Abstract
Purpose
Refinement currently offered in new sound processors may improve noise listening capability reducing constant background noise and enhancing listening in challenging signal-to-noise conditions. This study aimed to identify whether the new version of speech processor preprocessing strategy contributes to speech recognition in background noise compared to the previous generation processor.
Methods
This was a multicentric prospective cross-sectional study. Post-lingually deaf adult patients, with at least 1 year of device use and speech recognition scores above 60% on HINT sentences in quiet were invited. Speech recognition performance in quiet and in noise with sound processors with previous and recent technologies was assessed under four conditions with speech coming from the front: (a) quiet (b) fixed noise coming from the front, (c) fixed noise coming from the back, and (d) adaptive noise ratios with noise coming from the front.
Results
Forty-seven cochlear implant users were included. No significant difference was found in quiet condition. Performance with the new processor was statistically better than the previous sound processor in all three noisy conditions (p < 0.05). With fixed noise coming from the back condition, speech recognition was 62.9% with the previous technology and 73.5% on the new one (p < 0.05). The mean speech recognition in noise was also statistically higher, with 5.8 dB and 7.1 dB for the newer and older technologies (p < 0.05), respectively.
Conclusion
New technology has shown to provide benefits regarding speech recognition in noise. In addition, the new background noise reduction technology, has shown to be effective and improves speech recognition in situations of more intense noise coming from behind.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Cochlear implant is an outstanding device that offers the opportunity to hear for those who cannot benefit from conventional hearing aids, although it still does not mimic natural auditory system. The normal functions of the peripheral auditory system beyond the ganglion cells should ideally be represented by speech processor features to mimic natural events as much as possible. As compared with what happens in normal listeners, each structure of the outer and middle ear should ideally be able to contribute to the cochlear implant users’ ability to separate from the background noise what they want to hear and focus their attention on. Similarly, the controlling role of the outer hair cells and the olivo-cochlear system seemingly so effective in normal-hearing individuals should be represented as much as possible in the cochlear implant signal.
Several approaches to signal management have previously been implemented in Nucleus cochlear implant systems, ranging from speech coding strategies development to new microphone features that represent expressive improvement to CI recipients’ outcomes. Furthermore, technologies like Automatic Sensitivity Control (ASC) and adaptive dynamic range optimization (ADRO) are available to automatically improve the input signals received at the microphone [1,2,3]. These initiatives reflect the attempt to provide the profoundly hearing-impaired individuals with better filtering of sound, getting them thus closer to the performance of listeners with healthy peripheral auditory system.
Nevertheless, technology is moving at a fast pace and refinement currently offered in new sound processors, may allow not only device connectivity, but improved noise listening features. These improvements target to reduce constant background noise and enhance listening in challenging signal-to-noise conditions, with intelligent signal processing technology that allows automatic adaptation. Automatic scene classification aims to automatically deliver a selection of appropriate input processing technology to the user at each specific listening environment [4]. Signal management with the use of directional microphones with two microphones acting together and blocking sounds that are captured behind the head of the user, enables the listener to have a better configuration of speech information arriving from the front, while attenuating noise arriving from other directions [5]. ForwardFocus (FF), is a new background noise reduction technology introduced into the Nucleus 7® speech processor, which aims to reduce constant background noise and enhance listening in challenging signal-to-noise conditions [6]. FF was developed to provide additional benefits over directional microphones, as a spatial post-filter technology and is implemented on unilateral conventional behind the ear sound processors (SP) [6].
Another issue to be considered is that sound processors have a finite life and require periodic replacement after several years of use. In such cases, if the benefits derived from a new SP are significant, it could be interesting for users to opt for the most up-to-date device expecting not only a replacement, but a true upgrade. However, SP upgrade usually comes with significant cost implications. It is therefore important that potential real-life benefits of a new SP model over the previous version are objectively demonstrated to users, clinicians and stakeholders [5].
