Abstract
Microemulsion electrokinetic chromatography (MEEKC) is a special mode of capillary electrophoresis employing a microemulsion as carrier electrolyte. Analytes may partition between the aqueous phase of the microemulsion and its oil droplets which act as a pseudostationary phase. The technique is well suited for the separation of neutral species, in which case charged oil droplets (obtained by addition of an anionic or cationic surfactant) are present. A single set of separation parameters may be sufficient for separation of a wide range of analytes belonging to quite different chemical classes. Fine-tuning of resolution and analysis time may be achieved by addition of organic solvents, by changes in the nature of the surfactants (and cosurfactants) used to stabilize the microemulsion, or by various additives that may undergo some additional interactions with the analytes. Besides the separation of neutral analytes (which may be the most important application area of MEEKC), it can also be employed for cationic and/or anionic species. In this chapter, MEEKC conditions are summarized that have proven their reliability for routine analysis. Furthermore, the mechanisms encountered in MEEKC allow an efficient on-capillary preconcentration of analytes, so that the problem of poor concentration sensitivity of ultraviolet absorbance detection is circumvented.
Access provided by CONRICYT – Journals CONACYT. Download protocol PDF
Similar content being viewed by others
Key words
- Microemulsion
- Electrokinetic chromatography
- Capillary electrophoresis
- Pseudostationary phase
- Hydrophobic interaction
1 Introduction
Microemulsion electrokinetic chromatography (MEEKC) covers variants of capillary electrophoresis (CE) employing a microemulsion as carrier electrolyte. Contrary to other CE techniques, MEEKC allows the separation of neutral analytes. In addition, this technique is also suited for separation of charged species, whereby separation selectivities may be achieved which are significantly different from those obtained by commonly used CE techniques for separation of ionic analytes. Microemulsions have been discovered more than 70 years ago by Hoar and Schulman [1] and have been introduced for CE separation techniques in 1991 by Watarai [2]. Since then the numbers of applications of MEEKC have increased steadily, which has been documented in review papers that have been published regularly within the last few years [3–8].
1.1 Fundamentals
Microemulsions are dispersions of two immiscible liquids and may consist either of oil droplets suspended in water (oil-in-water [o/w] microemulsions) or of water droplets suspended in an oil phase (water-in-oil [w/o] microemulsions). MEEKC separations are mostly carried out in oil-in-water microemulsions. Typically, they consist of octane droplets dispersed in an aqueous buffer containing surfactants to coat the octane droplets and lower the surface tension between the two liquids. Furthermore, a short-chain alcohol like n-butanol (called a cosurfactant) is added which also lowers the surface tension. Under such conditions, a stable microemulsion is generated with droplet sizes below 10 nm. It is optically transparent and looks like a single-phase solvent although it is a two-phase system. As mentioned earlier, o/w microemulsions are the most common form of microemulsions used in MEEKC. Therefore, the following discussions will mostly focus on this type, and w/o microemulsions will be treated only shortly in part 1.4.
Sodium dodecyl sulfate (SDS) is commonly used as surfactant for stabilization of the microemulsion droplets. At the interface between the aqueous phase and the oil phase, the dodecyl chain is oriented toward the inner of the oil droplet, whereas the negatively charged sulfate group is oriented toward the aqueous phase. The cosurfactant such as n-butanol will also attach to the surface of the oil droplet with the butyl group toward the oil phase and the alcohol group toward the aqueous phase. As a result of the presence of the anionic surfactant, the oil droplets will acquire a negative charge and will exhibit an electrophoretic mobility in the direction of the anode. The aqueous phase is generally buffered at an alkaline pH. In fused-silica capillaries, alkaline buffers generate an electroosmotic flow (EOF) toward the cathode. Provided that the pH is high enough, the magnitude of the EOF exceeds the electrophoretic mobility of the oil droplets (which is directed against the EOF). Therefore, the EOF will sweep the oil droplets to the cathode. The apparent mobility of the oil droplets is directed to the cathode and has a magnitude that is lower than that of the EOF .
Highly hydrophilic neutral analytes injected at the anodic side of the capillary will reside predominantly in the aqueous phase so that they will be transported to a detector positioned at the cathodic side of the separation capillary by the EOF according to the electroosmotic mobility. The time at which they reach the detector after injection may be called t EOF. Conversely, highly hydrophobic analytes will reside predominantly in the oil droplets, will be transported to the cathodic detection side according to the apparent mobility of the droplets, and will reach the detector after the time t ME. Analytes of medium polarity will undergo partitioning equilibria between the aqueous phase and the oil phase, and will reach the detector at a time t, which is between t EOF and t ME. Obviously, MEEKC separates neutral analytes according to their hydrophobicities. The technique offers a limited separation time window governed by t EOF and t ME. These two parameters may be determined by injection of methanol as EOF marker (t EOF) and octanophenone or dodecyl benzene as microemulsion marker (t ME).
The partitioning equilibria of analytes established between the aqueous phase and the oil droplets indicate that chromatographic principles are involved in the separation (justifying the word “chromatography” in MEEKC). Therefore, the oil droplets may be called a pseudostationary phase . In analogy to chromatography, one can define retention factors k for the analytes:
In case of a true stationary phase as encountered in liquid chromatography, t ME would become infinite and t EOF would be the dead time. The equation given earlier would turn into the well-known definition of k being the ratio of net retention time to dead time.
A schematic presentation of the MEEKC separation process is given in Fig. 1. Additional details can be found in recently published review papers (see for example [9]). It should be pointed out that the separation mechanisms encountered in MEEKC are similar to those in micellar electrokinetic chromatography (MEKC) , which uses micelles (aggregates of surfactant molecules) as pseudostationary phase . Advantages of MEEKC over MEKC may include the fact that oil droplets exhibit a reduced rigidity compared to micelles so that hydrophobic analytes can more easily penetrate the surface and enter the core of the pseudostationary phase. Furthermore, MEEKC may offer a somewhat larger separation time window, because the total charge of the droplets (and thereby t ME) can be manipulated by employing mixed surfactants composed of charged and neutral species in different compositions.
