1 Basic Concepts of X-Ray Spectroscopy

1.1 Interaction of X-Rays with Matter

Electromagnetic radiation interacts with matter in ways that are characteristic for the type of matter, depending on the energy of the light. X-rays are generally defined as photons with energies between 0.1 keV and 200 keV (approximately 6 pm–12 nm, Fig. 28.1a). Such photons have the ability to interact with electrons that surround the cores of atoms, namely, core electrons [1,2,3,4,5]. Unlike valence electrons, which make up the outer shells of an atom, core electrons are tightly bound to the nucleus and do not participate in chemical bonding. Because of their tight binding to the nucleus, the energy required to excite them to an unoccupied orbital is relatively high. The closer the electrons are located to the positively charged core, the higher this energy requirement becomes [6]. The core electron shells are termed K, L, M, and N, depending on their vicinity to the core (Fig. 28.1b). Their nomenclature is based on quantum numbers, i.e., 1s corresponds to K, 2s and 2p to L, and so on. This means that within one element, electrons in the K shell require the highest amount of energy for excitation as this corresponds to the excitation of a 1s electron. The energy at which the excitation of a 1s electron occurs is called the “K-edge” (>1 keV for transition metals). The required energy for excitation decreases moving to the L- and M-edges (generally ~0.1–1 keV for (transition) metals) [7]. This is illustrated in Fig. 28.1c, where the absorption edges of the K, L, and M shells and their corresponding orbitals are shown as a function of energy. X-ray absorption spectroscopy (XAS) is used to measure these edge energies and to analyze the properties of the core electrons of an element under investigation. XAS is a bulk technique, and as such the spectrum that results from the acquired XAS data is that of the average atom in the sample.

Fig. 28.1
figure 1

(a) Wavelengths in m (upper number) and corresponding energies in eV (lower numbers) of electromagnetic radiation or light. The electromagnetic radiation can be divided into subgroups as indicated in the figure. X-rays constitute the range of 0.1–200 keV and in turn can be divided into hard and soft X-rays, where the first has higher energy than the latter; (b) schematic representation of core electron excitation from the K, L, and M shell by incoming light with sufficient energy (i.e., X-rays); (c) schematic representation of absorption edges. The highest energy is required to excite a core electron from the 1s or K shell, followed by the L (2s, 2p) and M (3s, 3p, 3d) shells; (d) typical example of an X-ray absorption spectrum, depicting the X-ray absorption near-edge structure (XANES) and the extended X-ray absorption fine structure (EXAFS); (e) schematic representation of X-ray absorption, where an incident X-ray photon excites a core electron, ejecting it from the atom (into the continuum). The excited atom can fill the core hole through several mechanisms to return to the ground state; (f) schematic representation of X-ray fluorescence. Electrons from higher shells (i.e., for K shell core holes, electrons from the L and M shells) fill the core hole, releasing X-ray photons with characteristic energies depending on the element and shell; (g) schematic representation of Auger electron emission. The core hole is filled by an electron from a higher shell, while emitting an electron of the same shell into the continuum

To obtain an X-ray absorption spectrum, the sample under study is exposed to an X-ray source, which scans along a range of relevant energies close to a known absorption edge of the element under study [6, 7]. An intensity jump in the spectrum can be observed at a specific energy, which relates to the absorption of photons by the sample as a function of energy. Three distinct energy-dependent events are of importance while scanning X-rays across an absorption edge. These three events give rise to distinct features in X-ray absorption spectra [2, 5]:

  1. 1.

    The pre-edge, the threshold of absorption which is distinct for excitations into the lowest unoccupied states.

  2. 2.

    The edge, where core transitions to quasi-bound states occur.

  3. 3.

    Scattering of electrons with neighboring atoms. In the high-kinetic-energy range, the constructive and destructive interference of the resonance of excited electrons at fixed positions of neighboring atoms gives rise to clearly distinguishable absorption features (oscillations).

A typical example of an X-ray absorption spectrum featuring the abovementioned events is shown in Fig. 28.1d, including the division of the spectrum in two separate regions. These two regions are referred to as the X-ray absorption near-edge structure (XANES), also known as near-edge X-ray absorption fine structure (NEXAFS), and the extended X-ray absorption fine structure (EXAFS). In XANES, events 1 and 2 can be observed. This technique focuses on the absorption edge, approximately 200 eV before the edge and approximately 1000 eV after it for K-edge spectra of transition metals. The scattering of electrons with neighboring atoms, event 3, is what is considered in EXAFS, which is found in the higher-kinetic-energy region of the X-ray absorption spectra [5,6,7].

The areas of a spectrum corresponding to these separate events hold different information on the material under study. First, the energy of the measured edge gives information on the characteristic absorption edge of the material and thus yields element-specific information [8]. For the K-edge, this involves excitation from the 1s ➔ p shells. Since there is only one electron in an s-orbital, there is only one (main) K-edge. There are however three L-edges, since these involve both an s and two p-orbitals. The L1-edge corresponds to s ➔ p transitions, while the L2- and L3-edges correspond to p ➔ d transitions. The former is excited at higher energy than the latter two, which generally occur closer to each other in energy (Fig. 28.1c), and can even overlap or be difficult to distinguish in a spectrum. This also complicates the scattering spectrum in the high-kinetic-energy range, requiring different modes of measurement. These different edges not only give information on which element is being studied, but also provide information on the average oxidation state.

When observing the profile of a K-edge spectrum in more detail, often a small peak can be observed in the lower-energy region of the spectrum, right before the intense increase in absorption. This is the pre-edge (point 1 described above), corresponding to the excitation of a 1s electron to an unoccupied d-orbital (this is a dipole-forbidden, quadrupole-allowed transition and therefore a much weaker feature than the electric dipole allowed 1s to p-orbital transitions). This pre-edge can already give information on the oxidation state of the average atom by revealing the occupancy of the d-orbitals [2, 9]. This includes the observation of ligand (geometry) and local distortions, e.g., Chen et al. reported strong geometry-dependent 3d molecular orbital energies, resulting in a broad 1s ➔ 3dx2-y2 transition and elongated M-R bonds in their studies of axial ligation of Ni metalloporphyrins [10, 11]. As mentioned, the pre-edge and the edge together are often referred to as the XANES or NEXAFS regions. The XANES region of an X-ray absorption spectrum is generally used to determine electronic (e.g., oxidation state) and (to a lesser extent) geometric information about the sample. Examples are estimations of ligand fields, spin states, or the charge on a metal carrier (oxidation state).

Above the edge, at high kinetic energies, we can observe the EXAFS as post-edge oscillations [5, 12]. These oscillations originate from scattering interactions of the emitted electrons with neighboring atoms. The resulting constructive and destructive interference results in the oscillations observed in the EXAFS and thereby provides information on the local surroundings of an atom [13]. The EXAFS is generally used to determine precise local geometric information on the average absorbing atom. First, the spacing between the oscillations or their frequency is inversely proportional to the distance between the atom that absorbed the X-ray photon and a neighboring atom that the emitted electron scattered off of, allowing determination of the average local atomic surroundings. For example, if a solid being measured is in its metallic form, the distance between two atoms will be larger than when the solid is oxidized. In the latter case, the scattering atoms (oxygen) would be found at a position that is closer to the atom than a metallic neighbor would be (i.e., at bond lengths of below 2 Å, whereas metal-metal bonds are found at around or above 2 Å).

