Abstract
Purpose
The goal was to investigate whether the application of antiseptics and fluorides during orthodontic treatment affects the biomechanics of the levelling of dental arches by changing the working properties of nickel–titanium (NiTi) archwires.
Methods
The sample consisted of 60 individuals aged 12–22 years (53% females). There were 20 individuals in each experimental group: (I) individuals conducted regular oral hygiene, (II) individuals used a high concentration of fluorides for intensive prophylaxis for the first month; and (III) individuals used chlorhexidine in the same manner. NiTi alloy archwires (dimensions 0.508 × 0.508 mm) were analyzed 3 months after intraoral exposure and compared to as-received NiTi wires. Elastic modulus, yield strength, springback ratio, and modulus of resilience were calculated. Dimensions of the dental arches were analyzed when NiTi alloy (T1) was placed intraorally and after 3 months (T2). Change was quantified as the difference in dimensions (T2–T1). Anterior width-to-length ratio was used as a measure for dental arch shape.
Results
Intraoral exposure reduced elastic modulus, yield strength, springback ratio, modulus of resilience, loading, and unloading forces of NiTi wires (p ≤ 0.021). Chlorhexidine mouthwash and gel with high concentration of fluorides did not change these properties more than saliva with regular hygiene. The amount of change of dental arch shape in the maxilla and mandible did not differ significantly between the experimental groups.
Conclusion
Using antiseptics or a high concentration of fluorides during orthodontic treatment does not significantly affect the mechanical properties of NiTi wires and would not have clinical implications in changing orthodontic biomechanics.
Zusammenfassung
Absicht
Das Ziel der Studie war es zu untersuchen, ob die Anwendung von Antiseptika und Fluoriden während der kieferorthopädischen Behandlung die Biomechanik der Nivellierung der Zahnreihen beeinflusst, indem die Arbeitseigenschaften von NiTi(Nickel-Titan)-Bögen verändert werden.
Methode
Die Stichprobe bestand aus 60 Personen im Alter von 12–22 Jahren, von denen 53 % weiblich waren. Es gab jeweils 20 Personen in jeder experimentellen Gruppe: (I) Personen, die regelmäßige Mundhygiene durchführten, (II) Personen, die im ersten Monat eine hohe Konzentration von Fluoriden für eine intensive Prophylaxe verwendeten, und (III) Personen, die Chlorhexidin auf die gleiche Weise verwendeten. NiTi-Legierungsbögen (Abmessungen 0,508 × 0,508 mm) wurden nach 3 Monaten intraoraler Exposition analysiert und mit ursprünglich erhaltenen NiTi-Bögen verglichen. Elastizitätsmodul, Streckgrenze, Rückfederungsfaktor und Resilienzmodul wurden berechnet. Die Dimensionen der Zahnbögen wurden analysiert, als die NiTi-Legierung intraoral eingesetzt wurde (T1) und nach 3 Monaten (T2). Die Änderung wurde als Unterschied in den Dimensionen (T2–T1) quantifiziert. Das Verhältnis von Breite zu Länge im vorderen Bereich wurde als Maß für die Form des Zahnbogens verwendet.
Ergebnis
Die Ergebnisse zeigten, dass die intraorale Exposition das Elastizitätsmodul, die Streckgrenze, den Rückfederungsfaktor, das Resilienzmodul sowie die Belastungs- und Entlastungskräfte der NiTi-Bögen reduzierte (p ≤ 0,021). Chlorhexidin-Mundspülung und Gel mit hoher Fluoridkonzentration veränderten diese Eigenschaften nicht mehr als Speichel bei regelmäßiger Hygiene. Die Veränderung der Form des Zahnbogens im Ober- und Unterkiefer unterschied sich nicht signifikant zwischen den experimentellen Gruppen.
Schlussfolgerung
Die Anwendung von Antiseptika oder einer hohen Konzentration von Fluoriden während der kieferorthopädischen Behandlung hat keine signifikanten Auswirkungen auf die mechanischen Eigenschaften von NiTi-Bögen und keine klinischen Auswirkungen auf die Veränderung der kieferorthopädischen Biomechanik.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
After many years of development and improvements in orthodontic treatment, certain fundamental aspects are important for successful outcome of orthodontic treatment. These include orthodontic biomechanics and the dental materials which comprise active and passive orthodontic elements.