Instead of alkaline buffers, acidic buffers are used occasionally. In such a case the EOF is very low and is no longer able to transport an anionic pseudostationary phase (droplets stabilized by SDS) to the cathodic detection side. Therefore, one has to switch the polarity and the detection must be at the anodic side of the capillary.
Besides anionic surfactants like SDS, also cationic surfactants like cetyltrimethylammonium bromide (CTAB) may be employed for stabilization of the oil droplets. Such surfactants will not only lead to positively charged droplets, but they also act as EOF modifiers due to the generation of a positively charged inner surface of the fused-silica capillary, resulting in a reversed direction of the EOF. Therefore, the detector must be positioned at the anodic end of the capillary when working with such cationic surfactants.
MEEKC separations of ionic analytes involve somewhat more complex mechanisms, because the apparent mobility of the analytes is governed by both their electrophoretic mobilities and their interactions with the pseudostationary phase . Generally, nonionic surfactants can be used leading to a neutral pseudostationary phase, but cationic or anionic pseudostationary phases may be suited as well. In the latter case, one has to take into account a possible repulsion of the charged analyte from the charged pseudostationary phase if both are anionic (or if both are cationic). In case of analytes with a charge opposite to the pseudostationary phase, additional ion-pairing equilibria at the surface of the droplets may have an impact on the separation. In addition, ion pairing between the charged analyte and excess of surfactant may occur in the aqueous phase, which may favor the partitioning reaction into the oil droplet.
Last, but not least, one should keep in mind that depending on the pH of the microemulsion the analytes may be in a neutral form or in a protonated/deprotonated form. Therefore, different types of microemulsions (neutral or charged) may be recommendable, and different separation selectivities can be expected. This is demonstrated in Fig. 2, which shows the separation of closely related methyl derivatives of quinoline that are used as raw materials for industrial production of agrochemicals and pharmaceuticals [10]. Chromatogram A presents the separation of methylquinolines at pH 9.4 (neutral analytes) using a negatively charged oil phase, and B presents the separation of the same set of analytes at pH 4.0 (protonated analytes) using a neutral oil phase [10]. In case A, the separation selectivity is solely governed by the partitioning between the aqueous phase and the oil droplets, whereas in case B separation selectivity is significantly different because it is affected by both partitioning and electrophoretic behavior of the analytes. It is worth mentioning that a buffer of pH 4 without oil droplets (corresponding to a pure capillary zone electrophoretic mode) would not lead to any satisfactory separation.
1.2 Optimization of the Separation of Neutral Analytes
Variables to be optimized with respect to manipulation of migration order and optimization of separation selectivity of neutral analytes include the kind of oil phase, the kind of surfactant and cosurfactant, the addition of water-miscible solvents, and the use of specific additives such as cyclodextrins , carbon nanotubes, and others, that introduce extra effects for the separation of certain analytes.
1.2.1 Oil Phase
The concentration of the oil phase in the carrier electrolyte is typically around 1 % or less. Frequently, n-alkanes like hexane, heptane, or octane are employed as oil phase, with octane often being preferred. As an alternative, ethyl acetate has been selected because of its lower surface tension which allows lower concentrations of surfactant for stabilization. Other compounds occasionally reported for preparation of microemulsions include cyclohexane, toluene, 1-chloropentane, alcohols of medium chain length like 1-hexanol or 1-octanol, and propylene glycol monomethylester acetate. More recently, ionic liquids have been investigated as oil phase, whereby 1-butyl-3-methylimidazolium hexafluorophosphate may be promising [11–13]. Even vegetable oils and artificial oils made of alkane and alcohol may have some potential [14]. Different partitioning coefficients provided by the different oil phases may lead to somewhat different separations, but major changes in migration order are not likely. Unfortunately, it is often still a matter of trial and error to find the best oil phase. In any case, octane may be a good start.
For separation of enantiomers , a chiral oil phase may be used. Chiral alkyl tartrates have been investigated for this purpose [15, 16]. Resolution between enantiomers was obtained if borate buffers were employed, whereas phosphate or Tris buffers did not lead to any enantioseparation. The authors attributed this phenomenon to the formation of a complex between borate and the alkyltartrate.
1.2.2 Surfactants
Surfactants are a key component in the microemulsion. They have a direct impact on stability of the oil droplets by lowering the surface tension, and they affect size and charge of the droplets, magnitude, and direction of the EOF . Anionic or cationic surfactants as well as mixtures of them with nonionic surfactants have been employed for separation of neutral analytes. One should keep in mind that the addition of ionic surfactants can lead to a significant increase of electric conductivity of the carrier electrolyte, which may limit the applied voltage in order to avoid excessive Joule heating.
The most common surfactant for MEEKC is sodium dodecyl sulfate (SDS), which is typically used at concentrations around 3 %. Alternative anionic surfactants include lithium dodecyl sulfate (which leads to somewhat lower electric currents), bile salts like sodium cholate, or sulfosuccinates like sodium bis(2-ethylhexyl) sulfosuccinate.
Cationic surfactants reported for use in MEEKC are based on quaternary ammonium salts like dodecyltrimethyl ammonium chloride, tetradecyltrimethyl ammonium bromide, or cetyltrimethyl ammonium chloride/bromide. As mentioned in the part on fundamentals, the behavior of these salts as EOF modifiers must be taken into account.
The use of mixtures of surfactants may provide various benefits. The combination of SDS and Brij-35 (a nonionic surfactant) allows the manipulation of the charge of the droplets and thereby manipulation of the separation time window.