Additionally, scattering probability generally increases with atomic mass and is element-dependent, meaning that we can extract information not only about the distance to neighboring atoms but also about their nature. Finally, the amplitude of the EXAFS oscillations reflects the number of scattering neighbors and thus can be used as a measure for the average coordination number of the element under study.

If the element of interest is present in low concentrations, insufficient signal might be obtained following XAS measurement configuration. In this case, X-ray emission spectroscopy, XES, techniques may be employed to collect spectra with good signal-to-noise ratios [2, 6, 7].

XES is closely related and complementary to XAS. Figure 28.1e illustrates how a core hole is created upon X-ray absorption. This core hole can be filled by an electron from a higher shell through the mechanisms shown in Fig. 28.1f, g. The emitted photons or electrons can be detected, leading to X-ray fluorescence spectroscopy (XRF) and Auger electron spectroscopy, respectively.

1.2 The EXAFS Equation

The description given so far has been purely qualitative, but it can be derived using quantum mechanics. This derivation results in what is often referred to as the EXAFS equation in which the EXAFS signal χ(k) is expressed as the sum of the different contributions. Note first that the EXAFS equation is given as a function of k, wave number, and not of E, energy, by the following conversion:

$$ k=1/\mathrm{\hslash}\surd \left(2{m}_{\mathrm{e}}\left(E-{E}_0\right)\right), $$

and with that:

$$ \chi (k)={S_0}^2\Sigma \left({N}_{\mathrm{i}}\right)\kern0.24em {f}_{\mathrm{i}}(k)/{kD_i}^{2\ast }{\mathrm{e}}^{-2D/\lambda (k)\ast }{\mathrm{e}}^{-2{k}^2{\sigma}^2}\sin \left(2{kD}_{\mathrm{i}}+{\delta}_{\mathrm{i}}(k)\right), $$

where S02 represents the amplitude reduction factor, which is an element-dependent (and sample-specific) constant that accounts for the increased influence that the electrons experience from the positively charged nucleus after the emission of a core electron. This results in the orbitals adapting to the decrease in shielding from the core, or in other words, it results in relaxation of all the other electrons in the absorbing atom. Because there is now a slight symmetry mismatch in orbitals compared to the initial atom, the amplitude of EXAFS oscillations is reduced by a factor of S02. This term is obtained for EXAFS analysis by measuring a standard for the specific experimental or beamline setup and generally lies (or is kept) between 0.7 and 1.0.

Ni in the EXAFS equation represents the degeneracy, or the number of neighbors of the absorbing atom. Upon inspection of the expression for the EXAFS absorption as a function of the wave number, χ(k), we see that if we know fi(k) and δi(k) we can determine the coordination number of the neighboring atom. fi(k) and δi(k) are both dependent on atomic number and represent the scattering amplitude and the phase shift, respectively. These factors can be calculated, given a valid approximation of the periodic structure of the sample, by software that was built to model EXAFS.

Aside from the degeneracy Ni, the EXAFS equation can also be used to determine Di (the distance from the absorber to the scatterer and back, often represented a R) and σ, the measure of disorder in neighboring atoms. These factors are the quantitative descriptors to what was described in the last paragraph of the previous section.

As the EXAFS equation implies, data is often represented in k-space to show dependence of momentum instead of energy. Additionally, data can be weighted (often by k2 or k3) to reveal weak signals at high kinetic energies, even though this approach generally also amplifies noise. Subsequently, the EXAFS data in the form of (k-weighted) χ(k) will often undergo a Fourier transform, resulting in the expression of χ in R-space, which is related to a radial distribution function. This can in turn be fitted using specialized EXAFS software to obtain information on bond distances. However, the trained eye can already gain preliminary understanding based on distinguishable features.

Ultimately, proper analysis of XAS results cannot be performed without the measurement of appropriate standards and reference materials. This requires some level of understanding of the system under study – often obtained through standard laboratory techniques – before conducting X-ray experiments. For example, transmission electron microscopy can be used to determine nanoparticle size and dispersion, and when combined with energy-dispersive X-ray spectroscopy, the correct edges to consider can be determined; UV-Vis spectroscopy can be used to get an idea of charge transfer mechanisms, and probe molecule (infrared) spectroscopy can reveal the type of sites present in a sample. This knowledge should be used to determine appropriate standards. Along with possible modeling and calculation of EXAFS using dedicated software (e.g., Artemis, Larch), such appropriate standards are essential to a successful XAS experiment and its in-depth analysis. For more information on data treatment and dedicated EXAFS software, we refer the reader to a number of textbooks [2, 5,6,7].

2 The X-Ray Absorption Experiment

2.1 The X-Ray Source

2.1.1 Synchrotron Radiation Sources

Because a range of energies at high photon flux in the X-ray energy range is required to measure XAS data, standard laboratory equipment could not be used for a long time. Instead, a tunable X-ray source is employed, often in the form of a synchrotron or particle accelerator, which produces light in a broad range of wavelengths and with a high photon flux by the circulation of electrons in a storage ring [14]. A schematic overview of a synchrotron and its workings is given in Fig. 28.2. After generation of electrons in an electron gun, they are accelerated to (nearly) the speed of light in a linear accelerator. They are then transferred to the booster ring, with the aim of increasing their energy to the desired level. This level differs per synchrotron radiation facility. For example, the Super Photon Ring-8 (SPring-8) in Japan operates at 8 GeV, while the Swiss Light Source (SLS) in Switzerland operates at 2.4 GeV, and the Advanced Light Source (ALS) in the United States operates at 1.9 GeV. Finally, the electrons are moved to the storage ring, where they can be used to generate electromagnetic radiation with high photon flux [15]. The trajectory of the electrons in the storage ring is determined by bending magnets, steering them in a circle instead of a straight line. This comes with the release of energy, i.e., light, with a broad range of wavelengths and high photon flux. The generated light can be extracted from the storage ring in beamlines, which are placed at specific locations in the storage ring and as such can be used for experiments. Furthermore, beamlines can be built behind undulators and wigglers, which are different types of equipment in the storage ring that induce movement in the electrons and thus produce X-rays. Light generated at bending magnets, undulators, and wigglers all have their own photon flux (with the latter two resulting in a more brilliant beam than bending magnets) and energy range which in turn all come with their own advantages and drawbacks.