Nickel–titanium (NiTi) wires have low stiffness and good springback, responding elastically to loading and return to their original shape when unloaded, with only low forces acting [1, 2]. Low stiffness is characterized by a low Young’s modulus of elasticity (E), while yield strength (YS) is the maximum stress at which a material begins to deform plastically. The combination of E and YS is used to characterize flexibility as a response to loading by deformation without fracture, and resilience as the ability of an archwire to absorb energy during straining while being engaged in the brackets and to release energy during unloading while regaining its original shape. NiTi archwires are active elements that can generate a constant and light force during unloading. Unloading forces are key in orthodontics because they stimulate bone remodeling mediated by the tooth roots and periodontal ligaments. The direction of tooth movement is conditioned by the shape of the archwire (“archwire driven”) and by the alveolar bone substrate, i.e., the alveolar envelope.
Clinical implications associated with corrosion of orthodontic appliances in the oral cavity include changes in archwire properties, e.g., roughness, friction, hardness, stiffness, flexibility, resilience, and strength [2]. Dental biofilm control is an ongoing struggle in dentistry and it has been confirmed that biofilm accumulation during 1 month of exposure in the oral cavity can reduce the unloading forces of NiTi archwires [3].
Orthodontists are trying to combat the most common side effect of orthodontic treatment, which is a significantly increased risk for developing enamel demineralization and gingival inflammation caused by poor biofilm control, as well as oral mucosal lesions [4, 5]. To counteract these side effects, it is advisable to prescribe fluorides for the enamel and antiseptics for the soft tissues [6,7,8].
Fluorides and antiseptics may also promote corrosion of NiTi archwires by changing their working properties, thus, interfering with the desired tooth movement and levelling of the dental arches [9,10,11]. However, the majority of existing evidence derives from in vitro research, which poses two problems—oversimplifying the simulation of the oral environment and exaggerating the effects of the tested substances [9].
Therefore, clinical trial and laboratory tests were combined in this research with the aim to investigate whether the application of antiseptics and fluorides during orthodontic treatment affects the biomechanics of dental arch levelling by altering the working properties of NiTi archwires. It was hypothesized that the intraoral application of antiseptics and high concentrations of fluorides would not significantly alter the flexibility and resilience of NiTi archwires and would not limit levelling of the dental arch.
Materials and methods
The local ethics committees approved the research and written informed consent was obtained from all subjects or their parents/legally authorized representatives for children under 18 years of age (Ethics Committee of the Clinical Hospital Centre Rijeka No. 2170-29-02/1-14‑2 and Ethics Committee of the University of Rijeka, Faculty of Medicine No. 2170-24-01-14-05).
The sample for analysis consisted of 60 individuals aged 12–22 years (median 15; interquartile range 14–17 years); 53% of whom were females. A Consolidated Standards of Reporting Trials (CONSORT) flow diagram is shown in Fig. 1. Recruitment was initiated in November 2017, but the trial was stopped due to the coronavirus disease 2019 (COVID-19) pandemic in May 2020. All patients were undergoing orthodontic treatment with a 0.022″/0.56 mm slot MBT fixed orthodontic appliance (American Orthodontics, Sheboygan, WI, US) at the University Dental Clinic in Rijeka, Croatia. There were 20 individuals in each of three parallel experimental group: (I) individuals conducting regular oral hygiene, (II) individuals using a fluoride gel for intensive prophylaxis for 1 month, and (III) individuals using chlorhexidine digluconate mouthwash in the same manner. Individuals were allocated to groups according to a randomization plan generated at http://www.randomization.com.
A NiTi alloy archwire (dimensions 0.020 × 0.020″/0.508 × 0.508 mm; BioForce PLUS, Dentsply GAC International, Islandia, NY, USA) was placed intraorally after completion of the alignment phase of orthodontic treatment (6 months using NiTi archwires 0.014″/0.35 mm and 0.016 × 0.022″/0.41 × 0.56 mm) and left in place for 90 days. The alloy composition was wt%(Ni) = 50.4% and wt%(Ti) = 49.6%.