Chiral surfactants have been introduced for MEEKC separations of enantiomers , such as R- and S-dodecoxycarbonylvaline (DDCV) [17–19]. This approach can also be combined with the use of a chiral oil phase (see previous part) which has been demonstrated for the use of DDCV together with dibutyltartrate or diethyltartrate [20, 21].
1.2.3 Cosurfactant
The variations of nature and concentration of the cosurfactant may be exploited for fine-tuning of the separation (see for example [22]). Short chain alcohols are frequently used as cosurfactants, with 1-butanol at a concentration of around 6 % being the most common one. It has been suggested that such solvents do not only act as cosurfactants, but that a significant portion of it can partition into the oil droplet, especially as the amount of cosurfactant present in the microemulsion exceeds that of the actual oil phase [23]. Thereby, the chromatographic properties of the pseudostationary phase are modified and with it the k values of the analytes affected. General rules for selection of appropriate cosurfactants are still difficult to establish.
Chiral separations may benefit from the use of chiral 2-alkanols like R(−)-2-pentanol, R(−)-2-hexanol or R(−)-2-heptanol as cosurfactants [24]. A synergistic effect has been observed when (S)-2-hexanol was employed together with a chiral surfactant [25]. In addition, even three-chiral-component microemulsions (R- or S-DDCV, S-2-hexanol, and R- or S-diethyltartate) have been investigated and compared with one- and two-chiral-component microemulsions [26].
Interestingly, it has also been claimed that a stable microemulsion prepared by hexane and SDS in an ammonium acetate solution can be generated without the use of any cosurfactant [27], but such an approach has not made its way to a wider range of applications.
1.2.4 Water-Miscible Solvents
For certain applications, water-miscible organic solvents may be added to the microemulsion [23]. In this way, the partitioning equilibria of the analytes between the aqueous phase and the oil phase may be manipulated. This aspect is of major significance when analytes with very poor solubility in water are separated. Such analytes would not partition at all into a purely aqueous phase and would therefore reach the detector after the time t ME. A typical example for the benefits of water-miscible solvents is the analysis of highly hydrophobic polymer stabilizers [28]. Depending on the type of water-miscible solvent, there are upper limits for its use in MEEKC. Exceeding these limits will result in a disintegration of the microemulsion. It has been reported that methanol may be used up to 8 % (v/v), acetonitrile up to 12 %, whereas 2-propanol may be used at considerably higher concentrations [29]. One should not forget the well-known side effect of organic solvents on the magnitude of the EOF which depends on the dielectric constant of the liquid phase, on the viscosity, and on the zeta potential of the capillary wall (all these parameters are directly affected by amount and type of an organic solvent in the aqueous phase of the microemulsion).
1.2.5 Other Additives
The addition of cyclodextrins to the carrier electrolyte is a well-established approach for chiral separations in capillary zone electrophoresis . The formation of transient diasteromeric complexes with cyclodextrins can also be exploited in MEEKC as an interaction in addition to the partition equilibrium between aqueous and oil phase, whereby either neutral cyclodextrins or cyclodextrins modified by charged groups (sulfated cyclodextrin) may be suited [30–32].
More recently, carbon nanotubes dispersed in the microemulsion have been investigated in order to establish additional interactions that might improve the separation selectivity in MEEKC [33–35].
Some experiments have been done with water-soluble ionic liquids as additives. In case of an anionic surfactant, the cation of the ionic liquid may interact and may partly neutralize the negative charge, thereby changing the properties of the pseudostationary phase [36].
1.3 Optimization of the Separation of Ionic Analytes
In the simplest case, ionic analytes are separated by using an oil phase stabilized by nonionic surfactants. In this case, the principles for optimization of the separation are similar to those mentioned earlier for separation of neutral analytes in a charged microemulsion. Nonionic surfactants most often employed are Brij-35, Tween-20, or Triton X-100. In addition, a less common nonionic surfactant, Pluronic F-127 has seen suggested [37] (which is an amphiphilic block copolymer consisting of ethylene oxide and propylene oxide), although so far only in combination with SDS for separation of neutral analytes. Most recently, zwitterionic surfactants like N-dodecyl-N,N-dimethyl-3-ammonio-1-propanesulfonate (DAPS) have been studied [38].
As mentioned in the Introduction, the use of charged microemulsions may lead to additional attraction or repulsion of ionic analytes to/from the droplets. This attraction/repulsion may be manipulated by using a mixture of a cationic and an anionic surfactant (see for example [39]). Oppositely charged analytes/droplets or analytes/excess surfactant in aqueous phase may also undergo interactions by ion-pair formation. A systematic treatment of such complex additional interactions is somewhat difficult so that generally valid strategies for optimization of separation selectivity are still limited.
1.4 Water-in-Oil Microemulsions
Although MEEKC is almost exclusively done in oil-in-water microemulsions, a few attempts have been made to apply water-in-oil microemulsions. Altria et al.[40, 41] introduced w/o microemulsions typically composed of 10 % SDS, 80 % butanol (or 78 % butanol and 2 % octane), and 10 % aqueous buffer (or slight modifications of this composition). Similar compositions have been used later by other groups [42–45], but up to now the number of applications of w/o microemulsions in MEEKC is quite limited.
