Abstract
Introduction
Ionizing radiation is used daily during endourological procedures. Despite the dangers of both deterministic and stochastic effects of radiation, there is a lack of knowledge and awareness among urologists. This study reviewed the literature to identify the radiation exposure (RE) of urologists during endourological procedures.
Methods
A literature search of the Medline, Web of Science, and Google Scholar databases was conducted to collect articles related to the radiation dose to urologists during endourological procedures. A total of 1966 articles were screened. 21 publications met the inclusion criteria using the PRIMA standards.
Results
Twenty-one studies were included, of which 14 were prospective. There was a large variation in the mean RE to the urologist between studies. PCNL had the highest RE to the urologist, especially in the prone position. RE to the eyes and hands was highest in prone PCNL, compared to supine PCNL. Wearing a thyroid shield and lead apron resulted in a reduction of RE ranging between 94.1 and 100%. Educational courses about the possible dangers of radiation decreased RE and increased awareness among endourologists.
Conclusions
This is the first systematic review in the literature analyzing RE to urologists over a time period of more than four decades. Wearing protective garments such as lead glasses, a thyroid shield, and a lead apron are essential to protect the urologist from radiation. Educational courses on radiation should be encouraged to further reduce RE and increase awareness on the harmful effects of radiation, as the awareness of endourologists is currently very low.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Ionizing radiation is used in daily practice during endourological procedures. It is well known that radiation has side effects on the human body, including deterministic and stochastic effects. Deterministic effects of radiation manifest after exposure to ionizing radiation surpasses a specific threshold. Examples of such effects include hair loss and dermal burns. [1]. Fortunately, these radiation thresholds are not reached during endourological procedures under normal conditions. Stochastic effects, conversely, occur in a linear manner, without a lower threshold for their manifestation. This suggests that higher levels of exposure elevate the likelihood of an effect rather than the type or severity of the radiation's impact. Examples include leukemia, multiple myeloma, and various types of cancers, such as thyroid, bladder, breast, lung, ovarian, and colon cancers. Stochastic effects do occur in endourological procedures where ionizing rays are used [2]. The development of cataracts is considered to be a consequence of both deterministic and stochastic effects of radiation [1]. Minor opacities of the eye lens result from stochastic effects of radiation and occur in a linear fashion. Visually impairing cataracts, however, result from exceeding a certain threshold of radiation, which is a deterministic effect [1].
Generally, there is lack of knowledge and awareness about the dangers of ionizing radiation among urologists. After conducting a survey among urology residents in 2012, it was found that half of the residents were unaware of the potential fatal cancer risk associated with ionizing radiation [3]. Urology residents exhibited notably low compliance with protective equipment during various procedures. Merely 30.6% consistently wore a thyroid shield, and a mere 4% consistently wore protective eyeglasses during endourological procedures [3]. In contrast, 52% of high-volume endourologists used protective lead eyewear with every procedure [4]. These results highlighted the large gap between residents and high-volume urologists. In a recent study, urologists with 15–20 years of experience wore a thyroid shield in 90% of cases and protective glasses in 16% of cases. Conversely, residents utilized these protective items in 93.5% and 8.7% of cases, respectively [5]. Only 15–35% of urologists always wore their dosimeter [3, 4, 6, 7]. Urologists who took an educational course on ionizing radiation, wore protective equipment significantly more frequently than those who never took a course on radiation safety [3].
Considering the knowledge gaps mentioned above, we aimed to identify the radiation exposure (RE) of urologists during endourological procedures, as reported in published literature, to further increase the awareness about the risks related to ionizing radiation for urologists.
Materials and methods
Two authors (LH and XM) conducted a systematic review of literature using the Medline, Web of Science and Google Scholar databases in August 2023. The search terms “fluoroscopy AND (urology OR ureteroscopy OR urs OR (retrograde intrarenal surgery) OR rirs OR (percutaneous nephrolithotomy) OR pcnl OR (extracorporeal shock wave lithotripsy) OR eswl OR lithotripsy)” were used and the filters “english” and “humans” were applied. Articles published since 1980 were considered. All articles related to the radiation dose to urologists during endourological procedures, were included. After full text assessment of these articles (L.H. and X.M.) and using the population, intervention, comparator, outcome (PICO) study design approach and Preferred Reporting Items for Systematic Reviews and Meta-analysis (PRISMA) standards, publications that met the inclusion criteria for this review were chosen (Fig. 1). Case reports, editorials and letters were excluded. Additional articles, identified through references lists, were also included. A narrative synthesis for analysis of the studies was used.