Charters technique of brushing teeth was demonstrated to all subjects and they were instructed to brush their teeth for 2 min three times a day. The first group was a control group and included subjects conducting regular oral hygiene, commonly recommended for orthodontic patients (toothbrush + toothpaste with a low fluoride concentration of 1450 ppm) throughout the study duration of 3 months. The second and third groups used adjuvant oral hygiene products during the first month: one group brushed their teeth three times per day without toothpaste and used a 0.12% chlorhexidine digluconate mouthwash (Curasept ADS 212, Curaden, Saronno, Italy) twice a day, while the other group brushed their teeth once a day without toothpaste and twice using a gel containing 6150 ppm of sodium fluoride (Mirafluor-K-gel, Hager Werken, Duisburg, Germany) for 30 days. For the next 2 months subjects performed standard hygiene (Charter’s technique, 3 times per day for 2 min with a toothpaste containing a low concentration of 1450 ppm fluoride).
A minimum sample size of 16 participants in each group was calculated as necessary if a change in arch shape (premolar width-to-length ratio) of 0.15 in one experimental group and 0.05 in the other group with a standard deviation (SD) of 0.10 in both groups is anticipated, with a power of 80% and significance of α = 0.05. For analysis of the mechanical characteristics, a sample size of 6 samples in each group was calculated to be sufficient to detect differences in Young modulus of 1.5 considering a standard deviation of 0.9.
Thus, archwires were collected from 18 participants and compared with as-received NiTi alloy wires. Six specimens from each experimental group, each 2.5 cm long, were used for mechanical three-point bending test in a thermal chamber at 37 °C on a universal machine Instron Model 1125/5500 (Instron, Binghamton, NY, USA). Each specimen was deflected to 3.1 mm and then dropped to zero deflection at a crosshead speed of 0.0167 mm/s. Force (N) and deflection (mm) were recorded every 5 ms for each specimen during loading and unloading, using Texture Exponent software (Stable Micro System, Godalming, UK). Force–deflection curves were generated from these data. The elastic modulus (E), yield strength (YS), springback ratio as measures of flexibility (YS/E), and the modulus of resilience (YS2/2E) were calculated. The surface of representative specimens was analyzed with a JSM-7800F field emission scanning electron microscope (FE-SEM; JEOL, Tokyo, Japan) using a secondary electron detector with an electron beam accelerating voltage of 10 kV and a working distance of 10 mm.
Plaster casts of the dental arches were made and analyzed before (T1) and after (T2 after 90 days) NiTi alloy archwire 0.508 × 0.508 mm were placed intraorally. Sliding caliper and Korkhaus caliper (Dentaurum GmbH, Ispringen, Germany) were used for measuring arch dimensions. Changes were quantified as the difference in dimensions (T2–T1) of anterior (premolar) and posterior (molar) arch width, and anterior (premolar) arch length in the maxilla and mandible. The following were measured: premolar arch width as the distance between the lower-most points of the transverse fissure of the maxillary first premolars, molar arch width as the distance between the point of intersection of the transverse fissure with the buccal fissure of the maxillary first permanent molar, and anterior arch length as a perpendicular from the most anterior labial surface of the central incisors to the connecting line of the reference points of premolar arch width. In the mandible, the vestibular contact points between the first and second premolar were used for the premolar width, and for the molar width the tip of the mediobuccal cusp of the mandibular first permanent molar were used. Changes were quantified as the difference in dimensions (T2–T1) in premolar and molar width of the dental arch, and lower premolar depth, separately for the maxilla and mandible. Premolar width-to-length ratio was used as a measure of the dental arch shape (a higher value indicated a more rectangular shape of the dental arch [wider and shorter] and a lower value indicated a more tapered or ovoid shape of the dental arch [narrower and elongated]). Persons who analyzed the plaster casts and mechanical properties of the archwires were blinded regarding group affiliations.
Normality of the data distribution was tested using the Shapiro–Wilk test. Statistical analyses included analysis of variance and Student–Newman–Keuls post hoc test for the archwire’s working properties, while Kruskal–Wallis and Mann–Whitney tests with Bonferroni correction were used to assess arch shape changes. The Wilcoxon test was used to compare arch dimensions before and after 3 months. Commercial statistical software SPSS version 22.0 (IBM Corp., Armonk, NY, USA) was used and a significance level was preset to p < 0.05.
Results
Baseline characteristics
Baseline demographic and clinical characteristics of each group are shown in Table 1. Age distribution of men and women and premolar width–length ratio did not differ significantly between the groups.