1.5 Sample Preconcentration by Sweeping
Spectroscopic detection techniques generally suffer from poor detection limits due to the short detection path length provided by the inner diameter of the separation capillary. Preconcentration effects occurring under proper injection conditions may help to improve detection limits. CE separation techniques based on pseudostationary phases may allow a preconcentration step called sweeping. It is generally defined as the picking and accumulating of analytes by a charged pseudostationary phase that penetrates the sample zone during application of a voltage. Most work on sweeping was done using micelles as pseudostationary phase (see, for example, the review in [46]), and the same principles work for microemulsions as well. Therefore, this chapter will not go into details regarding the theory of sweeping. In the simplest case, efficient preconcentration can be achieved with an microemulsion consisting of an oil phase stabilized by a negatively charged surfactant and an aqueous phase of low pH. The sample solution that does not contain the pseudostationary phase is injected hydrodynamically at the cathodic end of the capillary. After injection, the anionic pseudostationary phase will migrate from the cathodic carrier electrolyte vial into the capillary and through the sample zone (because of the low pH, the EOF can be neglected). In the sample zone, neutral analytes undergo partitioning and are focused into a narrow zone. As a result of the focusing effect, quite high volumes of sample may be injected without peak broadening (making possible a more than 1000-fold increase in sensitivity). Nevertheless, too long injection zones (without pseudostationary phases) may lead to instabilities of the system after applying voltage. Therefore, electrokinetic injection techniques have been used instead of hydrodynamic injection (which allows the selective injection of anions or cations without generating an excessively long zone of sample) followed by the sweeping step. Details of quite sophisticated combinations of injection techniques and sweeping would go beyond the scope of this chapter but can be found in recent review papers [3, 4, 7].
1.6 Detection
In common with other CE modes, the most widely used detection technique for MEEKC is UV–visible absorbance detection . Besides, fluorescence detection (with a xenon lamp or a laser as light source) may be the alternative for analytes that show native fluorescence or can be transformed into fluorescent derivatives prior to injection. A typical example for the latter approach is the separation of amino acids after derivatization with fluorescein isothiocyanate (FITC) [47].
Mass spectrometric (MS) detection may be most attractive as it provides the confirmation of peaks for target analytes or the structure elucidation of unknown peaks. Capillary zone electrophoresis can be hyphenated with MS via an electrospray ionization (ESI) source using a sheath–liquid interface which allows the realization of a makeup flow of a few μL/min to make flow rates better compatible with commercial ESI sources and at the same time allows the application of the high voltage of the CE separation [48]. Unfortunately, the high concentrations of surfactants used in MEEKC make the technique hardly suited for coupling with ESI which would suffer from severe ionization suppression. Instead of ESI, atmospheric pressure photoionization (APPI) was found to tolerate components of a microemulsion much better [49–51]. A microemulsion consisting of 0.8 % octane, 2 % SDS, 6.6 % butanol , and 90.6 % of 20 mM ammonium hydrogencarbonate buffer (pH 9.5) allowed the quantitative analysis by APPI-MS of various pharmaceuticals down to the sub-μg/ml range without dedicated sample preconcentration during injection [49].
More recently, MEEKC has also been hyphenated with MS detection by an inductively coupled plasma interface , thereby allowing element-selective detection. This approach has been used for the analysis of anticancer platinum complexes [52].
1.7 Applications
The following discussion cannot give an exhaustive compilation of applications reported so far, but intends to give an idea of the broad variety of classes than can be separated. In Table 1 the focus is put on those applications that demonstrate a separation of a larger number of analytes, whereas applications dealing with just a single analyte are not included. The separations done by MEEKC range from pharmaceutical drugs to vitamins, agrochemicals, polycyclic hydrocarbons, natural products, derivatized sugars, derivatized amino acids , proteins , fatty acids , nucleosides, and chiral compounds. Actually, it is possible to use a single set of operating conditions for different applications. A microemulsion consisting of 0.8 % (w/w) octane, 6.6 % (w/w) 1-butanol , 3.3 % SDS, and 89.3 % (w/w) 10 mM sodium tetraborate buffer may be successful for a large number of different analytes and is often a quite successful starting point. In cases where this composition does not lead to satisfactory results, fine-tuning is possible by variation of the components of the microemulsion according to the principles discussed earlier.
Besides its benefits for analytical chemistry, MEEKC has frequently been employed as a simple tool for assessment of hydrophobicity (expressed as octanol–water partition coefficient P o/w) [75–77]. The following linear relationship exists between P o/w and log k (k being the retention factor as mentioned earlier):
Slope and intercept of this line can be obtained from experiments with solutes of known octanol–water partition coefficients.
2 Materials
-
1.
Microemulsion for general applications using a negatively charged oil phase: mix 3.3 g SDS and 6.6 g 1-butanol , and then add 0.8 g n-octane and 89.3 g 10 mM borate buffer pH 9.4 (prepared from a 10 mM boric acid adjusted to pH 9.4 with NaOH). The mixture is placed in an ultrasonic bath for 30 min to obtain a clear solution. Afterward, the microemulsion is filtered through a 0.45 μm membrane filter.
-
2.
Microemulsion for highly hydrophobic analytes using a negatively charged oil phase: mix 2.25 g SDS, 0.75 g Brij 35 (see Note 1 ), and 6.6 g 1-butanol , and then add 0.8 g n-octane, 25 g 2-propanol, and 64.6 g 10 mM borate buffer pH 9.4 (prepared from a 10 mM boric acid adjusted to pH 9.4 with NaOH). The mixture is placed in an ultrasonic bath for 30 min to obtain a clear solution. Afterward, the microemulsion is filtered through a 0.45 μm membrane filter.
-
3.
Microemulsion for general applications using a neutral oil phase: mix 3.32 g Brij 35 and 6.62 g 1-butanol, and then add 0.82 g n-heptane and 89.2 g 25 mM acetate buffer pH 4.0 (prepared from a 25 mM acetic acid adjusted to pH 4.0 with NaOH). The mixture is placed in an ultrasonic bath for 30 min to obtain a clear solution. Afterward, the microemulsion is filtered through a 0.45 μm membrane filter (see Note 2 ).
-
4.
Microemulsion for on-capillary preconcentration by sweeping using a negatively charged oil phase: mix 3.3 g SDS and 6.6 g 1-butanol , and then add 0.8 g n-octane and 89.3 g 50 mM phosphoric acid pH 2.0. The mixture is placed in an ultrasonic bath for 30 min to obtain a clear solution. Afterward, the microemulsion is filtered through a 0.45 μm membrane filter.
