Introduction

Magnetic resonance (MR) imaging is well suited to examine bone and soft tissue, but metal artifacts from prostheses hamper the visualization of periprosthetic tissues. The metal leads to variations in the static magnetic field, superimposing all imaging gradients and leads on the through-plane artifact (because of corruption of the slice selection gradient) as well as the in-plane artifact (because of corruption of the frequency encoding gradient). This results in a curved, thickened, thinned or split slice, seen as distortion, signal loss and pile-up effects in the MR image [1]. In inversion recovery sequences, severe artifacts can arise if the excitation and inversion slices are curved differently, since uninverted parts of the image will have different contrasts. The degree of artifacts in an image depends on the different components in the hip prosthesis as well as the different metal artifact-reducing sequence (MARS) methods applied [2]. Since metal-on-metal (MoM) prostheses have more metal components than non-MoM prostheses, the artifacts in the MR images are more severe with this first type.

Since the 1990s, the MR examination technique for patients with metallic implants has improved [36]. With high-bandwidth readout (hiBW), a reduction of the in-plane artifacts can be achieved. Further reduction of the in- and through-plane artifacts requires more in-depth changes of the imaging sequence. Cho et al. [7] used view angle tilting (VAT) to compensate for the distortion of the magnetic field. Later, VAT in combination with increased strength and higher bandwidth gradients was used in MARS [8]. Recently, a method called slice encoding for metal artifact correction (SEMAC), which corrects through-plane distortions by adding a phase-encoding gradient in the z-direction, was introduced [9]. Since SEMAC includes the VAT technique, it provides both in- and through-plane distortion corrections.

To optimize examinations of patients with hip prostheses, it is essential to be able to assess the extent of the metal artifacts in order to evaluate to what extent parts of bone or soft tissue are not visualized. To our knowledge, the extent of distortion in clinical images has not been measured previously. We observed that the shape of anatomic structures near a prosthesis changes with different MARS types and different hip prostheses. The angle formed by the legs of the inner margins of the os ilium and acetabular roof, the acetabular roof angle (ARA), is diminished with an increasing through-plane artifact in coronal slices near the center of the prosthesis, as the slices are bent in the examined volume. We hypothesize that the ARA can be used to measure the distortion (caused mainly by combined through- and in-plane artifacts). The measurement of signal loss and pile-up (in-plane artifacts) in the central slice based on the largest artifact area was introduced previously by Sutter at al. [10].

In this study, we apply the new ARA measure and the established artifact area quantification method in order to compare the distortion and artifact area caused by MoM and non-MoM total hip prostheses when using different MARSs: T1-hiBW, T1 VAT and T1 VAT + SEMAC as well as STIR-hiBW and STIR with matched RF pulses (mRFp).

Materials and methods

Patients

The hospital’s Ethics Committee approved the study, and all participants gave their written consent.

Measurements were conducted in 36 patients with MoM THA (13 males, 23 females; mean age 57, range 33–72 years) and 15 patients with non-MoM THA (6 males, 9 females; mean age 65, range 40–82 years) as described in the flowchart (Fig. 1) . All recruited patients in both groups were routine clinical cases. Exclusion criteria were standard contraindications for MR investigation.

Fig. 1
figure 1

Flowchart

In the group with MoM prostheses, T1-hiBW and T1 VAT sequences were applied in 17 and T1-hiBW, T1 VAT and T1 VAT + SEMAC in 19 patients. In the non-MoM prosthesis group, T1-hiBW and T1 VAT sequences were applied in nine and T1-hiBW, T1 VAT and T1 VAT + SEMAC in six patients. Fewer patients were investigated with SEMAC because this sequence was not available at the time the first patients were examined.

The STIR-hiBW and STIR-mRFp sequences were applied in the first 14 and 8 patients with MoM and non-MoM prostheses, respectively, but thereafter only the STIR-mRFp was applied since the artifact size was indisputably better reduced using this latter sequence.

MR imaging

Imaging was conducted with a 1.5-T MRI system (MAGNETOM Avanto; Siemens AG, Healthcare Sector, Erlangen, Germany) using the spine matrix and body matrix coils. The imaging protocol (Table 1) comprised three T1-weighted (T1-w) sequences: TSE with a high-readout bandwidth (T1-hiBW), TSE with VAT (T1 VAT), and TSE with both VAT and SEMAC (T1 VAT + SEMAC) and two inversion recovery sequences: one with a high-readout bandwidth (STIR-hiBW) and one with matched RF pulses (STIR-mRFp). The parameters of the T1 VAT + SEMAC were set to equal the scan time of the T1 VAT sequence. In the STIR-mRFp sequence, equal curving of the excitation and inversion slices near the prosthesis was obtained by matching the bandwidths of the excitation and inversion RF pulses, i.e., the same slice gradient strength was used for both pulses. T1 VAT, T1 VAT + SEMAC and STIR-mRFp were prototype sequences (‘WARP’ works-in-progress sequence package, Siemens AG).