Data was summarized in tables, which were classified according to the body location of the dosimeter and according to the procedure type. Only the studies that reported the effective dose to the urologist in mSv were further discussed and analyzed, to be able to compare their results to the recommendations of ICRP, which also report RE in mSv [8, 9]. The studies that reported effective dose in other units were still included, but not further discussed, nor were they compared to the studies that reported effective radiation dose in mSv. The positions of the patient during percutaneous nephrolithotomy (PCNL) were divided into prone, supine, or not specified. When the RE of a limb of the urologist was measured, they were divided into right side, left side, dominant side or not specified.
Results
The literature search included 2036 records. After exclusion of the duplicates, 1966 records remained. Then, 1815 records were excluded based on title and abstract. Next, 151 full articles were assessed. 130 articles that only investigated radiation time or radiation dose to the patient and not to the urologist, were excluded. Finally, 21 articles were included in the review to examine the radiation dose to urologists during endourological procedures (Supplementary Fig. 1) [10].
Characteristics of the included studies are shown in Table 1 [6, 11,12,13,14,15,16,17,18,19,20,21,22,23,24,25,26,27,28,29,30]. A summary of the reported RE is shown in Table 2, categorized per body location and procedure [6, 11,12,13,14,15,16,17,18,19,20,21,22,23,24,25,26,27,28,29,30].
Considering the studies that reported effective dose (unit: mSv), the average radiation exposure (RE) to the eyes ranged from 0.02 mSv to 0.8 mSv per procedure, as outlined in Table 2. Similarly, the mean RE to the forehead varied between 0.03 mSv and 0.18 mSv per procedure (Table 2). Notably, there existed a distinction between the RE to the neck above the thyroid shield (ranging from 0.00222 mSv to 0.33 mSv per procedure) and the RE to the neck under the thyroid shield (ranging from 0.00084 mSv to 0.099 mSv) (Table 2). Regarding the chest, the mean RE over the lead apron varied from 0.0001 mSv to 1.12 mSv per procedure, while the mean RE under the lead apron ranged from 0 mSv to 0.02 mSv per procedure (Table 2). Moreover, the mean RE to the urologist's arm ranged from 0.31 mSv to 0.55 mSv (Table 2), whereas the mean RE to the urologist's hands varied between 0.008 mSv and 4.36 mSv per procedure (Table 2). Additionally, the mean RE to the urologist's foot ranged from 0.05 mSv to 0.1 mSv per procedure (Table 2). Lastly, the mean RE to the urologist's leg ranged from 4.1 µGy to 167 µGy (Table 2).
Five studies evaluated the reduction of RE when wearing a thyroid shield and a lead apron (Supplementary table) [17, 22, 26,27,28]. Wearing this protective equipment resulted in a 94.1 to 100% reduction of RE. Inoue et al. evaluated the influence of a lead curtain on RE [28]. They noticed a reduction of RE from 62.2% to 86.1% when wearing a thyroid and lead apron, respectively. When the lead apron was combined with the use of a lead curtain, this resulted in absolutely no measurable RE (0 mSv) under the lead apron [28].
A few studies did not differentiate RE by type of procedure [12, 15, 21]. For PCNL, most studies did not mention whether the procedure was performed with the patient in supine or prone position [11,12,13,14, 18, 21, 23, 24]. Finally, no study reported RE per urologist per year.
Discussion
This is the first systematic review in literature analyzing the RE to the urologist over a period of more than four decades. While RE generally remained very low during most endourological procedures, a few studies have reported rather high mean RE per procedure and deserve detailed analysis.
Generally, there was a large variation in mean RE to the urologist between studies. In studies that compared different procedure types, RE was highest during PCNL [6, 16, 20, 26]. The studies reporting the highest mean RE to the urologist ranged up to 0.8 mSv to the eyes, 0.18 mSv to the forehead, 4.36 mSv to hands and 0.1 mSv to feet, respectively [11, 24, 25]. In prone PCNL, highest RE was documented to the eyes and hands [6, 11]. Considering the latter, RE to the hands and eyes was higher during prone PCNL compared to supine PCNL [6]. For supine PCNL, highest RE was documented to the neck and chest [6]. Concerning comparisons between operating staff, highest RE in PCNL was measured to the person standing closest to the patient [14, 30]. As for URS, the RE was generally lower compared to PCNL, with the highest mean RE reaching 0.0427 mSv to the eyes, 0.10 mSv to the forehead, and 0.81 mSv to the hand, respectively [13, 24, 26]. During URS, highest RE under the thyroid shield and lead apron was reported to be 0.01 mSv and 0.02 mSv, respectively [26].