Fig. 28.2
figure 2

(a) Schematic illustration of a synchrotron light source. After generation, electrons are accelerated to nearly the speed of light in the accelerator and when entering the booster ring undergo an increase in their energy to the desired level. They are then transferred to the storage ring where they pass by bending magnets and insertion devices (i.e., undulators and wigglers) to generate the desired X-ray beam. Bending magnets change the trajectory of the electron bunch, while insertion devices like undulators consist of magnets with alternating poles; (b) schematic setup of an X-ray absorption experiment. After extraction from the storage ring, the X-rays pass through a slit to tune the beam size. Subsequently, a monochromator in combination with another slit is used to select the desired energy of the X-ray beam. The X-ray beam then passes through the first ionization chamber, I0, which is a gas-filled radiation integration detector, before passing through the sample. After the sample, the beam passes through another ionization chamber, I1. If a reference sample is measured alongside the sample, it will be placed after I1, and another ionization chamber, I2, will be employed; (c) X-ray fluorescence experiment. The setup is virtually the same as described in a, but a fluorescence detector (If) is placed at an angle θ with respect to the sample and usually contains more than one detection channel

Electrons travel together in small groups called bunches, and only when these bunches pass by the bending magnets, undulators, or wigglers, light is generated. The bunches are present in pulses in the GHz range, which means that for most practical purposes, the generated light can be deemed continuous [2, 7].

2.1.2 Lab-Based X-Ray Absorption Spectroscopy

Recently, lab-based X-ray spectroscopy methods have been developed that can meet the requirements for XAS, i.e., high photon flux over a range of X-ray energies. Lab-based X-ray sources generally comprise X-ray tubes, which are vacuum tubes consisting of a cathode and an anode side, as can be seen in Fig. 28.3. Electrons are generated at the cathode by applying a voltage, and X-rays are generated at the anode. Most of the energy is converted into heat; only about ~1% is converted into X-rays and can be used as light source for a spectroscopy experiment. To obtain sufficiently high photon flux, the voltage can be increased, but due to the high amount of heat generation, the melting point of the anode (often a metal like W) is a limiting factor of X-ray tubes. Therefore, with the aim of applying XAS in the lab, studies have focused on overcoming these limitations by implementing cooling techniques, alternative anode materials, and using techniques that lower the heat dissipation.

Fig. 28.3
figure 3

(a) Schematic and (b, c) photographs of a lab-based X-ray absorption spectroscopy setup to perform time-resolved long-duration operando spectroscopy of solid catalysts. (d, e) Comparison between lab-based X-ray absorption spectra (d) and synchrotron-based X-ray absorption spectra (e) [64] for cobalt reference materials measured: (a) Co, (b) CoTiO3, (c) Co3O4, and (d) CoO, thereby illustrating the overall quality of the performance of current lab-based setups. (Reproduced from [64], with permission from Wiley-VCH)

An example comprises rotation of the anode. As most of the heat is generated at the focal spot, which is typically very small, rotating the anode means a more homogeneous irradiation and thus lower average temperature generated. Alternatively, by focusing the electron beam as a line instead of a spot, heat dissipation becomes much more efficient. Such a source is called a line focus X-ray tube, LFXT, and is regarded a promising new development in the lab-based XAS community.

Apart from the developments in the X-ray source area of the setup, significant advances in detection equipment also contribute to the ongoing increase of implementation of lab-based XAS. Techniques like wavelength-dispersive X-ray (WDX) detection that originated at synchrotron beamlines can be used in a lab setup as well, increasing the energy resolution. WDX detectors exploit Bragg’s law to achieve spatial dispersion of photons, and they can be used both for soft and hard X-ray experiments, although with slightly different setups.

The advances in lab-based XAS methodologies have made it possible to obtain spectra in the lab that are similar to those obtained at synchrotrons. This has been shown by, e.g., Moya-Cancino et al., who demonstrate that despite a lower signal-to-noise ratio, the observed spectral features in their XAS spectra of the Co K-edge for a Co/TiO2 Fischer-Tropsch synthesis (FTS) catalyst closely resemble spectra obtained on the same system measured at a synchr a synchrotron radiation facility, which can be seen in Fig. 28.3c [16]. For a more detailed and comprehensive overview of various lab-based setups and their development, we refer the interested reader to the excellent review article recently published by Zimmerman et al. [17] The authors expect that lab-based XAS will become more routine and eventually will be used in a similar manner as we are currently doing with lab-based X-ray diffraction (XRD).

2.1.3 Free Electron Lasers

XAS experiments can also be carried out at facilities where free electron lasers (FELs) are available [18, 19]. In FELs, electrons are accelerated close to the speed of light and led by a number of undulators, producing light through the same physical principles as in storage rings. Through interaction of the produced radiation with the oscillation of the electrons, micro-bunches producing coherent radiation are formed. This results in intense X-ray pulses that can be used for ultrahigh time resolution XAS experiments [20]. The emission wavelength of FELs is tunable by adjusting the magnetic field strength of the undulators. The authors expect that a lot of exciting new time-resolved studies will become feasible with FELs, especially where repetition-type experiments can be conducted. The latter is possible when studying photocatalytic processes.

2.1.4 Soft Versus Hard X-Rays

X-rays are roughly divided into two types: soft (i.e., below ~5 keV) and hard X-rays (i.e., above 5 keV), although the region between 2 and 5 keV is sometimes also referred to as tender X-rays. Hence, soft X-rays have lower energy than hard X-rays and interact to a greater extent with absorbing materials they come across. On the one hand, the lower amount of energy they carry makes it easier for them to interact with light materials, while on the other hand the attenuation length in heavy materials is much lower than that for hard X-rays. These two together potentially make the collection of sufficient signal challenging in soft X-ray experiments. In that sense, it seems preferable to use hard X-rays that have a larger penetration depth, but the drawback is that these cannot detect some light materials with absorption edges at relatively low energies, like carbon. Generally speaking, hard X-rays are often applied to measure transition metals. The main reasons for this are ease of use, or increased practicality in experimental setups, and better signal. For these reasons, spectral acquisition times can currently be as low as tens of milliseconds at especially equipped beamlines (i.e., quick-XAS, QXAS). The large majority of XAS beamlines comprises hard X-ray beamlines for measurement of transition metal-based catalysts. Figure 28.4 shows two reactor designs, one that is typically used for soft X-ray techniques called a microelectromechanical system (MEMS) reactor. MEMS reactors can be used both for liquid and solid studies (provided that the attenuation length of the components to study does not lead to total absorption). The other is a capillary reactor that is – in different variations – often the choice for operando studies of heterogeneous catalyst powders reacting mainly with gases. It can be used, for example, in combination with hard X-ray microscopy and provides 3D information on, e.g., metals in a catalyst particle.