-
5.
CE instrument “7100 CE System” (Agilent, Waldbronn, Germany), or equivalent, equipped with an ultraviolet (UV) absorbance detector , high voltage supply up to +/−30 kV, and autosampler for both hydrodynamic and electrokinetic injection.
-
6.
Fused-silica capillaries (Polymicro Technologies, Phoenic, AZ) with inner diameter and outer diameter of 50 and 360 μm, respectively, a length from inlet to detector of 51.5 cm, and a length from inlet to outlet of 60 cm (see Note 3 ).
-
7.
Sample vials for autosampler of CE instrument.
3 Methods
3.1 General Procedure for Conditioning New Fused-Silica Capillaries
-
1.
Four vials are filled with 1 M NaOH, water, 0.1 M NaOH, and 0.2 M HCl, respectively.
-
2.
The vials are placed into appropriate positions of the autosampler for rinsing the capillary.
-
3.
The capillary is rinsed with 1 M NaOH for 10 min, with water for 5 min, with 0.2 M HCl for 10 min, with water for 1 min, with 0.1 M NaOH for 10 min, and with water for 10 min.
3.2 Separation of Neutral Analytes Using a Negatively Charged Oil Phase
-
1.
Two vials are filled with 0.1 M NaOH and microemulsion, respectively, for rinsing the capillary (the microemulsion is prepared according to the procedure given in Subheading 2, item 1.).
-
2.
Two carrier electrolyte vials (for inlet and outlet side) are filled with the microemulsion.
-
3.
Sample solutions and calibration solutions are filled into vials (see Note 4 ).
-
4.
All vials are put into appropriate positions of the autosampler.
-
5.
The capillary is rinsed with 0.1 M NaOH for 5 min and with microemulsion for 5 min.
-
6.
The first sample or calibration solution is injected using hydrodynamic injection at a pressure of 50 mbar for 5 s (see Note 5 ), and the separation is started by applying a voltage of +25 kV (see Note 6 ).
-
7.
The capillary is rinsed with 0.1 M NaOH for 1 min and with microemulsion for 1 min.
-
8.
Steps 6 and 7 are repeated for the next sample or calibration solution.
3.3 Separation of Highly Hydrophobic Analytes Using a Negatively Charged Oil Phase
-
1.
Two vials are filled with 0.1 M NaOH and microemulsion, respectively, for rinsing the capillary (the microemulsion is prepared according to the procedure given under Subheading 2, item 2).
-
2.
Two carrier electrolyte vials (for inlet and outlet side) are filled with the microemulsion.
-
3.
Fill vials with sample solutions and calibration solutions, prepared in the microemulsion as solvent.
-
4.
All vials are put into appropriate positions of the autosampler.
-
5.
The capillary is rinsed with 0.1 M NaOH for 5 min and with microemulsion for 5 min.
-
6.
The first sample or calibration solution is injected using hydrodynamic injection at a pressure of 50 mbar for 3 s, and the separation is started by applying a voltage of +30 kV.
-
7.
The capillary is rinsed with 0.1 M NaOH for 1 min and with microemulsion for 1 min.
-
8.
Steps 6 and 7 are repeated for the next sample or calibration solution.
3.4 Separation of Positively Charged Analytes Using a Neutral Oil Phase
-
1.
Two vials are filled with 0.1 M NaOH and microemulsion, respectively, for rinsing the capillary (the microemulsion is prepared according to the procedure given under Subheading 2, item 3).
-
2.
Two carrier electrolyte vials (for inlet and outlet side) are filled with the microemulsion.
-
3.
Vials are filled with sample solutions and calibration solutions.
-
4.
All vials are put into appropriate positions of the autosampler.
-
5.
The capillary is rinsed with 0.1 M NaOH for 5 min and with microemulsion for 5 min.
-
6.
The first sample or calibration solution is injected using hydrodynamic injection at a pressure of 50 mbar for 5 s (see Note 5 ), and the separation is started by applying a voltage of +25 kV (see Notes 6 and 7 ).
-
7.
The capillary is rinsed with 0.1 M NaOH for 1 min and with microemulsion for 1 min.
-
8.
Steps 6 and 7 are repeated for the next sample or calibration solution.
3.5 Separation of Neutral Analytes with On-Capillary Preconcentration by Sweeping
-
1.
Two vials are filled with 0.1 M NaOH and microemulsion, respectively, for rinsing the capillary (the microemulsion is prepared according to the procedure given in Subheading 2, item 4).
-
2.
Two carrier electrolyte vials (for inlet and outlet side) are filled with the microemulsion.
-
3.
Vials are filled with sample solutions and spiked sample solutions.
-
4.
All vials are put into appropriate positions of the autosampler.
-
5.
The capillary is rinsed with 0.1 M NaOH for 5 min and with microemulsion for 5 min.
-
6.
The first sample solution is injected using hydrodynamic injection at a pressure of 100 mbar for 150 s (see Note 8 ), and the separation is started by applying a voltage of −20 kV (see Note 6 ).
-
7.
The capillary is rinsed with 0.1 M NaOH for 1 min and with microemulsion for 1 min.
-
8.
Steps 6 and 7 are repeated for the next sample or spiked sample solution (see Note 9 ).
4 Notes
-
1.
The partial substitution of SDS by Brij 35 results in lower charge of the oil droplet and thereby in a lower velocity. This leads to a decrease of the analysis time. For a specific separation, one can try to vary the ratio of SDS/Brij 35 to achieve optimal analysis time.
-
2.
This microemulsion prepared in a buffer of pH 4.0 is suited for the separation of analytes that undergo protonation or deprotonation reactions at this pH, so that positively or negatively charged compounds are formed to some extent. The pH can be changed if necessary.
-
3.
Shorter or longer capillaries can be used if necessary to optimize resolution and analysis time.
