Introduction

Dual-energy computed tomography (DECT) has the potential to generate virtual monoenergetic images at specific kiloelectron volt (keV) levels ranging from 40 keV to 190 keV, reconstructed from the combination of two acquisitions with different tube potentials [1, 2]. Low keV level virtual monoenergetic images are characterized by higher contrast resulting from maximum iodine attenuation (close to the K edge of iodine at 33.2 keV), leading to improved liver lesions conspicuity [3,4,5]. Sudarski et al. [6] and Yamada et al. [7] have previously described the usefulness of standard virtual monoenergetic images (VMI), reporting that peak contrast-to-noise ratios (CNRs) for liver parenchyma and focal liver lesions occurred at 70 keV, resulting from the higher attenuation and lower image noise. Nevertheless, VMI datasets below 70 keV have been reported to be adversely affected by high image noise, limiting clinical application [8, 9].

Recently, an advanced noise-optimized virtual monoenergetic reconstruction algorithm (VMI+) has been developed to overcome this technical limitation [10]. The VMI+ algorithm combines the greater iodine attenuation of the standard VMI algorithm at low virtual photon energies with the lower image noise at higher virtual photon energies, ultimately resulting in a higher CNR at low virtual photon energies. Marin et al. confirmed the higher CNR at 40 keV for hypervascular liver lesions by using the VMI+ algorithm compared to standard VMI reconstruction [11]. However, whether these differences in image quality translate into an improved detection rate of hypervascular liver lesions remains unknown.

Thus, this study sought to assess the image quality and diagnostic accuracy of a noise-optimized VMI+ algorithm in comparison with standard VMI and linearly-blended reconstructions for the detection of hypervascular liver lesions in late hepatic arterial-phase DECT.

Materials and methods

Patient population

This prospective, single-centre study was approved by our local Institutional Review Board and conducted in compliance with the Health Insurance Portability and Accountability Act. All patients provided written informed consent before enrolment.

Between September 2014 and February 2015, consecutive patients with known or suspected hypervascular liver lesions who underwent a clinically indicated magnetic resonance imaging (MRI) examination of the liver were enrolled in the study. MRI was considered the reference standard. Additional details on the reference standard rationale and MRI technique are available in the Online Appendix. Inclusion and exclusion criteria are listed in Supplemental Table 1.

Radiation dose

The dose length product (DLP) of the late arterial phase, automatically provided in the scan protocol, was recorded for each examination. Effective radiation dose (ED) was calculated by multiplying the DLP with an abdomen-specific conversion coefficient k of 0.015 mSv·mGy−1 · cm−1 [12].

CT technique

A third-generation dual-source CT scanner was used for all CT studies (Somatom Force; Siemens Healthineers, Forchheim, Germany) in dual-energy (DE) mode. The acquisitions were conducted in a cranio-caudal direction from the diaphragmatic dome to the iliac crests. The following parameters were used for the DE arterial phase: tube settings 100/150Sn kVp and 180/90 mAs, respectively, equipped with an integrated tin filter (Selective Photon Shield II, SPS II, Siemens); pitch 0.7; collimation 2 × 64 × 0.6 mm for both detectors; field-of-view 350 mm. Automated real-time anatomical tube current modulation (CARE Dose 4D, Siemens) was selected in all acquisitions.

Contrast media (Omnipaque 350 mgI/ml, GE Healthcare, Milwaukee, WI, USA) was injected at a flow rate of 4 ml/s, followed by a 30-ml saline chaser using a dual-syringe power injector (Stellant D CT Injection System, Medrad, Inc., Warrendale, PA, USA) through an 18-gauge access placed in a superficial vein in the antecubital fossa, adapting the amount of contrast media to the BMI [13]. For subjects with a body mass index (BMI) less than 30 kg/m2, a contrast volume of 100 ml was administered, whereas 120 ml was administered for subjects with a BMI greater than 30 kg/m2. A dedicated bolus-tracking technique application (CARE Bolus, Siemens) was used to set the timing of acquisition with a region-of-interest (ROI) drawn in the abdominal aorta at the level of the diaphragmatic dome. A threshold of 100 Hounsfield units (HU) for the tracking scan (100 kVp, 23 mAs) was used with an 18-s delay for the late arterial phase and 70-s delay for the portal venous phase. Only the DE late arterial phase datasets were analysed in this research study.