Fig. 28.4
figure 4

(a) A microelectromechanical system (MEMS) reactor, which can be used for time-resolved long-duration in situ or operando soft X-ray spectroscopy experiments, as illustrated in Fig. 28.6. (b) A capillary reactor, which can be used for time-resolved long-duration in situ or operando hard X-ray absorption spectroscopy of solid catalysts, as illustrated in Fig. 28.7a. Both reactor designs have been used to study, for example, the Fischer-Tropsch synthesis (FTS) reaction over a Co/TiO2 catalyst system under realistic reaction conditions

2.2 Spectral Versus Time Resolution

An important parameter for all spectroscopic experiments is resolution, which is defined as the capability of the used equipment to distinguish between different unit scales (e.g., time or length). Ideally, a measurement has both good spectral resolution (e.g., the amount of points that is measured between two wave numbers) so that one can identify and separate all important features in a spectrum, as well as a good time resolution so that you can conduct measurements quickly. Quick measurements allow the ability to maximize the amount of samples measured during your limited beamtime and/or to distinguish important transitions during a time-resolved experiment. However, spectral and time resolution are a weigh-off, as generally the first increases with increasing measurement time (either by a longer acquisition time or by increasing the amount of accumulations that make up one spectrum). Therefore, a balance needs to be found between the two, based on the aim of the experiment. Note that although advancement in equipment and methods plays an important part in optimizing the spectral resolution, time resolution is ultimately the limiting factor. With this in mind, it is important to realize that nowadays in XAS, the spectral resolution can be increased greatly without sacrificing time resolution. This is because of the implementation of energy-dispersive X-ray or QXAS equipment, both of which severely reduce the acquisition time (as well as the risk of sample damage) [21]. An example of a QXAS beamline is the X10DA at the Swiss Light Source (SLS) with a special monochromator developed by Wuppertal University, that can oscillate at several tens of Hz frequency, and is shown in Fig. 28.5. Just like spectral resolution, time resolution is largely determined by equipment, since machinery needs to move, detectors need to be read out, data needs to be saved, etc. However, if we take these elements away, the factor that ultimately determines the maximum time resolution is the number of electrons (the bunches) in the storage ring. As mentioned earlier, light is generated in pulses in the GHz range, so for experiments with a timescale of seconds or longer, in all practicality, the X-ray light source can be deemed continuous. Still, the pulsed nature of the incoming X-ray radiation can be exploited to carry out transient experiments [2]. In the end, the time resolution of an experiment will be technically limited by the amount of electron bunches in the storage ring, the photon flux they generate (which differs per end-station/beamline type), the time required for mechanical movements of the equipment, and the readout speed of detectors and sensors. For more in-depth discussion of XAS theory, how synchrotrons work, and practical matters like how to prepare for X-ray spectroscopy experiments, we refer the reader to several excellent handbooks that are dedicated to XAS [1, 3, 4].

Fig. 28.5
figure 5

(a) Schematic overview of the X10DA beamline at the Swiss Light Source (SLS, Switzerland), one of the most advanced quick-X-ray absorption spectroscopy (QXAS) beamlines currently in operation, and which is used for the investigation of solid catalysts, including the example described in Fig. 28.9. (b) Photograph of the QEXAFS monochromator made by Wuppertal University, which enables the acquisition of an X-ray absorption spectrum in the up and down movements of the monochromator with several tens of Hz frequency. (c) X-ray absorption spectra of the Cu K-edge recorded at different frequencies, 10, 20, … 50 Hz [65]. (Reproduced from [65], with permission from Wiley-VCH)

2.3 Spatially Resolved X-Ray Absorption Spectroscopy

An in-depth discussion of spatially resolved XAS methods is not within the scope of this chapter, but we feel any work on XAS in catalysis would be lacking if this possibility was not mentioned as more and more we are moving toward the spatiotemporal characterization of catalyst systems, i.e., simultaneously measuring the system as function of space and time and in essence taking snapshots of a catalysts at work. In short, there are several techniques that can combine imaging with time-resolved XAS that can be roughly divided into microscopy and computational techniques [22]. Not all spatially resolved techniques are suitable for all types of in situ and operando measurements. Especially the computational measurements of “classic” heterogeneous catalysis processes with high spatial resolution are still very challenging. However, some beamlines are dedicated to the investigation of a niche form of catalysis, like the ATMOS beamline at the ALS for the investigation of batteries with X-ray ptychography or the BL36XU at SPring-8 for polymer electrolyte fuel cells (PEFCs) for computed tomography-XAS (CT-XAS) measurements [23]. For computed tomography techniques, the field of view and pixel sizes are (still) generally quite large for operando experiments (in the order of microns). Therefore, statistically relevant sampling sizes are generally obtained. However, for X-ray microscopy techniques, the obtained resolution is often much higher, and single particle approaches can be used, even under in situ and operando conditions. It is important to stress that one has to be aware that while X-ray microscopy can yield valuable insights, before drawing conclusions, statistically relevant amounts of samples or particles need to be analyzed. Conventional operando XAS is in principle a bulk technique and therefore does not generally suffer from this drawback. Using both soft and hard X-rays, several advances in in situ (and ex situ) spatially resolved X-ray microscopy have been published over the last decade. For example, using scanning transmission X-ray microscopy and soft X-rays, van Ravenhorst et al. were able to visually capture the genesis of an active Co/TiO2 Fischer-Tropsch synthesis (FTS) catalyst particle using a MEMS reactor, shown in Fig. 28.4a, during a study lasting approximately 2 days. This study showed that during an induction period, the TiO2 pores slowly fill with long-chain hydrocarbons, which takes several tens of hours and shows intraparticle hydrocarbon heterogeneities (Fig. 28.6) [24]. Such spatiotemporal differences in a single catalyst particle had previously not been brought to light and could only be captured by performing time-resolved long-duration operando XAS experiments. Cats et al. studied the same catalyst system with transmission X-ray microscopy (TXM) using hard X-rays, making use of the reactor cell shown in Fig. 28.4b. They were able to note similar intraparticle effects for Co as shown in the corresponding 2D TXM images (Fig. 28.7a). Furthermore, time-resolved long-duration XANES in 3D could be measured for an individual catalyst particle, and the study indicates that the catalyst material is under operando conditions mainly in its Co metallic state. This X-ray microscopy technique has also shown to be particularly relevant for a fluid catalytic cracking (FCC) particle (shown in Fig. 28.7b, c) where 3D reconstructions with voxels of down to 20–30 nm showed detailed information of metal (i.e., Fe and Ni) speciation throughout the FCC particle (Meirer et al. [25]) as a function of reaction time, and more recently, a 3D reconstruction of the carbon speciation throughout the FCC catalyst particle (Vesely et al. [26]), allowing a detailed view of the flow resistance of the crude oil feedstock through the pore network, the ~50-μm-sized FCC particle in the absence and presence of metal deposits (i.e., a long-term catalyst deactivation phenomenon) and carbon deposits (i.e., a short-term catalyst deactivation phenomenon).