-
4.
If the analytes are not easily soluble in water, the sample and calibration solutions can be prepared in the microemulsion as solvent. One should avoid pure organic solvents for the samples and the calibration solutions because these can disrupt the microemulsion adjacent to the zone of injected sample, leading to distorted peak shapes. It is recommended that an internal standard be added to both the sample and the calibration solutions.
-
5.
Somewhat longer injection times can be used to achieve lower detection limits. Peak distortion will occur at too long injection times.
-
6.
It may be advantageous to use somewhat lower or higher separation voltages depending on the length of the capillary.
-
7.
The positive voltage applied is suited for cationic analytes. In the case of anionic analytes, it may be necessary to use a negative voltage (depending on the electrophoretic mobility of the analyte in relation to the electroosmotic mobility).
-
8.
Depending on the analytes, this injection time may need to be decreased in order to avoid deterioration of peak shapes.
-
9.
The incorporation of the online preconcentration effect makes quantitation by standard addition instead of external standards preferable.
References
Hoar TP, Schulman JH (1943) Transparent water-in-oil dispersions: the oleopathic hydro-micelle. Nature 152:102–103
Watarai H (1991) Microemulsion capillary electrophoresis. Chem Lett 391–394
Yang H, Ding Y, Cao J, Li P (2013) Twenty-one years of microemulsion electrokinetic chromatography (1991-2012): a powerful analytical tool. Electrophoresis 34:1273–1294
Yu L, Chu K, Ye H, Liu X, Yu X, Chen G (2012) Recent advances in microemulsion electrokinetic chromatography. Trends Anal Chem 34:140–151
Ryan R, Altria K, McEvoy E, Donegan S, Power J (2013) A review of developments in the methodology and application of microemulsion electrokinetic chromatography. Electrophoresis 34:159–177
Ryan R, McEvoy E, Donegan S, Power J, Altria K (2011) Recent developments in the methodology and application of MEEKC. Electrophoresis 32:184–201
Ryan R, Donegan S, Power J, McEvoy E, Altria K (2009) Recent advances in the methodology, optimisation and application of MEEKC. Electrophoresis 30:65–82
McEvoy E, Marsh A, Altria K, Donegan S, Power J (2007) Recent advances in the development and application of microemulsion EKC. Electrophoresis 28:193–207
Ryan R, Donegan S, Power J, Altria K (2010) Advances in the theory and application of MEEKC. Electrophoresis 31:755–767
Schöftner R, Buchberger W (2003) Systematic investigations of different capillary electrophoretic techniques for separation of methylquinolines. J Sep Sci 26:1247–1252
Cao J, Qu H, Cheng Y (2010) The use of novel ionic liquid-in-water microemulsion without the addition of organic solvents in a capillary electrophoretic system. Electrophoresis 31:3492–3498
Wang Y, Li F, Yang FQ, Zuo HL, Xia ZN (2012) Simultaneous determination of α-, β- and γ-asarone in Acorus tatarinowii by microemulsion electrokinetic chromatography with [BMIM]PF6 as oil phase. Talanta 101:510–515
Li F, Yang FQ, Xia ZN (2013) Simultaneous determination of ten nucleosides and related compounds by MEEKC with [BMIM]PF6 as oil phase. Chromatographia 76:1003–1011
Siren H, Vesanen S, Suomi J (2014) Separation of steroids using vegetable oils in microemulsion elekctrokinetic capillary chromatography. J Chromatogr B 945–946:199–206
Hu S, Chen Y, Zhu H, Zhu J, Yan N, Chen X (2009) In situ synthesis of di-n-butyl L-tartrate-boric acid complex chiral selector and its application in chiral microemulsion electrokinetic chromatography. J Chromatogr A 1216:7932–7940
Hu S-Q, Chen Y-L, Zhu H-D, Shi H-J, Yan N, Chen X-G (2010) Effect of molecular structure of tartrates on chiral recognition of tartrate-boric acid complex chiral selectors in chiral microemulsion electrokinetic chromatography. J Chromatogr A 1217:5529–5535
Pascoe R, Foley JP (2002) Rapid separation of pharmaceutical enantiomers using electrokinetic chromatography with a novel chiral microemulsion. Analyst 127:710–714
Mertzman MD, Foley JP (2004) Effect of surfactant concentration and buffer selection on chromatographic figures of merit in chiral microemulsion electrokinetic chromatography. Electrophoresis 25:3247–3256
Mertzman MD, Foley JP (2004) Effect of oil substitution in chiral microemulsion electrokinetic chromatography. Electrophoresis 25:723–732
Kahle KA, Foley JP (2007) Two-chiral-component microemulsion electrokinetic chromatography-chiral surfactant and chiral oil: part 1. Dibutyl tartrate. Electrophoresis 28:1723–1734
Kahle KA, Foley JP (2007) Two-chiral component microemulsion EKC—chiral surfactant and chiral oil. Part 2: diethyl tartrate. Electrophoresis 28:2644–2657
Pomponio R, Gotti R, Luppi B, Cavrini V (2003) Microemulsion electrokinetic chromatography for the analysis of green tea catechins: effect of the cosurfactant on the separation selectivity. Electrophoresis 24:1658–1667
Klampfl C (2003) Solvent effects in microemulsion electrokinetic chromatography. Electrophoresis 24:1537–1543
Zheng ZX, Lin J-M, Chan W-H, Lee AWM, Huie CW (2004) Separation of enantiomers in microemulsion electrokinetic chromatography using chiral alcohols as cosurfactants. Electrophoresis 25:3263–3269