Table 1 Sequence parameters

Measurement of the distortion and artifact area

The angle formed by the legs of the inner margins of the os ilium and the acetabular roof in the coronal center slice of the hip, the acetabular roof angle (ARA), was used to measure the distortion (Fig. 2). Measurement of the right and left hip in MR examinations of ten patients without THA showed an ARA > 150° (Table 2). The artifact area was measured by visual outlining of the artifact (defined as the area with signal void and pile-up artifacts, i.e., low and high signal) in the location of the prosthesis and surroundings (Fig. 3). The artifact area was delineated in the central slice through the prosthesis with the largest area as described by Sutter at al. [10]. The examinations were reviewed by two experienced musculoskeletal radiologists, BL with >25 years and GM with 8 years of musculoskeletal radiology experience.

Fig. 2
figure 2

Measurement of the normal acetabular roof angle (ARA) in a patient without THA in a central coronal slice of a T1-w image. TSE MR sequence (a) and conventional radiography (b)

Table 2 Acetabular roof angle (ARA) angle in ten patients without metal implants in the T1-w sequence (7 males, 3 females)
Fig. 3
figure 3

The largest artifact area outlined in a STIR mRFp central coronal slice of a MoM prosthesis (a) and in the region of the stem in a prosthesis with stainless steel (b)

Statistical analysis

Statistical analyses were performed with statistical software (IBM SPSS Statistics, release 21.0, Chicago, IL). Descriptive statistics were used to calculate the mean and range of the distortion angle and the artifact area. A paired t-test was applied to evaluate the ARA and artifact areas using the different sequences. The distortion angle was compared between the MoM and non-MoM prosthesis groups using the Wilcoxon-Mann-Whitney test. The artifact areas could not be compared between the groups of MoM and non-MoM prostheses since the prostheses differed in size. The level for statistical significance was p ≤ 0.05.

Results

Analysis of the distortion and artifact area

The ARA varied with the different MARS sequences (Fig. 4). The angle was smaller than normal in T1-hiBW and STIR-mRFp in all prostheses and in the VAT in the MoM prostheses. The T1 VAT + SEMAC showed a normal angle in MoM as well as non-MoM prostheses. The statistics of the measured angles are listed in Table 3. Comparing the different prostheses, the ARA was smaller in the MoM compared to the non-MoM prostheses, i.e., showed more distortion in the first type. In both MoM and non-MoM prostheses, the t-test showed significantly less distortion in T1 VAT + SEMAC than T1-hiBW (p < 0.001 and = 0.006, respectively), in T1 VAT + SEMAC than T1 VAT (p < 0.001 and = 0.014) and in T1 VAT than T1-hiBW (p < 0.001) (Fig. 5). Comparing the distortion between MoM and non-MoM prostheses, the Wilcoxon-Mann-Whitney test revealed significantly more distortion in MoM prostheses (Table 4). The ARA could not be measured in STIR-hiBW because of the large artifact size.

Fig. 4
figure 4

The ARA is small in T1-hiBW (a + d), less reduced in T1 VAT (b + e) and normal in T1 VAT + SEMAC (c + f)

Table 3 Descriptive statistics of the distortion angle ARA
Fig. 5
figure 5

Coronal image of an MoM prosthesis with different T1-w. sequences; T1-hiBW (a) shows large metal artifacts, T1 VAT (b) with a reduced artifact area and T1 VAT + SEMAC (c) with no distortion

Table 4 Wilcoxon-Mann-Whitney test of the difference between the angles of MoM and non-MoM prostheses in the different sequences. The iliopectineal angle could not be measured in STIR-hiBW because of the large artifact size

The artifact area varied with the different types of prostheses as well as the material of the prosthesis components (Figs. 5 and 6). In the MoM prosthesis, the artifact area in the T1 VAT + SEMAC sequence was similar to that in the T1-hiBW sequence. Despite the better delineation of the MoM implant with SEMAC, it does, however, show an additional area of signal void adjacent to the cranial aspect of the acetabular component (Fig. 7), which is why the total artifact area is not improved compared to the VAT sequence without SEMAC. The STIR-mRFp showed significantly less arifact area compared to the STIR-hiBW (Fig. 8). The descriptive statistics of the artifact area are listed in Table 5.

