Abstract
We develop imaging (stigmatic) XUV spectrometers with the use of plane grazing-incidence varied line-space (VLS) diffraction gratings and focusing normal-incidence multilayer mirrors (MMs), including broadband aperiodic ones. A stigmatic 12–30 nm spectrometer with a resolving power of at least 500 is demonstrated.
Access provided by CONRICYT-eBooks. Download conference paper PDF
Similar content being viewed by others
Keywords
- Normal-incidence MM
- Varied Line-space (VLS)
- Aperiodic Ones
- Taylor Series Decomposition
- High Light-gathering Power
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.
1 Introduction
Grazing-incidence spectrometers with classical concave gratings for a wavelength range λ ~ 20–300 Å are astigmatic. This entails a drastic lowering in the irradiance of spectral lines and the loss of spatial resolution. Using focusing normal-incidence MMs in combination with varied line-space (VLS) diffraction gratings (DGs) makes it possible to remedy these defects.
Cornu [1] realized that monotonic variations in line spacing modified the curvature of the diffracted wavefronts and the spectral (horizontal) focal curve. Hettrick and Bowyer [2] proposed a way to obtain a stigmatic spectral image in the XUV by sending a converging homocentric beam onto a VLS DG whose local line density has to obey the local grating equation to bring the diffracted rays in the principal plane to a common focus. However, in this case, the spectral image is stigmatic at only one wavelength (Fig. 58.1).
The line density p(w) of a VLS grating is conveniently expressed as
Factor p 1 modifies the spectral curve and p 2 and p 3 remove meridional coma and spherical aberration. In the approach of [2], p 0 and φ 0 are free parameters.
2 Spectrometer Design
Let a slightly astigmatic beam be incident on a plane VLS grating (Fig. 58.2). Let L 1 be the distance of the vertical focus of the incident beam from the grating center and L 2 be that of the horizontal one (L 1 > L 2). Then, it is possible to cancel astigmatism at two wavelengths, λ 1 and λ 2, at a sacrifice of one free parameter (either p 0 or φ 0). The slightly astigmatic beam incident on the grating is produced by a near normal-incidence MM, which images a point source. The sequence of steps is as follows: we define the spectral range, and then determine λ 1 and λ 2 to minimize the defocusing in the range. If we define φ 0, then \(p_{0} = \sqrt {(L_{1} /L_{2} - 1)(\sin^{2} \varphi_{0} )/\lambda_{1} \lambda_{2} }\). If, alternatively, we define p 0, then \(\varphi_{0} = \arcsin (mp_{0} \sqrt {\lambda_{1} \lambda_{2} } /\sqrt {L_{1} /L_{2} - 1} )\). The grating parameters p i result from (58.2) after Taylor series decomposition:
We calculated several 1-m-long spectrometer versions for the 120–300 Å range, with λ 1 = 144 Å, λ 2 = 270 Å, and p 0 = 600 mm−1. For a grating area of 50 × 20 mm, the ray trace images of a point source are all confined to one detector pixel. The plate scales are about 5 Å/mm and, in view of the detector pixel size of 13 μm, typical practical resolution is expected at ~(0.05–0.07) Å throughout the range. The spectrometer may be equipped with a narrowband periodic multilayer mirror or a broadband aperiodic one. In the latter case, the operating range may span an octave in wavelength or more. In the instrument described below, the MM was mounted at an angle of incidence of 7.59° and the VLS grating was mounted at a grazing angle of 6.44°.
3 Spectrometer Implementation
The spectrometer layout is shown in Fig. 58.3. The XUV source was the plasma of a plane LiF target irradiated by a 1.06 μm, 0.5 J, 10 ns laser pulse. For a test, we took an aperiodic MM spanning a range of at least 125–250 Å [3].
The VLS grating (p 0 = 600 mm−1, p 1 = 2.32 mm−2) was fabricated by e-beam lithography (EBL) followed by inductively coupled plasma (ICP) etching. A 100 nm tungsten film was deposited on a glass substrate, which was next spin-coated with the positive-tone e-beam resist PMMA A4 (Microchem, 5000 rpm, 90 s). It was then exposed to EBL (beam energy 50 keV, current 15.5 nA, write field 600 × 600 μm, dwell time 0.14 ms). The resist was developed in MIBK:IPA (1:3) for 120 s and IPA for 60 s. Finally, the structure was formed with SF6 ICP etching followed by O2 plasma cleaning.
Due to its high light-gathering power, the spectrometer records the spectrum in one 0.5 J laser shot. The portion shown in Fig. 58.4 contains the lines of Li III and F V–F VII. The closest first-order lines resolved with a safety margin are the 163.138 Å line of F VI and the {163.456, 0.501, 0.558, 0.596 Å} unresolved line array of F V, yielding a conservative figure λ/δλ ~ 510. The line half-widths typically correspond to four detector pixels (52 μm). In view of the plate scale (5.5 Å/mm at 125 Å, 6.3 Å/mm at 200 Å), this corresponds to λ/δλ ~ 450 and 600, respectively. The strongest line arises from the 3d → 2p (127.653 Å, 127.796 Å) transitions in F VII, which saturates the CCD detector pixels corresponding to the near-surface portion of the space-resolved spectral image and broadens the apparent linewidth. Its second order is much weaker and is safely resolved, testifying to a resolving power of ~900.
References
Cornu, M.A.: Vérifications numériques relatives aux propriétés focales des réseaux diffringents plans. Comptes Rendus Acad. Sci. 117, 1032–1039 (1893)
Hettrick, M.C., Bowyer, S.: Variable line-space gratings: new designs for use in grazing incidence spectrometers. App. Opt. 22, 3921–3924 (1983)
Pirozhkov, A.S., Ragozin, E.N.: Aperiodic multilayer structures in soft X-ray optics. Phys. Usp. 58, 1095–1105 (2015)
Acknowledgements
This work was supported by the Russian Science Foundation (Grant No. 14-12-00506).
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2018 Springer International Publishing AG
About this paper
Cite this paper
Kolesnikov, A.O. et al. (2018). Broadband High-Resolution Imaging Spectrometers for the Soft X-Ray Range. In: Kawachi, T., Bulanov, S., Daido, H., Kato, Y. (eds) X-Ray Lasers 2016. ICXRL 2016. Springer Proceedings in Physics, vol 202. Springer, Cham. https://doi.org/10.1007/978-3-319-73025-7_58
Download citation
DOI: https://doi.org/10.1007/978-3-319-73025-7_58
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-319-73024-0
Online ISBN: 978-3-319-73025-7
eBook Packages: Physics and AstronomyPhysics and Astronomy (R0)