Kahle KA, Foley JP (2006) Chiral microemulsion electrokinetic chromatography with two chiral components: improved separations via synergies between a chiral surfactant and a chiral cosurfactant. Electrophoresis 27:896–904
Kahle KA, Foley JP (2007) Influence of microemulsion chirality on chromatographic figures of merit in EKC: results with novel three-chiral-component microemulsions and comparison with one- and two-chiral-component microemulsions. Electrophoresis 28:3024–3040
Nozal L, Arce L, Simonet BM, Rios A, Valcarcel M (2006) Microemulsion electrokinetic chromatography separation using hexane-in-water microemulsion without cosurfactant: comparison with MEKC. Electrophoresis 27:4439–4445
Hilder EF, Klampfl CW, Buchberger W, Haddad PR (2001) Separation of hydrophobic polymer additives by microemulsion electrokinetic chromatography. J Chromatogr A 922:293–302
Altria KD, Clark BJ, Mahuzier P-E (2000) The effect of operating variables in microemulsion capillary chromatography. Chromatographia 52:758–768
Borst C, Holzgrabe U (2013) Cyclodextrin-mediated enantioseparation in microemulsion electrokinetic chromatography. Meth Mol Biol 970:363–375
Borst C, Holzgrabe U (2008) Enantioseparation of DOPA and related compounds by cyclodextrin-modified microemulsion electrokinetic chromatography. J Chromatogr A 1204:191–196
Borst C, Holzgrabe U (2010) Comparison of chiral electrophoretic separation methods for phenethylamines and application on impurity analysis. J Pharm Biomed Anal 53:1201–1209
Cao J, Qu H, Cheng Y (2010) Separation of flavonoids and phenolic acids in complex natural products by microemulsion electrokinetic chromatography using surfactant-coated and carboxylic single-wall carbon nanotubes as additives. Electrophoresis 31:1689–1696
Cao J, Dun W, Qu H (2011) Evaluation of the addition of various surfactant-suspended carbon nanotubes in MEEKC with an in situ-synthesized surfactant system. Electrophoresis 32:408–413
Cao J, Li P, Chen J, Tan T, Dai H-B (2013) Enhanced separation of compound Xueshuantong capsule using functionalized carbon nanotubes with cationic surfactant solutions in MEEKC. Electrophoresis 34:324–330
Ni X, Yu M, Cao Y, Cao G (2013) Microstructure of microemulsion modified with ionic liquids in microemulsion electrokinetic chromatography and analysis of seven corticosteroids. Electrophoresis 34:2568–2576
Hsieh S-Y, Wang CC, Wu SM (2013) Microemulsion electrokinetic chromatography for analysis of phthalates in soft drinks. Food Chem 141:3486–3491
Cao W, Hu S-S, Li X-Y, Pang X-Q, Cao J, Ye L-H, Dai H-B, Liu X-J, Da J-H, Chu C (2014) Highly sensitive analysis of flavonoids by zwitterionic microemulsion electrokinetic chromatography coupled with light-emitting diode-induced fluorescence detection. J Chromatogr A 1358:277–284
Cao J, Dun WL (2011) Separation and sweeping of flavonoids by microemulsion electrokinetic chromatography using mixed anionic and cationic surfactants. Talanta 84:155–159
Altria KD, Broderick MF, Donegan S, Power J (2004) The use novel water-in-oil microemulsions in microemulsion electrokinetic chromatography. Electrophoresis 25:645–652
Broderick M, Donegan S, Power J, Altria K (2005) Optimisation and use of water-in-oil MEEKC in pharmaceutical analysis. J Pharm Biomed Anal 37:877–884
Nyunt KTN, Prutthiwanasan B, Suntornsuk L (2013) Microemulsion electrokinetic chromatography of β-carotene and astaxanthin. J Liq Chrom Rel Techn 36:671–686
Bitar Y, Holzgrabe U (2007) Impurity profiling of atropine sulfate by microemulsion electrokinetic chromatography. J Pharm Biomed Anal 44:623–633
Cao J, Chen J, Yi L, Li P, Qi L-W (2008) Comparison of oil-in-water and water-in-oil microemulsion electrokinetic chromatography as methods for the analysis of eight phenolic acids and five diterpenoids. Electrophoresis 29:2310–2320
Huang H-Y, Liu W-L, Singco B, Hsieh S-H, Shih Y-H (2011) On-line concentration sample stacking coupled with water-in-oil microemulsion electrokinetic chromatography. J Chromatogr A 1218:7663–7669
Terabe S (2009) Capillary separation: micellar electrokinetic chromatography. Ann Rev Anal Chem 2:99–120
Lin W-C, Liu W-L, Tang W-Y, Huang C-P, Huang H-Y, Chin T-Y (2014) Determination of amino acids by microemulsion electrokinetic chromatography laser induced fluorescence method. Electrophoresis 35:1751–1755
Klampfl CW, Buchberger W (2010) Recent advances in the use of capillary electrophoresis coupled to high-resolution mass spectrometry for the analysis of small molecules. Curr Anal Chem 6:118–125
Himmelsbach M, Haunschmidt M, Buchberger W, Klampfl CW (2007) Microemulsion electrokinetic chromatography with on-line atmospheric pressure photoionization mass spectrometric detection. Anal Chem 79:1564–1568
Himmelsbach M, Haunschmidt M, Buchberger W, Klampfl CW (2007) Microemulsion electrokinetic chromatography with on-line atmospheric pressure photoionization-mass spectrometric detection of medium polarity compounds. J Chromatogr A 1159:58–62
Schappler J, Guillarme D, Rudaz S, Veuthey J-L (2008) Microemulsion electrokinetic chromatography hyphenated to atmospheric pressure photoionization mass spectrometry. Electrophoresis 29:11–19
Bytzek AK, Reithofer MR, Galanski M, Groessl M, Keppler BK, Hartinger CG (2010) The first example of MEEKC-ICP-MS coupling and its application for the analysis of anticancer platinum complexes. Electrophoresis 31:1144–1150