A markable reduction of RE ranging from 94.1 to 100% was seen when wearing a thyroid shield and lead apron (Supplementary table) [6, 17, 22, 26,27,28]. This corresponds with reports on transmitted exposure through lead equivalent aprons: lead aprons of 0.25 mm and 0.5 mm thickness were shown to attenuate at least 90% and 98% of RE, respectively [31,32,33]. Also, lead glasses, including their lateral protection, have been proven useful. Without lateral protection, the eyes are still exposed to 50% of the radiation load [2, 34]. One study stated that the registered eye lens RE to the urologist does not seem to be related to the number of procedures, but rather to the use of lead glasses and the ALARA protocol, which will be further discussed below [35].
The maximum effective dose per year, determined by the ICRP, is 20 mSv to the lens of the eye and 500 mSv to the skin, hands and feet [8, 9]. On a hypothetical basis and considering studies with higher RE range, some urologists could only perform 25 PCNL procedures or 69 URS procedures per year before exceeding the ICRP maximum effective dose to the eyes [11, 14, 17]. Vano et al. compared their high RE to the eyes (0.296 mSv) and long mean fluoroscopy time (11.5 min) during PCNL with other disciplines in their hospital [14]. The radiation exposure (RE) to urologists was found to be 18.7 times higher than that experienced by radiologists and cardiologists, and 4.2 times higher than the values recorded for vascular surgeons [14]. Urologists seem to be more exposed to radiation when sitting down, due to the closer distance to the radiation source [36]. Therefore, some studies advised to perform endourological procedures in a standing position [36]. Other factors that increased RE during PCNL include standing closely to the patient, larger stone burden, increased number of accesses, larger sheath size, higher kV on fluoroscopy settings and lower stone Hounsfield units [14, 30, 34, 37].
To avoid exceeding dose limits and the associated negative long-term effects of ionizing radiation, it is recommended to act in accordance with the “as low as reasonably achievable (ALARA)” principle [38,39,40]. This can be achieved by reducing fluoroscopy time, using pulsed fluoroscopy, low-dose radiation, collimation to the zone of interest, increasing the distance between the radiation source and the urologist, and keeping hands out of the radiation beam [7, 38, 41]. Also, synchronizing the patient's respiration with the use of fluoroscopy can reduce RE [42]. Unfortunately, following a recent survey, 10% of urologists did not know this ALARA principle and 6% of urologists never took any additional radiation protection course [7]. Educational courses about the possible dangers of RE have been proven to raise awareness among urologists and to decrease RE [3, 43].
The exposure detection methods have changed over the years. In the oldest studies, film badges were used, which directly measure radiation dose [11, 27]. Some studies used optically stimulated luminescence dosimeters [14, 17, 26]. These passive dosimeters measure the scattered dose [14]. Other studies used electronic pocket dosimeters [14, 28] that measure the scattered dose and the dose rate received by the urologist every second. When worn on the chest over the lead apron, they can estimate RE to the eyes, according to the ICRP [14]. In the more recent articles, thermoluminescent dosimeters were the standard method of detection of RE. These dosimeters are compact and lightweight, which make them easy to wear on different parts of the body [11]. Only one study used educational direct dosimeters, which are not as compact, but can detect the radiation dose very accurately [13].
This review has some limitations. Due to methodological variations between studies, evolution of RE cannot be identified over the years. Some studies documented RE in other units than mSv. This impedes the comparison between studies and the contextualization of these results in relation to the annual maximum radiation doses established by the ICRP. Also, different methods of radiation detection were used, which might have caused variations in reported RE.
