI in patients primarily elected for CI due to severe tinnitus. This study may lead to the broadening of recommendations regarding tinnitus as the major reason for CI in conditions such as single-sided deafness. The risk of developing or worsening of tinnitus after implantation is low, but should be discussed in pre-implantation patient counseling [32].
Conclusions
Our study leads to the following conclusion: tinnitus suppression is a positive side effect of cochlear implantation. Many CI recipients experience pre-implant tinnitus which in most cases are reduced or totally suppressed following implantation. Patients are also less burdened of their tinnitus. The largest tinnitus reduction was observed within 4 months of CI use, but some patients continued improving up to 14 months after implantation. We found a sustained significant reduction in tinnitus and few incidences of new or worsened tinnitus following implantation. This research adds to the evidence that CI is an effective treatment for severe tinnitus in adults with moderate to profound SNHL, but further studies are needed. Understanding tinnitus suppression will assist defining future clinical recommendations and cochlear implantation criteria.
Code availability
All codes for data cleaning and analysis associated with the current submission are available in OSF at https://osf.io/25azh/?view_only=cd198a9c61bc46929b7359572db67523.
Availability of data and materials
All data generated or analysed during the study are included in this published article and presented in Table 2. The anonymized full datasets are available upon reasonable request from the corresponding author.
Abbreviations
- CI:
-
Cochlear implant
- dB HL:
-
Decibel hearing level
- HA:
-
Hearing aid
- NCIQ:
-
Nijmegen cochlear implant questionnaire
- PTA6:
-
Pure-tone audiometry average for six frequencies
- SNHL:
-
Sensorineural hearing loss
- SRS:
-
Speech reception score
- SSQ-12:
-
Speech, Spatial, and Qualities of Hearing Scale
- THI:
-
Tinnitus Handicap Inventory
References
Quaranta N, Wagstaff S, Baguley DM (2004) Tinnitus and cochlear implantation. Int J Audiol. 43(5):245–251. https://doi.org/10.1080/14992020400050033
Andersson G, Freijd A, Baguley DM, Idrizbegovic E (2009) Tinnitus distress, anxiety, depression, and hearing problems among cochlear implant patients with tinnitus. J Am Acad Audiol 20:315–319. https://doi.org/10.3766/jaaa.20.5.5
Pierzycki RH, Kitterick PT (2020) Insomnia, anxiety and depression in adult cochlear implant users with tinnitus. Ear Hear 42:235–243. https://doi.org/10.1097/aud.0000000000000900
Sarac ET, Ozbal Batuk M, Batuk IT, Okuyucu S (2020) Effects of cochlear implantation on tinnitus and depression. ORL J Otorhinolaryngol Relat Spec 82:209–215. https://doi.org/10.1159/000508137
Olze H, Szczepek AJ, Haupt H, Zirke N, Graebel S, Mazurek B (2012) The impact of cochlear implantation on tinnitus, stress and quality of life in postlingually deafened patients. Audiol Neurootol 17:2–11. https://doi.org/10.1159/000323847
Sullivan CB, Al-Qurayshi Z, Zhu V, Liu A, Dunn C, Gantz BJ, Hansen MR (2020) Long-term audiologic outcomes after cochlear implantation for single-sided deafness. Laryngoscope 130:1805–1811. https://doi.org/10.1002/lary.28358
Yuen E, Ma C, Nguyen SA, Meyer TA, Lambert PR (2021) The effect of cochlear implantation on tinnitus and quality of life: a systematic review and meta-analysis. Otol Neurotol 42:1113–1122. https://doi.org/10.1097/MAO.0000000000003172
Opperman E, le Roux T, Masenge A, Eikelboom RH (2021) The effect of tinnitus on hearing-related quality of life outcomes in adult cochlear implant recipients. Int J Audiol 60:246–254. https://doi.org/10.1080/14992027.2020.1828633
Biswas R, Lugo A, Akeroyd MA, Schlee W, Gallus S, Hall DA (2022) Tinnitus prevalence in Europe: a multi-country cross-sectional population study. Lancet Reg Health Eur 12:100250. https://doi.org/10.1016/j.lanepe.2021.100250
