Introduction

Diffusion-weighted imaging (DWI) has become an essential part of the imaging of many oncological diseases [1]. It is able to sensitively measure differences in diffusivity of water molecules and is therefore a valuable tool in detecting and characterizing various malignant lesions [2]. Cancer therapy can cause cell swelling or apoptosis which influences water diffusivity [3]. Thus, it has been shown that DWI may be helpful in the evaluation of therapy response [4]. DWI is technically demanding, and was first established in neuroimaging [5]. Further developments were necessary to transfer DWI to regions outside the brain so that it could be introduced into the daily routine of diagnostic imaging of individual organs [6, 7]. The introduction of the background body signal suppression technique [8] allowed DWI to be applied to whole-body examinations with acceptable acquisition times and image quality. However, DWI is still one of the most time-consuming sequences in whole-body protocols, hampering its broad clinical application.

The simultaneous excitation and acquisition of multiple slices (SMS) is one possible approach to reducing examination times. This technique was introduced in the late 1980s [9, 10], and is based on different sensitivities of multicoil arrays for separation of multiple slices. However, low sensitivity in separating closely spaced slices for a long time hindered its widespread application. More recently, advances such as the adaptation of parallel imaging technology [11] and the introduction of the “controlled aliasing in parallel imaging results in higher acceleration” (CAIPIRINHA) technique [12] and its modified version for the echoplanar imaging readout [13], has enabled today’s application of DWI-SMS in whole-body imaging [14, 15]. Being a functional imaging technique, DWI is often combined with or compared to positron emission tomography (PET) for oncological imaging [16,17,18,19]. For several years it has been possible to acquire PET and MRI simultaneously [20]. This has enabled the acquisition of multiple tissue parameters derived from PET and MRI at the same time with very good alignment [21] and has also opened up new technical possibilities such as using the MR information for motion correction of PET images [22,23,24]. However, with still rather long whole-body PET/MRI scan times, patient comfort and therefore also patient compliance are markedly reduced [25, 26].

Thus, the aim of this study was to compare image quality, lesion conspicuity and apparent diffusion coefficient (ADC) values in tumorous lesions and organs obtained using DWI-SMS with those obtained using standard DWI (DWI-STD) for multiparametric whole-body PET/MRI examination in oncological patients.

Materials and methods

This prospective study was approved by the local ethics committee. Written informed consent was obtained from all volunteers and patients regarding the examination and the scientific evaluation of their data.

MRI protocol

All examinations were performed on a fully integrated 3-T PET/MRI scanner (Biograph mMR; Siemens Healthineers, Erlangen, Germany). The specifications and performance of the MR component of the Biograph mMR system used are comparable to those of the Magnetom Verio system (Siemens Healthineers) [20]. The scan parameters for DWI-SMS (Multi-Band EPI package, release 13; Center for Magnetic Resonance Research, University of Minnesota, MN, USA) were adapted to the DWI-STD protocol used in our department for whole-body PET/MRI examinations (Table 1). In both techniques, an inversion recovery technique (spectral attenuated inversion recovery, SPAIR) was used for fat suppression. Two b-values were acquired: b = 50 s/mm2 and b = 800 s/mm2. All examinations were performed with the subject in free breathing. For anatomical correlation, a T2-weighted half Fourier acquisition single shot turbo spin echo (T2-HASTE) sequence with continuous table movement was acquired in all examinations with the subject in free breathing with the following parameters: echo time (TE) 76 ms, repetition time (TR) 1,200 ms, number of averages 1, flip angle 120°, and resolution 0.7 × 0.7 × 0.6 mm. Depending on the clinical indication, further sequences were performed in patients, but were not part of this study.

Table 1 Acquisition parameters of DWI-SMS and DWI-STD in patients/volunteers and in the phantom study (values in parentheses)