Based on the results of this review, wearing lightweight thermoluminescent dosimeters during every procedure on different parts of the body, as well as wearing a lead apron, a thyroid shield and lead glasses with lateral protection, would provide adequate protection during endourological procedures. The recommendation of wearing gloves is contradicted by the high annual dose limit to the extremities, i.e. 500mSv, as well as the risk of increasing fluoroscopy time due to the lesser tactile function when gloves are worn [8, 44]. Also, it remains important to use pulsed fluoroscopy and to keep distance from the patient’s body, as RE to the urologist is reduced approximately by 4 when the distance between urologist's and patient’s bodies is doubled [44]. Furthermore, it would be of interest to gain knowledge of total RE to the urologist on an annual basis, as well as detailing the number of endourological interventions per operator. This would provide information between high and low volume centers. Also, recent studies found reduced RE when including ultrasound during endourological procedures without compromising outcomes [45,46,47,48,49,50]. Some studies have even shown that fluoroscopy is not needed at all to perform a successful endourological procedure [51,52,53,54]. The role of a radiation technologist has also been studied to a limited extent and could contribute to a lower RE [55, 56]. Together with the influence of lead gloves, the role of caseload and surgical experience, this should be evaluated in future prospective studies to further reduce RE.
Conclusions
There is a large variation in RE to the urologist during endourological procedures. Highest RE is observed during PCNL, especially in prone position. Considering studies with highest RE, urologists would be limited to 25 PCNL or 69 URS procedures per year before reaching the ICRP maximum effective dose of 20 mSv to the lens of the eye. Wearing a thyroid shield and lead apron resulted in a reduction of RE ranging between 94.1 and 100%. Educational courses on radiation have been proven to reduce RE and increase the awareness on the harmful effects of radiation.
Data availability statement
The data that support the findings of this study are available on request from the corresponding author.
References
Hamada N, Azizova TV, Little MP (2020) An update on effects of ionizing radiation exposure on the eye. Br J Radiol 93(1115):20190829. https://doi.org/10.1259/bjr.20190829
Miller DT, Semins MJ (2021) Safety During Ureteroscopy: Radiation, Eyes, and Ergonomics. Front Surg. https://doi.org/10.3389/fsurg.2021.737337
Söylemez H, Sancaktutar AA, Silay MS et al (2013) Knowledge and attitude of European urology residents about ionizing radiation. Urology 81(1):30–35. https://doi.org/10.1016/j.urology.2012.07.097
Dudley AG, Semins MJ (2015) Radiation Practice Patterns and Exposure in the High-volume Endourologist. Urology 85(5):1019–1024. https://doi.org/10.1016/j.urology.2014.11.060
Tzelves L, Somani B, Knoll T et al (2020) Level of knowledge on radiation exposure and compliance to wearing protective equipment: where do endourologists stand? An ESUT/EULIS survey. World J Urol 38(3):761–768. https://doi.org/10.1007/s00345-019-02807-6
Galonnier F, Traxer O, Rosec M et al (2016) Surgical Staff Radiation Protection During Fluoroscopy-Guided Urologic Interventions. J Endourol 30(6):638–643. https://doi.org/10.1089/end.2016.0022
Henderickx M, Baard J, Beerlage HP, Kamphuis GM (2021) Fluoroscopy-use during ureterorenoscopy: are urologists concerned about radiation exposure? A nationwide survey in Belgium and The Netherlands. Acta Chir Belg 121(3):170–177. https://doi.org/10.1080/00015458.2019.1693150
Rehani MM, Ciraj-Bjelac O, Vañó E et al (2010) ICRP Publication 117 Radiological protection in fluoroscopically guided procedures performed outside the imaging department. Ann ICRP 40(6):1–102. https://doi.org/10.1016/j.icrp.2012.03.001
EU-OSHA. Directive 2013/59/Euratom: protection against ionising radiation. Secondary Directive 2013/59/Euratom: protection against ionising radiation 07/2019 2014. https://osha.europa.eu/en/legislation/directives/directive-2013-59-euratom-protection-against-ionising-radiation.