Surface morphology
SEM images of the surface of the NiTi wires showing roughness and porosity in the as-received archwires and in the three experimental groups are presented in Fig. 2. Corrosion resulted in a smoothening of the surface under all conditions, which is due to a dissolution of surface peaks. However, the samples exposed to the fluoride medium demonstrated the highest number of surface defects.
Elastic properties
Intraoral exposure reduced the elastic properties of the NiTi wires (p ≤ 0.021), but fluorides and antiseptics did not change the properties more than saliva combined with regular oral hygiene (Figs. 3, 4, and 5). Findings were the same for all measured loading and unloading forces at any deflection and also for the elastic modulus, yield strength, springback ratio, and modulus of resilience.
Changes in dental arch shape
Maxillary premolar length increased significantly in the control and in the group using antiseptics (p ≤ 0.025) with medians of 0.5 and 1 mm, respectively. Maxillary premolar width increased only in the group using antiseptics (median = 0.5 mm; p = 0.008). There was a decrease in the width–length ratio in the maxilla (median = −0.02 in the group using antiseptics and −0.06 in the control group), with the dental arch becoming more tapered (more in controls) or ovoid (more in the group using antiseptics). When the changes in the dimensions and shape of the maxillary arch (ΔT2–T1) were compared between the experimental groups, there were no statistically significant differences (Table 2, Fig. 6).
There was a significant increase in premolar mandibular width in each of the three groups (p ≤ 0.008) with a median change of 1 mm. Mandibular premolar length did not change significantly in any of the groups. Width–length ratio in the mandible mostly increased in the group using fluoride (median 0.08; p = 0.023). In the control group and in the group using antiseptics, mandibular width–length ratio also increased a median of 0.07 and 0.01, respectively, i.e., it reshaped to a more rectangular form. There were no significant differences in the amount of change of arch dimensions and shape in the mandible (ΔT2–T1) between the three experimental groups.
Discussion
The present research implies that application of fluorides and antiseptics during orthodontic treatment to control dental biofilm and enamel demineralization causes minor changes in the clinical performance of NiTi archwires.
Microstructure, pH of saliva, and the presence of fluorides cannot be acknowledged as possible reasons of corrosion of NiTi alloy, although published results vary greatly [12]. Qualitative surface analysis of the wires in the present research also showed a damaged surface after the use of high concentration fluoride gel, although our previous quantitative analysis implied that corrosion probably dissolves peaks, thus, smoothing the surface intraorally [13]. Intraoral corrosion seems inevitable. The presence of pre-existing defects on the surface of NiTi archwires can be the predilected locations for corrosion, while a titanium oxide layer protects against corrosion [14]. Longer in vivo exposure (2 months) decreased the corrosion susceptibility [15]. However, fluorides had a tendency to dissolve the titanium oxide layer causing localized, pitting corrosion [16]. The nickel ions are mostly released by corrosion during the first 3 days after which the corrosion tendency decreased [11].
Our previously published results revealed that general corrosion is the predominant type of corrosion of intraoral aging of NiTi wires, but localized corrosion may also occur [13]. Corrosion behavior is similar when fluorides and antiseptics are used, but their use did not additionally increase the corrosion rate. Furthermore, surface roughness did not change significantly. The present research demonstrates that NiTi archwires after intraoral exposure under the influence of saliva and regular oral hygiene were less elastic and produced lower forces than as-received, unused archwires. Similar results were found in in vitro studies [17,18,19]. The load–deflection properties of archwires are among the most important parameters in determining the biological response during tooth movement [20]. A three-point bending test used in this research is the gold standard to assess the elastic properties of orthodontic archwire materials. Forces generated by orthodontic appliances are difficult to measure intraorally which is why elastic properties were measured on wires after removal from the mouth. This laboratory testing was able to measure unloading forces for different deflections which are equivalent to the resulting tooth displacement under clinical conditions.
According to our research, although some in vitro studies claim that topical fluoride agents affected the mechanical properties of NiTi wires during unloading of stress [10, 21], neither high concentration of fluorides nor daily use of a chlorhexidine mouthwash for 1 month during orthodontic treatment further impaired the elastic properties of the wires.
In addition, straining of NiTi wires in the oral environment in the presence of fluoride is followed by a significant risk for corrosion–induced fractures [22, 23]. Earlier research reported no significant reduction in the load or deflection forces of NiTi wires in vitro [24].
