Abstract
Heterogeneity of cell populations is a major obstacle for understanding complex biological processes. In order to have a more comprehensive quantitative comprehending of cellular processes, it is necessary to quantify the distribution of behavior in a population of individual cells. Analysis of single-cell behaviors requires efficient single-cell capture, controllable single-cell culture performance as well as reliable analysis techniques. The microfluidic system provides advanced technology for single-cell culture and observation. This chapter gives a brief account of single-cell capture by microfluidic methods, long-term single-cell culture on both two-dimensional models and three-dimensional microfluidic systems, as well as single-cell growth and differentiation in a microfluidic environment. Furthermore, the advanced methods used for characterizing on-chip single-cell culture were also discussed.
Access provided by Autonomous University of Puebla. Download chapter PDF
Similar content being viewed by others
Keywords
3.1 Introduction
Single-cell analysis holds much promise to better understand cell behaviors and cell metabolism [1, 2]. Individual isogenic cells are not identical even in the same culture environment [3, 4] known as heterogeneity. Heterogeneity among cell populations is a major obstacle to understand complex biological processes. The averaged results from the sample of large cell populations are insufficient when considering individual cell behaviors and usually obscure the variable response of individual cells. These problems are highlighted in the research of normal and malignant stem and progenitor cell populations, because of the lack of cell purification methods [5] and the inherent stochastic nature of cells’ self-renewing and differentiation [6, 7]. Cell-to-cell interactions grown in a monolayer cell culture microenvironment are considered as one of the vital factors which influence large distributions in behavior. In this case, both contact and diffusible elements seem to be working. Therefore, single-cell analysis is increasingly being employed to a number of different biomedical fields, such as genetic analysis [8], cancer progression [9, 10], developmental biology [11], and fundamental biological studies on rare stem cells, or cancer cells [12].
The accompanying need with single-cell analysis is cell culture method to interrogate phenotypes through a culture of the rare amount of cells or single cells [13,14,15,16,17]. Single-cell culture is based on multiple generations from single-cell starting points. However, it is technically difficult to establish the single-cell culture platform. Technical difficulties are mainly reflected in (1) precise manipulation of cells and minimal loss of samples; (2) contact with non-biological surfaces and minimum shear stress or other damage stresses; (3) sustainable growth of target cell number and total cell number (need enough culture space, replaceable medium, etc.). In order to solve these difficulties, the development of cell culture platform requires the inherent advantages of microfluidics. Microfluidic platforms have recently been widely applied to realize cell-based assays, for example, manipulation of single cells [18], automated media perfusion [19, 20], providing cellular microenvironment and external stimuli to study cellular responses [21,22,23,24], and establishment of long-term cell culture systems.
In general, single-cell analysis requires (1) a large number of captured individual cells in order to obtain statistical significance of single-cell properties, (2) long-term clonal cell culture originated from a start of a single cell, (3) continuous observation of multiple generations of cells on their growth, differentiation, phenotype, metabolism, and other aspects. Single-cell culture is a crucial step in single-cell level analysis followed by single-cell isolation. In this chapter, we mainly present the advances, limitations, and outlooks in microfluidics based on the exploration of single-cell culture, categorized as two-dimensional (2D) and three-dimensional (3D) devices for the cultivation of a variety of cells, and analysis at single-cell level. We first introduce single-cell capture including conventional methods and microfluidic methods, and then we shift to single-cell culture based on microfluidics. In the last section, we focus on the advanced analysis techniques for single-cell culture.
3.2 Single-Cell Capture
Single-cell capture is the first step for single-cell analysis. Conventional methods for single-cell capture include micromanipulation , laser capture microdissection (LCM), and fluorescence activated cell sorting (FACS) (Fig. 3.1). Micromanipulation usually achieves through manual identification and selection with a microscope, which has the advantages of being low cost and easy-to-implement and the disadvantages of being laborious and low throughput [25]. LCM utilizes laser to cut and collect individual cells from tissues also under a microscope [26, 27]. LCM usually requires fixed tissue or cells and high laboratory skills. FACS is a high-throughput single cell sorting system, which could collect single cell into tubes or microwells for further experiments; while FACS is generally lack of visual inspection of the cells. Conventional micromanipulation , FACS , and LCM , could be referred to previous reviews [28,29,30,31], and here are only briefly described.
Modern approaches are based on microfluidic systems for single-cell capture and analysis, including droplet , microwell, hydrodynamic traps, live single-cell extractor (LSCE), electrical traps , optical traps (Fig. 3.1). Microfluidic devices have more advantages than conventional approaches in single-cell capture. Microfluidic systems offer a closed platform for integrating downstream experiments, which reduce the risk of sample contamination. Miniaturized compartments for single cell reduce the experimental cost and enrich biomolecules with low concentration [32]. Moreover, most microfluidic systems could be automated to save labor, which means easy to achieve high throughput. In this part, we focus on the microfluidic design of single-cell capture for further downstream culture or analysis.
3.2.1 Conventional Methods
3.2.1.1 Micromanipulation
Micromanipulation relies on the manual selection technique typically through a micropipe combined with a microscope [33]. As shown in Fig. 3.1a, target single-cell identification is via microscope visual inspection. The micropipe moves in close proximity and picks a specific suspended cell in a petri dish or well plate. This process is usually performed manually. Micromanipulation can obtain specific cells accurately, but cannot achieve high throughput. Except single-cell capture, the micromanipulation is widely applied in the bacterial analysis [34], reproductive medicine, and forensics.
3.2.1.2 Laser Capture Microdissection
Laser capture microdissection (LCM ) is a mature technique to capture single cells or tissue compartments typically from fixed tissue or cells samples [33]. As shown in Fig. 3.1b, the working process of LCM consists of laser cutting procedure and extraction of dissected tissue. Target cell or compartment is observed via a microscope and marked to be cutoff. The focused laser cuts the selected section, and then the operator extracts the dissected tissue. A particular advantage of LCM is that the access to cells in situ is beneficial for the spatial information of single cells [35, 36]. Typical samples are cryo-fixed, fixed in formalin, or embedded in paraffin, which causes cell death. Some modern LCM systems allow the extraction of living cell or tissue for downstream cell culture or analysis, such as Leica LMD7000 with live cell cutting (LCC). However, if the target cells are not precisely cut, they may be contaminated by adjacent cells. LCM allows very limited throughput.
3.2.1.3 Fluorescence-Activated Cell Sorting
Fluorescence-activated cell sorting (FACS) is a high-throughput and automatic method for cell sorting, and large amount of cells can be sorted in a very short time [37]. FACS uses laser excitation and provides various analytical choices (Fig. 3.1c). Cell characteristics such as relative size and granularity can be obtained as forward scattering (FSC) and side scattering (SSC), respectively. In addition, cells’ functional properties can be measured by fluorescence staining. The samples for FACS cover almost every cell type from blood, bone marrow, tumor, plants, protoplasts, yeast, bacteria, and viruses. Thus, FACS has a diverse spectrum of applications, such as quantification of soluble molecules and subpopulations [38], cancer diagnostics, microbial analysis [39], cell cycle analysis, hematopoietic stem cells , DNA content analysis, and apoptosis. FACS has been recognized as a worldwide standard for cell population analysis and sorting. However, the drawback to this approach is that the morphological information is limited to cell size and complexity. In addition, the open environment of FACS system increases the risk of introducing pollution.
3.2.2 Microfluidic Methods
3.2.2.1 Droplets
Droplet -based microfluidics use oil to keep aqueous droplets separated. The droplet like a separate microreactor encapsulates single cells, thus single cells can be collected [40, 41]. Droplet systems for single-cell capture typically generate droplets from a statistically dilute suspension of cells [42]. Single-cell capture via droplets is a feasible and high-throughput method. This method provided a platform for various biological assays on the single-cell level, such as single-cell culture [43, 44], single-cell counting [45], antibody detection [46], drug screening [47], RNA sequencing (RNA-seq) [48, 49], PCR [50, 51], whole genome amplification [52, 53], enzyme screening [54, 55].
Droplet-based single-cell isolation is usually consistent with Poisson distribution, which means that more than half of the droplets are cell-free [56]. Stochastic cell loading is the inherent variability of droplet contents. For single-cell analysis, it is important to improve the droplet encapsulation efficiency of a single cell, while reduce empty droplets and droplets encapsulating multiple cells. Much effort had been made to improve the efficiency of single-cell encapsulation, such as optimizing channel designs [57, 58]. Improving the efficiency of single-cell loading is another method to maximize the number of droplets containing a single cell. Edd et al. proposed a method to controllably load single cells into drops by utilizing inertial sorting to generate two trains of single cells in the loading channel (Fig. 3.1d) [59]. This method caused self-organization of cells, which conquered the intrinsic limitations arose from Poisson distribution, and ensuring that almost every drop contains a single cell. Kemna et al. introduced Dean forces by long helical microchannels and the final cell encapsulation efficiency of ~80% because cells were ordered in the channel before being encapsulated into droplets [60]. Droplet-based microfluidics has becoming a popular technology with extensive applications, such as biotechnology [61], chemical analysis [62]. Droplet -based high-throughput single-cell biology is promising, but not yet realized.
3.2.2.2 Microwells
Microwell -based single-cell capture generally relies on sized well or modified surface, and cells are separated by the physical boundaries of multi-well. With the microwell array, cells that are attracted to the microwells can be preserved, and other cells outside the microwells would be washed away. Microwells provide a simple method enabling high-throughput single-cell analysis. The prominent advantages of microwells are that the microdevice is simple to operate and easy to achieve the capture of plenty cells. However, the disadvantage is that single-cell capture may occur only in part of the microwell array, and there may be a microwell with multiple cells or without any cells. Adjusting the size and shape of the microwell and optimizing the concentration of the cell suspension are the common solutions to improve the efficiency of single-cell capture.
Microfabrication technology allows the simple preparing of parallel microwells and enables high-throughput single-cell analysis. Liu et al. molded self-assembled polystyrene microspheres in poly (dimethylsiloxane) (PDMS) to produce microwell arrays for single-cell collection (Fig. 3.1e) [63]. For improving the efficiency of single-cell capture, Huang et al. constructed a truncated cone-shaped microwell array, which realized ~90% single-cell occupancy within a few seconds [64]. Surface modification achieved cell collecting without complicated fabrication of cell-sized microwells [65,66,67]. The combination of surface modification and specific size of microwells could improve the efficiency of cell occupancy. Specific aptamer and protein micropattern could also be modified in the microwell for capturing single cells from a mixture [68].