Objective
This research aims to evaluate whether there is a contribution of combining automatic noise reduction pre-processing strategies with fixed microphone directionality in the speech recognition in adult cochlear implant users.
Methods
This study was a multicentric, prospective and cross-sectional study, approved by all individual Ethics Review Boards (main protocol no 3.118.749), in which each subject served as their own control. Assessments were carried out at six different Cochlear Implants centers.
Patients
Selection criteria included adult CI users of both genders, 18 or more years of age with post-lingual deafness, users of at least an unilateral cochlear implant with Nucleus 5® (CP 810) speech processor for at least 1 year, with a minimum score of 60% in HINT in quiet condition with their own SP, native speakers of the local language, with no cognitive, neurological or other alterations that would prevent to carry out the procedures involved in the investigation. Participants were invited and selected after signing the informed consent.
A total of forty-seven subjects meeting the selection criteria were recruited, with ages ranging from 19 to 70 years. Demographical data are shown in Table 1. Among the participants, seven were bilateral CI users and four were bimodal users. However, during test sessions, all participants were tested unilaterally with the CI side or, in the case of bilateral users, the ear with the best outcome was considered in this evaluation.
Subjective listening benefits and satisfaction with their cochlear implants assessed by the SSQ12. Considering the maximum expected score in the SSQ for each of the subscales is ten, median results of 5.4, 5.7 and 6.5 revealed the difficulty in challenging situations for most of the patients, reinforcing the need of improvements in their quality of hearing.
Materials and procedures
Evaluation involved the HINT Test—Hearing in Noise Test (HINT) in the local language [8] presented from one or two loudspeakers positioned at zero azimuth and 180°, 1 m from the patient in a sound attenuated booth, in four conditions (Fig. 1).
-
Quiet: speech presented at 0° azimuth at 65 dBSPL
-
In noise
-
o
Fixed noise S0N0: speech presented at 65 dBSPL with SNR = + 10 dB
-
o
Adaptive noise S0N0: noise presented at 55 dBSPL with variable speech presentation levels with SNR targeted to 75% correct words
-
o
Fixed noise S0N180: speech presented at 65 dBSPL with SNR = 0 dB. In this situation, to assess the contribution of the forward focus (FF) feature, it was enabled in the N7®
-
o
Both processors were evaluated in all conditions. Bilateral CI patients were assessed in unilateral fitting with both Nucleus 5® (N5®) and Nucleus 7® (CP1000) (N7®) processors. Test order with N5® and N7® was randomized (www.randomizer.org).
Subjective listening benefits and satisfaction with their cochlear implants were assessed using the SSQ-12 questionnaire (Speech, Spacial and Qualities) in the local language [7] in their daily life situations with their regular sound processor (CP 810).
In order to perform the evaluations with functioning speech processors, the patient´s regular routine maps (maps in use) were copied to a new N5® speech processor, and also converted to a new processor N7® (Fig. 2). The subjects were fitted with the N7® sound processor using the same settings and same program of the map in use with N5®. No home experience was provided since evaluations were all conducted in the same session. Clinicians also enabled Forward Focus option on N7® speech processors.
Statistical analysis
Collected variables involved the percentage of speech recognition in quiet and in noise and the SRT (in dB) for the adaptive noise situation. The SRT was collected in dB of the signal to noise ratio (SNR) to achieve 75% correct words on sentences at S0N0. Statistical analysis included the mean values comparison with two-tailed paired t test.
Results
Speech recognition mean scores data for N5® and N7® speech processors in quiet and noise conditions are shown in Fig. 3.
Two-tailed paired t test showed no significant difference between N5® and N7® in quiet conditions (Table 2), however, for speech in noise significant difference were observed for all conditions. Results indicate significant improvements in the percentage of correct words and also in the mean SNR with N7® when compared with N5® on a two-tailed paired t test (p < 0.05). In the condition that the speech comes from the front and noise from the back, performance with N7® with ForwardFocus enabled was significantly better than with N5® with Beam (p < 0.05).