Fig. 28.6
figure 6

Schematic overview of the operando microelectromechanical system (MEMS) reactor and related data acquisition method by scanning transmission X-ray microscopy (STXM) for measuring a single Co/TiO2 Fischer-Tropsch synthesis (FTS) catalyst particle as a function of both space and time. (a) Schematic representation of the MEMS reactor. Bottom-left, optical microscopy photograph of heating spiral with 0.1 μm overall thickness, with SiNx windows of 15 nm thickness containing the single Co/TiO2 catalyst particles. Bottom, data acquisition method, the focused soft X-ray beam is changed in energy by the monochromator to measure an X-ray absorption spectrum for each pixel as a function of increasing time-on-stream [24]. (b) Time-resolved long-duration X-ray absorption spectra show the C-edge spectra of the catalyst particle, thereby capturing the organic part of an active Co/TiO2 catalyst particle [56]. (c) Time-resolved long-duration spatial maps of an active Co/TiO2 catalyst particle [56]. The different spectra represent different chemometrically resolved C-edge spectra, thereby gradually showing the different carbon species formed, including aromatic species, with increasing time-on-stream. (Reproduced with permission from [56])

Fig. 28.7
figure 7

(a) Two time-resolved long-duration 2D X-ray operando nanotomography reconstructions of a single Fischer-Tropsch synthesis (FTS) Co/TiO2 catalyst particle [62]. The in situ X-ray absorption near-edge spectroscopy (XANES) data of the cobalt K-edge are an example of the data quality of K-edges studied with this type of technique and measured as a function of reaction time. These results indicate that cobalt mainly was found in the metallic state under FTS conditions with minor contribution of CoTiO3 [66]. (b) 3D X-ray nanotomography reconstruction of a fluid catalytic cracking (FCC) catalyst particle of approximately 50 μm in diameter, visualizing the distribution of Fe and Ni in the FCC catalyst particle as a function of its age (i.e., days). (c) 3D X-ray nanotomography reconstruction of an FCC catalyst particle, with both Fe and Ni distribution and the pore network shown on its right-hand side the pore network includes streamlines, which illustrate the connectivity between the different macropores within the catalyst particle [25]. The same system was examined with X-ray holotomography allowing the 3D imaging of different carbon deposits formed during the FCC process. The carbon K-edge spectra are shown to the right [26]. (Reproduced from [25, 62, 66], with permission from Wiley-VCH and AAAS)

2.4 Choosing the Correct Experimental Mode

In order to choose the right mode of measurement (i.e., transmission or fluorescence), it is important to have a clear idea of the kind of information that should be contained in the acquired data. This includes the element of interest, more specifically its absorption edge energies, its concentration and distribution, and the physical state of the sample. Transmission experiments involve simply measuring the X-rays before and after passing through a sample (that should be uniform) in a straight line. Samples should not be too thick or too thin, with a rule of thumb of ~4 absorption lengths. Fluorescence mode can be employed either when the concentration of the element of interest is low in thick samples or high in thin samples. In fluorescence experiments not absorption but emission is detected, and the detector is placed at an angle. This angle is often chosen to be 90° of the incident beam, while the sample is placed at 45° to maximize emission yield and minimize noise from scattering. The sample angle can also be decreased to a grazing incidence-type setup. In this mode of measurement, the angle is chosen such that the penetration depth of the X-rays is limited to a few nanometers. This makes it possible to selectively investigate the surface of flat samples by detecting the X-ray fluorescence at 90°.

3 X-Ray Absorption Spectroscopy in Catalysis

In the previous section, we have described both qualitatively and quantitatively that synchrotron-based XAS is an extremely powerful characterization technique and capable of studying catalysts in a spatiotemporal manner. By selecting the proper incident energy, important characteristics of the material of interest can be studied, and even very small distortions in crystal structure can be detected. This not only holds for static samples but also for dynamic samples or systems when time resolution is high enough. XAS therefore enables the elucidation of how introduced species change the (average) state of the catalyst [27, 28]. This is exactly what we are interested in when we want to study catalysis. In catalysis research, important goals include the elucidation of active sites and (subsequent) structure-performance relationships [29]. For example, an in-depth understanding of reaction and deactivation mechanisms can lead to the design of new and improved catalytic processes in terms of activity, selectivity, stability, and as such can have major effects on sustainability and economic factors. Taking into account that 80% of all manufactured goods have encountered at least one catalyst in their production process [30], small improvements in such processes can be greatly advantageous both from an economical and a societal point of view.

In situ and operando spectroscopy offers the possibility to investigate various aspects of such catalytic processes in great detail as it is aimed at characterization of catalysts in their active state. Catalysts may also change their structure upon heating or when subjected to increased pressures, for example, metal nanoparticles are known to be dynamic in structure particularly at the elevated temperatures and pressures relevant to industrially applied catalysis [31,32,33,34]. Therefore, ex situ analysis will never be sufficiently capable of capturing the full complexity of a catalyst at work, and in situ or operando analysis is essential to generate relevant structure-performance relationships [35, 36]. The term operando spectroscopy is used when simultaneously analyzing reaction products made by the catalyst at work (usually by gas chromatography, GC, and/or mass spectrometry (MS)). More specifically, when using operando spectroscopy, one makes sure that the catalyst of interest is actually representative of the catalytic process aimed to study in terms of activity and (side) products. For example, in situ or operando UV-Vis spectroscopy can monitor the buildup of organics in a catalyst during reaction (e.g., polyaromatics or coke) [37], while in situ or operando infrared spectroscopy can provide a fingerprint-like molecular map of a reaction [38]. Such lab-based techniques are relatively easy to apply even in operando mode – assuming a suitable reactor is available – but cannot always provide a complete picture. Sometimes, to unravel relevant important details of a (catalytic) process, information on the system is required that can only be obtained by using (advanced) X-ray-based characterization methods, for example, to yield highly specific and precise information on the local electronic environment of the catalyst materials (under reaction conditions) [39]. It is important to note that the design or use of a suitable operando reactor or cell is not trivial and requires serious consideration regarding your reaction, research goal, and the type of X-rays you will use [40]. Apart from the possibility to bring your own reactor, some beamlines also offer the possibility of using theirs [39, 41, 42].

As discussed above, XAS allows for the identification of parameters like coordination number, oxidation state, and local bond distances. Aside from these, the features in the spectra containing information on the different constituents all show up at different energies, both with regard to different elements as well as the different phases a specific element may boast in a sample. That means that XAS is element-specific (i.e., an element shows up in a XAS spectrum within a predetermined set range of energies) and that it can distinguish between different valences or coordination numbers of an element within this specific range of energies. This is especially useful for catalysis as a heterogeneous catalyst typically consists of an active material in the form of transition metal (oxide) nanoparticles29 carried on a porous, metal oxide (e.g., SiO2, Al2O3, and TiO2) or carbonaceous support (e.g., active carbon and carbon nanofiber), most of which are materials suitable for (hard) X-ray operando studies [43] (recall that soft X-rays easily interact with light materials, which include gases, products, and even the operando cell, making operando soft X-ray studies more challenging). Therefore, in principle, the entire catalyst (i.e., both active material and support) can be characterized in depth, both in ex situ and in situ/operando modes. When introduced species, such as the aforementioned gaseous reactants, bind to the surface of the catalyst, corresponding changes in the average electronic structure can be observed in the spectra, and this information can be used to identify the oxidation state of the active phase. This holds true for many relevant processes like catalyst activation (e.g., reduction), catalytic conversion, and deactivation of the catalyst material. As such, after thorough analysis of the observed changes, a clear picture of the entire catalytic reaction can be obtained [44].