3.2.2.3 Hydrodynamic Traps
Hydrodynamic trap is a very common method for single-cell capture. The working principle of hydrodynamic trap system is that physical barriers or hydrodynamic tweezers stop the cell and remove it from the flow of cell suspension. In order to trap single cell, physical barriers are designed microscale structures, and hydrodynamic tweezers are specific and complex fluid flow profiles [69]. The advantages of using this method are that it does not require complex experimental steps and operations and could handle a large number of cells in a short time.
U-shaped traps as a popular physical barrier have been widely applied to obtain single cell. One of the typical single-cell U-shaped traps was demonstrated by Di Carlo et al. to collect single cells in large arrays (Fig. 3.1f) [70]. They used this U-shaped traps to determine single-cell enzyme kinetics for three different cell line (HeLa, 293T, Jurkat). Pattern flow channels have been introduced into the U-shaped traps structure, which allowed the fluid to fill the channel and provide a suitable environment for cell culture [71]. Another type of hydrodynamic trap system is hydrodynamic tweezers. Lutz et al. used four vortices fabricated from cyclic oscillations of the sound waves around a cylinder to trap single cell in the vortex center [72]. In addition, hydrodynamic tweezers also have been used to trap particles [73, 74].
3.2.2.4 Live Single-Cell Extractor
Live single-cell extractor (LSCE) is a newly developing tool for collecting single adhered cell in a tissue culture. LSCE consists of injection part and aspiration part (Fig. 3.1g) [75]. The target cell is digested by the trypsin injected around the cell and then is extracted by aspiration part for further culture or analysis. A prominent advantage of LSCE is to extract live single-cell in situ without damage. Thus, it holds the spatial information of live single cells to their specific positions in the tissue. This method is usually combined with a microscope, which allows for the selecting of the target cell. LSCE takes only a few seconds to extract a single cell each time, without complicated operation. This method will provide a valuable novel tool for studying biology on the single-cell level.
The concept of LSCE was proposed by Mao et al. for revealing the correlation between adhesion strength and viability at single-cell resolution [75]. They used LSCE to extract adhered U87-MG cells (U87) and human hepatoma (HepG2) cells cultured in a culture dish. The microscopy recorded the information of individual cells’ morphology and stained metabolites. The relationship among cell adhesion strength, cell morphologies as well as intracellular metabolites were explored, which revealed cell heterogeneity. Deepened on LSCE, Mao et al. clarified the cell-matrix adhesion strength on the single-cell level and revealed the effects of biomaterials on cell-matrix adhesion and heterogeneity of cell-matrix adhesion for adherent cell culture [76]. Moreover, LSCE also clarified the drug effects on the adhesion strength of individual circulating tumor cells (CTCs) on endothelial cells (ECs), which has broad prospects in drug screening for cancer therapy [77].
3.2.2.5 Electrical Traps
The cell membranes generally exhibit negative electronegativity at neutral pH, so the suspended cells are attracted toward the positive electrode. Cells have different charge, sizes, and masses, resulting different electric field force of cells; therefore, different cells can be distinguished. This is the working principle of electrophoresis (EP) [78]. However, the specificity of different cells in electrophoretic migration is not efficient. Electrical traps for single-cell capture generally include electric field-directed adhesion and dielectrophoresis (DEP) . The method of electric field-directed adhesion needs cell surface modification. Toriello et al. demonstrate a microfluidic cell capture system comprised of interdigitated gold electrodes covered in an oxide layer within the PDMS channel (Fig. 3.1h) [79]. The cell surface was modified with thiol functional groups by endogenous RGD receptors. The exposed gold pads bonded the single cell with modified thiol functional group through applying a driving electric potential. However, DEP could electrically capture cells without cell surface modification. DEP uses alternating current and generates a non-uniform electric field. DEP has been widely used for cell trap and further cell analysis. For example, Das et al. coated DEP electrodes on the microscope slide, which could collect different kinds of cells on the slide [80]. Thomas et al. designed concentric ring negative DEP (nDEP) for trapping single cell in the flow system [81]. In addition, DEP combined single-cell Raman spectra (SCRS) to capture single cells for downstream Raman detection [82, 83].
3.2.2.6 Optical Traps
Tightly focused laser beam acts as laser tweezers to trap single cell in aqueous solution. The laser beam could also act as an excitation source to generate a Raman spectrum of samples, known as laser tweezers Raman spectroscopy (LTRS) [84]. Laser tweezers could be integrated into microfluidic devices for transmission, identification, and simultaneous sorting of single cells [85, 86]. Adrian et al. presented the integrated optofluidic Raman-activated cell sorting (RACS) platforms for label-free cell sorting [87]. The specific single cells were identified by Raman spectroscopy and then captured and removed from cell suspensions (Fig. 3.1i). This method achieved automated sampling and high-throughput cell sorting. Another type of optical traps is optoelectronic tweezer , a combination of laser tweezers and DEP , which has been applied to manipulate and select single cells [88, 89].
3.3 Single-Cell Culture on Microfluidics
Many creative microfludic platforms have been proposed for single-cell culture. Except for allowing the collecting of single cells, microfluidic platforms for single-cell culture require sufficient space for cell division and growth, timely nutritional supplementation, and appropriate microenvironment with minimal shear stress or other damage stress. In addition, access to cells of interest should be considered for subsequent study. Here, we classified microfluidic platforms by two dichotomous characteristics: two-dimensional (2D) and three-dimensional (3D ). The methods and applications of single-cell culture carried out on these platforms have been described, with multiple types of cells including stem cells, tumor cells, immune cells, and so on.
3.3.1 Two-Dimensional Single-Cell Culture on Microfluidics
2D single-cell culture platforms based on microfluidics divided into two categories: closed and open microfluidics. Closed systems typically integrate hydrodynamic trapping structures and physical barriers into the platforms, which are isolated from the external physical stimulation. In contrast, open microfluidic systems , such as traditional microplate, microwells, and micropatterned surfaces, expose cells to the external physical, which allows direct access to single cells. Closed and open microfluidics with diverse designs have their own merits and demerits.
3.3.1.1 Closed Microfluidic Systems
Closed systems typically rely on hydrodynamic trapping structures and physical barriers. For example, Carlo et al. designed U-shaped hydrodynamic trapping structures on the bottom of the flow channel to trap single cells and cultured the cell in the original position for 24 h (Fig. 3.2) [90]. One of the important advantages of this single-cell culture array is controllable cell–cell interaction by both contact and diffusible elements.
To improve the single-cell capture efficiency, Kobel et al. optimized trap geometries with U-shaped hydrodynamic regions on the edge/wall of a microchannel (Fig. 3.3a) [91]. The single-cell capture efficiency of this device was nearly 100%. And this device allowed a long-term culture of individual non-adherent T-cell lymphoma cell in high throughput without a significant decrease in cell viability. Except U-shaped hydrodynamic trapping structures as efficient trap geometries, Lin et al. [92] fabricated sieve-like traps on adhesive protein micropatterns to capture single cells and cultured the trapped Hela cells on the micropatterns (Fig. 3.3b). The sieve-like trap positioning device was detachable, allowing easy access to individual cells of interest for subsequent manipulations. Moreover, the micropattern with multiple protein pattern clusters could adhere different cell populations, which provided a promising opportunity for studying cell–ECM interactions and cell–cell interactions.
To investigate single cell–cell contact, Frimat et al. reported a highly parallel microfluidic method combined differential fluidic resistance trapping with cellular valving (Fig. 3.4) [93]. They applied differential fluidic resistance to sequential single-cell arraying. Continuous single-cell arrangement of the second cell type relied on the reversal of the fluid. The trapped single-cell pairs could contact with each other through the aperture. This microfluidic system is promising to study homotypic/heterotypic co-culture at the level of a pair of single cells. In addition, the long-term microfluidic culture of mammalian cells was considered difficult for reduced growth rates and deviations from normal phenotypes [94]. Lecault et al. exploited a high-throughput microfluidic system with automated medium exchange to investigate hematopoietic stem cell (HSC) proliferation control at the single-cell level [95]. Closed systems have the advantages of protecting samples from contamination and solvent (e.g., medium) evaporation. However, the disadvantages are the limited accessibility to samples of interest and increased costs from the integrated components (e.g., pumps, tubes).
3.3.1.2 Open Microfluidic Systems
Microwell chips as a kind of common open microfluidic systems have been useful in single-cell culture and analysis [96]. It is easy to achieve large-scale single-cell capture and culture and get the information of single-cell metabolism. A large number of optimization methods for efficient single-cell capture were proposed [97, 98]. Cell suspensions are normally introduced manually into the microwells, and cells are randomly positioned in the wells. Cells outside the well are then washed away. Single non-adherent cells and adherent cell both could be collected by microwell chips, such as single epithelial cells [99], hematopoietic stem and progenitor cells [100], blood cells, and lymphocyte. It is more difficult to fabricate microfluidic devices for the investigation of non-adherent cells, because of restricting non-adherent cells to a known position in the microdevice. Cell seeding efficiency is the most important parameter to evaluate microwell chips. The number of wells containing cells and the number of wells containing single cells, corresponding to well occupancy and single-cell occupancy respectively, in addition, the shape, size, the number of the microwells and material are all parameters should be taken into consideration. Square, hexagonal, and round shapes are common for microwells. Larger wells provide enough space for long-term culture, while smaller wells are propitious to the collection of single cells and instant analysis (hours and days). The number of the microwells (density of wells) need considers two aspects, harboring many cells and communicating with other cells to keep normal functions (without those conditions, cells may subsequently change their normal functions, e.g., diminished viability). Various materials have been used for the fabrication of microwell chips, such as etching/drilling silicon and glass [101], polymer PDMS [102]. The surface of silicon and glass with favorable mechanical properties could be grinded and polished chemically to be suitable for cell adhesion. Besides the mechanical properties, the optical properties should be taken into consideration. Transparent materials have good satisfactory performance, which facilitates acquiring cell images, such as flat glass surfaces and conventional microscopic slips. Polymer PDMS is another material commonly applied for well-based single-cell chips with the advantages of easy fabrication, facilitated sealing onto other materials, good light transmittance, and biologically inert property. Open microwell chips allowed cell manipulation and handling.