Baguley D, McFerran D, Hall D (2013) Tinnitus. Lancet 382:1600–1607. https://doi.org/10.1016/s0140-6736(13)60142-7
World Health Organization (2021) World report on hearing. Geneva. ISBN 978-92-4-002048-1
Nondahl DM, Cruickshanks KJ, Huang GH, Klein BE, Klein R, Nieto FJ, Tweed TS (2011) Tinnitus and its risk factors in the Beaver Dam offspring study. Int J Audiol. 50(5):313–320. https://doi.org/10.3109/14992027.2010.551220
Bovo R, Ciorba A, Martini A (2011) Tinnitus and cochlear implants. Auris Nasus Larynx 38:14–20. https://doi.org/10.1016/j.anl.2010.05.003
Baguley DM, Atlas MD (2007) Cochlear implants and tinnitus. In: Tinnitus: pathophysiology and treatment; progress in brain research, pp347–355
Levy DA, Lee JA, Nguyen SA, McRackan TR, Meyer TA, Lambert PR (2020) Cochlear implantation for treatment of tinnitus in single-sided deafness: a systematic review and meta-analysis. Otol Neurotol 41:e1004–e1012. https://doi.org/10.1097/MAO.0000000000002711
Giraud AL, Price CJ, Graham JM, Frackowiak RSJ (2001) Functional plasticity of language-related brain areas after cochlear implantation. Brain 124:1307–1316. https://doi.org/10.1093/brain/124.7.1307
Kim DK, Moon IS, Lim HJ, Yoo SY, Heo KW, Bae SC, Moon KR, Lee JJ, Choung YH, Park SN (2016) Prospective, multicenter study on tinnitus changes after cochlear implantation. Audiol Neurootol 21:165–171. https://doi.org/10.1159/000445164
Tyler RS, Rubinstein J, Pan T, Chang SA, Gogel SA, Gehringer A, Coelho C (2008) Electrical stimulation of the cochlea to reduce tinnitus. Semin Hear 29:326–332. https://doi.org/10.1055/s-0028-1095892
Kloostra FJJ, Verbist J, Hofman R, Free RH, Arnold R, van Dijk P (2018) A prospective study of the effect of cochlear implantation on tinnitus. Audiol Neurootol 23:356–363. https://doi.org/10.1159/000495132
Kim DK, Bae SC, Park KH, Jun BC, Lee DH, Yeo SW, Park SN (2013) Tinnitus in patients with profound hearing loss and the effect of cochlear implantation. Eur Arch Otorhinolaryngol 270:1803–1808. https://doi.org/10.1007/s00405-012-2193-2
van Zon A, Smulders YE, Ramakers GG, Stegeman I, Smit AL, Van Zanten GA, Stokroos RJ, Hendrice N, Free RH, Maat B et al (2016) Effect of unilateral and simultaneous bilateral cochlear implantation on tinnitus: a prospective study. Laryngoscope 126:956–961. https://doi.org/10.1002/lary.25493
Kompis M, Pelizzone M, Dillier N, Allum J, DeMin N, Senn P (2012) Tinnitus before and 6 months after cochlear implantation. Audiol Neurootol 17:161–168. https://doi.org/10.1159/000335126
Mavrogeni P, Maihoub S, Tamas L, Molnar A (2022) Tinnitus characteristics and associated variables on Tinnitus Handicap Inventory among a Hungarian population. J Otol 17:136–139. https://doi.org/10.1016/j.joto.2022.04.003
Danish Society of Otorhinolaryngology (2013) H.a.N.S. Udredning af voksne til CI. 2013, 3, http://dsohh.dk/wp-content/uploads/2015/04/DSOHH-KKR-CI-voksne1.pdf
Rasmussen KMB, West NC, Bille M, Sandvej MG, Caye-Thomasen P (2022) Cochlear implantation improves both speech perception and patient-reported outcomes: a prospective follow-up study of treatment benefits among adult cochlear implant recipients. J Clin Med. https://doi.org/10.3390/jcm11082257
Newman CW, Jacobson GP, Spitzer JB (1996) Development of the Tinnitus Handicap Inventory. Arch Otolaryngol Head Neck Surg 122:143–148. https://doi.org/10.1001/archotol.1996.01890140029007
Zachariae R, Mirz F, Johansen LV, Andersen SE, Bjerring P, Pedersen CB (2000) Reliability and validity of a Danish adaptation of the Tinnitus Handicap Inventory. Scand Audiol 29:37–43. https://doi.org/10.1080/010503900424589
Wewers ME, Lowe NK (1990) A critical review of visual analogue scales in the measurement of clinical phenomena. Res Nurs Health 13:227–236. https://doi.org/10.1002/nur.4770130405