The events that a catalytic reaction encompasses take place simultaneously albeit on different timescales, ranging from picoseconds for bond breaking to (typically, depending on the process) weeks-months-years for deactivation [34, 45,46,47]. The different timescales can all be investigated using XAS but require different approaches. For example, when trying to figure out the equilibrium state of the catalyst at work, often steady-state operation is used as the system then is most representative for industrial operation [46]. In principle, due to the equilibrium, the timescales involved are such that we do not need to sacrifice spectral resolution for time resolution. This is especially true due to the advent of QXAS. In contrast, when we want to investigate the exact moment a reactant first comes into contact with the catalyst, kinetics, or reaction intermediates, we want to look at shorter timescales which may also require different equipment or a different beamline altogether. Such alternate approaches may, for example, require introducing reactants in a pulsed fashion. It is important to realize that under such conditions, even when the reaction products are being monitored by GC/MS, the system is inherently not representative for its industrial counterpart (should there be one) even though such experiments are typically still referred to as operando.

Although technically you could measure steady-state or deactivation processes with sub-second or sub-microsecond resolution, the amount of data generated and that would require analysis would be very large (in the order of terabytes). If the scope is to learn more about deactivation, obtaining such an amount of data would not increase the chances of achieving your goal. Similarly, if you want to study a process taking place at a sub-second timescale, it would not make sense to use a time resolution of seconds or more since the information will be averaged out or not even recorded at all. Hence, a careful choice of the proper beamtime setup is important to realize the goals of a particular characterization study, including the type of catalysis (e.g., photocatalysis), and both short-duration and long-duration time-resolved studies have their merits to capture the transient behavior of catalysts under relevant reaction conditions.

4 Showcases from the Field of Heterogeneous Catalysis

4.1 Automotive Catalysis

Zeolites are an archetypal class of heterogeneous catalysts, combining shape selectivity originating from their ordered crystalline and microporous structure with catalytic activity originating from single sites, like Brønsted acid or metal ion sites. SSZ-13, a zeolite with the chabazite (CHA) topology meaning it consists of cages connected by 6- and 8-membered rings (Fig. 28.8a), is such a material, and its copper-exchanged derivative known as Cu-SSZ-13 has received a great deal of attention over its activity in selective catalytic reduction (SCR) of NOx with NH3 as reducing agent. NH3-SCR is currently already being applied in trucks and other vehicles that run on diesel. Cu-SSZ-13-catalyzed NH3-SCR has been investigated intensively with numerous analytical methods, including time-resolved XAS. Figure 28.8 summarizes some important aspects of this catalytic process. First, in Fig. 28.8c, we see that this process involves two operational regimes: one at low temperature (standard), for example, when the engine is just being turned on, and one at high temperature (fast) [48, 49]. The distinct regimes favor different mechanisms [50,51,52] and therefore different reaction products. The different sites involved in catalysis comprise the different oxidation states (+1 and +2) Cu ions can adopt and where they are located within the zeolite framework (Fig. 28.8b), which can be identified using operando XAS. The (type of) species or sites these distinct Cu ions prefer to bind or coordinate to can be identified in the EXAFS, also revealing catalyst deactivation due to binding to poisons in the gas feed [53]. All these mechanisms were investigated with operando XAS on different timescales and conditions accommodating to the aim of the experiment.

Fig. 28.8
figure 8

Selective catalytic reduction (SCR) of NOx over Cu-loaded zeolite Cu-SSZ-13 catalysts. (a) Schematic of the chabazite (CHA) topology, which is the framework structure of the SSZ-13 zeolite; (b) zoom-in of the CHA framework showing the possible positions of the Cu ions in SSZ-13, namely, in the 8-membered ring (red) and the 6-membered ring (blue) [49, 51]; (c) the two typical NOx reduction reactions during SCR operation; (d) k2-weighted extended X-ray absorption fine structure (EXAFS) spectra [54]; (e) operando X-ray absorption near-edge spectroscopy (XANES) spectra of Cu-SSZ-13 during low-temperature SCR of NOx under various conditions [67]; (f) operando XANES of the Cu-SSZ-13-catalyzed SCR recorded at various reaction temperatures [54]; (g) temperature-dependent evolution of Cu species related to NO conversion and NH3 temperature-programmed desorption [52]. (Reproduced from [49, 51, 52, 54, 67], with permission from Wiley-VCH and Springer Nature Publications)

The different temperature regimes for NH3-SCR operation over zeolite Cu-SSZ-13 have been observed in time-resolved operando Cu K-edge XAS experiments at a number of different beamlines [49,50,51,52, 54]. During the low-temperature regime run at steady conversion, typically below 200 °C, lower NO conversion and N2 production are observed in MS measurements compared to the high-temperature regime. Inspection of the XANES spectra at this low T shows both a pre-edge peak originating from Cu(II) (1s ➔ 3d) and an edge-rising peak attributed to Cu(I) (1s ➔ 4p). Comparison with reference spectra and application of linear combination fit (LCF) reveals that these species are present in around a 1:1 ratio, albeit that Cu(I) is only present as NH3-solvated mobile species, whereas Cu(II) is present as both mobile NH3-solvated species and framework-bound species. This distinction can also be observed in the k2-weighted FT-EXAFS spectra, where part of the Cu(II) is observed as framework-coordinated species by an established, well-defined second shell peak in Fig. 28.8d. The other species predominantly show a first shell coordination number of 2 for Cu(I) and 4 for Cu(II) and induce perturbation in the second shell peak through their mobile nature (recall that more crystalline environments result in well-defined peaks, and vice versa). Upon moving to the high-T regime, these perturbations decrease through the formation of more framework-coordinated Cu(II) species, as is also observed in the XANES.

During the reaction in the low-T regime, XANES measurements revealed that the mobile Cu(II) species are reduced to Cu(I) by ammonia, before the latter species are oxidized back to Cu(II) to complete the redox cycle [50]. Transient operando XAS experiments in the form of LCF analysis of time-resolved QEXAFS spectra were conducted to track the dynamic Cu speciation induced by the addition of NO to the gas feed. The results confirmed that in this operational regime, the reoxidation of twofold coordinated NH3-solvated Cu(I) species to form the active fourfold coordinated Cu(II)-(NH3)4 species is the rate-limiting step of the reaction [52]. At temperatures above 350 °C, most Cu(II) is present as coordinated to zeolite framework, indicating a different mechanism is at play [48, 52].

In the high-temperature regime, Cu(I) species are still observed, even though Cu(II) species are postulated to be the active site. The presence of stable Cu(I) species is therefore not desired, and insight into their continued presence is required in order to increase NH3-SCR activity. Cu ions dispersed over and coordinated to well-defined materials like zeolites may seem similar activity-wise, but the fact is that the chemical environment in a 6- and in an 8-membered ring leads to differences in reducibility of Cu(II) in the presence of NH3 [49]. Through fundamental alteration of the catalyst ensuring the presence of Cu ions only in 6-membered rings, operando XANES measurements revealed that the active Cu(II) species did not suffer from reduction into stable Cu(I) with subsequent loss of high-T regime SCR activity in this position. This indicates that merely Cu(II) present in 8-membered rings is sufficiently reducible to be converted into the stable Cu(I) under influence of NH3 and high T.

