Abstract
Liquid blood phantoms are a tool to calibrate, test and compare near-infrared spectroscopy (NIRS) oximeters. They comprise a mixture of saline, blood and Intralipid, which is subsequently oxygenated and deoxygenated to assess the entire range of tissue oxygen saturation (StO2) from 0% to 100%. The aim was to investigate two different deoxygenation methods: yeast versus nitrogen (N2) bubbling. The phantom was oxygenated with pure O2 in both experiments, but deoxygenated by bubbling N2 in the first and by addition of yeast and glucose in the second experiment. A frequency domain NIRS instrument (OxiplexTS) was used as reference and to monitor changes in the reduced scattering coefficient (μs’) of the phantom. Both deoxygenation methods yielded comparable StO2 values. The deoxygenation was slower by a factor 2.8 and μs’ decreased faster when bubbling N2. The constant bubbling of N2 mechanically stresses the Intralipid emulsion and causes a decrease in μs’, probably due to aggregation of lipid droplets. Deoxygenation by N2 requires a more complex, air tight phantom. The gas flow cools the liquid and temperature needs to be monitored more closely. Consequently, we recommend using yeast for phantom deoxygenation.
Access provided by CONRICYT-eBooks. Download chapter PDF
Similar content being viewed by others
1 Introduction
Since a few decades several near infrared spectroscopy (NIRS) devices have been developed and approved for clinical applications [1, 2]. Devices from different companies rely on different algorithms delivering varying StO2 values for the same measurement [1, 3]. Absolute values to be used as treatment guidelines require different thresholds for each NIRS device [4,5,6,7]. Therefore, we recently developed a liquid phantom to compare and validate different NIRS devices [4,5,6, 8]. To achieve this, a liquid phantom with optical properties similar to the targeted human tissue is sequentially deoxygenated and oxygenated (i.e., StO2 runs from 100% to 0% and back to 100%) while StO2 is measured by different NIRS devices simultaneously.
The aim was to compare two deoxygenation methods with each other: (i) deoxygenation by bubbling nitrogen (N2) versus (ii) deoxygenation by addition of yeast. Both methods are compared regarding the StO2 values they deliver and on the phantom’s scattering property.
2 Methods
The setup employed was previously utilized for similar studies [5, 8]. Essentially, a black container was filled with a liquid phantom, mimicking the optical properties of neonatal brain. This container was equipped with four windows, i.e. holes in the black wall were covered watertightly with a silicone layer, which mimicked the optical properties of skull bone in the near infrared region: absorption coefficient μa (830 nm) = 0.11 cm−1, reduced scattering coefficient μs’ (830 nm) = 8.3 cm−1, thickness = 2.5 mm. Sensors were placed on these windows and their near-infrared light passed through the windows into the phantom.
2.1 Liquid Phantom and Measurement Procedure
The black container was filled with liquid ingredients to resemble the optical properties of a neonatal brain. The ingredients were phosphate buffered saline (PBS), Intralipid (IL), human blood, sodium bicarbonate buffer (SBB); their quantities are listed in Table 1. PBS matches the osmolarity and ion concentration of the human body, whereas the IL adds scattering to the mixture. Human blood was added as erythrocyte concentrate containing haemoglobin. The SBB stabilized the liquid phantom’s pH-level at ~7.4. The container was placed on a heating plate with a magnetic stirrer. The liquid phantom’s temperature was kept constant at 37 °C throughout the measurements. The magnetic stirrer was set to a speed of 500 rpm, ensuring homogeneity of the liquid phantom during the measurement . The liquid phantom setup is depicted in Fig. 1a.
With four windows available, the StO2 measured by three devices of interest were compared to the StO2 measured by the reference device occupying the fourth window. To test the devices on the full measurement range, the liquid phantom was sequentially fully oxygenated and deoxygenated. To achieve StO2 = 100%, O2 was bubbled into the liquid phantom through a tube attached to the liquid container at a flow rate of 2 l/min during ~10 s. Once the liquid phantom was fully oxygenated, two different methods were applied to completely deoxygenate (StO2 = 0%) it.