Previously, we reported that fluorides tended to reduce stiffness and hardness more than saliva or antiseptics, but this effect was not significant [13]. In that previous paper, only parts of the analysis of the mechanical and chemical properties of the NiTi wires were reported (roughness, friction, hardness, stiffness, and corrosion behavior). The present paper reports about the analysis of additional results from different samples from the same experiment (elasticity). In addition, clinical effects of the whole sample were calculated. Contrary to our findings, one clinical trial indicated that topical fluoride regimens reduced the unloading properties of NiTi wires in patients undergoing fixed orthodontic treatment, but only at higher deflections [25].
The in vitro experiments revealed no significant effect of the chlorhexidine mouthwash on the roughness of NiTi archwires and on frictional resistance. Furthermore, no further deterioration in the elastic properties of the NiTi wires compared to the effects of saliva could be found [11, 26]. In contrast, another in vivo study found that both roughness and friction increased after the use of chlorhexidine mouthwash [27]. These opposite results may be due to the chlorhexidine molecules in the in vivo situation, where binding to the teeth and oral mucosa may have caused prolonged exposure. In vitro results suggested that a change in the pH of saliva had only minor influence on the hardness and modulus of elasticity of NiTi archwires [28]. In addition, consumption of soft drinks did not significantly lead to degradation of the physical and chemical properties of NiTi archwires [29].
Corrosion does not necessarily lead to a major deterioration of the working properties of a NiTi alloy. A decrease in the elastic properties resulted in only smaller effects on dental arch levelling. For the assessment of dental arch shape changes, we defined dental arch shape as a ratio of premolar width and length to annul large interindividual variability in arch width and arch length. Changes in the shape of the dental arches were expected as these were defined by the original archwire form. As a consequence, the maxillary arch transformed towards a semi-ellipse, and the mandibular towards a parabola. This pattern was the same in all three experimental groups and was not significantly altered by fluorides or chlorhexidine application. To our knowledge, this is the first study that related laboratory tests of mechanical properties of orthodontic wires to clinical findings. The wires were analyzed after they had been in the mouth for 3 months. This is usually the time when they are replaced with new wires as part of orthodontic treatment. Gel with high fluoride concentration is usually not used for more than 1 month. Chlorhexidine is also not used for a longer periods of time, as it causes discoloration of the teeth and tongue as well as a change in taste. Therefore, this experiment is a good approximation of the real and typical clinical situation. Alignment of the teeth had been done before the experiment started, but the leveling of the curve of Spee and also expansion of the arches was not completed. Thus, a possible association between any changes in wire properties due to experimental exposure to fluorides and antiseptics, and the resulting biomechanics could be established. The observed clinical effect is the result of the activity of the archwire in the mouth—it changes the shape of the dental arch. These were the advantages of this study. The limitations are that the extent of the levelling of the dental arches may have been determined by the alveolar envelope. However, NiTi arches do not control the vestibulo-oral position of the roots very well, but mostly tip the crown of the teeth. Therefore, we do not expect this to be a major limitation. Future studies should analyze the changes induced by thinner round wires in the initial alignment phase.
Conclusion
Intraoral exposure deteriorated the elastic properties of nickel–titanium (NiTi) archwires but the use of fluorides or antiseptics during orthodontic treatment had no significant effect on the elastic properties of the wires and should have no major clinical implications in changing the orthodontic biomechanics.