Finally, catalyst deactivation can be caused by several factors, for instance, the presence of steam or impurities like SO2 in the feedstock [53]. The first causes deactivation of zeolite-based materials by extracting Al from the framework, with a subsequent loss of Brønsted acid sites. However, as Ye et al. showed using XANES-based STXM, even when extra-framework Al species were generated upon treatment with steam, the Cu(II) sites proved to be very stable and remained active [55]. Whereas XANES shows that its presence can cause the formation of Cu bisulfate species which is detrimental for the activity of the catalyst, EXAFS also indicates that at low T the presence of SO2 inhibits the formation of active sites. Such a combination of operando XAS studies of the different aspects of Cu-SSZ-13-catalyzed NH3-SCR showcases that time-resolved XANES and EXAFS, especially when combined with other spectroscopic techniques, can provide fundamental understanding of the active sites, reaction mechanism, and deactivation of heterogeneous catalysts [49, 54].

4.2 Hydrogenation Catalysis

Heterogeneous catalysts often consist of metal nanoparticles supported on oxidic or carbonaceous material. The support materials can take on different crystal phases (e.g., titania can be present in the rutile, anatase, and brookite phase) or be amorphous. Porosity often varies throughout the support material, as large-scale preparation of such materials is not conducted with structure-inducing templates. Deposition of catalytic material on these supports is not always easy to control, leading to nanoparticles showing a size distribution and/or an inhomogeneous dispersion. Furthermore, metal nanoparticles are facetted, consequently supporting many kinds of sites available for interaction with reactants. The activity of certain sites can be enhanced by the addition of promotors or, oppositely, blocked to prevent deactivation or poisoning. Additionally, for many catalysts supported on metal oxide supports, strong metal support interactions (SMSI) can take place. All of this complicates the analysis of such industrially relevant materials, especially under operando conditions where the catalyst material is in a dynamic state and changes continuously.

In conglomeration with (operando) characterization data from other analytical techniques and advanced data analysis, invaluable insights can be obtained on such systems through the use of time-resolved operando XAS. For instance, Vogt et al. studied Ni/SiO2 catalysts with different particle sizes prepared by an industrial partner for CO2 reduction with operando QXAS [38]. In their operando FT-IR spectra, they observed different reaction mechanisms occurring over the various Ni particle sizes, indicating structure sensitivity. They then conducted operando QXAS experiments probing the catalysts with alternating pulses of the reactant, CO2, and a reducing gas, H2. By recording spectra with a very high time resolution (100 ms), it was possible to study the dynamics of changes in nanoparticles at relevant timescales. Nevertheless, such experiments yielded enormous datasets (>70.000 spectra), and the authors set out to study very subtle changes (1–2% of absorption). To this end, the XANES spectra were analyzed by principal component analysis (PCA), which significantly reduced the noise level. Subsequent least square linear combination fitting with Ni and NiO references revealed subtle changes (1–2%) in oxidation state of the Ni catalysts as a result of the alternating pulses, as shown in Fig. 28.9. Among the different particle sizes, the subtle changes varied in accordance with FT-IR spectroscopy results that had shown the presence of different reaction mechanisms taking place depending on catalyst size. By relating these results, these/such distinct reaction pathways could be related to specific sites on the Ni nanoparticles, thus explaining the observed structure sensitivity. This study is an excellent example of high time resolution XAS and its relevance to millisecond dynamics of solid catalysts.

Fig. 28.9
figure 9

A time-resolved short-duration operando X-ray absorption spectroscopy of a set of Ni/SiO2 methanation catalysts . (a) Spectrum recorded at 100 ms time resolution in a study where kinetic details of the CO2 methanation over Ni/SiO2 catalyst materials with increasing Ni metal nanoparticle size were investigated [38]. (b) A selection of transmission electron microscopy (TEM) images of these catalyst. In several Ni/SiO2 catalysts with differing mean particle size (1–6 nm), were studied with operando quick X-ray absorption spectroscopy (QXAS) by pulsing CO2 and H2, and comparing the QXAS data of the different particle size responses to those reactant pulses. The large amount of QXAS data recorded during each experiment was refined by several (multivariate) data analysis steps allowing the discrimination of several interesting particle-size-dependent phenomena of the CO2 methanation reaction. (c) Differences between the different time-resolved metallic Ni contribution as a function of the pulse duration when switching between H2 and CO2. Inserts a and b as well as c and d show the dynamics of the Ni metal surface under different reaction environments and its dependency of the metal nanoparticle size. (Reproduced from [38], with permission from Nature Springer Publishers)

On the other hand, as mentioned before, catalytic processes have many relevant timescales. Recently, van Ravenhorst et al. published a time-resolved long-duration (i.e., 47 h) operando hard X-ray XAS study of Fischer-Tropsch synthesis (FTS) catalysis. A selection of the corresponding XANES data is shown in Fig. 28.10 [56]. Here the authors also used PCA and cluster analysis (CA) to reduce the 47 spectra into 5 clusters representing the largest variation in spectral changes. These five clusters happened to align chronologically, indicating a gradual change in the Co spectra during the FTS reaction which the authors linked to the formation of Co2C, a commonly proposed deactivation mechanism. Interestingly, the formation of this cobalt carbide did not correlate to a decrease in activity, which does not support a common theory that the formation of Co2C is linked to the deactivation mechanism of this catalyst material.

Fig. 28.10
figure 10

(a) A time-resolved long-duration operando X-ray absorption spectroscopy study of a Co/TiO2 Fischer-Tropsch synthesis (FTS) catalyst at 16 bar and 220 °C. (b) By performing principal component analysis (PCA) and cluster analysis (CA), the large number of X-ray absorption spectra obtained during 2 days of operation could be downsized to five different clusters representing the spectral changes observed, where the formation of cobalt carbide (a commonly proposed deactivation mechanism in FTS) could not be linked to a decrease in observed catalyst activity [56]. (b, c) A zoom-in of the areas with the highest amount of spectral changes. (d) The time frame of the different clusters with increasing time-on-stream. The colored boxes represent the cluster and the number represents the time in hours. The different clusters were fitted with the spectra of post-reduction, post-H2 treatment (post-H2) and the Co2C reference, and the linear combination result with the residual is shown in (e). (f) With increasing time-on-stream, the amount of Co2C phase increased, while the amount of metallic Co decreased. (Reproduced from [56], with permission from Wiley-VCH)

4.3 Electrocatalysis

Electrocatalytic processes are also a major point of interest for time-resolved XAS studies. Polymer electrolyte fuel cells (PEFCs) aimed at H2 conversion for energy production, for instance, have been studied by a number of groups. The PEFCs often consist of a Pt catalyst supported on carbon as both the anode (H2 dissociation) and the cathode (oxygen reduction reaction, ORR) catalyst. Especially the cathode side presents an interesting topic of study since the ORR over Pt catalysts suffers from slow kinetics and agglomeration of the Pt NPs that results in deactivation [57]. Due to dedicated operando PEFC setups representative for large-scale application that have been built at various beamlines, these materials are uniquely suited for advanced operando XAS studies. As such, the ORR over Pt/C under working conditions has been extensively investigated. For example, a study by the Tada group used time-resolved operando XAS to reveal oxidation kinetics at various potentials relevant for PEFC operation (i.e., 0.4–1.0 V) [58]. Additionally, they compared the pure Pt/C catalysts to bimetallic Pt3Co/C catalysts. TEM and EXAFS confirmed that the latter consisted of an alloy Pt-Co core and a Pt-rich shell as the characteristics of Pt-Co bonding did not change upon oxidation, whereas the contribution of Pt-Pt decreased and a characteristic Pt-O shell was observed. Furthermore, they observed that upon alloying with Co (and Ni [59]), kinetic control of the Pt oxidation rate could be achieved. However, in a study involving accelerated degradation tests, they found that the long-term stability of Pt3Co/C as cathode catalysts was not suitable for long-term successful PEFC operation [60]. This degradation was further investigated with computed tomography XAS, also denoted as CT-XAS. By rotating the sample 161° and recording XANES spectra at all angles, the catalyst could be reconstructed. Both Pt and Co were imaged, and this revealed differences between their degradation mechanisms that seemed to be largely influenced by cracks in the carbon support [61].