Phantom study

To assess the reliability of the quantification of diffusion restriction (ADC values) without subject-related artefacts, we performed a phantom study. Therefore, a dedicated sucrose DWI phantom was constructed to model different diffusivities (Fig. 1). Five tubes with a diameter of 25 mm were filled with sucrose solutions of different concentrations (0%, 10%, 20%, 30%, 40%) in distilled water and were placed in a cylindrical acrylic glass container filled with distilled water. DWI-STD and DWI-SMS sequences were performed using a multichannel head coil with the protocol parameters shown in Table 1. Each DWI sequence was applied twice successively with the same protocol parameters. All image datasets were transferred to a stand-alone workstation for offline postprocessing using MATLAB (MathWorks Inc, Natick, MA). Diffusion-weighted trace images ITrace for two measured b-values (b = 50 and 800 s/mm2) were calculated according to the following equation:

$$ {I}_{\mathrm{Trace}}={I}_0{\mathrm{e}}^{-\mathrm{ADC}\cdot b}={I}_0{\mathrm{e}}^{-\frac{D_1+{D}_2+{D}_3}{3}\cdot b}={\left({I}_1\cdot {I}_2\cdot {I}_3\right)}^{\frac{1}{3}} $$
(1)

where I1, I2 and I3 are the measured diffusion-weighted images with three orthogonal gradient directions and D1, D2 and D3 are the corresponding diffusion coefficients. Their arithmetic mean results in the isotropic diffusion coefficient, namely ADC. The ADC maps were calculated from Eq. 1 using a log-linear fitting algorithm.

Fig. 1
figure 1

Top left: DWI phantom and the corresponding ADC map (DWI-STD). Five tubes were filled with sucrose solutions of different concentration of sacrose solutions (0%, 10%, 20%, 30%, 40% in distilled water) and were placed in a cylindrical acrylic glass container filled with distilled water. Top right: ADC values acquired with different sucrose concentrations from ADC-STD and ADC-SMS maps (ADC values in 10−3 mm2/s). Bottom: Bland-Altman plot of ADC values from the phantom measurements from the ADC-STD and ADC-SMS maps. Differences between pairs of ADC values (expressed as percentages, y-axis) are plotted against the means of the pairs of ADC values (x-axis). The solid line indicates the mean absolute difference, the dashed black lines indicate ±1.96 × standard deviation, the dashed grey lines indicate 95% CI of the mean difference (limits of agreement), and the dotted grey line is the line of equality. Arithmetic mean 0.51% CI 38.9 to 40.0 p=0.973

To minimize the impact of low-frequency spatially varying signal intensity, five rectangular regions of interest (ROIs), each of 21 pixels, were placed in each of the tubes to assess the ADC values. The ROIs were placed in the central region along the tube axis congruently in all reconstructed images. This was performed with the software Imagine (Imagine 2.0, Matlab Central File Exchange; MathWorks Inc., Natick, MA). The resulting ADC values in different sucrose solutions were calculated as the mean value over all five ROIs.

Volunteers

Between March 2016 and June 2016, ten healthy volunteers (aged 29 ± 9.3 years, five women) underwent PET/MRI examination in the supine position, arms positioned alongside the abdomen. All volunteers were examined from the cranial vertex to the thighs using head, body and spine multichannel coils. The MRI protocol was as described above.

The following evaluations were performed in the volunteers using the software syngo.via (syngo.via VA30; Siemens Healthineers, Erlangen, Germany).

ADC values and signal intensities in organs

For anatomical correlation, the T2-HASTE images in all volunteers were coregistered to the ADC maps and b50 images from DWI-SMS (ADC-SMS) and DWI-STD (ADC-STD). The ROIs were set congruently in all images in the white matter, the right liver lobe, the spleen, the kidney and the psoas muscle, avoiding image artefacts and organ borders.

Patients

Between May 2016 and September 2016, 20 patients (aged 62.3 ± 10.3 years, seven women) with the following oncological diseases underwent a whole-body PET/MRI examination with both DWI-SMS and DWI-STD: melanoma (eight patients), lymphoma (four), neuroendocrine tumour (grade G2, three), prostate cancer (Gleason score 7b, one), lung cancer (one), breast cancer (one), renal cell carcinoma (one) and cancer of unknown primary (one). Inclusion criteria were a history of oncological disease with known or suspected metastases. Exclusion criteria were the typical contraindications for MRI, i.e. metal implants including pacemakers, and claustrophobia.