Page MJMJ, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD et al (2021) The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ. https://doi.org/10.1136/bmj.n71
Rao PN, Faulkner K, Sweeney JK, Asbury DL, Sambrook P, Blacklock NJ (1987) Radiation dose to patient and staff during percutaneous nephrostolithotomy. Br J Urol 59(6):508–512. https://doi.org/10.1111/j.1464-410x.1987.tb04864.x
Taylor ER, Kramer B, Frye TP, Wang S, Schwartz BF, Köhler TS (2013) Ocular radiation exposure in modern urological practice. J Urol 190(1):139–143. https://doi.org/10.1016/j.juro.2013.01.081
Hristova-Popova J, Zagorska A, Saltirov I, Petkova K, Vassileva J (2015) Risk of radiation exposure to medical staff involved in interventional endourology. Radiat Prot Dosimetry 165(1–4):268–271. https://doi.org/10.1093/rpd/ncv089
Vano E, Fernandez JM, Resel LE, Moreno J, Sanchez RM (2016) Staff lens doses in interventional urology A comparison with interventional radiology, cardiology and vascular surgery values. J Radiol Prot 36(1):37–48. https://doi.org/10.1088/0952-4746/36/1/37
Medici S, Pitzschke A, Cherbuin N, Boldini M, Sans-Merce M, Damet J (2017) Eye lens radiation exposure of the medical staff performing interventional urology procedures with an over-couch X-ray tube. Phys Med 43:140–147. https://doi.org/10.1016/j.ejmp.2017.11.002
Hartmann J, Distler F, Baumüller M, Guni E, Pahernik S, Wucherer M (2018) Risk of Radiation-Induced Cataracts: Investigation of Radiation Exposure to the Eye Lens During Endourologic Procedures. J Endourol 32(10):897–903. https://doi.org/10.1089/end.2018.0324
Park IW, Kim SJ, Shin D, Shim SR, Chang HK, Kim CH (2021) Radiation exposure to the urology surgeon during retrograde intrarenal surgery. PLoS ONE. https://doi.org/10.1371/journal.pone.0247833
Hellawell GO, Mutch SJ, Thevendran G, Wells E, Morgan RJ (2005) Radiation exposure and the urologist: what are the risks? J Urol 174(3):948–952. https://doi.org/10.1097/01.ju.0000170232.58930.8f
Safak M, Olgar T, Bor D, Berkmen G, Gogus C (2009) Radiation doses of patients and urologists during percutaneous nephrolithotomy. J Radiol Prot 29(3):409–415. https://doi.org/10.1088/0952-4746/29/3/005
Majidpour HS (2010) Risk of radiation exposure during PCNL. Urol J 7(2):87–89
Patel R, Dubin J, Olweny EO, Elsamra SE, Weiss RE (2017) Use of Fluoroscopy and Potential Long-Term Radiation Effects on Cataract Formation. J Endourol 31(9):825–828. https://doi.org/10.1089/end.2016.0454
Yecies TS, Fombona A, Semins MJ (2017) Single Pulse-per-second Setting Reduces Fluoroscopy Time During Ureteroscopy. Urology 103:63–67. https://doi.org/10.1016/j.urology.2017.01.019
Bowsher WG, Blott P, Whitfield HN (1992) Radiation protection in percutaneous renal surgery. Br J Urol 69(3):231–233. https://doi.org/10.1111/j.1464-410x.1992.tb15518.x
Ritter M, Krombach P, Martinschek A et al (2012) Radiation exposure during endourologic procedures using over-the-table fluoroscopy sources. J Endourol 26(1):47–51. https://doi.org/10.1089/end.2011.0333
Amirhasani S, Daneshdoost R, Mousavibahar S, Ghazikhanlou-Sani K, Raeisi R (2021) Reduction of Radiation Dose Received by Surgeons and Patients During Percutaneous Nephrolithotomy: A New Shielding Method. Urol J 18(3):271–276. https://doi.org/10.22037/uj.v16i7.5200
Kim CH, Kim SJ, Kim MH et al (2016) Are Urologists Performing Semi-rigid Ureteroscopic Lithotripsy Safe From Radiation Exposure? A Guidance to Reduce the Radiation Dose. Urology 95:54–59. https://doi.org/10.1016/j.urology.2016.06.001
Bagley DH, Cubler-Goodman A (1990) Radiation exposure during ureteroscopy. J Urol 144(6):1356–1358. https://doi.org/10.1016/s0022-5347(17)39739-2
Inoue T, Komemushi A, Murota T et al (2017) Effect of Protective Lead Curtains on Scattered Radiation Exposure to the Operator During Ureteroscopy for Stone Disease: A Controlled Trial. Urology 109:60–66. https://doi.org/10.1016/j.urology.2017.07.036
