Abstract
Purpose
The purpose of this study is to evaluate a new device providing real-time monitoring on radiation exposure during fluoroscopy procedures intending to reduce radiation in an interventional radiology setting.
Materials and Methods
In one interventional suite, a new system providing a real-time radiation dose display and five individual wireless dosimeters were installed. The five dosimeters were worn by the attending, fellow, nurse, technician, and anesthesiologist for every procedure taking place in that suite. During the first 6-week interval the dose display was off (closed phase) and activated thereafter, for a 6-week learning phase (learning phase) and a 10-week open phase (open phase). During these phases, the staff dose and the individual dose for each procedure were recorded from the wireless dosimeter and correlated with the fluoroscopy time. Further subanalysis for dose exposure included diagnostic versus interventional as well as short (<10 min) versus long (>10 min) procedures.
Results
A total of 252 procedures were performed (n = 88 closed phase, n = 50 learning phase, n = 114 open phase). The overall mean staff dose per fluoroscopic minute was 42.79 versus 19.81 µSv/min (p < 0.05) comparing the closed and open phase. Thereby, anesthesiologists were the only individuals attaining a significant dose reduction during open phase 16.9 versus 8.86 µSv/min (p < 0.05). Furthermore, a significant reduction of total staff dose was observed for short 51 % and interventional procedures 45 % (p < 0.05, for both).
Conclusion
A real-time qualitative display of radiation exposure may reduce team radiation dose. The process may take a few weeks during the learning phase but appears sustained, thereafter.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
The exposure to radiation is a growing concern due to its cumulative effect on the interventional working environment and is most extensive in procedures using fluoroscopy [1–3]. Occupational dose reduction relies on two basic processes: X-ray output reduction and proper radiation protection [4]. X-ray reduction has matured over the last decade and includes optimization of fluoroscopy technique using lower dose protocols, improved sensitivity of hardware, and advanced image processing [5]. However, proper radiation hygiene and protection may be warranted further scrutiny to reduce radiation dose to the operators since the team dose is predominantly a product of scatter radiation [6]. There are various measures for radiation protection including lead aprons, shields, optimizing X-ray projection angle, the distance of subject to detector, the proper use of magnifications, collimators, and wedges [7]. In addition, there are behavioral components to radiation protection including the amount of fluoroscopy, the number of digital subtraction acquisitions, and the location of team members during the procedure. For control purposes, most institutions apply legislated monthly feedback methods [8]. The thermoluminescent dosimeter badges record the radiation dose for the particular month as well as cumulative annual radiation exposure [9]. Based on these radiation recordings, a written report with corrective actions is typically only instituted if a monthly or annual dose limit is exceeded.
Lately however, a system providing real-time monitoring on radiation exposure during interventional procedures has been introduced. The purpose of the present study was to evaluate whether the use of a real-time feedback system on radiation exposure may reduce radiation dose in an interventional radiology setting.
Materials and Methods
Study Design
This study was approved by the IRB and all the data were recorded in a HIPAA compliant manner. The duration of the present study was 22 weeks consisting of three phases:
-
Closed phase display for qualitative radiation dose not activated but evaluated (6 weeks).
-
Learning phase display for qualitative radiation dose activated but not evaluated (6 weeks)
-
Open phase display for qualitative dose activated and evaluated (10 weeks)
The dose readings for all phases were stored safely but not analyzed prior to completion of the study period. Accordingly, there was no demonstration of any preliminary data on radiation dose from DoseAware to the staff members. In line, there were no teaching or behavioral inputs provided.