All patients undergoing PET/MRI were scanned from the cranial vertex to the thighs. Patients were positioned in the supine position, arms positioned alongside the abdomen and the MRI scan was performed as described above. 18F-FDG (average dose 315 MBq) was used as the PET tracer in patients with melanoma, lymphoma, lung cancer, breast cancer, renal cell carcinoma and cancer of unknown primary. The somatostatin receptor-specific radiotracer 68Ga-DOMITATE (average dose 196 MBq) was used in patients with neuroendocrine tumours, and 68Ga-PSMA (dose 172 MBq) was used in the patient with prostate cancer. PET images were reconstructed using the vendor’s software and the following parameters: 3D ordered subsets expectation maximization algorithm, 21 subsets, two iterations, matrix size 256 × 256, voxel size 2.8 × 2.8 × 2mm3, 4-mm gaussian filter. For the vendor-provided attenuation correction, a segmentation-based attenuation map was generated with a T1-weighted spoiled gradient echo sequence in end-expiratory breath-hold with Dixon-based fat–water separation with the following parameters: TR 3.6 ms, TE1 1.23 ms, TE2 2.46 ms, flip angle 10°, pixel size 2.6 × 2.6 × 2.6 mm3. The attenuation map of each patient was carefully checked for artefacts.

Subjective image quality and artefacts

Subjective image quality was evaluated by two independent radiologists each with 5 years experience in DWI-MRI (F.S. and J.T.) by reading the b800 images (b800-SMS) and subsequently the ADC maps from the DWI-SMS in all patients, and 6 weeks later reading the b800 images (b800-STD) and subsequently the ADC maps from the DWI-STD. The readings were performed in a blinded fashion. The readers were therefore not able to identify the respective imaging technique from any kind of image description. Subjective image quality was assessed in five different body regions which also corresponded to typical examination regions on PET: head, neck, thorax, abdomen and pelvis. This was performed using a five-point Likert scale relating to the delineation of organs and the overall diagnostic validity (5 excellent, 4 good, 3 satisfactory, 2 poor, and 1 nondiagnostic).

Image artefacts were evaluated in a consensus reading (F.S. and J.T.) in each body region separately and categorized into four different groups: (1) distortion artefacts, (2) ghosting artefacts, (3) inhomogeneous fat saturation, and (4) other artefacts (caused by, for example, metal implants or interslice intensity variation). Artefacts were only assessed if they affected the investigated body region. The degree of observed artefacts was rated using a five-point Likert scale relating to the impact on the diagnostic validity of the image (5 low impact, 1 high impact).

Evaluation of tumorous lesions

All examinations were evaluated in a clinical setting in a consensus reading by a radiologist and a nuclear medicine physician, both with more than 10 years experience in hybrid imaging. Based on this report, two radiologists with 4 and 12 years experience in hybrid imaging (F.S. and C.S) evaluated the PET/MRI images from each patient in consensus, blinded to the respective DWI technique, and with the images opened side-by-side: the PET images, the T2-HASTE images, the b800-SMS and the b800-STD images, and the ADC-SMS and ADC-STD maps.

In the first reading, the conspicuity of PET-positive tumorous lesions was evaluated on the b800-SMS and b800-STD images using a five-point Likert scale (5 very good, 1 not visible). Care was taken in the characterization of PET-positive tracer accumulations not to mistake possible inflammatory lesions for metastatic lesions. A maximum of ten lesions per organ were evaluated. For lymph nodes, a maximum of ten lymph nodes per body region (neck, thorax, abdomen, pelvis) were included. For peritoneal and pleural carcinomatosis, only circumscribed nodules were included. The longest diameter of a lesion was assessed in the axial plane in the T2-HASTE images. Cerebral metastases were not included in the evaluation.

In the second reading, the precision of ADC values of PET-positive tumorous lesions was evaluated. In order to only acquire possible deviations in ADC values derived from the respective acquisition technique, other influencing factors, such as partial volume effects and image artefacts, were eliminated. Therefore, only lesions with an axial diameter of >1 cm in the first reading were included in the evaluation of ADC values. The ADC values were assessed in a blinded fashion by drawing a free-hand intralesional ROI in the region of the lesion with the largest axial diameter on one of the ADC maps. The ROI was copied to the other ADC map; in individual cases, the ROI had to be relocated slightly but was not changed in size or shape. Care was taken to avoid blurred lesion borders on both ADC maps. Lesions were excluded from the evaluation if they were affected by artefacts on one of the ADC maps.

These evaluations were performed using the software syngo.via (syngo.via VA30; Siemens Healthineers).

Statistical analysis

Statistical analyses were performed using SPSS version 23.0 (IBM Corporation, Armonk, NY). The normality of distributions was tested using the Kolmogorov-Smirnov test. Likert scores and artefacts were compared using the Friedman test. For quantitative comparison of both sequences, ADC values were compared using Bland-Altman plots in percentages with limits of agreement ±1.96 standard deviations (MedCalc for Windows, version 17.8.6; MedCalc Software, Ostend, Belgium). Equivalence was defined by the inclusion of the line of equality in the 95% confidence interval (CI) of the mean difference [27]. Analysis of variance with Bonferroni post hoc correction was performed if the comparisons showed significant differences. A p value <0.05 was considered statistically significant.