2.1.1 Deoxygenation Through N2 Bubbling
For deoxygenation 1, the liquid phantom was deoxygenated by bubbling N2 through a tube attached to the liquid phantom container. N2 washes-out O2 in a diffusion process and consequently reduces the blood’s oxygenation. This process is nonlinear and time consuming. To completely deoxygenate the liquid phantom, the N2 bubbling was performed in steps of increasing N2 flow at time points of 50, 94, 112 and 136 minutes after starting N2 bubbling. After complete deoxygenation, N2 bubbling was stopped.
2.1.2 Deoxygenation Through the Addition of Yeast
Before deoxygenation 2, reoxygenation was performed by bubbling O2 into the liquid phantom through another tube. Subsequently, 3 g of yeast and 3 ml of glucose were added. Glucose fed the yeast, which consumed the O2 and deoxygenated the haemoglobin. The preparation of the second measurement cycle took <20 min.
2.2 Measurement Devices
Three different commercially available tissue oximeters were tested, because they are often deployed in clinics. All devices were equipped with paediatric sensors and were acquiring StO2 in the phantom simultaneously.
-
FORE-SIGHT Elite tissue oximeter (CAS Medical Systems, Inc., CASMED, Branford USA). The FORE-SIGHT system employs five wavelengths (690, 730, 770, 810, and 870 nm). The small sensor has two source-detector separations (SDS): 1.0 and 2.5 cm.
-
NIRO-200 tissue oximeter (Hamamatsu Photonics K.K., Japan). The NIRO-200 employs three wavelengths (735, 810 and 850 nm). The small sensor has two SDS: 3.0 and 4.0 cm.
-
INVOS 5100C tissue oximeter (Medtronic, Minneapolis, USA). The INVOS applies two wavelengths (730 and 810 nm). The Oxyalert Neonatal NIRSensor has detectors at a SDS of 3.0 and 4.0 cm.
-
OxiplexTS (ISS, Champaign, Illinois USA) frequency-domain NIRS device works with two frequency-modulated light sources (690 and 830 nm), measures absolute μa and μs’ and therefore was our reference. A rigid sensor was employed with four SDS (2.5, 3.0, 3.5 and 4.0 cm).
3 Results
The complete deoxygenation process took 144 min for the N2 bubbling and 52 min for the yeast. Thus, yeast is deoxygenating the phantom 2.8 times faster than N2.
For each deoxygenation, the StO2 measured by the OxiplexTS was related to the corresponding simultaneous values of the other tissue oximeters in the scatter plots shown in Fig. 1b–d. The values drawn in the scatter plots show data from the decreasing slope. Depicted are the StO2 values of the OxiplexTS in the range from 95% to 5%, with linear fits based for this range. The StO2 was not affected by the deoxygenation method (Fig. 1).
The absolute μs’ at 690 and 830 nm and the StO2 are shown in Fig. 2. Both were measured with OxiplexTS. μs’ (830 nm) decreased during the N2 cycle from 5.4 to 4.9 cm−1 (8.4% reduction in 163 min, 3.1% per hour). A reduction from 5.0 to 4.9 cm−1 (2.3% reduction in 71 min, 1.9% per hour) was observed during the yeast cycle. The decrease during the yeast cycle was corrected for the additional scattering induced by incorporating the yeast. The reduction in μs’ was five times smaller for yeast than for N2 bubbling.
4 Discussion and Conclusion
We compared two deoxygenation methods for liquid blood phantoms and observed a good agreement in measured StO2. Especially INVOS and NIRO-200 demonstrate a nearly perfect overlap between both methods, whereas FORE-SIGHT Elite displays consistently but only slightly lower StO2 values when employing N2 bubbling .