DoseAware Equipment
In a single interventional suite at a tertiary care vascular center, we installed a real-time radiation dose display screen and five individual wireless personal dosimeters (PDMs) (DoseAware, Raysafe, Hopkinton, MA). The dose screen adjacent to the fluoroscopic display showed which PDMs are in the room and provided real-time monitoring for every PDM on the current radiation dose exposed. For that reason, the individual dose rate per PDM was visually displayed as a color bar that increases in size and changes color as the radiation thresholds increase from 0.2 (green), 2 (orange), and 20 (red) mSv/h. The minimal reading was set to 0.04 Sv/h. Of note, the monitor only displayed snapshot information of radiation exposure for every PDM during fluoroscopy or digital subtraction angiography imaging but no information on the accumulated procedural dose was displayed, they were only recorded. The absolute radiation doses were retrospectively downloaded to a computer for evaluation purposes. The PDMs were worn by the attending, fellow, nurse, technician, and anesthesiologist for every procedure performed in that one suite at collar height and facing the front. For that purpose, every PDM was strictly assigned to one specific subgroup (attending, fellow, nurse, technician, anesthesiologist). If one subgroup did not attend a procedure, the zero reading for that procedure was not taken into account for radiation evaluation.
The PDMs were meant to measure the effective dose on the operators. For that purpose and to include the total body scatter dose, the personal dose equivalent Hp(d) was regarded the dose equivalent in tissue at a depth of d = 10 mm; Hp(10) according to the International Commission on Radiation Units and Measurements [10].
Measurements and Documentation
The accumulated radiation dose for every single procedure included in that trial was recorded and locked in the PDM. On a weekly basis, dose readings were exported from the PDMs to the DoseManager software and stored in a password secured manner by a blinded physician for further analysis. Of note, all dose readings were de-identified and thus it was not possible to assign one or the other reading to the specific staff member i.e., for teaching, demonstration, or behavioral adaption purposes. In addition, the fluoroscopy time was recorded for all procedures. Based on that, the radiation dose of every single procedure was correlated with fluoroscopy time [dose per fluoroscopic minute (PFM)] and procedural radiation dose [dose area product (DAP)]. In addition, every procedure performed during the study period was documented on a case report form. Table 1 illustrates case variables to be obtained for every procedure. Thereby and among others, the type of procedure and its duration were documented. Based on that, diagnostic and interventional as well as long (>10 min) and short (<10 min) procedures were differentiated.
Statistics
Continuous variables are reported as means and differences were assessed using a student t test. A p value <0.5 was considered to indicate statistical significance. Statistical analysis was performed using SPSS.
Results
A total of 252 procedures were performed during the study period (closed phase 88 procedures, learning phase 50 procedures, open phase 114 procedures). There were 5 procedures that were regarded erroneous and thus omitted as they had extraordinary high readings (at least 10 times higher than average) that could not be explained after evaluation of the procedure type, fluoroscopy time, and DAP which were all below average.
The total staff dose PFM during closed phase was higher compared with the total staff dose PFM during open phase (42.79 vs 19.81 µSv/min; p < 0.05) as illustrated in Fig. 1. With respect to individual dose PFM, only a statistically significant reduction from 16.9 to 8.9 µSv/min was attained for the anesthesiologist (p < 0.05) when comparing the open and the closed phase (Fig. 2). Of interest, none of the individual procedure types yielded a statistical difference comparing the closed and the open phase (Fig. 3). However, a significant reduction in staff dose PFM was achieved during open phase for interventional procedures (Fig. 4). The average staff dose PFM for interventional procedures was 31.8 and 17.5 µSv/min (p < 0.05) during closed and open phase and 52.1 and 16.4 µSv/min (p > 0.05) for diagnostic procedures, respectively. Furthermore, a significant reduction in total staff dose (PFM) was attained for short procedures (closed phase 38.6 µSv/min, open phase 19.3 µSv/min, p < 0.05), whereas this did not account for long procedures (closed phase 31.5 µSv/min, open phase 20.2 µSv/min, p > 0.09) as shown in Fig. 5. For all DAP measurements, no statistical significance was obtained using the real-time monitoring system.
Discussion
The present study demonstrated that a real-time monitoring system on radiation allows for an overall reduction of radiation dose PFM. Further subgroup analysis, however, revealed only anesthesiologists to benefit significantly, thereof. In addition, a lower radiation dose was obtained for the subset of interventional as well as short procedures.