Discussion
Since the introduction of the first cochlear implants (ICs) in the 1980s, manufacturers have been refining and upgrading internal components, particularly with regard to electrode design, and signal pre-processing algorithm [4, 5]. Although the IC provides excellent quality of hearing, facilitating communication and speech recognition for its users, speech comprehension remains a challenge in complex real-world acoustic environments. In these situations, background noise can negatively interfere with the understanding of a conversation, as reported by the results from the SSQ in our sample, revealing low scores especially in the speech and quality domains. This study aimed to determine whether technology can improve listening in these challenging situations.
Our results showed significant improvements with no interference in quiet. Indeed, De Ceulaer et al. [9] in a study with adults verified that the average speech recognition in silence was not significantly different among the processors. In fact, the idea is to offer in noise a similar performance to silence.
On the other hand, there was a significant contribution in noise, either in the fixed noise condition (10%) and the adaptive noise mean SRT (1.3 dB) with N7®. In this situation, the automatic scene classifier (SCAN) includes SNR-NR (signal to noise ration noise reduction) technology over the ADRO and ASC (autosensitivity control) used by N5® processor. The SNR-NR is designed to attenuate constant type of signals such as noise, which mask the low energy cues of speech, maintaining the transient information of acoustic signals. Mauger et al. [4] and Plasmans et al. [10] in studies in adults and children using N6 with the automatic algorithm provided (SCAN), showed an improvement of 1.2–1.7 dB in SRT, which would be equivalent to approximately 10–20% increase in speech recognition scores for adults. Although the speech reception threshold (SRT), defined as the signal-to-noise ratio (SNR) at which 50% of the speech is correctly understood has been proposed as a common measure of the ability of a listener to understand speech in noise, in the present study we intended to raise the difficulty and assess the SNR with adaptive noise where 75% of the speech was correctly understood. Even in this more challenging condition, the contribution of the noise reduction capability of the new technology (SNR-NR) could be evidenced. Moreover, the adaptive directional microphone (BEAM) is able to enhance the attenuation of noise by steering its null towards a dominant noise location, while Forward Focus is superior to attenuate noise locations coming from behind Hey et al. [6].
The present study compared speech recognition obtained using the N7® Sound Processor (CP 1000) and N5® (CP 800 series), which incorporates the new single integrated processing chip, enabling further miniaturization, but its basic signal processing is equivalent to that in the predicate Nucleus six series sound processors. The main innovation in quality sound is the incorporation of FowardFocus, which enhances hearing in noise environments with fixed directionality that can be used in combination with automatic input processing technology of ASC, ADRO, SNR/NR.
Our results may be applicable to all CI users, regardless of the speech recognition scores. Nevertheless, based on results of the study from Nascimento and Bevilacqua [10], we predicted that patients with 60% of speech recognition in quiet situations might decline to around 30% at SNR + 10 dB. Thus, we preferred not to recruit patients with less than 60% to avoid the risk of 0% scores during the assessment.
Comparisons of new CI sound processors (SPs) with their old versions have been made in a number of previous studies [9, 11,12,13]. Such studies are important as a means of confirming that the technological development results in measurable clinical benefits. This information may be useful for reimbursement purposes as well as providing evidence to clinical professionals and stakeholders.
Conclusion
Cochlear implant users showed significant improvements in speech perception in noise from the use of noise reduction technologies implemented in the Nucleus 7® sound processor, even without going through home experience with the new technology.
Code availability
Not applicable.