Lowe FC, Auster M, Beck TJ, Chang R, Marshall FF (1986) Monitoring radiation exposure to medical personnel during percutaneous nephrolithotomy. Urology 28(3):221–226. https://doi.org/10.1016/0090-4295(86)90047-6
Kumari G, Kumar P, Wadhwa P, Aron M, Gupta NP, Dogra PN (2006) Radiation exposure to the patient and operating room personnel during percutaneous nephrolithotomy. Int Urol Nephrol 38(2):207–210. https://doi.org/10.1007/s11255-005-4972-9
Bushberg JT, Boone JM (2011) The essential physics of medical imaging: Lippincott Williams & Wilkins
Simon SL (2011) Organ-specific external dose coefficients and protective apron transmission factors for historical dose reconstruction for medical personnel. Health Phys 101(1):13
Christodoulou EG, Goodsitt MM, Larson SC, Darner KL, Satti J, Chan HP (2003) Evaluation of the transmitted exposure through lead equivalent aprons used in a radiology department, including the contribution from backscatter. Med Phys 30(6):1033–1038
Doizi S, Audouin M, Villa L et al (2019) The eye of the endourologist: what are the risks? A review of the literature. World J Urol 37(12):2639–2647. https://doi.org/10.1007/s00345-019-02667-0
Bravo-Balado A, Fontanet S, Skolarikos A et al (2023) Estimated Radiation Dose to the Lens During Endourologic Procedures: The Role of Leaded Glasses and the ALARA Protocol-An ESU/ESUT-YAU Endourology Group Collaboration. J Endourol 37(8):935–939. https://doi.org/10.1089/end.2023.0061
Horsburgh BA, Higgins M (2016) A Study of Occupational Radiation Dosimetry During Fluoroscopically Guided Simulated Urological Surgery in the Lithotomy Position. J Endourol 30(12):1312–1320. https://doi.org/10.1089/end.2016.0596
Balaji SS, Vijayakumar M, Singh AG, Ganpule AP, Sabnis RB, Desai MR (2019) Analysis of factors affecting radiation exposure during percutaneous nephrolithotomy procedures. BJU Int 124(3):514–521. https://doi.org/10.1111/bju.14833
A. Skolarikos HJ, A. Neisius, A. Petřík, B. Somani, T. Tailly, G. Gambaro,. EAU Guidelines on Urolithiasis. Secondary EAU Guidelines on Urolithiasis march 2023 2023. https://d56bochluxqnz.cloudfront.net/documents/full-guideline/EAU-Guidelines-on-Urolithiasis-2023.pdf.
Assimos D, Krambeck A, Miller NL et al (2016) Surgical Management of Stones: American Urological Association/Endourological Society Guideline. PART II. J Urol 196(4):1161–1169. https://doi.org/10.1016/j.juro.2016.05.091
Bhanot R, Hameed ZBM, Shah M, Juliebø-Jones P, Skolarikos A, Somani B (2022) ALARA in Urology: Steps to Minimise Radiation Exposure During All Parts of the Endourological Journey. Curr Urol Rep 23(10):255–259. https://doi.org/10.1007/s11934-022-01102-z
Ozaki K, Fukawa T, Kawanaka T et al (2022) Appropriate management reduces radiation exposure in daily urological practice. Int J Urol 29(10):1207–1212. https://doi.org/10.1111/iju.14971
Tzelves L, Juliebø-Jones P, Manolitsis I et al (2023) Radiation protection measures during endourological therapies. Asian J Urol 10(3):215–225. https://doi.org/10.1016/j.ajur.2022.12.001
Kumar V, Kumar Pal A, Ks S et al (2021) Effect of Structured Educational Program on Practices of Radiation Safety Measures Among Health Care Providers in Urology Operation Theater. Cureus. https://doi.org/10.7759/cureus.15765
Vassileva J, Zagorska A, Karagiannis A et al (2021) Radiation Exposure of Surgical Team During Endourological Procedures: International Atomic Energy Agency-South-Eastern European Group for Urolithiasis Research Study. J Endourol 35(5):574–582. https://doi.org/10.1089/end.2020.0630
Arabzadeh Bahri R, Maleki S, Shafiee A, Shobeiri P (2023) Ultrasound versus fluoroscopy as imaging guidance for percutaneous nephrolithotomy: A systematic review and meta-analysis. PLoS ONE. https://doi.org/10.1371/journal.pone.0276708
Chau HL, Chan HC, Li TB, Cheung MH, Lam KM, So HS (2016) An Innovative Free-Hand Puncture Technique to Reduce Radiation in Percutaneous Nephrolithotomy Using Ultrasound with Navigation System Under Magnetic Field: A Single-Center Experience in Hong Kong. J Endourol 30(2):160–164. https://doi.org/10.1089/end.2015.0296
