Keywords

1 Introduction

It is well-known that reference materials (RMs) perform the role of measurement standards in chemistry. RMs of chemical element solutions have been widely used in modern quantitative chemical analysis in constructing calibration dependences of the output signals of, e.g., spectral measurement instruments on the amount of substance. The ease-of-use of RM multicomponent solutions is characterized by the versatility of modern instrumental methods of analysis, which allows calibration of measurement instruments for simultaneous determination of several elements by one or several signals.

2 Literature Review

Over the past eight years, the Ural Scientific Research Institute for Metrology (UNIIM) has accumulated a repository of pure substances, comprising over 70 compounds of 55 individual elements. Having undergone a full purity assessment, these substances are candidates for comparison measurement standards [1]. Such comparison measurement standards in international practice are called primary reference materials [2], primary calibration substances (materials) [3], primary standards for elemental determination [4], etc. The work [5] shows that high-purity materials may act as the material realization of a mole unit in the International System of Units (SI); their gravimetrically prepared solutions are primary calibration solutions and can be used in a variety of methods for quantitative chemical analysis. Having a sufficiently large collection of pure substances, it is reasonable to use their solutions (primary calibration solution [5]) to characterize the RMs of element solutions (secondary calibration solution [5]) using high-performance inductively-coupled plasma spectrometry with high precision [6]. In this case, considering that the relative expanded uncertainty of a mass fraction of the basic component in a solution of comparison measurement standard will be in the order of 0.05%, the relative expanded uncertainty of this component mass fraction in the RM will be in the order of (0.1–0.5)% depending on its stability or the assigned shelf life. Thus, a scheme for ensuring the traceability of RMs of mono-element solutions can be implemented.

As for the production of an RM of a multicomponent solution, it is apparent that the simplest way is to mix mono-element solutions with a well-known analyte content. For the estimation of the uncertainty induced by the preparation method, a method set out in ISO Guide 35: 2017 and based on the use of weighted-mean estimates [7, 9.3.4] is proposed, where the content of a measured component in a mixture is determined using the formulas:

$$ \bar{x}_{i} = \mathop \sum \limits_{j = 1}^{N} w_{j} \cdot x_{ij} , $$
(1)
$$ w_{j} = m_{j} /\mathop \sum \limits_{j = 1}^{N} m_{j} , $$
(2)

where

  • \( \bar{x}_{i} \) is the weighted mean of the content of the i-th component in the mixture;

  • \( x_{ij} \) is the content of the i-th component in the j-th component in the mixture;

  • \( w_{j} \) is weight coefficients;

  • \( m_{j} \) is the mass of the j-th component in the mixture.

ISO Guide 35: 2017 also offers a formula for estimating the standard uncertainty due to characterization using the gravimetric preparation procedure of a binary mixture [7, (13)], for the case when N = 2.

However, within the framework of the GUM methodology [8], the work [9] shows that the square of the combined standard uncertainty of the output value \( x_{j} \) can be represented for an unlimited number of components \( (N < \infty ) \) in general:

$$ u_{{\bar{x}_{i} }}^{2} = {{\mathop \sum \limits_{{j = 1}}^{N} \left[ {m_{j}^{2} \cdot u_{{x_{{ij}} }}^{2} + \left( {x_{{ij}} - \bar{x}_{i} } \right)^{2} \cdot u_{{m_{j} }}^{2} } \right]} \mathord{\left/ {\vphantom {{\mathop \sum \limits_{{j = 1}}^{N} \left[ {m_{j}^{2} \cdot u_{{x_{{ij}} }}^{2} + \left( {x_{{ij}} \bar{ - }\bar{x}_{i} } \right)^{2} \cdot u_{{m_{j} }}^{2} } \right]} {\left( {\mathop \sum \limits_{{j = 1}}^{N} m_{j} } \right)^{2} ,}}} \right. \kern-\nulldelimiterspace} {\left( {\mathop \sum \limits_{{j = 1}}^{N} m_{j} } \right)^{2} ,}} $$
(3)

where

  • \( u_{{x_{ij} }}^{{}} \) is the standard uncertainty of the content of the j-th component in the mixture;

  • \( u_{{m_{j} }}^{{}} \) is the standard measurement uncertainty of the mass of the j-th component in the mixture.