Image reconstruction

All raw data were transferred to a three-dimensional workstation (syngo.via, version VA30, Siemens). DECT datasets were reconstructed using an abdominal soft tissue convolution kernel (Qr40) with a section thickness of 3.0 mm and an increment of 2.0 mm. Three image series were reconstructed for the purpose of the study. The first image series was designed to simulate the traditional clinical routine single-energy 120 kVp by using vendor-recommended settings to create linearly-blended images (M_0.6), combining 60% of low tube voltage and 40 % of high tube voltage data. The second and the third image series were reconstructed as a standard VMI and noise-optimized VMI+ dataset, respectively, using a clinically available DE application (syngo.via Dual Energy workflow, version VA30, Siemens), at energy levels ranging from 40 keV to 100 keV with 5-keV increments. Higher keV levels were not investigated mainly due to the lower attenuation previously reported [4, 10]. For each dataset, iterative reconstruction (ADMIRE, Siemens Healthcare) was used with a strength level of 3.

Objective image analysis

All measurements were performed on a multi-modality workstation (syngo.via, version VA30, Siemens). Liver attenuation values in HU were recorded by one radiologist with 7 years of experience in abdominal imaging, by drawing three 1-cm2 circular ROIs in the liver parenchyma (segments II, VII and VIII) avoiding great vessels and lesions. Additionally, two 1-cm2 circular ROIs were drawn within the paired para-spinal musculature at the level of L2. The liver parenchyma and paravertebral muscle ROIs were separately averaged. Image noise was recorded as the standard deviation in the left-sided pararenal fat. The DE application enabled automatic recording of all attenuation numbers and image noise values within a single ROI for all keV levels, consequently reducing the variance in multiple measurements. Liver CNR was obtained using the following formula [14]:

$$ Liver\ CNR=\frac{\left({HU}_{liver}-{HU}_{muscle}\right)}{SD_{fat}} $$

The same radiologist (not blinded to the patient records) analysed all hypervascular lesions reported by drawing an as large as possible circular ROI without exceeding the lesion size. CNR of hypervascular lesions was obtained as previously described [15]:

$$ Lesion\ CNR=\frac{\left({HU}_{lesion}-{HU}_{liver}\right)}{SD_{fat}} $$

Subjective image analysis

Based on the objective image quality results, subjective image quality was assessed for M_0.6, VMI at 70 and 75 keV, and VMI+ at 40–60 keV every 5 keV, since we observed the highest CNR values without statistical differences among these image series. To avoid recall bias, image series were randomized and evaluated at least 2 weeks between reading sessions. Two other independent readers with 10 and 8 years of experience in abdominal imaging, blinded to MRI results, patient records and image reconstructions, performed image quality assessment and lesion detection analysis. Image evaluation was performed individually in separate reading sessions. Window settings were manually adjustable by the radiologist if needed. Image quality was assessed, according to reader perception, with a 5-point Likert scale [16] based on the following score system: 1= insufficient for diagnosis; 2= substandard; low diagnostic confidence; 3= acceptable; 4= above average, good diagnostic confidence; 5= excellent. Overall diagnostic image quality was represented as the sum of the single variables score (sharpness, image noise and lesion conspicuity; Supplemental Table 2). An inter-reader agreement analysis was also performed.

Diagnostic accuracy

After a 3-month interval, the two readers who had previously performed the subjective image assessment were asked to independently evaluate the presence of hypervascular liver lesions stratified by size (< 1 cm and ≥ 1 cm). For the evaluation of diagnostic accuracy, the keV reconstructions with the highest objective and subjective image quality for each dataset were selected. Readers were blinded to patient records and image reconstructions. Lesion-based sensitivity and specificity were calculated for both readers and averaged between them. Eventually, areas under the receiver operating characteristic (ROC) curves (AUCs) of the three datasets were calculated.

An inter-reader agreement analysis was performed to assess reproducibility of the technique.

Statistical analysis

Statistical analysis was performed using MedCalc Statistical Software version 16.8.4 (MedCalc Software bvba, Ostend, Belgium). Continuous variables are displayed as mean ± standard deviation; categorical variables are conveyed as frequencies and percentages. A Kolmogorov-Smirnov test was used to assess data distribution.