Revzin et al. fabricated microwells composed of PEG hydrogel walls and square glass attachment pads and modified the glass pads with T-cell specific anti-CD5 to capture T-lymphocytes with the single-cell occupancy of ~95% (Fig. 3.5a) [103]. They used laser capture microdissection (LCM) for the retrieval of individual cells from the microwell array, which was prepared for downstream experiments (e.g., genomic or proteomic analysis). Tokimitsu et al. picked the single antigen-specific B-cells from cell-sized microwells by using micromanipulation [104]. Microwell chips should be designed for further cell study with the possibility to retrieve samples of interest. Since most of the existing analysis techniques for cell retrieval are suitable for open microfluidic systems , thus it can be integrated with most microwell chips. Although most microwells are roofless, the wells can be sealed by the addition of a roof (e.g., a glass slide) to capture the cell in the wells from the external environment (Fig. 3.5b) [105]. The roof could be modified as a functionalized seal to pick the cells or the components out of the wells for further analysis, for example, the analysis of the secreted cytokines or antibodies [106].
Single-cell culture is required to confine single cells to a specific location on the substrate. Chemically micropatterned surface with cytophilic and cytophobic regions (promoting and suppressing cell growth) is another strategy existing for controlling the position of single cells. Micropatterned surfaces for single-cell culture have been helpful in studying the effects of cell morphology and microenvironment on movement, migration, proliferation, and differentiation of cells [107]. Similar to the most microwell-based assay, the micropatterned surface is also a kind of open microfluidic systems . Various methods have been used to pattern cytophilic and cytophobic chemicals on the substrate for single-cell culture and analysis. Cheng et al. combined plasma-assisted surface chemical modification, soft lithography, and protein-induced surface activation to accomplish surface patterning for single-cell culture (Fig. 3.6) [108]. A polydimethylsiloxane membrane mask was put on the polystyrene film. The patterning was formed by oxygen plasma treatment which could produce hydrophilic areas. Then, the patterned film was incubated with either Pluronic F108 solution or a mixture of Pluronic F108 solution and fibronectin. Protein loading enhanced selective cell attachment on patterned dishes. Long-term (>2 weeks) single-cell culture experiments showed the influence of surface patterning on both cell and nucleus shape and also confirmed the stability of the produced single-cell molds in serum medium. Ye et al. proposed an alternative approach based on micromolding in capillaries (MIMIC) to achieve chemical surface patterning for single-cell culture [109]: the PDMS micropattern with grid pattern was placed on the cytophilic substrate; cytophobic chemicals filled the microchannels under capillary action; discrete cell adhesion regions surrounded by cytophobic chemicals; removed the PDMS mold and seeded cells in the cytophilic pattern. The micropatterned surface for single-cell culture could be used to investigate the physiological activity of specific cells and employed for further analysis (e.g., drug screening).
3.3.2 Three-Dimensional Single-Cell Culture on Microfluidics
Traditional 2D single-cell culture models generally suspend cells freely in culture medium or randomly allowed cells adhere onto the support surface. Although 2D models have advantages of low cost, reproducibility, accessibility, and easy operation, they are insufficient to recapitulate physiological systems. The poor physiological correlation of 2D models may cause misunderstandings of the cell proliferation , differentiation, cytotoxicity, and metabolism, leading to inaccurate prediction of in vivo behaviors. In addition, most two-dimensional models are faced with some common limitations, such as nutrient depletion, toxin accumulation, influence of the complicated fluid shear, and random cellular migration. Thus, three-dimensional (3D ) single-cell culture microfluidic platform mimics the real 3D microenvironment in vivo to reflect cell function and the geometry of tissues. Large amount of microfluidic devices featuring long-term 3D single-cell culture has sprung up. Microgel provides a promising strategy for single-cell 3D culture. Microgel allows long-term single-cell analysis, including heterogeneity of cellular proliferation, differentiation, and drug cytotoxicity. With the advantages of repeatability, flexibility, and highly controllable supporting microenvironment, cells could be independently packaged, cultured, monitored, or manipulated in the microgel . The structures of microgel for single-cell 3D culture are generally divided into three types: hydrogel droplets, microgel column, and microgel arrays.
In order to produce homogeneous hydrogel droplets for encapsulation of single cell, Utech et al. proposed a novel method to form alginate microgels in a highly controlled manner [110]. They added acetic acid into the continuous oil phase to dissociate Ca2+-EDTA into Ca2+, thus releasing Ca2+ to react with alginate chains (Fig. 3.7i). They used RGD-functionalized alginate to encapsulate single mesenchymal stem cells , as RGD provided integrin binding sites for cell attachment. The results showed that 25% of the droplets generated carrying single cells, while the majority of drops (70%) remained an empty and very small number of drops in contained more than one cell (Fig. 3.7ii). Improving the single-cell encapsulation efficiency has always been one of the challenges of microgel droplet technology. The encapsulated cells in the droplets kept high viability for 2 weeks. The cells proliferate inside the microstructure and maintained spherical morphology (Fig. 3.7iii). Similarly, Dolega et al. presented a method for 3D single-cell culture based on a flow-focusing microfluidic system that encapsulates epithelial cells in matrigel beads to analyze clonal acinar development [111]. They cultured single prostatic and breast cells in each individual bead, and the cells proliferated and differentiated into a single acinus per bead. Compared to conventional protocols of bulky 3D culture, this method generated more uniform acini population and recorded acinar development from the very first division to the final stage. Moreover, it provided easy recovery of 3D structures for further analysis, such as the combination with large particle FACS, fundamental genomics.
Hydrogel microcolumn inside microchannels is also one of the microgel types for single-cell culture. Liu et al. developed an photopolymerization approach for controlled encapsulation of single cells in poly(ethylene glycol) diacrylate (PEG-DA) precursor [112]. As shown in Fig. 3.8, cells were suspended in PEG-DA precursor solution in the microchannel. A fluorescence microscope projection provided UV source to generate photopolymerization of PEG-DA precursor. The photopolymerization only occurred where the target cell suspended. Thus, the cell was chosen for encapsulation inside the hydrogel microstructures. The remaining freestanding cells in uncross-linked precursor were removed by TE buffer. Single cells encapsulated inside the microgel structure remained their viability for hours to days, which allowed long-term culture for further single-cell analysis [113].
Microgel arrays provide a feasible and high-throughput 3D mode for long-term single-cell culture. Large-scale microgel arrays carry out high-throughput single-cell analysis, and each unit of the arrays acts as an individual 3D cell culture room. Guan et al. developed a microcollagen gel array (μCGA) for 3D single-cell culture [114]. The process of μCGA fabrication was not complex: soaked the PDMS microwell array in the mixture of collagen solution and cell suspension; pressed the PDMS membrane with cells in-wells with another flat PDMS membrane; removed the cover layer after gelatinization; placed the fabricated μCGA with cells encapsulated into culture medium; added fresh medium to supply nutrients for cell growth and proliferation (Fig. 3.9i). In order to analyze the cell proliferation heterogeneity under 3D culture conditions, they continuously recorded the proliferation of some random single cells for 11 days, which showed growth of five typical cell clones with different proliferation ability during the whole culture process (Fig. 3.9ii). Furthermore, they retrieved the cells of interest in μCGA by integration with a microscale manipulation platform, as open culture conditions based on the microwell array allowed for more convenient retrieve of the target cells.
3.4 Analysis Techniques for Single-Cell Culture
The final step for single-cell culture is to characterize and screen cell biophysical and biochemical information, such as morphology, behaviors, metabolism, and so on. A variety of technologies has been emerging for obtaining the single-cell information during a series of processes, including single-cell capture, single-cell short-term/long-term culture, single-cell lysis, and drug monitoring. Here, we covered optical, electrochemical, and mass spectrometric techniques for physical, chemical, molecular biology, and gene analysis of single cells.
3.4.1 Optical Characterization Techniques
Optical characterization is very commonly used in single-cell analysis, due to the direct visual observation of cells. Fluorescence methods play a vital role in optical characterization , providing images with high contrast and chemical-specific properties. Through fluorescence imaging, it is easy to reveal the distribution of intracellular molecules and dynamic states of biological processes, and the trace of biomolecules. However, fluorescence imaging requires the target molecule to generate fluorescence. As to the non-luminescent molecule, fluorescent labeling or chemical modification is needed, which limits the use of fluorescence methods. Thus, label-free optical methods without the need for fluorescent labels or staining have been emerging. In addition, super-resolution microscopy allows the resolution down to 10 nm [115] and permits the observation of the submicroscopic structure of cells. It is necessary to briefly introduce recent advances in optical characterization techniques, including fluorescence techniques and label-free methods.
3.4.1.1 Fluorescence Techniques
Fluorescence resonance energy transfer (FRET) as a specialized fluorescence technique is extensively applied to biochemical reactions. The principle of FRET is that the energy transfer from an excited donor fluorophore to an acceptor molecule leading to the acceptor emitting fluorescence. FRET sensors with different excitation and emission wavelengths can be designed to analyze parameters in parallel. Ng et al. designed multi-color Förster resonance energy transfer (FRET)-based enzymatic substrates in a microfluidics platform to measure multiple specific protease activities from water-in-oil droplets encapsuling single cells [116]. They first mixed suspended cells with multiple modified FRET substrates for cell encapsulation in water-in-oil droplets (Fig. 3.10a). The secreted proteases cleaved multi-color FRET substrates to yield multiple fluorescent signals in cell-encapsulated droplets (Fig. 3.10a). They successfully used four FRET sensors with different excitation and emission wavelengths to measure different metalloproteinases of several breast cancer cell lines on the microdroplet-based platform (Fig. 3.10a). Metalloproteinases (MMPs) have been known as important biomarkers of cancer diagnosis and treatment. For example, MMP9 degrades the basement membrane of the extracellular matrix (ECM), which facilitates cancer cell invasion and metastasis [117]. Metalloproteinases were also regarded as the target in research by Son et al., who designed a micropatterned photodegradable hydrogel array integrated with reconfigurable microfluidics to enable cell-secreted metalloproteinases analysis and specific cell retrieval at the single-cell level. They also monitored the activity of protease molecules secreted from single cells through FRET peptides entrapped inside microfabricated compartments. Moreover, the gel islands could be degraded by UV exposure, which easily allowed to release specific single cells of interest.