Exposure to radiation within occupational settings is a rising concern. There are various strategies and attempts to reduce radiation while applying intensified training and education. The findings of the present study, however, illustrate the extent of unawareness to radiation exposure. The unawareness of exposure to radiation within the setting of interventionists was demonstrated earlier and particularly for shorter procedures [11]. Accordingly, the lack of awareness may explain the benefit to radiation exposure for the subgroup of anesthesiologists within the present study. In addition, anesthesiologists are consistently close to the patients for the purpose of monitoring, thus allowing for easier detection of significance. Of interest however, that the subgroup of attending physicians did not show a significant reduction in radiation dose and that the fellows showed even increased radiation exposure during the open phase. Commonly, the attending physician is the closest to the patient and the radiation beam, and was shown earlier to profit from real-time radiation monitoring within a pediatric setting [12]. We assume, however, that our observations for the attending physicians and fellows may be affected by the start time of the trial and that of the fellowship. The fellows at our institution start in July whereas this trial was launched in August. Therefore, most fellows may have been mainly observing procedures during the closed phase of August/September with less exposure to radiation, whereas they started to do more hands-on and interventional work throughout the open phase. Thereby, the fellows may have enjoyed close guidance and observation from the attending physicians once they started to perform interventions during the open phase. We assume that this might explain the lack of dose reduction for the attending physicians and the increased radiation dose for our fellows during the open phase along with the prolonged fluoroscopy time for some less complex or diagnostic procedures. Of note, however, the increase of PFM values for aorto-peripheral angiograms without interventions may be explained by reduced fluoroscopy times for that specific procedure and an increase in the DSA imaging. Besides awareness, training, education, and behavioral adaption are important to reduce radiation exposure. However, and due to the above-mentioned limitations based on the start of our study, radiation training and behavioral adaption were very limited for the attending physicians and the fellows within the present study. Nevertheless, the present trial included a learning period between the open and closed phases of 6 weeks. This time allowed the staff to familiarize themselves with the display panel and to begin establishing their own behavioral changes including its effect on radiation exposure. Among others (i.e., fellows starting to do more hands-on and interventional work, slightly more complicated procedures), we believe that the establishing of behavioral adaption may be one of reason for the increased radiation dose during the learning phase. In line, we believe that the behavioral adaption of the anesthesiologists, such as the use of shields, more distance from the patient and the radiation beam while monitoring the patients, may have contributed to their significant dose reduction during the open phase. However, it is difficult to clearly detect behavioral changes within the present trial for various reasons. First, operators and team members varied throughout the study period; second, there was no evaluation of preliminary data (i.e., after the learning phase) providing any feedback to the staff members; third, behavioral adaption of the staff was not systematically evaluated.
In addition to personal subgroup analysis, a significant reduction of radiation was attained for interventional and short procedures. Previous studies reported the unawareness of true dose exposure and the associated lack of preventive measures, particularly for shorter interventions [11]. Therefore, the present study underlined the importance of awareness regarding radiation and, in line, the benefit of a real-time monitoring mechanism for short procedures. The findings on interventional procedures within the present study represent the higher radiation dose applied when compared with diagnostic procedures. Accordingly, interventional procedures provide a wider scope for improvement on radiation, which was shown to benefit from real-time monitoring.
Between the open and closed phases there was no statistically significant reduction in system dose (DAP) even if normalized per procedural time. This may be related to a number of factors including the number and variability of cases and operators, which may be expected in a large institution like ours. There was a trend towards less total radiation dose PFM, but this was not shown as being statistically significant and it was difficult to prove within this study design.
Limitations
The present study contains various limitations. The effect of real-time monitoring on dose exposure was only validated by objective measurements but not for individual behavioral changes of the operators and the staff. By normalizing dose readings with fluoroscopy time, there was an attempt of comparing a large range of procedure types and different operators with a single variable to normalize for procedural complexity and time of procedure. The DAP is much more variable as it takes into account the total radiation dose including digital subtraction acquisitions which may not truly assess the time spent by the team member adjacent to the patient during the procedure. Fluoroscopy time should serve as a better surrogate for the time when team members were exposed to scatter radiation during a procedure. Interestingly, the average total fluoroscopy time increased by 4.1 min, between the closed and open phase likely related to the fellows doing more hands-on and interventional work and additionally slightly more complicated cases during the open phase. However due to the number of different operators, different patient size, and different procedural specifics it is difficult to accurately compare staff dose. Individual procedures did not yield a statistically significant reduction in team dose that are likely related to the small case numbers, and larger studies on specific procedures may be needed.