For interobserver agreement, kappa values were calculated with the following levels of agreement: 0.00 no agreement, 0.01–0.20 slight, 0.21–0.40 weak, 0.41–0.60 moderate, 0.61–0.80 strong, and 0.81–1.00 almost perfect [28]. Intraobserver agreement was calculated using the intraclass coefficient according to the method described by Fleiss [29] with the following levels of agreement: 0.00–0.40 poor, 0.041–0.75 fair to good, and 0.76–1.00 excellent.

Results

The DWI-STD and DWI-SMS scan times were 140 s and 84 s per table position, respectively, corresponding to a reduction in total scan time by 40%. The interobserver agreement in the evaluation of image quality was high (kappa 0.62 ± 0.03), and the intraobserver agreement in the evaluation of image quality was excellent (0.85, CI 0.823–0.875).

Phantom study

The results of the phantom study are shown in Fig. 1. Overall, the ADC values from the two techniques did not differ significantly (arithmetic mean 0.506%, CI −38.9 to 40.0%, p = 0.973 in the Bland-Altman plot). The highest difference in ADC values between the two techniques was found with the 20% sucrose solution (1.4%).

Volunteers

All examinations in volunteers were performed with five bed positions. The DWI scan times for the whole-body examinations were therefore 420 s for DWI-SMS and 700 s for DWI-STD. The ADC values in various organs obtained from the two techniques are given in Table 2. The ADC values in the liver and in the psoas muscle were significantly lower with DWI-SMS. Besides these regions, ADC values obtained with the two methods were considered equivalent. Bland-Altman plots of the body regions examined are shown in Fig. 2. Overall, a relative underestimation of ADC values with DWI-SMS was seen in tissues with low signal intensities in b50 images (Fig. 3).

Table 2 Mean ADC values and standard deviations (×10−3 mm2/s) acquired in different body regions
Fig. 2
figure 2

Bland-Altman plots of ADC values measured from the ADC-STD and ADC-SMS maps (ADC values given in x 10-3mm2/s) in various organs. Differences between pairs of ADC values (expressed as percentages, y-axis) are plotted against the means of the pairs of ADC values (x-axis). The solid line indicates the mean absolute difference, the dashed black lines indicate ±1.96 × standard deviation, the dashed grey lines indicate 95% CI of the mean difference (limits of agreement), and the dotted grey line is the line of equality. For the liver and the psoas muscle, the line of equality does not lie within the interval of the 95% CI of the mean difference. For the kidney, the spleen and lesions, the line of equality lies within the interval of the 95% CI of the mean difference

Fig. 3
figure 3

Deviation of ADC values expressed as percentages [100 × (ADC-STD − ADC-SMS)/ADC-STD] and the respective signal intensity on the DWI-STD b50 images, together with the mean value (solid line) and standard deviation (dashed lines). Lower signal intensities in the b50 image (i.e. in tissues with shorter T2 relaxation times) are associated with higher ADC value deviations in the SMS-accelerated sequence

Patients

The number of table positions varied between four and six, depending on the patient’s height. The mean scan times were therefore 403 ± 44 s (range 336–504 s) for DWI-SMS and 672 ± 73 s (range 560–840 s) for DWI-STD.

Subjective image quality and artefacts

Subjective image quality ratings in the patient studies are shown in Table 3. No significant differences in subjective image quality ratings were found between the b800-SMS and b800-STD images. The subjective image quality ratings for the thoracic and abdominal regions were significantly lower on the ADC-SMS maps than on the ADC-STD maps.

Table 3 Mean values and standard deviations calculated from image quality rankings as rated by two independent radiologists for the b800 images and the ADC maps

More artefacts were found in the b800-SMS images than in the b800-STD images (Table 4). The diagnostic impact of artefacts in terms of the average of the artefact ratings did not differ significantly between the two techniques (3.8 vs. 3.7, p = 0.28). In the corresponding ADC maps, artefacts were almost evenly distributed: 100 artefacts in the ADC-SMS maps and 104 artefacts in the ADC-STD maps (mean diagnostic impact values 3.8 vs. 3.8, p = 0.28).