McCombe A, Baguley D, Coles R, McKenna L, McKinney C, Windle-Taylor P, British Association of Otolaryngologists, H, Neck, S (2001) Guidelines for the grading of tinnitus severity: the results of a working group commissioned by the British Association of Otolaryngologists, Head and Neck Surgeons. Clin Otolaryngol Allied Sci 26:388–393. https://doi.org/10.1046/j.1365-2273.2001.00490.x
Zeman F, Koller M, Figueiredo R, Aazevedo A, Rates M, Coelho C, Kleinjung T, de Ridder D, Langguth B, Landgrebe M (2011) Tinnitus handicap inventory for evaluating treatment effects: which changes are clinically relevant? Otolaryngol Head Neck Surg 145:282–287. https://doi.org/10.1177/0194599811403882
RStudio Team RStudio (2019) Integrated development for R. RStudio Inc, Boston
Amoodi HA, Mick PT, Shipp DB, Friesen LM, Nedzelski JM, Chen JM, Lin VY (2011) The effects of unilateral cochlear implantation on the tinnitus handicap inventory and the influence on quality of life. Laryngoscope 121:1536–1540. https://doi.org/10.1002/lary.21851
Olze H, Szczepek AJ, Haupt H, Förster U, Zirke N, Gräbel S, Mazurek B (2011) Cochlear implantation has a positive influence on quality of life, tinnitus, and psychological comorbidity. Laryngoscope 121:2220–2227. https://doi.org/10.1002/lary.22145
Punte AK, Meeus O, Van de Heyning P (2011) Cochlear implants and Tinnitus. In: Textbook of Tinnitus, pp 619–624
Mallen JR, Chiu J, Marquis H, Ottochian A, Perez E, Kuo CL, Otto S, Ryan T, Roberts DS (2020) Quantifying tinnitus suppression in cochlear implantation using tinnitus interval-limited tracking. Laryngoscope 130:2047–2052. https://doi.org/10.1002/lary.28414
Quaranta N, Fernandez-Vega S, D’Elia C, Filipo R, Quaranta A (2008) The effect of unilateral multichannel cochlear implant on bilaterally perceived tinnitus. Acta Otolaryngol 128:159–163. https://doi.org/10.1080/00016480701387173
Lazard DS, Vincent C, Venail F, Van de Heyning P, Truy E, Sterkers O, Skarzynski PH, Skarzynski H, Schauwers K, O’Leary S et al (2012) Pre-, per- and postoperative factors affecting performance of postlinguistically deaf adults using cochlear implants: a new conceptual model over time. PLoS ONE 7:e48739. https://doi.org/10.1371/journal.pone.0048739
Ramakers GGJ, van Zanten GA, Thomeer H, Stokroos RJ, Heymans MW, Stegeman I (2018) Development and internal validation of a multivariable prediction model for tinnitus recovery following unilateral cochlear implantation: a cross-sectional retrospective study. BMJ Open 8:e021068. https://doi.org/10.1136/bmjopen-2017-021068
Dixon PR, Crowson M, Shipp D, Smilsky K, Lin VY, Le T, Chen JM (2020) Predicting reduced tinnitus burden after cochlear implantation in adults. Otol Neurotol 41:196–201. https://doi.org/10.1097/MAO.0000000000002481
Rodvik AK, Myhrum M, Larsson ELA, Falkenberg ES, Kvaerner KJ (2022) Sustained reduction of tinnitus several years after sequential cochlear implantation. Int J Audiol 61:322–328. https://doi.org/10.1080/14992027.2021.1939448
West N, Bille M, Caye-Thomasen P (2020) Cochlear implantation for adults. Ugeskr Laeger 182. https://www.ncbi.nlm.nih.gov/pubmed/33215588
Funding
K.R. has received a research grant from the William Demant Foundation.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare no conflict of interest.
Institutional review board statement
The study was approved by The Danish National Committee on Health Research Ethics (H-17034918) and The Danish Data Protection Agency (RH-2017-308).
Informed consent statement
Informed consent was obtained from all subjects involved in the study.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Rasmussen, K.D., West, N.C., Bille, M. et al. Tinnitus suppression in a prospective cohort of 45 cochlear implant recipients: occurrence, degree and correlates. Eur Arch Otorhinolaryngol 280, 4073–4082 (2023). https://doi.org/10.1007/s00405-023-07921-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00405-023-07921-1