Fluorescence in situ hybridization (FISH) relies on fluorescence labeled nucleic acid probes to localize specific sequences of DNA or RNA molecules in single cells or tissues. FISH has many advantages, such as high economy and safety without radioactive isotopes, high stability and specificity, simultaneous detection of multiple sequences through different colors displayed in the same nucleus, yet with the disadvantages of limited throughput. FISH is widely applied to identify heterogeneities in gene expression within single cells or tissues samples. For example, Perez-Toralla et al. presented a protocol for quantitative characterization of ERBB2 gene by FISH based on the microfluidic chip made of cyclic olefin (Fig. 3.10b) [118]. This protocol allowed cell immobilization with minimal dead volume and performed characterization in the liquid phase. The target ERBB2 gene as a biomarker for the monitoring of HER2+ breast cancer progression was measured quantitatively with a tenfold reduction of sample consumption and decreases the assay time by a factor of two [118].
Fluorescent super-resolution microscopy (SRM) breaks through the original optical far-field diffraction limit in principle. With the help of fluorescent molecules, it could exceed the limit of optical resolution and reach nanometer resolution. This technology is widely applied in biology, chemistry, medicine, and so on. SRM can be used to track target intracellular molecules in single-cell research. Spectacular two- and three-dimensional images of subcellular components have been obtained by SRM techniques. Although SRM could provide images with high resolution, it remains challenging to obtain multiplexed images for a large number of distinct target species. Jungmann et al. used the transient binding of short fluorescently labeled oligonucleotides (DNA-PAINT, a variation of point accumulation for imaging in nanoscale topography) for images with ultra-high resolution that achieves sub-10 nm spatial resolution in vitro on synthetic DNA structures [119]. They experimentally demonstrated spectacular imaging of proteins in fixed cells (Fig. 3.10c).
3.4.1.2 Label-Free Optical Methods
Laser tweezers Raman spectroscopy (LTRS) is promising in the label-free analysis of individual cells, particularly suited to the research of living cells, because it allows cells to be captured and analyzed in an aqueous environment. However, the most reported LTRS methods involved manual trapping of cells, which is time consuming. Casabella et al. proposed a LTRS with an automated microfluidic platform for single-cell Raman spectroscopy (Fig. 3.10d) [120]. They introduced simple microfluidic channel to realize an alternating flow: cells trapped by optical tweezers during low flow intervals and successfully removed once fluid flow increases. During each capture period, Raman spectroscopy was used to measure individual cells. Results showed discrimination in the Raman signals of live epithelial prostate cells and lymphocytes. This method improved the throughput and reduced the manual work of the single-cell Raman measurements.
Surface plasmon resonance imaging (SPRi) is also a popular method used for the label-free measurement of biomolecular interactions. Stojanovic et al. applied SPRi to screen and quantify antibody production from individual hybridoma cells (Fig. 3.10e) [121]. The cells from a hybridoma produced monoclonal antibodies recognizing epithelial cell adhesion molecule (EpCAM) that was preimmobilized on the SPR sensor surface. Through this method, an excreted antibody from individual cells ranged from 0.02 to 1.19 pg per cell per hour.
Interferometric scattering microscopy (iSCAT) is another optical microscopy technique without fluorescent labels. The principle of iSCAT is that light is scattered by an object leading to the change of light intensity based on interference with a reference light field. Agnarsson et al. developed evanescent light-scattering microscopy for label-free interfacial imaging [122]. The core technology of the evanescent light-scattering microscopy is a waveguide chip consisting of a flat silica core embedded in a symmetric organic cladding with a refractive index matching that of water (Fig. 3.10f). Measurements and theoretical analysis showed that the size of single surface-bound lipid vesicles could be characterized by the light-scattering signals without employing fluorescent lipids as labels.
Live cell tomography is another new label-less super-resolution microscopy technique allowing direct imaging of unstained living biological specimens. Cotte et al. recently presented this microscopic method that 3D imaging of living cells with a resolution less than 100 nm could be obtained by phase contrast of unlabeled single cells [123]. Through this method, they realized longtime neuronal observations for synaptic remodeling in 3D (Fig. 3.10g) and direct study of bacteria (Escherichia coli).
3.4.2 Electrochemical Analysis
In recent years, electrochemical analysis has been applied in the field of single cells. Electrochemistry analysis can be integrated on a miniaturized platform, and microelectrodes are suitable for miniaturization of signal acquisition. Microelectrodes can be easily fabricated on various material substrates (e.g., polymers, silicon, or glass). A wide range of electrochemically active molecules could be detected by microelectrodes. Thus, microelectrodes are very useful in the quantitative measurement of neuronal communication and related neuroscience research. One of the hotspots is the fabrication of ultrasmall electrodes for measuring the neurotransmitter’s release of single vesicle from living cells. Anderson et al. used carbon fiber microelectrodes to penetrate into individual cells and cell nuclei. The changes of electrode impedance with cell and nuclear penetration were measured [124]. They also monitored transmitter release from single vesicles of individual cells. However, the carbon fiber microelectrodes have to be punched into cells, which requires high experimental operation skills and is very low in throughput. In order to improve the throughput and reduce the cell damage, Robinson et al. fabricated vertical nanowire electrode array (VNEA) for parallel electrical interfacing to multiple mammalian neurons . Depending on VNEA, neuronal activity of rat cortical neurons was intracellularly recorded and stimulated. It was also possible to map multiple individual synaptic connections (Fig. 3.10h) [125]. Another common application of electrochemical methods at the single-cell level is to monitor oxidative stress. Jeffrey E. Dick recently presented a macroscopic setup for the detection of reactive oxygen species (ROS) at single cells level [126]. He monitored the consumption of a single cell’s contents upon its collisions with a microelectrode under the presence of surfactants. He obtained large difference with two orders of magnitude between acute lymphoblastic lymphoma T-cells and healthy thymocytes .
Electrochemical analysis is sensitive. The introduction of the enzymatic labeling amplification step further increases the detection limit and gets faster and more reliable signals. For example, Safaei et al. [127] used velocity valley (VV) chip to capture CTCs tagged with magnetic nanoparticles modified with the anti-EpCAM antibody. And then, they enzymatically labeled alkaline phosphatase on the cells, which catalyzed the reaction of paminophenyl phosphate to an electrochemically active reagent p-aminophenol to be measured. This method shows sensitive performance in analyzing the whole blood samples. The above methods are all based on fixed electrodes, but if the electrode is freely movable, it can be used for the two-dimensional scanning of the electrochemical properties of the underneath samples. Scanning electrochemical microscopy (SECM) is based on this principle. Koch et al. combined SECM with fast-scan cyclic voltammetry (CV) for the simultaneous measurement of impedance and amperometric current of PC12 cells [128]. Amperometric signals were converted into topographical images, because the current density depended on the distance between the tip and the cell (Fig. 3.10i). CV revealed the spatial distribution of minimum oxygen consumption. This method made it possible to achieve chemically and spatially resolved measurements along with imaging topography. However, one unavoidable disadvantage of the electrochemical method is that testing many substances simultaneously is difficult.
3.4.3 Mass Spectrometric Analysis
Mass spectrometry is a powerful tool for single-cell analysis, with the advantages of high sensitivity, simultaneous detection of multiple substances, and the ability to structure the molecules of interest. In particular, mass spectrometry can differentiate hundreds of biological molecules from the sample without labeling, thus making it attractive. However, low abundant cellular targets with any labels are still challenging in single-cell analysis through mass spectrometry. This is why most studies focus on highly abundant metabolites. Moreover, quantification by mass spectrometry generally needs internal references. Various ionization methods have been developed used for desorption/ionization of different kinds of samples in single-cell analysis field, such as electrospray/nano-electrospray ionization (ESI/Nano-ESI), laser ablation/laser desorption ionization (LA/LDI), and secondary ionization mass spectrometry (SIMS). These ionization methods provide multiple choices to ionize and analyze a wide range of substances, such as nucleotide, peptides, proteins, carbohydrates, esters, and small molecular metabolites.
3.4.3.1 Electrospray Ionization MS
Electrospray ionization MS (ESI-MS) is very suitable to investigate the samples with small volumes. In order to be supplied to ESI-MS, the single cell must be lysate. Huang et al. reported a cell ordering platform induced by Dean flow to isolate single cells from cellular suspension. The platform has been integrated with ESI-MS (Fig. 3.10j) [129]. In this method, individual cells were lysate one after another and then analyzed ESI-MS to identify a single-cell into a subpopulation by lipid profiling. This platform made cells in suspension evenly distributed and significantly improved the efficiency of single-cell mass spectrometry. Lipid is the high abundant target to detect, and some researches more focused on other low abundant targets. For example, Gong et al. developed a probe ESI-MS spectrometry with capillary microsampling to monitor single living leave cells and analyzed the cellular stress of healthy and damaged plant leave cells [130]. The results showed the clear differences in the levels of abscisic acid as compared to other MS analysis methods. However, this method had a disadvantage that the microneedles might hamper normal cellular functions.
3.4.3.2 Matrix-Assisted Laser Desorption/Ionization MS
Matrix-assisted laser desorption/ionization MS (MALDI-MS) utilizes the power of a laser to make the sample embedded in matrix crystals desorption and ionization. This is a soft ionization technique, thus the majority of molecules maintain their original size and weight without fragment. MALDI-MS was employed in single-cell analysis over the last few years. Many efforts have been made to improve the spatial resolution and decrease the interference from the matrix material. For example, Krismer et al. screened different strains of Chlamydomonas reinhardtii through MALDI-MS [131]. They embedded microwells to the stainless steel plate for separating individually spotted cells. And then, they fast freezed the samples and deposited the matrix to obtain the mass spectra of single cells (Fig. 3.10k). The MS results showed that the native strain with two different kinds of chlorophyll could be distinguished from the mutant strain lacking one of these chlorophyll subtypes.