In Fig. 2 we see a higher noise level for StO2 and μs’ at low StO2. This originates from the high μa at 690 nm due to the high concentration of deoxyhaemoglobin, which absorbs strongly at 690 nm. This leads to low light levels and higher noise of the OxiplexTS.
The decrease in μs’ is five times larger for deoxygenation 1 than for deoxygenation 2. The influence of employing N2 on μs’ might be explained by two reasons: (i) the deoxygenation cycle took 2.8 times longer; and (ii) the continuous bubbling of N2 may cause mechanical stress leading to an aggregation of lipid droplets and hence a drop in μs’. Additionally, yeast is less costly and does not create optical inhomogeneities in the phantom through bubble formation.
In conclusion, although both methods provide comparable StO2 results, we recommend using yeast for phantom deoxygenation due to higher speed and phantom stability.
References
Wolf M, Naulaers G, van Bel F et al (2012) Review: a review of near infrared spectroscopy for term and preterm newborns. J Near Infrared Spec 20:43–55
Scholkmann F, Kleiser S, Metz AJ et al (2014) A review on continuous wave functional near-infrared spectroscopy and imaging instrumentation and methodology. NeuroImage 85(Part 1):6–27
Van Bel F, Lemmers P, Naulaers G (2008) Monitoring neonatal regional cerebral oxygen saturation in clinical practice: value and pitfalls. Neonatology 94:237–244
Kleiser S, Hyttel-Sorensen S, Greisen G et al (2016) Comparison of near-infrared oximeters in a liquid optical phantom with varying Intralipid and blood content. In: Elwell EC, Leung ST, Harrison KD (eds) Oxygen transport to tissue XXXVII. Springer New York, New York, pp 413–418
Kleiser S, Nasseri N, Andresen B et al (2016) Comparison of tissue oximeters on a liquid phantom with adjustable optical properties. Biomed Opt Express 7:2973–2992
Hyttel-Sorensen S, Kleiser S, Wolf M et al (2013) Calibration of a prototype NIRS oximeter against two commercial devices on a blood-lipid phantom. Biomed Opt Express 4:1662–1672
Hessel TW, Hyttel-Sorensen S, Greisen G (2014) Cerebral oxygenation after birth – a comparison of INVOS and FORE-SIGHT near-infrared spectroscopy oximeters. Acta Paediatr (Oslo, Norway : 1992) 103:488–493
Nasseri N, Kleiser S, Ostojic D et al (2016) Quantifying the effect of adipose tissue in muscle oximetry by near infrared spectroscopy. Biomed Opt Express 7:4605–4619
Acknowledgments
This work was supported by Swiss Cancer Research grant KFS-3732-08-2015, the Clinical Research Priority Programs (CRPP) Tumor Oxygenation TO2 and Molecular Imaging Network Zurich MINZ of University of Zurich, Danish Council for Strategic Research (grant number 00603-00482B), the Nano-Tera projects ObeSense, ParaTex and NewbornCare and the Swiss National Science Foundation (project 159490). We further thank CASMED and Hamamatsu who provided us with their instruments and sensors for this experiment.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2018 Springer International Publishing AG, part of Springer Nature
About this chapter
Cite this chapter
Isler, H., Kleiser, S., Ostojic, D., Scholkmann, F., Karen, T., Wolf, M. (2018). Liquid Blood Phantoms to Validate NIRS Oximeters: Yeast Versus Nitrogen for Deoxygenation. In: Thews, O., LaManna, J., Harrison, D. (eds) Oxygen Transport to Tissue XL. Advances in Experimental Medicine and Biology, vol 1072. Springer, Cham. https://doi.org/10.1007/978-3-319-91287-5_61
Download citation
DOI: https://doi.org/10.1007/978-3-319-91287-5_61
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-319-91285-1
Online ISBN: 978-3-319-91287-5
eBook Packages: Biomedical and Life SciencesBiomedical and Life Sciences (R0)