4.4 Photocatalysis

The field of photocatalysis is particularly interesting for time-resolved XAS studies with high resolutions, since photocatalytic processes occur on the sub-microsecond timescale. For example, charge transfer processes taking place on the photocatalytically active metal center can be investigated using pump-probe techniques. This approach involves excitation of the catalyst under pulsed UV-Vis light typically in the order of ns, while the storage ring of the beamline operates in single or few bunches mode, resulting in X-ray pulses in the order of ps, separated (and limited) by the bunch frequency. Such an exploitation of the electron bunches when appropriately equilibrated with the probe can result in ns time resolution. This is demonstrated by Baran et al., who studied water splitting over a model α-Fe2O3/IrOx catalyst using Ir-LIII XANES spectra, pulse-activated by 400 nm LED light in synchronization with the incoming X-rays. Comparison between illuminated and dark samples revealed that more holes were present in IrOx during irradiation, inducing a cascade of charge transfers from the Fe2O3 layer to the IrOx layer, resulting in more photo-generated holes. These holes in the IrOx layer were observed through analyzing the density of states obtained from the XANES spectra, while the catalyst was operating under a realistic potential. A change in potential showed a change in the charge transfer process, whereby minor amounts of Ir(IV) were reduced to Ir(III), while transiently (in the order of ns) maintaining Ir(IV) coordination distances. In the XANES spectra, this resulted in a shift of the spectrum to lower energies [2]. Such a study showcases that time resolutions suitable for the detection of transient species, in this case 600 ns, can be achieved on catalysts when the capabilities of the equipment are properly exploited and the experiment is designed appropriately.

5 Toward Ultrafast X-Ray Spectroscopy of Catalysts

To push the time resolution down even further and into the direction of ps-fs, a different X-ray source can be employed, namely, FELs. This is demonstrated by the investigation of an elementary step in many heterogeneously catalyzed reactions: the dynamics of CO adsorption and desorption on metal surfaces in the absence and presence of, for example O2. This exciting field of frontier research has been summarized in a recent article by Nilsson and co-workers [62]. A model Ru(0001) surface was employed, and CO was probed with soft X-rays to inspect the C-K-edge with XAS, and particularly the 1s ➔ π* transition [63]. As CO adsorbs linearly through the C atom, preferably on a top site, a change in this transition is indicative of a change in orientation of the CO molecule on the surface. Additionally, both in-plane and out-of-plane polarized beams were used to investigate the behavior of Ru(0001)-CO. Such changes in behavior of CO were brought about by pulsing the Ru(0001) surface with 400 nm light, heating it up. As such, XAS spectra revealed a redshift in the peak position upon of the surface. This redshift is attributed to an increase in population of the π* orbital. This indicates CO moving from a top to a bridge or hollow site, a phenomenon taking place on the ps timescale. In the ps that follow, the polarization-dependent XAS spectra show peaks closely resembling gas-phase CO, but still interacting with the surface in several configurations. Upon subsequent cooling down of the Ru(0001) substrate, only some CO desorbs completely, while most readsorbs to the surface. This indicates that the short-lived, transient phase is in fact a precursor state. The entire process described here takes place over a period of 25 ps, illustrating the potential of time-resolved XAS measurements conducted at FELs, an experimental approach which will become more frequently used in the upcoming decade.

6 Conclusions and Outlook

Time-resolved in situ and operando X-ray absorption spectroscopy (XAS) is a powerful analytical tool for heterogeneous catalysis research to shed new insights in their (local) structure and functioning, including their reaction and deactivation mechanism. XANES and EXAFS spectra allow in-depth characterization of solid catalysts under various conditions (hydrated, dehydrated, and reduced state) and, most importantly, in their working state (operando characterization approach) on different timescales, very short (milliseconds and faster) as well as very long (hours, days, and longer). Both timescales are of importance in the field of catalysis; especially the latter is sometimes forgotten as solid catalysts evolve toward their active state (“genesis” of the catalyst) and has to be evaluated as well during its deactivation phase to shed insight in deactivation phenomena in order to extend the lifetime of solid catalysts. The XAS methodology can be applied to a plethora of catalytic reactions ranging from photocatalysis over electrocatalysis to thermocatalysis. Processes on different timescales can be investigated by smart exploitation of equipment and design of experiments. While a true operando experiment investigates the catalyst working at industrially relevant conditions in terms of gas feed composition, temperature, pressure, and reaction products formed, not all timescales relevant to catalysis can be investigated under the same conditions. Kinetics, intermediate species, and transient states are generally investigated on shorter timescales (ns-ps), using, for example, pulsed feed or excitation, lower conversion rates, or alternate catalyst compositions. Such measurements provide time-resolved XAS with the capability to investigate fundamental and short-lived phenomena occurring during catalytic reactions. However, as XAS is element-specific and can therefore only look at a single material at a time, it is important to combine time-resolved XAS with results from other spectroscopic and analytical methods, including theoretical modelling and chemometrics. Only through combing the insights obtained from a multi-technique characterization approach can we truly get to understand a catalytic process in depth.

In the coming decade, we expect that technological advancements in various fields will push down the time resolution for (Q)XAS even further without sacrificing spectral resolution or sample integrity. Next to that, we expect that developments in imaging methods, like operando X-ray microscopy and ptychography, will play a key role in revealing structure-performance relationships in all fields of catalysis with both high spatial and temporal resolution, especially combined with techniques from data science like unsupervised learning and data mining. It is expected that long-duration experiments at synchrotron radiation facilities, which can last for months or years, will become feasible by building setups which can be temporarily moved in and out of the X-ray beam, while still fully functional. Finally, the aforementioned technological advancements will not only play an important role at synchrotron radiation facilities and FELs, but they will also make lab-based XAS more viable as a standard laboratory technique, similar to lab-based [XRD] studies. This will make XAS so accessible that in the future we will be able to simply measure our XAS experiments whenever we want and under almost identical reaction conditions as we are currently doing in lab-based catalysis experiments, thereby allowing for time-resolved (e.g., an entire year or even longer) long-duration operando XAS measurements to fully capture the genesis, life, and death of a solid catalyst. In this manner, the operando concept can be fully captured as the experimental conditions of an industrial reactor can be best mimicked in a lab environment with the ultimate goal to make a molecular movie of the working catalyst.