Conclusion
Real-time dose exposure monitoring may allow for behavioral changes and reduce staff exposure. However, further scrutiny on real-time monitoring systems is warranted within larger studies.
References
Sanchez R, Vano E, Fernandez JM, Gallego JJ (2010) Staff radiation doses in a real-time display inside the angiography room. Cardiovasc Intervent Radiol 33(6):1210–1214. doi:10.1007/s00270-010-9945-4
Vano E, Gonzalez L, Guibelalde E, Fernandez JM, Ten JI (1998) Radiation exposure to medical staff in interventional and cardiac radiology. Br J Radiol 71(849):954–960. doi:10.1259/bjr.71.849.10195011
Bartal G, Vano E, Paulo G, Miller DL (2014) Management of patient and staff radiation dose in interventional radiology: current concepts. Cardiovasc Intervent Radiol 37(2):289–298. doi:10.1007/s00270-013-0685-0
Kim KP, Miller DL, Balter S, Kleinerman RA, Linet MS, Kwon D et al (2008) Occupational radiation doses to operators performing cardiac catheterization procedures. Health Phys 94(3):211–227. doi:10.1097/01.HP.0000290614.76386.35
Jensen K, Zangani L, Martinsen AC, Sandbaek G (2011) Changes in dose-area product, entrance surface dose, and lens dose to the radiologist in a vascular interventional laboratory when an old X-ray system is exchanged with a new system. Cardiovasc Intervent Radiol 34(4):717–722. doi:10.1007/s00270-010-0017-6
Le Heron J, Padovani R, Smith I, Czarwinski R (2010) Radiation protection of medical staff. Eur J Radiol 76(1):20–23. doi:10.1016/j.ejrad.2010.06.034
Miller DL, Vano E, Bartal G, Balter S, Dixon R, Padovani R et al (2010) Occupational radiation protection in interventional radiology: a joint guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology. J Vasc Interv Radiol 21(5):607–615. doi:10.1016/j.jvir.2010.01.007
Cuaron JJ, Hirsch AE, Medich DC, Hirsch JA, Rosenstein BS (2011) Introduction to radiation safety and monitoring. J Am Coll Radiol 8(4):259–264. doi:10.1016/j.jacr.2010.08.020
Vano E, Gonzalez L, Fernandez JM, Haskal ZJ (2008) Eye lens exposure to radiation in interventional suites: caution is warranted. Radiology 248(3):945–953. doi:10.1148/radiol.2482071800
External Dosimetry: Operational Quantities and their Measurement, 11th International Congress of the International Radiation Protection Association (IRPA) Sess. (May, 2004)
Williams J (2004) Is there a benefit in promoting the concept of radiation risk? Br J Radiol 77(919):545–546
Racadio J, Nachabe R, Carelsen B, Racadio J, Hilvert N, Johnson N et al (2014) Effect of real-time radiation dose feedback on pediatric interventional radiology staff radiation exposure. J Vascul Interv Radiol 25(1):119–126. doi:10.1016/j.jvir.2013.08.015
Conflict of interest
Frederic Baumann: no conflict of interest. Barry T. Katzen: member of Advisory Board for Philips. Bart Carelsen: employee of Philips Healthcare who distributes the DoseAware technology. Nicolas Diehm: no conflict of interest. James Benenati: no conflict of interest. Constantino Peña: no conflict of interest.
Statement of Informed Consent
This study focused on the review of data acquired during the quality assessment process. IRB permission for the retrospective use of this HIPAA compliant data was obtained and informed consent was waived.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Baumann, F., Katzen, B.T., Carelsen, B. et al. The Effect of Realtime Monitoring on Dose Exposure to Staff Within an Interventional Radiology Setting. Cardiovasc Intervent Radiol 38, 1105–1111 (2015). https://doi.org/10.1007/s00270-015-1075-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00270-015-1075-6