3 Materials and Methods

In order to test the proposed approach, a multicomponent RM was prepared, consisting of a mix of solutions of chemical elements (V, Cr, Mn, Fe, Co, Ni, Cu, Zn, Cd) in a matrix of 5% nitric acid (blank). Mono-element solutions were gravimetrically prepared from pure metals corrected for the air buoyancy in weighing the metal sample intake and final solution. For the preparation of solutions, containers of high density polyethylene were used, purified using non-boiling distillation of nitric and hydrofluoric acids and deionized water. The mass fraction of the main component in the basic material was determined according to the scheme “100% minus combined impurities” using the results of measurements by the mass spectrometry method with inductively coupled plasma (ICP MS, PerkinElmer Inc. NexION 300D). For the determination of gas-forming impurities (N, O, H), the reducing melting method (CGHE, Eltra GmbH ELEMENTRAC ONH-p) was used. The evaluation of this value and associated uncertainty was tested by the following formulas:

$$ w = 100 - \mathop \sum \limits_{i = 1}^{n} z_{i} - \mathop \sum \limits_{j = 1}^{r} \frac{{y_{j} }}{2}, $$
(4)
$$ U(w) = 2 \cdot u_{c} (w) = 2 \cdot \sqrt {\sum\nolimits_{i = 1}^{n} {u(z_{i} )^{2} + \sum\nolimits_{j = 1}^{r} {u(y_{j} )^{2} } } } , $$
(5)

where

  • uc(w) is the combined standard uncertainty of the content of the main component in the starting material;

  • \( z_{i} \) is the mass fraction of the i-th detected impurity;

  • n is the number of detected impurities;

  • \( y_{j} \) is the detection limit established for the j-th impurity according to the 3σ criterion;

  • r is the number of detection limits;

  • \( u(z_{i} ) \) is the standard measurement uncertainty of the j-th impurity;

  • \( u(y_{j} ) = y_{j} /2 \) is the standard uncertainty caused by the establishment of a detection limit in the j-th impurity.

When n + r = 91, the most complete assessment of purity in the characterization of comparison standards is achieved.

The mass fractions of the target impurity components in each solution were determined in the same way using ICP MS and the mass fraction of the major component according to the formula:

$$ x = K_{stab} \cdot K_{homo} \cdot K_{evap} \cdot \left( {w_{0} + \frac{{m_{n} \cdot w \cdot \left( {\frac{{\rho_{a} }}{{\rho_{r} }} - 1} \right)}}{{m_{r} \cdot \left( {\frac{{\rho_{a} }}{{\rho_{n} }} - 1} \right)}}} \right), $$
(6)

where

  • \( K_{stab} ,K_{homo} ,K_{evap} \) are coefficients associated with temporal stability, homogeneity and possible evaporation of the solution during the preparation process, respectively (in the calculations they are taken to be equal to 1, and their uncertainties are determined experimentally);

  • \( w_{0} \) is the mass fraction of the main component in the form (in the example below, the value is taken to be equal to 0 with the uncertainty expressed in the form of the detection limit);

  • \( m_{n} \) is the mass of the starting metal sample intake (pure metal);

  • \( m_{r} \) is the mass of the solution;

  • \( w \) is the mass fraction of the main component in the starting material;

  • \( \rho_{a} ,\rho_{r} ,\rho_{n} \) are the densities of air, solution and starting material, respectively.

The characteristic budget of uncertainty for the analyte content in such a solution is assembled in Table 1. It can be seen that the relative expanded uncertainty of the element mass fraction in such a solution is lower than 0.05%.

Table 1 Uncertainty budget of the primary calibration solution preparation

The original data of the mixture preparation are presented in Table 2.

Table 2 Mixture preparation data

For the estimation of the uncertainty of the mass fraction of the target components in the mixture by formula (3), the standard mass measurement uncertainty was taken to be 0.0005 g, and the uncertainties of the mass fractions of the components listed in Table 3 were used. The values of the standard uncertainties of the main component mass fractions in each of the solutions are taken to be equal to 0.1%, since such tasks commonly imply the use of the so-called secondary calibration solutions, rather than reference solutions.