Chi T, Masic S, Li J, Usawachintachit M (2016) Ultrasound Guidance for Renal Tract Access and Dilation Reduces Radiation Exposure during Percutaneous Nephrolithotomy. Adv Urol 2016:3840697. https://doi.org/10.1155/2016/3840697
Goren MR, Goren V, Ozer C (2017) Ultrasound-Guided Shockwave Lithotripsy Reduces Radiation Exposure and Has Better Outcomes for Pediatric Cystine Stones. Urol Int 98(4):429–435. https://doi.org/10.1159/000446220
Baralo B, Samson P, Hoenig D, Smith A (2020) Percutaneous kidney stone surgery and radiation exposure: a review. Asian J Urol 7(1):10–17. https://doi.org/10.1016/j.ajur.2019.03.007
Usawachintachit M, Tzou DT, Washington SL 3rd, Hu W, Li J, Chi T (2017) Ultrasound-Guided Renal Access and Tract Dilation. Videourology (New Rochelle). https://doi.org/10.1089/vid.2016.0046
Aliaga A, Avalos AV, Sanchez R, Rojas S, Aguila F, Marchant F (2023) Fluoroscopy-free semirigid ureteroscopy for ureteral stone treatment: A prospective single-arm study of feasibility, efficacy, and safety. Urol Ann 15(3):304–310. https://doi.org/10.4103/ua.Ua_181_20
Bragaru M, Popescu RI, Munteanu AM, Cozma C, Geavlete P, Geavlete B (2023) Flexible Ureteroscopy Without Radiation Exposure. Maedica (Bucur) 18(2):203–208. https://doi.org/10.26574/maedica.2023.18.2.203
Chi T, Usawachintachit M, Weinstein S et al (2017) Contrast Enhanced Ultrasound as a Radiation-Free Alternative to Fluoroscopic Nephrostogram for Evaluating Ureteral Patency. J Urol 198(6):1367–1373. https://doi.org/10.1016/j.juro.2017.07.074
Smith HE, Bryant DA, KooNg J, Chapman RA, Lewis G (2016) Extracorporeal shockwave lithotripsy without radiation: Ultrasound localization is as effective as fluoroscopy. Urol Ann 8(4):454–457. https://doi.org/10.4103/0974-7796.192104
Elkoushy MA, Morehouse DD, Anidjar M, Elhilali MM, Andonian S (2012) Impact of radiological technologists on the outcome of shock wave lithotripsy. Urology 79(4):777–780. https://doi.org/10.1016/j.urology.2011.09.013
Logarakis NF, Jewett MA, Luymes J, Honey RJ (2000) Variation in clinical outcome following shock wave lithotripsy. J Urol 163(3):721–725
Acknowledgements
None.
Funding
None.
Author information
Authors and Affiliations
Contributions
Vincent De Coninck: Protocol/project development, Data collection or management, Data analysis, Manuscript writing/editing. Laura Hendrickx: Data collection or management, Data analysis, Manuscript writing/editing. Xavier Mortiers: Data collection or management, Data analysis. Thomas Tailly: Manuscript writing/editing. Bhaskar Somani: Manuscript writing/editing. Esteban Emiliani: Manuscript writing/editing. Emre Tarik Sener: Manuscript writing/editing. Amelia Pietropaolo: Manuscript writing/editing. Patrick Jones: Manuscript writing/editing. Andreas Skolarikos: Manuscript writing/editing. Stefan De Wachter: Manuscript writing/editing. Olivier Traxer: Manuscript writing/editing. Etienne Xavier Keller: Data analysis, Manuscript writing/editing.
Corresponding author
Ethics declarations
Conflict of interest
Vincent De Coninck is a speaker and/or consultant for BD, Coloplast, and Karl Storz, and has no specific conflicts relevant to this study. Olivier Traxer is a consultant for Coloplast, Karl Storz, Rocamed, Quanta Systems, Ambu, Boston Scientific, and IPG Medical, and has no specific conflicts relevant to this study. Etienne Xavier Keller is a speaker and/or consultant for Coloplast, Olympus, Boston Scientific, Recordati, Debiopharm and Alnylam, and has no specific conflicts of interest relevant to this work. All other authors have no conflicts of interest.
Ethics approval
Not applicable.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
De Coninck, V., Hendrickx, L., Mortiers, X. et al. Radiation exposure of urologists during endourological procedures: a systematic review. World J Urol 42, 310 (2024). https://doi.org/10.1007/s00345-024-05023-z
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s00345-024-05023-z