References
Kapila S, Sachdeva R (1989) Mechanical properties and clinical applications of orthodontic wires. Am J Orthod Dentofacial Orthop 96:100–109. https://doi.org/10.1016/0889-5406(89)90251-5
Brantley WA (2020) Evolution, clinical applications, and prospects of nickel-titanium alloys for orthodontic purposes. J World Fed Orthod 9:S19–S26. https://doi.org/10.1016/j.ejwf.2020.08.005
Sapata DM, Ramos AL, Sábio S, Normando D, Pascotto RC (2020) Evaluation of biofilm accumulation on and deactivation force of orthodontic Ni-Ti archwires before and after exposure to an oral medium: A prospective clinical study. J Dent Res Dent Clin Dent Prospects 14:41–47. https://doi.org/10.34172/joddd.2020.006
Walsh LJ, Healey DL (2019) Prevention and caries risk management in teenage and orthodontic patients. Aust Dent J 64(Suppl 1):S37–S45. https://doi.org/10.1111/adj.12671
Manuelli M, Marcolina M, Nardi N, Bertossi D, De Santis D, Ricciardi G et al (2019) Oral mucosal complications in orthodontic treatment. Minerva Stomatol 68:84–88. https://doi.org/10.23736/S0026-4970.18.04127-4
Sonesson M, Brechter A, Abdulraheem S, Lindman R, Twetman S (2020) Fluoride varnish for the prevention of white spot lesions during orthodontic treatment with fixed appliances: a randomized controlled trial. Eur J Orthod 42:326–330. https://doi.org/10.1093/ejo/cjz045
Benson PE, Parkin N, Dyer F, Millett DT, Germain P (2019) Fluorides for preventing early tooth decay (demineralised lesions) during fixed brace treatment. Cochrane Database Syst Rev 2019:CD3809. https://doi.org/10.1002/14651858.CD003809.pub4
Takenaka S, Ohsumi T, Noiri Y (2019) Evidence-based strategy for dental biofilms: Current evidence of mouthwashes on dental biofilm and gingivitis. Jpn Dent Sci Rev 55:33–40
Fragou S, Eliades T (2010) Effect of topical fluoride application on titanium alloys: a review of effects and clinical implications. Pediatr Dent 32:99–105
Gupta AK, Shukla G, Sharma P, Gupta AK, Kumar A, Gupta D (2018) Evaluation of the effects of fluoride prophylactic agents on mechanical properties of nickel titanium wires using scanning electron microscope. J Contemp Dent Pract 9:283–286
Rincic Mlinaric M, Karlovic S, Ciganj Z, Acev DP, Pavlic A, Spalj S (2019) Oral antiseptics and nickel-titanium alloys: mechanical and chemical effects of interaction. Odontology 107:150–157. https://doi.org/10.1007/s10266-018-0387-9
Močnik P, Kosec T (2021) A Critical appraisal of the use and properties of nickel-titanium dental alloys. Materials (Basel) 14:7859. https://doi.org/10.3390/ma14247859
Zibar Belasic T, Pejova B, Otmacic Curkovic H, Kamenar E, Cetenovic B, Spalj S (2021) Influence of intraoral application of antiseptics and fluorides during orthodontic treatment on corrosion and mechanical characteristics of nickel-titanium alloy in orthodontic appliances. Angle Orthod 91:528–537. https://doi.org/10.2319/052620-480.1
Huang HH, Chiu YH, Lee TH, Wu SC, Yang HW, Su KH et al (2003) Ion release from NiTi orthodontic wires in artificial saliva with various acidities. Biomaterials 24:3585–3592. https://doi.org/10.1016/s0142-9612(03)00188-1
Primozic J, Hren M, Mezeg U, Legat A (2022) Tribocorrosion susceptibility and mechanical characteristics of as-received and long-term in-vivo aged nickel-titanium and stainless-steel archwires. Materials (Basel) 15:1427. https://doi.org/10.3390/ma15041427
Abbassy MA (2016) Fluoride influences nickel-titanium orthodontic wires’ surface texture and friction resistance. J Orthod Sci 5:121–126. https://doi.org/10.4103/2278-0203.192114
Hosseinzadeh Nik T, Ghadirian H, Ahmadabadi MN, Shahhoseini T, Haj-Fathalian M (2012) Effect of saliva on load-deflection characteristics of superelastic nickel-titanium orthodontic wires. J Dent (Tehran) 9:171–179
Ramazanzadeh BA, Ahrari F, Sabzevari B, Zebarjad SM, Ahrari A (2011) Effects of a simulated oral environment and sterilization on load-deflection properties of superelastic nickel titanium-based orthodontic wires. Int J Orthod Milwaukee 22:13–21
Harris EF, Newman SM, Nicholson JA (1988) Nitinol arch wire in a simulated oral environment: changes in mechanical properties. Am J Orthod Dentofacial Orthop 93:508–513. https://doi.org/10.1016/0889-5406(88)90080-7