Table 4 Overview of numbers and types of observed artefacts in the body regions examined and estimated impact on diagnostic validity in b800 images and ADC maps

Evaluation of tumorous lesions

PET-positive tumorous lesions were found in 19 of the 20 patients. In total, 116 lesions (diameter 2.0 ± 1.4 cm) were included in the first reading: 34 metastatic lymph nodes, 25 liver lesions, 16 bone lesions, 15 soft-tissue lesions, 13 lung lesions, 7 peritoneal lesions, 5 pleural lesions and 1 pancreatic lesion. The lesions were assigned to the body regions corresponding to those used in the evaluation of image quality (Table 5).

Table 5 Mean values and standard deviations of lesion conspicuity as rated by two radiologists in consensus for the b800 images and mean ADC values with standard deviations for lesions with an axial diameter of >1 cm

With both DWI techniques, lesion conspicuity was rated highest in the pelvis and lowest in the neck and thoracic region, as shown in the examples given in Fig. 4. Overall, lesion conspicuity did not differ significantly between the b800-SMS images and the b800-STD images. In the b800-STD images, 20 PET-positive lesions could not be delineated due to artefacts (Likert score 1): 5 in the neck region, 11 in the thoracic region, 2 in the abdomen (liver), and 3 in the upper thigh. Of these lesions, 6 were depicted in b800-SMS images. The average diameter of the missed lesions was 1.0 cm (range 0.5–2.0 cm). In the b800-SMS images, 16 PET-positive lesions could not be delineated due to artefacts: 3 in the neck region, 10 in the thoracic region, 2 in the abdomen (liver), and 1 in the pelvis. Of these lesions, 2 were depicted in the b800-STD images. The average diameter of these lesions was 1.3 cm (range 0.5–4.5 cm). Examples are given in Fig. 5.

Fig. 4
figure 4

Examples of lymph-node metastases in the upper thoracic region (a) and pelvis (b) positive on 18F-FDG PET/MRI. Due to artefacts, the lesion conspicuity in the upper thoracic/lower neck region (white arrows) is impaired while the lesion conspicuity in the pelvis (white arrows) is good on both b800-STD and b800-SMS images

Fig. 5
figure 5

Examples of impaired lesion conspicuity due to artefacts on both DWI-STD (a) and DWI-SMS (b). a A 64-year-old man with metastases from melanoma of the lung positive on 18F-FDG PET/MRI (white arrows) which is not visible on the DWI-STD image. b A 72-year-old woman with metastases from melanoma of the liver positive on 18F-FDG PET/MRI (white arrows) which can hardly be delineated on the DWI-SMS image

For the evaluation of ADC values in lesions, 26 lesions (those with a diameter ≥1 cm in a region not affected by artefacts on DWI-SMS or DWI-STD) were evaluated, 9 metastatic lymph nodes, 5 bone lesions, 4 peritoneal lesions, 3 lung lesions, 2 liver lesions, 2 soft-tissue lesions and 1 pancreatic lesion. The average diameter of the lesions included in this evaluation was 3.3 cm (±1.7 cm). Overall, no significant difference was found in the intralesional ADC values between the two DWI techniques (Fig. 6, Table 5).

Fig. 6
figure 6

A 48-year-old woman with breast cancer and lymph-node metastases in the mediastinum positive on 18F-FDG PET/MRI (white arrows). Note the good correlation between the mean ADC values (ADC-STD 1.505 × 10−3 mm2/s, ADC-SMS 1.521 × 103 mm2//s) in this lesion.

Discussion

In this study, the DWI-SMS technique provided image quality in most body regions comparable to that provided by the DWI-STD technique, and reliable ADC values of tumorous lesions for multiparametric whole-body PET/MRI examinations in oncological patients. The acquisition time with DWI-SMS as performed in this study (TR 3,000 ms) was 40% less than with DWI-STD (TR 6,000 ms).