3.4.3.3 Secondary Ion Mass Spectroscopy
Secondary ion mass spectroscopy (SIMS) provides a tool to enable and analyze the composition of solid surfaces and thin films. The principle of SIMS is that MS investigates the release of secondary ions from the surface sputtered with an ion beam. Although SIMS has the advantages with the low detection limit and the high spatial resolution, yet samples must be placed in the ultra-high vacuum for analysis, thus inhibiting the study of living organisms. Bobrowska et al. have published a protocol of single cells preparation for the time of flight secondary ion mass spectrometry (TOF-SIMS), including the steps for pretreatment of single-cell samples (fixation, washing, dehydration) and detection by SIMS [132]. TOF-SIMS has frequently been employed for single-cell analysis. For example, it was applied to uncover differences among breast cancer cell lines, and the results showed an 18-carbon chain fatty acid only presenting in the BT-474 cell line [133]. In addition, 3D TOF-SIMS allowed for multiple MS images taken at different heights, thus creating Z-scans for samples [134].
3.5 Conclusions and Outlooks
Single-cell research is a newly established field and develops rapidly. The rapid development of this field is inseparable from the contributions of emerging microfluidic technologies. Although conventional methods for single-cell capture, in particularly FACS, are mature and used routinely, they have one or more following disadvantages: high laboratory skills, low throughput, lack of visual inspection, and the risk of introducing pollution. Emerging microfluidics for single-cell capture overcomes these shortcomings and becomes increasingly mature. However, high-throughput and efficient single-cell separation are still an important goal of microfluidics. In addition, open microfluidic techniques (e.g., microwells) provide convenience for subsequent single-cell manipulation, but also increase the risk of sample contamination; therefore, the work of microfluidic methods for single cell sorting needs continuous development. Microfluidic designers must take full account of the throughput of cell sorting, the efficiency of single-cell separation, the subsequent extraction of single cells, and sample contamination, and visual observation of cells, easy operation, and low cost are other aspects that need to be taken into account.
Single-cell culture is especially important for rare cells. Multiple generations from single-cell starting points are different from the cell populations derived randomly. Single-cell dynamic culture process provides real-time evidences for better understanding of cell behaviors and interrogation of cell phenotypes. Compared to microfluidic methods for single-cell capture, single-cell culture microfluidic platforms need bigger space for single-cell division, timely nutrition supplement for long-term cell growth, and minimum shear stress or other damage stresses. In addition, the single-cell culture platform should consider the easy acquisition of cells of interest. Similar to microfluidic single-cell capture, microfluidics provides 2D and 3D molds for single-cell culture. Although 2D models have the advantages with low cost, reproducibility, and easy operation, 3D molds simulate cell function and the geometry of tissues and organs, which is an irreplaceable advantage of 2D models. We also briefly presented advanced analytical technologies to identify heterogeneities among large cell populations, including optical characterization techniques, electrochemical analysis techniques, and mass spectrometric analysis techniques. Optical characterization techniques based on microscopy and fluorescence have the advantages of high sensitivities and spatial resolutions. Electrochemical and mass spectrometric techniques play unique roles in single-cell analysis in recent years, due to the detection of label-free biomolecules. Single-cell analysis methods are not limited to these technologies, and more novel technologies are emerging to extend our knowledge of cellular processes. However, the reliability of quantitative results is often questioned due to the limited accuracy of many analytical approaches and the lack of references. In addition, combining two or more analysis technologies can provide complementary information, because of the complexity of cellular processes. We believe that further development of microfluidics and innovations of analytical instruments will cast light on single-cell heterogeneity and potentially promote the development of individual therapy.
References
Faley SL, Copland M, Wlodkowic D, Kolch W, Seale KT, Wikswo JP, Cooper JM (2009) Microfluidic single cell arrays to interrogate signalling dynamics of individual, patient-derived hematopoietic stem cells. Lab Chip 9(18):2659–2664. https://doi.org/10.1039/b902083g
Rao CV, Wolf DM, Arkin AP (2002) Control, exploitation and tolerance of intracellular noise. Nature 420(6912):231–237. https://doi.org/10.1038/nature01258
Fukano Y, Tsuyama N, Mizuno H, Date S, Takano M, Masujima T (2012) Drug metabolite heterogeneity in cultured single cells profiled by pico-trapping direct mass spectrometry. Nanomedicine 7(9):1365–1374. https://doi.org/10.2217/nnm.12.34
Jiang Y, Zhao H, Lin YQ, Zhu NN, Ma YR, Mao LQ (2010) Colorimetric detection of glucose in rat brain using gold nanoparticles. Angew Chem Int Edit 49(28):4800–4804. https://doi.org/10.1002/anie.201001057
Kent DG, Copley MR, Benz C, Wohrer S, Dykstra BJ, Ma E, Cheyne J, Zhao Y, Bowie MB, Zhao Y, Gasparetto M, Delaney A, Smith C, Marra M, Eaves CJ (2009) Prospective isolation and molecular characterization of hematopoietic stem cells with durable self-renewal potential. Blood 113(25):6342–6350. https://doi.org/10.1182/blood-2008-12-192054
Dykstra B, Kent D, Bowie M, McCaffrey L, Hamilton M, Lyons K, Lee SJ, Brinkman R, Eaves C (2007) Long-term propagation of distinct hematopoietic differentiation programs in vivo. Cell Stem Cell 1(2):218–229. https://doi.org/10.1016/j.stem.2007.05.015
Yamazaki S, Nakauchi H (2009) Insights into signaling and function of hematopoietic stem cells at the single-cell level. Curr Opin Hematol 16(4):255–258. https://doi.org/10.1097/MOH.0b013e32832c6705
Lasken RS, McLean JS (2014) Recent advances in genomic DNA sequencing of microbial species from single cells. Nat Rev Genet 15(9):577–584. https://doi.org/10.1038/nrg3785
Pardal R, Clarke MF, Morrison SJ (2003) Applying the principles of stem-cell biology to cancer. Nat Rev Cancer 3(12):895–902. https://doi.org/10.1038/nrc1232
Prasetyanti PR, Medema JP (2017) Intra-tumor heterogeneity from a cancer stem cell perspective. Mol. Cancer 16(1):41. https://doi.org/10.1186/s12943-017-0600-4
Boiani M, Scholer HR (2005) Regulatory networks in embryo-derived pluripotent stem cells. Nat Rev Mol Cell Biol 6(11):872–884. https://doi.org/10.1038/nrm1744
Lutolf MP, Doyonnas R, Havenstrite K, Koleckar K, Blau HM (2009) Perturbation of single hematopoietic stem cell fates in artificial niches. Integr Biol (Camb) 1(1):59–69. https://doi.org/10.1039/b815718a
Singhvi R, Kumar A, Lopez GP, Stephanopoulos GN, Wang DI, Whitesides GM, Ingber DE (1994) Engineering cell shape and function. Science 264(5159):696–698. https://doi.org/10.1126/science.8171320
Parker KK, Brock AL, Brangwynne C, Mannix RJ, Wang N, Ostuni E, Geisse NA, Adams JC, Whitesides GM, Ingber DE (2002) Directional control of lamellipodia extension by constraining cell shape and orienting cell tractional forces. Faseb J 16(10):1195–1204. https://doi.org/10.1096/fj.02-0038com
Chua CW, Shibata M, Lei M, Toivanen R, Barlow LJ, Bergren SK, Badani KK, Mc Kiernan JM, Benson MC, Hibshoosh H, Shen MM (2014) Single luminal epithelial progenitors can generate prostate organoids in culture. Nat Cell Biol 16(10):951–961. https://doi.org/10.1038/ncb3047
Gracz AD, Williamson IA, Roche KC, Johnston MJ, Wang F, Wang Y, Attayek PJ, Balowski J, Liu XF, Laurenza RJ, Gaynor LT, Sims CE, Galanko JA, Li L, Allbritton NL, Magness ST (2015) A high-throughput platform for stem cell niche co-cultures and downstream gene expression analysis. Nat Cell Biol 17(3):340–349. https://doi.org/10.1038/ncb3104
Pushkarsky I, Tseng P, Black D, France B, Warfe L, Koziolwhite CJ, Jester WF, Trinh RK, Lin J, Scumpia PO (2018) Elastomeric sensor surfaces for high-throughput single-cell force cytometry. Nat Biomed Eng 2(2):124–137. https://doi.org/10.1038/s41551-018-0193-2
Roman GT, Chen YL, Viberg P, Culbertson AH, Culbertson CT (2007) Single-cell manipulation and analysis using microfluidic devices. Anal Bioanal Chem 387(1):9–12. https://doi.org/10.1007/s00216-006-0670-4