Krishnan V, Kumar KJ (2004) Mechanical properties and surface characteristics of three archwire alloys. Angle Orthod 74:825–831. https://doi.org/10.1043/0003-3219(2004)074%3C0825:MPASCO%3E2.0.CO;2
Mane P, Ganiger CR, Pawar R, Phaphe S, Ronad YA, Valekar S et al (2021) Effect of fluoride on mechanical properties of NiTi and CuNiTi orthodontic archwires: an in vitro study. Dental Press J Orthod 26:e212020. https://doi.org/10.1590/2177-6709.26.2.e212020.oar (eCollection 2021)
Kassab E, Gomes JP (2021) Corrosion induced fracture of NiTi wires in simulated oral environments. J Mech Behav Biomed Mater 116:104323. https://doi.org/10.1016/j.jmbbm.2021.104323
Kaneko K, Yokoyama K, Moriyama K, Asaoka K, Sakai J (2004) Degradation in performance of orthodontic wires caused by hydrogen absorption during short-term immersion in 2.0 % acidulated phosphate fluoride solution. Angle Orthod 74:487–495. https://doi.org/10.1043/0003-3219(2004)074%3C0487:DIPOOW%3E2.0.CO;2
Srivastava K, Chandra PK, Kamat N (2012) Effect of fluoride mouth rinses on various orthodontic archwire alloys tested by modified bending test: an in vitro study. Indian J Dent Res 23:433–434. https://doi.org/10.4103/0970-9290.102253
Vo J, Chudasama DN, Rinchuse DJ, Day R (2010) A clinical trial to evaluate the effects of prophylactic fluoride agents on the superelastic properties of nickel-titanium wires. World J Orthod 11:135–141
Hosseinzadeh Nik T, Hooshmand T, Farazdaghi H, Mehrabi A, Razavi ESE (2013) Effect of chlorhexidine-containing prophylactic agent on the surface characterization and frictional resistance between orthodontic brackets and archwires: an in vitro study. Prog Orthod 14:48. https://doi.org/10.1186/2196-1042-14-48
Razavi EE, Nik TH, Hooshmand T, Farazdaghi H, Arefi AH (2021) Surface characterization and frictional force between stainless steel brackets and archwires in orthodontic patients using chlorhexidine- and persica-containing mouthrinses: A randomized controlled trial. Dent Res J (Isfahan) 18:21
Jusufi Osmani Z, Poljšak B, Zelenika S, Kamenar E, Marković K, Perčić M et al (2022) Ion release and surface changes of nickel-titanium archwires induced by changes in the pH value of the saliva-significance for human health risk assessment. Materials (Basel) 15:1994. https://doi.org/10.3390/ma15061994
Parenti SI, Guicciardi S, Melandri C, Sprio S, Lafratta E, Tampieri A et al (2012) Effect of soft drinks on the physical and chemical features of nickel-titanium-based orthodontic wires. Acta Odontol Scand 70:49–55. https://doi.org/10.3109/00016357.2011.575083
Acknowledgements
Special gratitude to Goran Bosanac for help in the three-point bending test.
Funding
The part of the work is made possible by using the equipment funded via the European Regional Development Fund project No. RC.2.2.06-0001 (Development of Research Infrastructure at the University of Rijeka Campus—RISK). Grants: Granted by Croatian Science Foundation No. IP-2014-09-7500 (IMUNODENT) and IP-2020-02-4027 (DENTALBIOFILM) and DOK-2021-02-6438.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
T.Z. Belasic, M. Zigante, M. Uhac, S. Karlovic, I.J. Badovinac and S. Spalj declare that they have no competing interests.
Ethical standards
All procedures performed in studies involving human participants or on human tissue were in accordance with the ethical standards of the institutional and/or national research committee and with the 1975 Helsinki declaration and its later amendments or comparable ethical standards. Informed consent was obtained from all individual participants included in the study. The local ethics committees approved the research, with written informed consent obtained from all subjects or their parents/legally authorized representatives in children under the age of 18 years (Ethics Committee of the Clinical Hospital Centre Rijeka No. 2170-29-02/1-14‑2 and Ethics Committee of the University of Rijeka, Faculty of Medicine No. 2170-24-01-14-05). Trial registration: ClinicalTrials.gov Identifier: NCT03334461.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Zibar Belasic, T., Zigante, M., Uhac, M. et al. Effect of use of antiseptics and fluorides during orthodontic treatment on working properties of NiTi archwires in levelling dental arches. J Orofac Orthop 85 (Suppl 1), 63–72 (2024). https://doi.org/10.1007/s00056-023-00475-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00056-023-00475-1