Fig. 6
figure 6

Artifacts in the T1 VAT images of (a) an MoM prosthesis with large artifacts in the cup region with low signal (*) and pile-up effect (arrowhead) in the femoral region and distortion (arrow), (b) a prosthesis with a cemented metal cup with a polyethylene lining and cemented stainless steel stem with a large artifact in the cup and stem region but less distortion since the lining is not metal and (c) a prosthesis with a cemented cup with non-MoM and uncemented titanium steel showing fewer artifacts in the cup and stem region

Fig. 7
figure 7

T1 VAT + SEMAC sequence (a) with a characteristic large artifact subsequent to the prosthesis cup (*) in a non-MoM prosthesis with a titanium cup and long titanium steel and (b) in surface replacement

Fig. 8
figure 8

Coronal images of an MoM prosthesis; STIR-hiBW showing large artifacts (a) and STIR-mRFp with reduced artifacts (b)

Table 5 Descriptive statistics of the artifact area

In the patients with MoM prostheses, the t-test showed a significantly smaller artifact area in the T1 VAT compared to T1-hiBW (p = 0.00, n = 36) and to the T1 VAT + SEMAC sequence (p = 0.00, n = 19), but no difference in artifact size between the T1-hiBW and T1 VAT + SEMAC sequence (p = 0.93, n = 19). In the patients with non-MoM prostheses, a significant difference was revealed between the T1 VAT and T1-hiBW sequence (p = 0.041, n = 15), but no difference between the T1 VAT and T1 VAT + SEMAC sequence (p = 0.18, n = 6) or between the T1-hiBW and T1 VAT + SEMAC sequence (p = 0.41, n = 6) was revealed.

The t-test showed a significant difference in the artifact area between the STIR-mRFp and STIR-hiBW in patients with MoM prostheses (p < 0.001, n = 14) and non-MoM prostheses (p < 0.001, n = 8) with smaller artifacts in the STIR-mRFp.

Discussion

The variation in the magnetic field around metal prostheses results in the in- and through-plane distortion of the image [1], seen as a deformation of the anatomic shape of the surrounding tissues. This is particularly the case for the bone near the prostheses; therefore, the angle formed by the legs of the inner margins of the os ilium and acetabular roof in the coronal center slice of the hip, the acetabular roof angle (ARA), is well suited to measure the distortion. The ARA will diminish with increasing distortion and is therefore reciprocal to the distortion. To our knowledge, a measurement of image distortion in clinical MR examinations of hip prostheses has not been previously reported.

Regarding the artifacts in the prostheses groups in our study, there was more distortion in the MoM prosthesis group, but artifact areas were overall larger in the non-MoM prosthesis group, with wide ranges of the measurement of distortion and artifact area in both groups. These variations were caused by the difference in material in the prosthesis components between and within the groups. The acetabular component of the MoM prostheses consisted of a cobalt-chromium alloy and of the non-MoM of a polyethylene liner. The stem in MoM prostheses varied from a short cobalt-chromium alloy base in the surface replacements to a long titanium stem in non-surface replacement prostheses. In the non-MoM prosthesis, the stem was of stainless steel or titanium. Since we compared the sequences in the same patients, the differences of the prosthesis parts did not affect the results.

Our results are similar to those of a recent study by Sutter et al. [10] regarding the small artifact area using a modified STIR sequence optimized for metal artifact reduction, but differ from this study as we found a smaller artifact area also in T1 VAT compared to the T1-hiBW and T1 VAT + SEMAC. These authors did not specify the type of prosthesis in their study. However, we think the different experiences regarding the latter type of sequence could be due to different cup materials in the studies, with more MoM prostheses in our material.

As expected, there was less distortion on average when imaging any type of prosthesis using both the in- and through-plane correcting T1 VAT + SEMAC sequence. In the non-MoM prostheses, the distortion in T1 VAT was also low but still higher compared with T1 VAT + SEMAC. The T1 VAT showed a significantly smaller artifact area than T1-hiBW in MoM and non-MoM prostheses as well as T1 VAT + SEMAC in MoM prostheses. The latter result of an increased artifact area with SEMAC may seem surprising; however, this is a consequence of a wrap-around artifact in the slice selection direction due to a trade-off in scan time, making the number of z-encoding steps in T1 VAT + SEMAC insufficient for a full correction of the curved slice.

Consequently, a limitation of our study is that the T1 VAT + SEMAC was designed to have a similar scan time as the other T1-weighted sequences. The scan time for each sequence was slightly more than 6 min; if further increased, the additional motion artifacts would be a potential drawback. The scan time in T1 VAT + SEMAC increases proportionally to the number of z-encoding steps, and the time restraint restricted the number of such steps to 12, which in some cases was not sufficient to restore the curved slice and resulted in severe signal voids due to artifact folding in the z-direction (slice selection direction); a further optimized sequence would require about 16 min. Thus, in this specific context the artifact area measure was influenced not only by an in-plane but also a through-plane effect. In the future, if further technical improvements reduce the acquisition time of the T1 VAT + SEMAC sequence and allow more z-encoding steps to be applied, this sequence may provide an overall superior performance in artifact reduction.

In summary, the new measure ARA can be used to evaluate metal-induced image distortion. Applying ARA as well as a previously described artifact area quantification method, we found the metal artifacts significantly reduced in MARS T1 VAT + SEMAC, T1 VAT and STIR-mRFp compared to MARS with only high bandwidth in both MoM and non-MoM prostheses.