Titmarsh D, Hidalgo A, Turner J, Wolvetang E, Cooper-White J (2011) Optimization of flowrate for expansion of human embryonic stem cells in perfusion microbioreactors. Biotechnol Bioeng 108(12):2894–2904. https://doi.org/10.1002/bit.23260
Kim L, Toh YC, Voldman J, Yu H (2007) A practical guide to microfluidic perfusion culture of adherent mammalian cells. Lab Chip 7(6):681–694. https://doi.org/10.1039/b704602b
Wu MH, Huang SB, Lee GB (2010) Microfluidic cell culture systems for drug research. Lab Chip 10(8):939–956. https://doi.org/10.1039/b921695b
Chen HY, Cornwell J, Zhang H, Lim T, Resurreccion R, Port T, Rosengarten G, Nordon RE (2013) Cardiac-like flow generator for long-term imaging of endothelial cell responses to circulatory pulsatile flow at microscale. Lab Chip 13(15):2999–3007. https://doi.org/10.1039/c3lc50123j
Gao D, Liu HX, Jiang YY, Lin JM (2012) Recent developments in microfluidic devices for in vitro cell culture for cell-biology research. Trac-Trend. Anal Chem 35(35):150–164. https://doi.org/10.1016/j.trac.2012.02.008
Titmarsh DM, Hudson JE, Hidalgo A, Elefanty AG, Stanley EG, Wolvetang EJ, Cooper-White JJ (2012) Microbioreactor arrays for full factorial screening of exogenous and paracrine factors in human embryonic stem cell differentiation. PLoS ONE 7(12):e52405. https://doi.org/10.1371/journal.pone.0052405
Lee LM, Liu AP (2014) The application of micropipette aspiration in molecular mechanics of single cells. J Nanotechnol Eng Med 5(4):0408011–0408016. https://doi.org/10.1115/1.4029936
Datta S, Malhotra L, Dickerson R, Chaffee S, Sen CK, Roy S (2015) Laser capture microdissection: big data from small samples. Histol Histopathol 30(11):1255–1269. https://doi.org/10.14670/HH-11-622
Espina V, Wulfkuhle JD, Calvert VS, VanMeter A, Zhou W, Coukos G, Geho DH, Petricoin EF 3rd, Liotta LA (2006) Laser-capture microdissection. Nat Protoc 1(2):586–603. https://doi.org/10.1038/nprot.2006.85
Hu P, Zhang W, Xin H, Deng G (2016) Single cell isolation and analysis. Front Cell Dev Biol 4:116–127. https://doi.org/10.3389/fcell.2016.00116
Hodne K, Weltzien FA (2015) Single-cell isolation and gene analysis: pitfalls and possibilities. Int J Mol Sci 16(11):26832–26849. https://doi.org/10.3390/ijms161125996
Orfao A, RuizArguelles A (1996) General concepts about cell sorting techniques. Clin Biochem 29(1):5–9. https://doi.org/10.1016/0009-9120(95)02017-9
Andersson H, van den Berg A (2004) Microtechnologies and nanotechnologies for single-cell analysis. Curr Opin Biotechnol 15(1):44–49. https://doi.org/10.1016/j.copbio.2004.01.004
Livesey FJ (2003) Strategies for microarray analysis of limiting amounts of RNA. Brief Funct Genomic Proteomic 2(1):31–36. https://doi.org/10.1093/bfgp/2.1.31
Gross A, Schoendube J, Zimmermann S, Steeb M, Zengerle R, Koltay P (2015) Technologies for single-cell isolation. Int J Mol Sci 16(8):16897–16919. https://doi.org/10.3390/ijms160816897
Brehmstecher BF, Johnson EA (2004) Single-Cell Microbiology: tools, technologies, and applications. Microbiol Mol Biol R 68(3):538–559. https://doi.org/10.1128/mmbr.68.3.538-559.2004
Fink L, Kwapiszewska G, Wilhelm J, Bohle RM (2006) Laser-microdissection for cell type- and compartment-specific analyses on genomic and proteomic level. Exp Toxicol Pathol 57(6):25–29. https://doi.org/10.1016/j.etp.2006.02.010
Fink L, Bohle RM (2005) Laser microdissection and RNA analysis. Methods Mol Biol 293(293):167–185. https://doi.org/10.1385/1-59259-853-6:167
Adan A, Alizada G, Kiraz Y, Baran Y, Nalbant A (2017) Flow cytometry: basic principles and applications. Crit Rev Biotechnol 37(2):163–176. https://doi.org/10.3109/07388551.2015.1128876
Valet G (2003) Past and present concepts in flow cytometry: a European perspective. J Biol Regul Homeost Agents 17(3):213–222. https://doi.org/10.1089/104303403322124828
Davey HM, Kell DB (1996) Flow cytometry and cell sorting of heterogeneous microbial populations: the importance of single-cell analyses. Microbiol Rev 60(4):641–696. https://doi.org/10.1006/mpat.1996.0079
Joensson HN (2012) Droplet Microfluidics—A Tool for Single-Cell Analysis. Angew Chem Int Edit 51(49):12176–12192. https://doi.org/10.1002/anie.201200460
Guo MT, Rotem A, Heyman JA, Weitz DA (2012) Droplet microfluidics for high-throughput biological assays. Lab Chip 12(12):2146–2155. https://doi.org/10.1039/c2lc21147e
Kã¶Ster S, Angil㨠FE, Duan H, Agresti JJ, Wintner A, Schmitz C, Rowat AC, Merten CA, Pisignano D, Griffiths AD (2008) Drop-based microfluidic devices for encapsulation of single cells. Lab Chip 8(7):1110–1115. https://doi.org/10.1039/b802941e
Wang BL, Ghaderi A, Zhou H, Agresti J, Weitz DA, Fink GR, Stephanopoulos G (2014) Microfluidic high-throughput culturing of single cells for selection based on extracellular metabolite production or consumption. Nat Biotechnol 32(5):473–478. https://doi.org/10.1038/nbt.2857
Chen F, Zhan Y, Geng T, Lian H, Xu P, Lu C (2011) Chemical transfection of cells in picoliter aqueous droplets in fluorocarbon oil. Anal Chem 83(22):8816–8820. https://doi.org/10.1021/ac2022794
Lu H, Caen O, Vrignon J, Zonta E, El Harrak Z, Nizard P, Baret JC, Taly V (2017) High throughput single cell counting in droplet-based microfluidics. Sci Rep 7(1):1366. https://doi.org/10.1038/s41598-017-01454-4
Mazutis L, Gilbert J, Ung WL, Weitz DA, Griffiths AD, Heyman JA (2013) Single-cell analysis and sorting using droplet-based microfluidics. Nat Protoc 8(5):870–891. https://doi.org/10.1038/nprot.2013.046
Brouzes E, Medkova M, Savenelli N, Marran D, Twardowski M, Hutchison JB, Rothberg JM, Link DR, Perrimon N, Samuels ML (2009) Droplet microfluidic technology for single-cell high-throughput screening. Proc Natl Acad Sci USA 106(34):14195–14200. https://doi.org/10.1073/pnas.0903542106
Macosko EZ, Basu A, Satija R, Nemesh J, Shekhar K, Goldman M, Tirosh I, Bialas AR, Kamitaki N, Martersteck EM, Trombetta JJ, Weitz DA, Sanes JR, Shalek AK, Regev A, McCarroll SA (2015) Highly Parallel genome-wide expression profiling of individual cells using nanoliter droplets. Cell 161(5):1202–1214. https://doi.org/10.1016/j.cell.2015.05.002
Klein AM, Mazutis L, Akartuna I, Tallapragada N, Veres A, Li V, Peshkin L, Weitz DA, Kirschner MW (2015) Droplet barcoding for single-cell transcriptomics applied to embryonic stem cells. Cell 161(5):1187–1201. https://doi.org/10.1016/j.cell.2015.04.044
Geng T, Novak R, Mathies RA (2014) Single-cell forensic short tandem repeat typing within microfluidic droplets. Anal Chem 86(1):703–712. https://doi.org/10.1021/ac403137h
Zhang W, Li N, Koga D, Zhang Y, Zeng H, Nakajima H, Lin JM, Uchiyama K (2018) Inkjet printing based droplet generation for integrated online digital polymerase chain reaction. Anal Chem 90(8):5329–5334. https://doi.org/10.1021/acs.analchem.8b00463
Fu Y, Li C, Lu S, Zhou W, Tang F, Xie XS, Huang Y (2015) Uniform and accurate single-cell sequencing based on emulsion whole-genome amplification. Proc Natl Acad Sci USA 112(38):11923–11928. https://doi.org/10.1073/pnas.1513988112
Leung K, Zahn H, Leaver T, Konwar KM, Hanson NW, Page AP, Lo CC, Chain PS, Hallam SJ, Hansen CL (2012) A programmable droplet-based microfluidic device applied to multiparameter analysis of single microbes and microbial communities. Proc Natl Acad Sci USA 109(20):7665–7670. https://doi.org/10.1073/pnas.1106752109
Sjostrom SL, Bai Y, Huang M, Liu Z, Nielsen J, Joensson HN, Andersson Svahn H (2014) High-throughput screening for industrial enzyme production hosts by droplet microfluidics. Lab Chip 14(4):806–813. https://doi.org/10.1039/c3lc51202a
Zinchenko A, Devenish SRA, Kintses B, Colin PY, Fischlechner M, Hollfelder F (2014) One in a million: flow cytometric sorting of singlecell-lysate assays in monodisperse picolitre double emulsion dropletsfor directed evolution. Anal Chem 86(5):2526–2533. https://doi.org/10.1021/ac403585p
Lim SW, Tran TM, Abate AR (2015) PCR-activated cell sorting for cultivation-free enrichment and sequencing of rare microbes. PLoS ONE 10(1):e0113549. https://doi.org/10.1371/journal.pone.0113549
Jing TY, Ramji R, Warkiani ME, Han J, Lim CT, Chen CH (2015) Jetting microfluidics with size-sorting capability for single-cell protease detection. Biosens Bioelectron 66:19–23. https://doi.org/10.1016/j.bios.2014.11.001
Park SY, Wu TH, Chen Y, Teitell MA, Chiou PY (2011) High-speed droplet generation on demand driven by pulse laser-induced cavitation. Lab Chip 11(6):1010–1012. https://doi.org/10.1039/c0lc00555j
Edd JF, Carlo DD, Humphry KJ, Köster S, Irimia D, Weitz DA, Toner M (2008) Controlled encapsulation of single cells into monodisperse picoliter drops. Lab Chip 8(8):1262–1264. https://doi.org/10.1039/b805456h
Kemna EWM, Schoeman RM, Wolbers F, Vermes I, Weitz DA, van den Berg A (2012) High-yield cell ordering and deterministic cell-in-droplet encapsulation using Dean flow in a curved microchannel. Lab Chip 12(16):2881–2887. https://doi.org/10.1039/c2lc00013j