Pedersen-Bjergaard S, Naess O, Moestue S, Rasmussen KE (2000) Microemulsion electrokinetic chromatography in suppressed electroosmotic flow environment. Separation of fat-soluble vitamins. J Chromatgr A 876:201–211
Oledzka I, Kowalski P, Baluch A, Baczek T, Paradziej-Lukowicz J, Taciak M, Pastuszewska B (2014) Quantification of the level of fat-soluble vitamins in feed based on the novel microemulsion electrokinetic chromatography (MEEKC) method. J Sci Food Agric 94:544–551
Yin C, Cao Y, Ding S, Wang Y (2008) Rapid determination of water- and fat-soluble vitamins with microemulsion electrokinetic chromatography. J Chromatogr A 1193:172–177
Altria KD (1999) Application of microemulsion electrokinetic chromatography to a wide range of pharmaceuticals and excipients. J Chromatogr A 844:371–386
Miksik I, Gabriel J, Deyl Z (1997) Microemulsion electrokinetic chromatography of diphenylhydrazones of dicarbonyl sugars. J Chromatogr A 772:297–303
Miksik I, Deyl Z (1998) Microemulsion electrokinetic chromatography of fatty acids as phenacyl esters. J Chromatogr A 807:111–119
Abromeit H, Werz O, Scriba GKE (2013) Separation of 5-lipoxygenase metabolites using cyclodextrin-modified microemulsion electrokinetic chromatography and head column field-amplified sample stacking. Chromatographia 76:1187–1192
Sun S-W, Yeh P-C (2005) Analysis of rhubarb anthraquinones and bianthrones by microemulsion electrokinetic chromatography. J Pharm Biomed Anal 36:995–1001
Chen Z, Lin Z, Zhang L, Cai Y, Zhang L (2012) Analysis of plant hormones by microemulsion electrokinetic capillary chromatography coupled with on-line large volume sample stacking. Analyst 137:1723–1729
Darji V, Boyce MC, Bennett I, Breadmore MC, Quirino J (2010) Determination of food grade antioxidants using microemulsion electrokinetic chromatography. Electrophoresis 31:2267–2271
Huang H-Y, Chuang C-L, Chiu C-W, Yeh J-M (2005) Application of microemulsion electrokinetic chromatography for the detection of preservatives in foods. Food Chem 89:315–322
Huang H-Y, Chuang C-L, Chiu CW, Chung M-C (2005) Determination of food colorants by microemulsion electrokinetic chromatography. Electrophoresis 26:867–877
Javor T, Buchberger W, Tanzcos I (2000) Determination of low-molecular-mass phenolic and non-phenolic lignin degradation compounds in wood digestion solutions by capillary electrophoresis. Microchim Acta 135:45–53
Klampfl CW, Leitner T (2003) Quantitative determination of UV filters in sunscreen lotions using microemulsion electrokinetic chromatography. J Sep Sci 26:1259–1262
Jiang T-F, Lv Z-H, Wang Y-H, Yue M-E, Lian S (2009) Separation and determination of nitrofuran antibiotics in turbot fish by microemulsion electrokinetic chromatography. Anal Sci 25:861–864
Wei S, Lin J, Li H, Lin J-M (2007) Separation of seven fluoroquinolones by microemulsion electrokinetic chromatography and application to ciprofloxacin, lomefloxacin determination in urine. J Chromatogr A 1163:333–336
Macia A, Borrull F, Calull M, Aguilar C (2005) Separation and on-column preconcentration of some nonsteroidal anti-inflammatory drugs by microemulsion electrokinetic capillary chromatography using high-speed separations. Electrophoresis 26:970–979
Fogarty B, Dempsey E, Regan F (2003) Potential for microemulsion electrokinetic chromatography for the separation of priority endocrine disrupting compounds. J Chromatogr A 1014:129–139
Lin Z, Zhang J, Cui H, Zhang L, Chen G (2010) Determination of phthalate esters in soil by microemulsion electrokinetic chromatography coupled with accelerated solvent extraction. J Sep Sci 33:3717–3725
Li R-H, Liu D-H, Yang Z-H, Zhou Z-Q, Wang P (2012) Vortex-assisted surfactant-enhanced-emulsification liquid-liquid microextraction for the determination of triazine herbizides in water samples by microemulsion electrokinetic chromatography. Electrophoresis 33:2176–2183
Huang H-Y, Wei M, Lin Y-R, Lu P-H (2009) Determination of organic impurities in mother liquors from oxidative terephthalic acid synthesis by microemulsion electrokinetic chromatography. J Chromatogr A 216:2560–2566
Altria KD (2000) Background theory and applications of microemulsion electrokinetic chromatography. J Chromatogr 892:171–186
Poole SK, Durham D, Kibbey C (2000) Rapid method for estimating the octanol-water partition coefficient by microemulsion electrokinetic chromatography. J Chromatogr B 745:117–126
Henchoz Y, Romand S, Schappler J, Rudaz S, Veuthey J-L, Carrupt P-A (2010) High-throughput log P determination by MEEKC coupled with UV and MS detections. Electrophoresis 31:952–964
Jiang X, Xia Z, Deng L, Wei W, Chen J, Xu J, Li H (2012) Evaluation of accuracy for the measurement of octanol–water partition coefficient by MEEKC. Chromatographia 75:347–352
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2016 Springer Science+Business Media New York
About this protocol
Cite this protocol
Buchberger, W. (2016). Microemulsion Electrokinetic Chromatography. In: Schmitt-Kopplin, P. (eds) Capillary Electrophoresis. Methods in Molecular Biology, vol 1483. Humana Press, New York, NY. https://doi.org/10.1007/978-1-4939-6403-1_6
Download citation
DOI: https://doi.org/10.1007/978-1-4939-6403-1_6
Published:
Publisher Name: Humana Press, New York, NY
Print ISBN: 978-1-4939-6401-7
Online ISBN: 978-1-4939-6403-1
eBook Packages: Springer Protocols