Table 3 Information on the uncertainty of input data

4 Results

The proposed approach was validated using the Monte Carlo simulation (MMC) [10] for 1 × 106 values of the content of each element having a normal distribution of input values, with the standard deviation being numerically equal to their total standard uncertainty. The output value has a distribution close to normal with a coverage factor of about 1.96, depending on the input data (Fig. 1).

Fig. 1
figure 1

Distribution of the output value in the MMC simulation

Table 4 displays the results of calculations using formulas (13), the averaged data obtained by MMC and the relative deviation of these assessments.

Table 4 Processing results

The comparison showed the compatibility of the obtained estimates and, consequently, the correctness of the weighted means of the components in the mixture and calculation of their standard uncertainty according to (13). At the same time, it is shown that the most significant contribution to the uncertainty of certified values is made by the uncertainty associated with the mass fractions of the certified components in their initial solutions (the relative contribution is more than 90%). Thus, the uncertainty of the certified values is composed of uncertainties associated with the gravimetric preparation of pure substance solutions calculated by the “100% minus combined impurities” scheme and those induced by the stability of these solutions and the results of comparative measurements using high-performance inductively coupled plasma optical emission spectroscopy (ICP OES, PerkinElmer Inc. OPTIMA 7300DV).

Then, if the mass of mixture components is determined sufficiently accurately, in order to avoid loss of precision of the final results, the sum of the impurity of the target element in the components of the mixture should not significantly exceed the value of the standard uncertainty of the target component content in the main solution. In this case, the accuracy of determining the composition of the impurity is not significant. Conversely, when the minimum accumulation of the target component impurities is impossible to achieve, the uncertainty of the combined content of impurities must not exceed ½ of the value of the standard uncertainty of the content of the target component in its main solution.

Estimates of components received by the preparation procedure are confirmed by measurements of these quantities by the ICP OES method (\( x_{l} \), L = 2) (Fig. 2).

Fig. 2
figure 2

Comparison of calculated data and ICP OES results

5 Discussion and Conclusions

It is worth noting that the question remains what to consider as the RM certified value: (1) the results obtained according to the preparation procedure; (2) the results of ICP OES measurements; (3) their mean or weighted mean.

There exists a widespread opinion that the uncertainty of the certified RM value cannot be better than the uncertainty associated with the measurement of this quantity, which can be demonstrated by the producer of that RM. We believe the approach, according to which the results of direct measurements, rather than the calculated data, are taken into account in calculations, to be rather unreasonable. The thing is that any additional knowledge about the measured (certified) value can bring the researcher closer to more reliable estimates. Therefore, such information should never be wasted.

It seems sensible to consider the final result to be the weighted mean of the abovementioned estimates \( (x_{r} ) \) with weighting coefficients associated with their standard uncertainties \( (u_{l} ) \) and the square of the combined standard uncertainty of the quantity \( x_{r} \) as a component of the uncertainty of the certified RM value due to characterization:

$$ x_{r} = {{\mathop \sum \limits_{l = 1}^{L} \frac{{x_{l} }}{{u_{l}^{2} }}} \mathord{\left/ {\vphantom {{\mathop \sum \limits_{l = 1}^{L} \frac{{x_{l} }}{{u_{l}^{2} }}} {\mathop \sum \limits_{l = 1}^{L} \frac{1}{{u_{l}^{2} }}}}} \right. \kern-0pt} {\mathop \sum \limits_{l = 1}^{L} \frac{1}{{u_{l}^{2} }}}}, $$
(7)
$$ u_{{x_{r} }}^{2} = {1 \mathord{\left/ {\vphantom {1 {\mathop \sum \limits_{l = 1`}^{L} \frac{1}{{u_{l}^{2} }}}}} \right. \kern-0pt} {\mathop \sum \limits_{l = 1`}^{L} \frac{1}{{u_{l}^{2} }}}}. $$
(8)

This approach is generally similar to that used in the characterization of mono-element RMs [11, 12], where the certified value is obtained as the weighted mean of the results of gravimetric preparation of a solution from a high-purity compound (with the established content of the main component) and those obtained using ICP OES (the spectrometer being calibrated by four primary standards from a high-purity compound with the established content of the main component).