References
Clark G (2003) Cochlear implants: fundamentals and applications. Springer-Verlag, New York
Frederigue NB, Bevilacqua MC (2003) Otimização da percepção da fala em deficientes auditivos usuários do sistema de implante coclear multicanal. Rev Bras Oto 69(2):227–233
Dawson PW, Mauger SJ, Hersbach AA (2011) Clinical evaluation of signal-to-noise ratio based noise reduction in nucleus cochlear-implant recipients. Ear Hear 32:382–390
Mauger SJ, Warren CD, Knight MR, Goorevich M, Nel E (2014) Clinical evaluation of the Nucleus 6 cochlear implant system: performance improvements with smart sound iQ. Int J Audiol 53(8):564–576
Warren CD, Nel E, Boyd PJ (2019) Controlled comparative clinical trial of hearing benefit outcomes for users of the Cochlear™ Nucleus(®) 7 sound processor with mobile connectivity. Cochlear Implants Int 20(3):116–126
Hey M, Hocke T, Böhnke B, Mauger SJ (2019) ForwardFocus with cochlear implant recipients in spatially separated and fluctuating competing signals - introduction of a reference metric. Int J Audiol 58(12):869–878
de Miranda-Gonsalez EC, de Almeida K (2017) Incapacidade auditiva medida por meio do questionário Speech, Spatial and Qualities of Hearing Scale (SSQ): estudo piloto da versão reduzida em Português Brasileiro. Audiol Commun Res 22:e1709
Bevilacqua MC, Banhara MR, Costa EA, Vignoly AB, Alvarenga KF (2008) The brazilian portuguese hearing in noise test (HINT). Int J Audiol 47(6):364–365
De Ceulaer G, Swinnen F, Pascoal D, Philips B, Killian M, James C, Govaerts PJ, Dhooge I (2015) Conversion of adult Nucleus® 5 cochlear implant users to the Nucleus® 6 system. Cochlear Implants Int 16(4):222–232
Nascimento LT, Bevilacqua MC (2005) Evaluation of speech perception in noise in cochlear implanted adults. Braz J Otorhinolaryngol 71(4):432–438
Plasmans A, Rushbrooke E, Moran M, Spence C, Theuwis L, Zarowski A, Offeciers E, Atkinson B, McGovern J, Dornan D, Leigh J, Kaicer A, Hollow R, Martelli L, Looi V, Nel E, Del Dot J, Cowan R, Mauger SJ (2016) A multicentre clinical evaluation of paediatric cochlear implant users upgrading to the Nucleus® 6 system. Int J Pediatr Otorhinolaryngol 83:193–199
Magalhães AT, Goffi-Gomez MV, Hoshino AC, Tsuji RK, Bento RF, Brito R (2013) Converted and upgraded maps programmed in the newer speech processor for the first generation of multichannel cochlear implant. Otol Neurotol 34(7):1193–1200
Mosnier I, Mathias N, Flament J, Amar D, Liagre-Callies A, Borel S, Ambert-Dahan E, Sterkers O, Bernardeschi D (2017) Benefit of the UltraZoom beamforming technology in noise in cochlear implant users. Eur Arch Otorhinolaryngol 274(9):3335–3342
Acknowledgements
The authors would like to thank the CI groups involved in the study, especially Prof. Dr. Ricardo Bento, Prof. Dr. Mariana Leal, Prof. Dr. Rogerio Hamerschmidt, Dr. Arthur Castilho, Prof. Dr. Shiro Tomita, Prof. Dr. Fayez Bahmad for their clinical and scientific support for this study.
Funding
Authors have no financial disclosure to declare.
Author information
Authors and Affiliations
Contributions
MVSGG and BC conceived the research idea and design. MVSGG and BC performed interpretation and analysis. All authors participated in data collection, preparing and reviewing the manuscript content.
Corresponding author
Ethics declarations
Conflict of interest
All authors have no conflict of interest to declare related to the scope of this manuscript, except BC who is a Research Manager for Cochlear Latin America.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Goffi-Gomez, M.S., Muniz, L., Wiemes, G. et al. Contribution of noise reduction pre-processing and microphone directionality strategies in the speech recognition in noise in adult cochlear implant users. Eur Arch Otorhinolaryngol 278, 2823–2828 (2021). https://doi.org/10.1007/s00405-020-06372-2
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00405-020-06372-2