Per-lesion sensitivity and specificity of VMI+, VMI and M_0.6 datasets were calculated with 95 % confidence intervals and compared using McNemar’s test. The capability of the datasets for the identification of hypervascular liver lesions was evaluated on a per-lesion basis using the AUCs with MRI being the reference standard. Comparison among the AUCs was performed using DeLong’s method [17]. A repeated-measures analysis of variance test (ANOVA) was used to assess differences in objective image quality, tested between M_0.6 images and VMI and VMI+ reconstructions and evaluated at each keV level. A post-hoc Bonferroni test was performed when significant differences were observed. Subjective image quality was analysed by averaging the two readers’ 5-point scores and using the nonparametric Friedman test with post-hoc tests.Inter-reader agreement on subjective image quality scores and number of detected lesions was evaluated using the Cohen kappa (κ) statistics explained as follow: ≤ 0.20, slight agreement; 0.21–0.40, fair agreement; 0.41–0.60, moderate agreement; 0.61-0.80, good agreement; ≥ 0.81, excellent agreement. Two-tailed p-values < 0.05 were considered statistically significant.

Results

Patient population

A total of 30 patients (17 males and 13 females, mean age 58.8 years) were prospectively enrolled in this study. The average time between the MRI and DECT examination was 17 days (range 1–50 days). The average number of lesions per patient was 3.6 (range 1–8). No patient experienced early adverse reaction to the administration of MRI or CT contrast media. Figure 1 and Table 1 show further patient details.

Fig. 1
figure 1

Flow diagram of patient recruitment. MRI was used as reference standard. DECT dual-energy computed tomography, HCC hepatocellular carcinoma, HLL hypervascular liver lesions

Table 1 Patient characteristics

Radiation dose

The mean DLP was 207.1 ± 90.3 mGy · cm (range 74.3–371.5 mGy · cm), for an estimated mean ED of 3.0 ± 1.4 mSv.

Objective image analysis

The highest CNR for liver parenchyma was observed at 40 keV for VMI+ images (5.50 ± 2.65) (Fig. 2a), showing significant improvement compared to highest parenchymal CNR for VMI at 75 keV (p =0.021). No significant differences in liver parenchyma with VMI+ were observed between 40–60 keV (all p ≥0.11), whereas all CNR values were significantly higher compared to the standard M_0.6 (3.36 ± 2.82; p <0.001). Instead, VMI images showed significantly inferior liver CNR values compared to the standard M_0.6 (p ≤0.021), reported in Table 2, influenced by image noise. M_0.6 dataset had a significantly lower noise compared with VMI at 75 keV (9.75 vs. 12.38, p =0.001) and VMI+ at 40 keV (24.74, p= 0.001), whereas VMI+ at 40 keV resulted in an image noise significantly higher than VMI at 75 keV (p =0.007) (Fig. 2b).

Fig. 2
figure 2

(a) Liver parenchyma CNR, (b) image noise and (c) hypervascular liver lesion contrast-to-noise ratios (CNRs) at different keV levels. Plots compare standard (VMI) and noise-optimized (VMI+) monoenergetic images. The horizontal line represents the linearly-blended images (M_0.6). VMI+ images are characterized by an overall lower image noise than VMI, moreover VMI+ dataset at low keV provided substantially higher liver- and lesion-to-liver CNR than VMI at 70 keV

Table 2 CNR values of liver parenchyma and hypervascular liver lesions for noise-optimized (VMI+) and standard (VMI) monoenergetic algorithms as well as for linearly-blended images (M_0.6)

Hypervascular liver lesion CNRs reached a peak at 40 keV VMI+ (15.25 ± 19.95), significantly greater than the highest CNR value for VMI at 70 keV (8.98 ± 8.50, p <0.001) and M_0.6 (9.72 ± 8.24, p <0.001) (Fig. 2c). No statistical differences were observed between 70 keV VMI and M_0.6 CNR (p =0.18). As previously observed for liver parenchyma, the CNR of hypervascular liver lesions for VMI+ was significantly higher from 40 to 60 keV compared to the other keV intervals (all p ≤0.033), whereas for VMI no significant differences were found between 70 keV and 75 keV (p =0.87) (Fig. 3). Full results of liver lesion CNRs are listed in Table 2.

Fig. 3
figure 3

Dual-energy CT study of the liver in a 63-year-old man with HCC. A hypervascular liver lesion is shown on transverse sections at the most suitable energy levels for VMI+ (a, b, c, d and e; 40 keV, 45 keV, 50 keV, 55 keV and 60 keV, respectively) and VMI (f and g; 70 keV and 75 keV, respectively), as well as M_0.6 (h) datasets. Low keV VMI+ images are characterized by higher image contrast in comparison to VMI and M_0.6 datasets. The window width and level were held constant at 450 and 100, respectively

Subjective image analysis

Table 3 shows the comparison of the averaged scores for both readers among the different datasets. The determinant values of the subjective image quality are reported in Table 4. According to objective image analysis results, subjective image analysis was performed within levels of keV ranging from 40 to 60 for VMI+ and at 70–75 keV for VMI datasets.