A recent study by Kenkel et al. [30] showed that whole-body DWI-SMS at 3 T was technically feasible and showed promising results in eight volunteers and one patient. Nevertheless, there is still the need to investigate the diagnostic performance of whole-body DWI-SMS in oncological imaging in a larger patient cohort. Thus, we adapted a prototype DWI-SMS sequence for use with whole-body DWI, matching it to our whole-body DWI-STD protocol to perform a systematic comparison. The prototype SMS sequence used in this study has been proven to be technically feasible and has already shown promising results in liver imaging at 1.5 T [31, 32]. Although we applied a higher magnetic field strength, the DWI parameters (i.e. ADC) were expected to be unaffected by these changes [33]. Indeed, the ADC values measured were within the range of previously published values [30, 32, 34,35,36] and DWI-SMS provided reliable ADC values with no significant differences compared with DWI-STD in the phantom study, in most tissue types in the volunteer study and in the investigated tumorous lesions. However, ADC values measured with the SMS-accelerated sequence were found to be lower in tissues with a short T2 relaxation time (i.e. the liver) and/or a long T1 relaxation time (i.e. the psoas muscle) compared with those obtained with DWI-STD. A possible explanation for this finding is the shorter TR with DWI-SMS (3,000 ms vs. 6,000 ms with DWI-STD) resulting in signal loss, an effect that has also been described by Kenkel et al. [30]. Thus, lower signal intensities in a tissue under study (e.g, hepatic tissue) due to a reduced TR (which is the case with SMS-DWI) automatically translates into lower ADC values. The relationship between ADC values and their signal intensities in the b50 images is given in Fig. 2.

In order to minimize DWI-SMS acquisition times, a TR as low as possible is chosen, resulting in only partial signal recovery causing a reduced signal in the acquired images [32]. Thus, generally lower signal intensity in b50 images (i.e. in tissue with a low T2) ultimately translates into lower ADC values and a larger deviation from ADC values measured with the standard sequence. However, as most malignant lesions possess T2 values higher than those of liver and muscle tissue, a significant deviation in lesion ADC values was not observed. A possible further explanation might be the divergent noise statistics or alteration in signal processing in the different sequences [31, 32]. The slightly reduced signal in DWI-SMS may account for the lower subjective image quality. This is of particular relevance in organs with low signal on DWI (e.g. the lung or the liver), and explains the lower subjective image quality found in the thoracic and abdominal regions with DWI-SMS. However, it needs to be stressed that DWI-STD also did not deliver optimal results in these body regions.

The b800-SMS sequence was more susceptible to artefacts than the b800-STD sequence. The source and strength of artefacts depends on various factors such as field homogeneity, gradient directions and the hardware installation [37], and their interpretation is complex. Yet, in spite of the discussed drawbacks of DWI-SMS, lesion conspicuity was preserved. Contributing factors are certainly the fact that malignant lesions generally show higher signal intensity on T2-weighted images so that a pronounced signal loss is not expected. With both DWI-STD and DWI-SMS, the presence of artefacts reduced the conspicuity of individual lesions (Fig. 5).

There were some limitations to our study. First, the small number of patients allowed only a preliminary investigation of the use of DWI-SMS in multiparametric whole-body PET/MRI. Furthermore, only PET-positive lesions were evaluated in this analysis and set as the reference standard. PET-negative, yet possibly malignant, lesions were not included in the evaluation. The segmentation-based method used for attenuation correction can result in an underestimation of tracer uptake in regions with a high amount of bone and might therefore have affected the detectability of bone lesions [38]. A dedicated evaluation of brain metastases was beyond the scope of this work, but previous studies have demonstrated the usefulness of DWI-SMS techniques also in brain imaging [11, 39].

The sequence parameters for DWI-SMS were adapted to our DWI-STD. Thus, when applying DWI-SMS only, further optimization of parameters including b-values, number of averages, voxel size and slice thickness, is possible. Radiologists were blinded in the reading of the diffusion-weighted images but with time and with increasing experience, the differences between DWI-STD and DWI-SMS images became obvious, especially as the DWI-SMS images were presented first. However, the authors tried to minimize this effect by choosing a subsequent evaluation scheme. Moreover, the conspicuity of PET-positive metastases was evaluated in a consensus reading which hampered the evaluation of interrater agreement. Finally, we did not obtain test–retest ADC values (as performed, for example in a previous study [40]) which might have given an indication of the repeatability of SMS-DWI and STD-DWI.

In conclusion, DWI-SMS seems to provide reliable lesion conspicuity and consistent ADC values of PET-positive tumorous lesions. Therefore, although suffering from slightly reduced image quality in comparison with DWI-STD, especially in the thoracoabdominal region, DWI-SMS has the potential to distinctly reduce scan times in oncological patients undergoing multiparametric whole-body PET/MRI examinations. Studies with larger numbers of patients should be encouraged to confirm our results.