Margulies M, Egholm M, Altman WE, Attiya S, Bader JS, Bemben LA, Berka J, Braverman MS, Chen YJ, Chen Z, Dewell SB, Du L, Fierro JM, Gomes XV, Godwin BC, He W, Helgesen S, Ho CH, Irzyk GP, Jando SC, Alenquer ML, Jarvie TP, Jirage KB, Kim JB, Knight JR, Lanza JR, Leamon JH, Lefkowitz SM, Lei M, Li J, Lohman KL, Lu H, Makhijani VB, McDade KE, McKenna MP, Myers EW, Nickerson E, Nobile JR, Plant R, Puc BP, Ronan MT, Roth GT, Sarkis GJ, Simons JF, Simpson JW, Srinivasan M, Tartaro KR, Tomasz A, Vogt KA, Volkmer GA, Wang SH, Wang Y, Weiner MP, Yu P, Begley RF, Rothberg JM (2005) Genome sequencing in microfabricated high-density picolitre reactors. Nature 437(7057):376–380. https://doi.org/10.1038/nature03959
Theberge AB, Courtois F, Schaerli Y, Fischlechner M, Abell C, Hollfelder F, Huck WTS (2010) Cheminform abstract: microdroplets in microfluidics: an evolving platform for discoveries in chemistry and biology. Angew Chem Int Edit 41(45):5846–5868. https://doi.org/10.1002/anie.200906653
Liu C, Liu J, Gao D, Ding M, Lin JM (2010) Fabrication of microwell arrays based on two-dimensional ordered polystyrene microspheres for high-throughput single-cell analysis. Anal Chem 82(22):9418–9424. https://doi.org/10.1021/ac102094r
Huang L, Chen Y, Chen Y, Wu H (2015) Centrifugation-assisted single-cell trapping (CAScT) in a truncated cone-shaped microwell array (TCMA) chip for the real-time observation of cellular apoptosis. Anal Chem 87(24):12169–12176. https://doi.org/10.1021/acs.analchem.5b03031
Lin LI, Chao SH, Meldrum DR (2009) Practical, microfabrication-free device for single-cell isolation. PLoS ONE 4(8):e6710. https://doi.org/10.1371/journal.pone.0006710
Jang K, Xu Y, Tanaka Y, Sato K, Mawatari K, Konno T, Ishihara K, Kitamori T (2010) Single-cell attachment and culture method using a photochemical reaction in a closed microfluidic system. Biomicrofluidics 4(3):032208. https://doi.org/10.1063/1.3494287 1539
Shi X, Lin LI, Chen SY, Chao SH, Zhang W, Meldrum DR (2011) Real-time PCR of single bacterial cells on an array of adhering droplets. Lab Chip 11(13):2276–2281. https://doi.org/10.1039/c1lc20207c
Chen Q, Wu J, Zhang Y, Lin Z, Lin JM (2012) Targeted isolation and analysis of single tumor cells with aptamer-encoded microwell array on microfluidic device. Lab Chip 12(24):5180–5185. https://doi.org/10.1039/c2lc40858a
Karimi A, Yazdi S, Ardekani AM (2013) Hydrodynamic mechanisms of cell and particle trapping in microfluidics. Biomicrofluidics 7(2):21501. https://doi.org/10.1063/1.4799787
Carlo DD, Nima Aghdam A, Lee LP (2006) Single-cell enzyme concentrations, kinetics, and inhibition analysis using high-density hydrodynamic cell isolation arrays. Anal Chem 78(14):4925–4930. https://doi.org/10.1021/ac060541s
Chen H, Sun J, Wolvetang E, Cooper-White J (2015) High-throughput, deterministic single cell trapping and long-term clonal cell culture in microfluidic devices. Lab Chip 15(4):1072–1083. https://doi.org/10.1039/c4lc01176g
Lutz BR, Chen J, Schwartz DT (2005) Microscopic steady streaming eddies created around short cylinders in a channel: flow visualization and Stokes layer scaling. Phys Fluids 17(2):023601. https://doi.org/10.1063/1.1824137 3430
Lieu VH, House TA, Schwartz DT (2012) Hydrodynamic tweezers: impact of design geometry on flow and microparticle trapping. Anal Chem 84(4):1963–1968. https://doi.org/10.1021/ac203002z
Tanyeri M, Schroeder CM (2013) Manipulation and confinement of single particles using fluid flow. Nano Lett 13(6):2357–2364. https://doi.org/10.1021/nl4008437
Mao S, Zhang W, Huang Q, Khan M, Li H, Uchiyama K, Lin JM (2017) In situ scatheless cell detachment reveals connections between adhesion strength and viability at single-cell resolution. Angew Chem Int Edit 130(1):236–240. https://doi.org/10.1002/ange.201710273
Mao S, Zhang Q, Li H, Huang Q, Khan M, Uchiyama K, Lin J-M (2018) Measurement of cell–matrix adhesion at single-cell resolution for revealing the functions of biomaterials for adherent cell culture. Anal Chem 90(15):9637–9643. https://doi.org/10.1021/acs.analchem.8b02653
Mao S, Zhang Q, Li HF, Zhang WL, Huang QS, Khan M, Lin JM (2018) Adhesion analysis of single circulating tumor cells on a base layer of endothelial cells using open microfluidics. Chem Sci 9(39):7694–7699. https://doi.org/10.1039/c8sc03027h
Voldman J (2006) Electrical forces for microscale cell manipulation. Annu Rev Biomed Eng 8(8):425–454. https://doi.org/10.1146/annurev.bioeng.8.061505.095739
Toriello NM, Douglas ES, Mathies RA (2005) Microfluidic device for electric field-driven single-cell capture and activation. Anal Chem 77(21):6935–6941. https://doi.org/10.1021/ac051032d
Das CM, Becker F, Vernon S, Noshari J, Joyce C, Gascoyne PR (2005) Dielectrophoretic segregation of different human cell types on microscope slides. Anal Chem 77(9):2708–2719. https://doi.org/10.1021/ac048196z
Thomas RSW, Mitchell PD, Oreffo ROC, Morgan H (2010) Trapping single human osteoblast-like cells from a heterogeneous population using a dielectrophoretic microfluidic device. Biomicrofluidics 4(2):022806. https://doi.org/10.1063/1.3406951 11
Chrimes AF, Khoshmanesh K, Tang SY, Wood BR, Stoddart PR, Collins SS, Mitchell A, Kalantar-zadeh K (2013) In situ SERS probing of nano-silver coated individual yeast cells. Biosens Bioelectron 49(11):536–541. https://doi.org/10.1016/j.bios.2013.05.053
Zhang P, Ren L, Zhang X, Shan Y, Wang Y, Ji Y, Yin H, Huang WE, Xu J, Ma B (2015) Raman-activated cell sorting based on dielectrophoretic single-cell trap and release. Anal Chem 87(4):2282–2289. https://doi.org/10.1021/ac503974e
Chan JW, Esposito AP, Talley CE, Hollars CW, Lane SM, Huser T (2004) Reagentless identification of single bacterial spores in aqueous solution by confocal laser tweezers Raman spectroscopy. Anal Chem 76(3):599–603. https://doi.org/10.1021/ac0350155
Chen K, Qin YJ, Zheng F, Sun MH, Shi DR (2006) Diagnosis of colorectal cancer using Raman spectroscopy of laser-trapped single living epithelial cells. Opt Lett 31(13):2015–2017. https://doi.org/10.1364/Ol.31.002015
Singh GP, Volpe G, Creely CM, Grötsch H, Geli IM, Petrov D (2010) The lag phase and G1 phase of a single yeast cell monitored by Raman microspectroscopy. J Raman Spectrosc 37(8):858–864. https://doi.org/10.1002/jrs.1520
Lau AY, Lee LP, Chan JW (2008) An integrated optofluidic platform for Raman-activated cell sorting. Lab Chip 8(7):1116–1120. https://doi.org/10.1039/B803598A
Pei YC, Ohta AT, Wu MC (2005) Massively parallel manipulation of single cells and microparticles using optical images. Nature 436(7049):370–372. https://doi.org/10.1038/nature03831
Huang KW, Wu YC, Lee JA, Chiou PY (2013) Microfluidic integrated optoelectronic tweezers for single-cell preparation and analysis. Lab Chip 13(18):3721–3727. https://doi.org/10.1039/c3lc50607j
Di Carlo D, Wu LY, Lee LP (2006) Dynamic single cell culture array. Lab Chip 6(11):1445–1449. https://doi.org/10.1039/b605937f
Kobel S, Valero A, Latt J, Renaud P, Lutolf M (2010) Optimization of microfluidic single cell trapping for long-term on-chip culture. Lab Chip 10(7):857–863. https://doi.org/10.1039/b918055a
Lin L, Chu YS, Thiery JP, Lim CT, Rodriguez I (2013) Microfluidic cell trap array for controlled positioning of single cells on adhesive micropatterns. Lab Chip 13(4):714–721. https://doi.org/10.1039/c2lc41070b
Frimat JP, Becker M, Chiang YY, Marggraf U, Janasek D, Hengstler JG, Franzke J, West J (2011) A microfluidic array with cellular valving for single cell co-culture. Lab Chip 11(2):231–237. https://doi.org/10.1039/c0lc00172d
Korin N, Bransky A, Dinnar U, Levenberg S (2009) Periodic “flow-stop” perfusion microchannel bioreactors for mammalian and human embryonic stem cell long-term culture. Biomed Microdevices 11(1):87–94. https://doi.org/10.1007/s10544-008-9212-5
Lecault V, Vaninsberghe M, Sekulovic S, Knapp DJ, Wohrer S, Bowden W, Viel F, McLaughlin T, Jarandehei A, Miller M, Falconnet D, White AK, Kent DG, Copley MR, Taghipour F, Eaves CJ, Humphries RK, Piret JM, Hansen CL (2011) High-throughput analysis of single hematopoietic stem cell proliferation in microfluidic cell culture arrays. Nat Methods 8(7):581–586. https://doi.org/10.1038/nmeth.1614
Lindstrom S, Andersson-Svahn H (2011) Miniaturization of biological assays—overview on microwell devices for single-cell analyses. Biochim Biophys Acta 1810(3):308–316. https://doi.org/10.1016/j.bbagen.2010.04.009
Taylor LC, Walt DR (2000) Application of high-density optical microwell arrays in a live-cell biosensing system. Anal Biochem 278(2):132–142. https://doi.org/10.1006/abio.1999.4440