Table 3 Full comparison of overall subjective image quality averaged between the two readers and expressed as mean ± standard deviation (SD)
Table 4 Subjective image quality assessment and inter-reader agreement of both readers

The highest overall image quality score averaged between the two readers was significantly greater for 50 keV VMI+ images (4.40 ± 0.54) compared to the highest value of VMI at 70 keV (3.76 ± 0.55; p <0.001) and M_0.6 (3.93 ± 0.44; p <0.001). No significant differences were reported between VMI at 70 keV and M_0.6 (p >0.99).

The inter-reader overall agreement was excellent for VMI+ images at 45, 50 and 55 keV and for VMI images at 70 and 75 keV (all κ>0.8), whereas good agreement was reached for M_0.6 images (κ=0.64). Complete results are reported in Table 4.

Assessment of diagnostic accuracy

Comprehensive details of diagnostic accuracy are available in Table 5. Detailed results for both readers are listed in Supplemental Table 3. A total of 61 hypervascular liver lesions were detected in 30 patients by the reference standard (n=36 for lesions with < 1 cm diameter and n= 25 for lesions with ≥ 1 cm) (Fig. 4). All CT examinations were included in the analysis.

Table 5 Comprehensive diagnostic accuracy for the identification of hypervascular liver lesion of VMI+, VMI, and M_0.6 datasets
Fig. 4
figure 4

Transverse sections from a dual-energy CT study of the liver in a 46-year-old man with hepatocellular carcinoma. Both readers correctly identified the ≥1 cm (arrow-head) and < 1cm (arrow) diameter lesions at 50 keV VMI+ (a), and 70 keV VMI (b) dataset, while the smaller lesion was missed by both readers at M_0.6 (c). As an outside reference standard, contrast-enhanced MRI (d) confirmed the two hypervascular lesions. An intrahepatic portal vein branch is highlighted in all the datasets (asterisk)

All lesions

The VMI+ dataset provided the highest sensitivity (79 %) in comparison to the VMI (71 %, p=0.004) and the M_0.6 algorithm (67 %, p <0.001) for the detection of all hypervascular liver lesions in comparison with MRI as the reference standard. Both the VMI and M_0.6 dataset reported comparable sensitivity (p =0.125). The specificity was similar across the three datasets (95 % for VMI+ and VMI, 97 % for M_0.6; all p =1). The VMI+ dataset obtained the largest AUC (0.871), significantly larger than VMI and M_0.6 (0.834 and 0.821, respectively, all p =0.002) while no difference was found between VMI and M_0.6 (p =0.325) (Fig. 5a).

Fig. 5
figure 5

Area under the receiver operating characteristic (ROC) curve (AUC) of the VMI+, VMI and M_0.6 datasets for all lesions (a), lesions ≥ 1 cm (b), and lesions < 1 cm (c). VMI+ always provided the largest AUC for lesion <1 cm (all p ≤ 0.003)

Overall inter-reader agreement between the two readers was excellent for all image datasets (κ =0.89, 0.86 and 0.83 for VMI+, VMI and M_0.6, respectively) (Supplemental Table 3).

Lesions ≥ 1 cm diameter

For lesions with a diameter ≥1cm, the VMI+ dataset provided no significant increment in sensitivity and specificity (92 % and 94 %) in comparison to VMI and M_0.6 datasets (90–94 % and 88–97 %, respectively, all p ≥0.5). The AUCs showed no significant differences among the three protocols (all p ≥0.317) (Fig. 5b).

Lesions < 1 cm diameter

For lesions with a diameter < 1 cm, the VMI+ dataset demonstrated the highest sensitivity (69 %) in comparison to VMI and M_0.6 datasets (58 % and 53 %, respectively, all p ≤0.008). The sensitivity of VMI and M_0.6 datasets was not statistically different (p =0.219). All datasets accounted for the same specificity (97 %, p =1). The VMI+ dataset showed a significantly larger AUC (0.833) compared to the VMI and M_0.6 datasets (0.777 and 0.749, respectively; p =0.003 and p <0.001, respectively). No significant differences were found between the AUCs of VMI and M_0.6 (p =0.098) (Fig. 5c).

Discussion

The aim of our study was to assess the image quality and diagnostic accuracy of DECT images reconstructed with VMI+, VMI, and linearly-blended M_0.6 algorithms for the detection of hypervascular liver lesions, utilizing MRI as the reference standard. Our findings demonstrate that the 50 keV VMI+ dataset provides both greater objective and subjective image quality and also an improved diagnostic accuracy for lesions with a diameter < 1 cm, compared to VMI and M_0.6 datasets. Thus, the results of this prospective study indicate that 50-keV VMI+ images may be beneficial in the detection of hypervascular liver lesions and should be preferred over the other DECT reconstruction techniques and settings.