Wood DK, Weingeist DM, Bhatia SN, Engelward BP (2010) Single cell trapping and DNA damage analysis using microwell arrays. Proc Natl Acad Sci U S A 107(22):10008–10013. https://doi.org/10.1073/pnas.1004056107
Dusseiller MR, Schlaepfer D, Koch M, Kroschewski R, Textor M (2005) An inverted microcontact printing method on topographically structured polystyrene chips for arrayed micro-3-D culturing of single cells. Biomaterials 26(29):5917–5925. https://doi.org/10.1016/j.biomaterials.2005.02.032
Kurth I, Franke K, Pompe T, Bornhauser M, Werner C (2009) Hematopoietic stem and progenitor cells in adhesive microcavities. Integr Biol 1(5–6):427–434. https://doi.org/10.1039/b903711j
Deutsch M, Deutsch A, Shirihai O, Hurevich I, Afrimzon E, Shafran Y, Zurgil N (2006) A novel miniature cell retainer for correlative high-content analysis of individual untethered non-adherent cells. Lab Chip 6(8):995–1000. https://doi.org/10.1039/b603961h
And JRR, Folch A (2005) Large-scale single-cell trapping and imaging using microwell arrays. Anal Chem 77(17):5628–5634. https://doi.org/10.1021/ac0505977
Revzin A, Sekine K, Sin A, Tompkins RG, Toner M (2005) Development of a microfabricated cytometry platform for characterization and sorting of individual leukocytes. Lab Chip 5(1):30–37. https://doi.org/10.1039/b405557h
Tokimitsu Y, Kishi H, Kondo S, Honda R, Tajiri K, Motoki K, Ozawa T, Kadowaki S, Obata T, Fujiki S (2010) Single lymphocyte analysis with a microwell array chip. Cytom Part A 71A(12):1003–1010. https://doi.org/10.1002/cyto.a.20478
Prakadan SM, Shalek AK, Weitz DA (2017) Scaling by shrinking: empowering single-cell ‘omics’ with microfluidic devices. Nat Rev Genet 18(6):345–361. https://doi.org/10.1038/nrg.2017.15
Love JC, Ronan JL, Grotenbreg GM, van der Veen AG, Ploegh HL (2006) A microengraving method for rapid selection of single cells producing antigen-specific antibodies. Nat Biotechnol 24(6):703–707. https://doi.org/10.1038/nbt1210
Thakar RG, Cheng Q, Patel S, Chu J, Nasir M, Liepmann D, Komvopoulos K, Li S (2009) Cell-shape regulation of smooth muscle cell proliferation. Biophys J 96(8):3423–3432. https://doi.org/10.1016/j.bpj.2008.11.074
Cheng Q, Komvopoulos K (2010) Integration of plasma-assisted surface chemical modification, soft lithography, and protein surface activation for single-cell patterning. Appl Phys Lett 97(4):043705. https://doi.org/10.1063/1.3462326
Fang Y, Jin J, Chang H, Li X, Deng J, Ma Z, Yuan W (2015) Improved single-cell culture achieved using micromolding in capillaries technology coupled with poly (HEMA). Biomicrofluidics 9(4):044106. https://doi.org/10.1063/1.4926807
Utech S, Prodanovic R, Mao AS, Ostafe R, Mooney DJ, Weitz DA (2015) Microfluidic generation of monodisperse, structurally homogeneous alginate microgels for cell encapsulation and 3D cell culture. Adv Healthc Mater 4(11):1628–1633. https://doi.org/10.1002/adhm.201500021
Dolega ME, Abeille F, Picollet-D’hahan N, Gidrol X (2015) Controlled 3D culture in Matrigel microbeads to analyze clonal acinar development. Biomaterials 52(1):347–357. https://doi.org/10.1016/j.biomaterials.2015.02.042
Liu JJ, Gao D, Mao SF, Lin JM (2012) A microfluidic photolithography for controlled encapsulation of single cells inside hydrogel microstructures. Sci China Chem 55(4):494–501. https://doi.org/10.1007/s11426-012-4538-5
Liu J, Gao D, Li H-F, Lin J-M (2009) Controlled photopolymerization of hydrogel microstructures inside microchannels for bioassays. Lab Chip 9(9):1301–1305. https://doi.org/10.1039/B819219G
Guan Z, Jia S, Zhu Z, Zhang M, Yang CJ (2014) Facile and rapid generation of large-scale microcollagen gel array for long-term single-cell 3D culture and cell proliferation heterogeneity analysis. Anal Chem 86(5):2789–2797. https://doi.org/10.1021/ac500088m
Sydor AM, Czymmek KJ, Puchner EM, Mennella V (2015) Super-resolution microscopy: from single molecules to supramolecular assemblies. Trends Cell Biol 25(12):730–748. https://doi.org/10.1016/j.tcb.2015.10.004
Ng EX, Miller MA, Jing TY, Chen CH (2016) Single cell multiplexed assay for proteolytic activity using droplet microfluidics. Biosens Bioelectron 81:408–414. https://doi.org/10.1016/j.bios.2016.03.002
Lukaszewicz-Zajac M, Mroczko B, Szmitkowski M (2009) The significance of metalloproteinases and their inhibitors in gastric cancer. Postepy Hig Med Dosw 63(835515):258–265. https://doi.org/10.1016/j.lfs.2009.03.019
Perez-Toralla K, Mottet G, Guneri ET, Champ J, Bidard FC, Pierga JY, Klijanienko J, Draskovic I, Malaquin L, Viovy JL, Descroix S (2015) FISH in chips: turning microfluidic fluorescence in situ hybridization into a quantitative and clinically reliable molecular diagnosis tool. Lab Chip 15(3):811–822. https://doi.org/10.1039/c4lc01059k
Jungmann R, Avendano MS, Woehrstein JB, Dai M, Shih WM, Yin P (2014) Multiplexed 3D cellular super-resolution imaging with DNA-PAINT and exchange-PAINT. Nat Methods 11(3):313–318. https://doi.org/10.1038/nmeth.2835
Casabella S, Scully P, Goddard N, Gardner P (2016) Automated analysis of single cells using laser tweezers raman spectroscopy. Analyst 141(2):689–696. https://doi.org/10.1039/c5an01851j
Stojanovic I, van der Velden TJG, Mulder HW, Schasfoort RBM, Terstappen LWMM (2015) Quantification of antibody production of individual hybridoma cells by surface plasmon resonance imaging. Anal Biochem 485:112–118. https://doi.org/10.1016/j.ab.2015.06.018
Agnarsson B, Lundgren A, Gunnarsson A, Rabe M, Kunze A, Mapar M, Simonsson L, Bally M, Zhdanov VP, Hook F (2015) Evanescent light-scattering microscopy for label-free interfacial imaging: from single sub-100 nm vesicles to live cells. ACS Nano 9(12):11849–11862. https://doi.org/10.1021/acsnano.5b04168
Cotte Y, Toy F, Jourdain P, Pavillon N, Boss D, Magistretti P, Marquet P, Depeursinge C (2013) Marker-free phase nanoscopy. Nat Photonics 7(2):113–117. https://doi.org/10.1038/nphoton.2013.116
Anderson SE, Bau HH (2015) Carbon nanoelectrodes for single-cell probing. Nanotechnology 26(18):185101. https://doi.org/10.1088/0957-4484/26/18/185101
Robinson JT, Jorgolli M, Shalek AK, Yoon MH, Gertner RS, Park H (2012) Vertical nanowire electrode arrays as a scalable platform for intracellular interfacing to neuronal circuits. Nat Nanotechnol 7(3):180–184. https://doi.org/10.1038/Nnano.2011.249
Dick JE (2016) Electrochemical detection of single cancer and healthy cell collisions on a microelectrode. Chem Commun 52(72):10906–10909. https://doi.org/10.1039/c6cc04515d
Safaei TS, Mohamadi RM, Sargent EH, Kelley SO (2015) In situ electrochemical ELISA for specific identification of captured cancer cells. ACS Appl Mater Inter 7(26):14165–14169. https://doi.org/10.1021/acsami.5b02404
Koch JA, Baur MB, Woodall EL, Baur JE (2012) Alternating current scanning electrochemical microscopy with simultaneous fast-scan cyclic voltammetry. Anal Chem 84(21):9537–9543. https://doi.org/10.1021/ac302402p
Huang QS, Mao SF, Khan M, Zhou L, Lin JM (2018) Dean flow assisted cell ordering system for lipid profiling in single-cells using mass spectrometry. Chem Commun 54(21):2595–2598. https://doi.org/10.1039/c7cc09608a
Gong XY, Zhao YY, Cai SQ, Fu SJ, Yang CD, Zhang SC, Zhang XR (2014) Single cell analysis with probe ESI -mass spectrometry: detection of metabolites at cellular and subcellular levels. Anal Chem 86(8):3809–3816. https://doi.org/10.1021/ac500882e
Krismer J, Sobek J, Steinhoff RF, Fagerer SR, Pabst M, Zenobi R (2015) Screening of chlamydomonas reinhardtii populations with single-cell resolution by using a high-throughput microscale sample preparation for matrix-assisted laser desorption ionization mass spectrometry. Appl Environ Microb 81(16):5546–5551. https://doi.org/10.1128/Aem.01201-15
Bobrowska J, Pabijan J, Wiltowska-Zuber J, Jany BR, Krok F, Awsiuk K, Rysz J, Budkowski A, Lekka M (2016) Protocol of single cells preparation for time of flight secondary ion mass spectrometry. Anal Biochem 511:52–60. https://doi.org/10.1016/j.ab.2016.06.011
Robinson MA, Graham DJ, Morrish F, Hockenbery D, Gamble LJ (2015) Lipid analysis of eight human breast cancer cell lines with ToF-SIMS. Biointerphases 11(2):02A303. https://doi.org/10.1116/1.4929633
Passarelli MK, Newman CF, Marshall PS, West A, Gilmore IS, Bunch J, Alexander MR, Dollery CT (2015) Single-cell analysis: visualizing pharmaceutical and metabolite uptake in cells with label-free 3d mass spectrometry imaging. Anal Chem 87(13):6696–6702. https://doi.org/10.1021/acs.analchem.5b00842
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2019 Springer Nature Singapore Pte Ltd.
About this chapter
Cite this chapter
Li, W., Lin, JM. (2019). Single-Cell Culture and Analysis on Microfluidics. In: Lin, JM. (eds) Microfluidics for Single-Cell Analysis. Integrated Analytical Systems. Springer, Singapore. https://doi.org/10.1007/978-981-32-9729-6_3