On a lesion-based analysis, 50 keV VMI+ demonstrated a 64 % higher lesion CNR compared to the standard 70 keV VMI and a 52 % higher CNR in comparison with M_0.6 images. The VMI+ algorithm provided higher lesion CNR in comparison to VMI images at every other energy level assessed, showing comparable values in an intra-dataset analysis for energy levels ranging from 40–60 keV, while VMI datasets provided the highest CNR at 70 keV and 75 keV. It should be noted that there was no significant difference between VMI and M_0.6 series regarding liver lesion CNRs despite the higher liver lesion attenuation observed for VMI, because of the significant increment in image noise of the VMI dataset. Despite a CNR peak at 40 keV, both readers preferred the 50 keV VMI+ images in subjective image quality analysis, with significant differences when compared to other VMI+ energy levels, 70 keV VMI and M_0.6 images. Different investigators have demonstrated a progressive increase of liver parenchyma and hypervascular liver lesion CNRs from high to lower energy levels in monoenergetic images, reaching the peak at 40 keV [11, 18, 19]. In particular, Marin et al. demonstrated that the maximal tumour-to-liver CNR for hypervascular lesion occurred at 40 keV in phantom and in vivo [11]. However, the study suggests, similar to our results, that 55 keV may be a more conservative choice for liver imaging applications when noise is considered as an independent variable of image quality. These findings obtained with a second-generation scanner are in accordance with our observations; in addition, we further validated the value of low-keV data-sets for the detection of small hypervascular liver lesion using liver MRI as gold-standard. Similar to our study, Shuman et al. reported both best CNR and subjective scores at 50 keV [5], similar divergences among objective and subjective quality evaluation have already been described in previous papers in several anatomical districts [12, 20,21,22,23,24,25]. A possible explanation for these differences is that 50 keV demonstrated lower noise in comparison with 40 keV, resembling conventional images routinely used in clinical practice. Beyond the different study designs characterizing all the aforementioned studies, a common result is represented by the higher diagnostic performance of low keV virtual monoenergetic images compared to the traditional polyenergetic spectrum for the identification of solid parenchymal lesions, and the superior performances of VMI+ compared to VMI.CT accuracy in hypervascular lesions detection is significantly affected by lesion size. In a recent meta-analysis by Lee et al. [26], CT showed a summarized sensitivity for hypervascular liver lesion detection of 82 % for lesions ≥ 1 cm and only 31 % for lesions < 1 cm. In our study, 50 keV VMI+ provided the highest sensitivity for lesions < 1 cm (69 %) among the three datasets, outperforming both VMI and M_0.6 datasets (58 % and 53 %, respectively) (Table 5). The improved diagnostic performance of VMI+ is presumably related to the higher VMI+ CNR, which allows for improved lesion conspicuity. Likewise, the lack of significant differences in the detection rates between VMI and M_0.6 datasets may be a consequence of the non-significant differences in liver lesion CNR between the two series. These results concur with previous published studies that analysed the standard VMI algorithm. Yu et al. [19] reported that 70 keV traditional monoenergetic images showed higher detectability of HCCs in comparison to conventional polyenergetic images. Similar results were reported by Lv et al. [27], who found that monoenergetic energy levels from 40 to 70 keV increased the detectability of small HCC lesions compared to conventional 140 kVp images, with the highest CNR found at 70 keV.Certain limitations of our study should be considered. First, the patient population of our prospective study was relatively small but with a large prevalence of hypervascular liver lesions; thus, larger studies are necessary to confirm our results. Second, no histological information was used to classify the lesion detected. This study, however, was exclusively designed to evaluate the effect of monoenergetic images on the detection of hypervascular liver lesions, irrespective of histopathology. Third, our results are currently only valid for DECT and post-processing techniques using dual-source CT technology, and cannot be automatically translated to DECT solutions by other vendors. Finally, we acknowledge that other keV data-sets demonstrating no significant difference in CNRs of hypervascular lesions could have provided similar diagnostic accuracy to the selected ones, since the difference in subjective image quality was small.

In conclusion, our results indicate that a 50-keV VMI+ dataset provides superior image quality and diagnostic accuracy for the detection of hypervascular liver lesions with a diameter < 1 cm, compared to VMI or M_0.6 reconstructions.